← Product Code [LZA](/productcode/LZA) · K190080

# Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black) (K190080)

_Comfort Rubber Gloves Industries Sdn. Bhd. · LZA · Jun 27, 2019 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K190080

## Device Facts

- **Applicant:** Comfort Rubber Gloves Industries Sdn. Bhd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jun 27, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black) is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use chemotherapy drugs. Tested drugs are as follows: Cisplatin 1.0 mg/ml (≥ 240 mins), Cyclophosphamide (Cytoxan) 20 mg/ml (≥ 240 mins), Dacarbazine (DTIC) 10.0 mg/ml (≥ 240 mins), Doxorubicin Hydrochloride 2.0 mg/ml (≥ 240 mins), Etoposide (Toposar) 20.0 mg/ml (≥ 240 mins), Fluorouracil 50.0 mg/ml (≥ 240 mins), Paclitaxel (Taxol) 6.0 mg/ml (≥ 240 mins). Please note that the following drugs have extremely low permeation time for Carmustine (BCNU) 3.3mg.ml - 54.1 (mins) Thiotepa 10.0 mg/ml - 16.0 (mins)

## Device Story

Disposable nitrile examination glove; worn by healthcare personnel to prevent cross-contamination and protect against chemotherapy drug exposure. Input: physical barrier to chemical agents. Operation: material properties provide resistance to drug permeation. Output: physical protection for the wearer. Used in clinical settings. Benefit: reduces risk of skin contact with hazardous chemotherapy agents during handling.

## Clinical Evidence

No clinical data. Bench testing only, including physical property testing (ASTM D412, D573, D3767), watertight testing (ASTM D5151), residual powder testing (ASTM D6124), chemotherapy drug permeation testing (ASTM D6978), and biocompatibility (ISO 10993-5, ISO 10993-10).

## Technological Characteristics

Nitrile material; powder-free; black color. Complies with ASTM D6319-10(2015) for dimensions, physical properties, and thickness. Permeation resistance tested per ASTM D6978-05. Biocompatibility per ISO 10993-5 and ISO 10993-10. Single-use, non-sterile.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) ([K180476](/device/K180476.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 27, 2019

Comfort Rubber Gloves Industries Sdn. Bhd. Chan Men QA & QMS Manager Lot 821, Jalan Matang Matang, 34750 My

Re: K190080

Trade/Device Name: Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black) Regulation Number: 21 CFR 880.6250

Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: March 28, 2019 Received: April 01, 2019

Dear Chan Men:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

### 510(k) Number (if known)

K190080

### Device Name

Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black)

### Indications for Use (Describe)

The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black) is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use chemotherapy drugs. Tested drugs are as follows:

|                                     | Average Breakthrough Detection Time (minutes) |
|-------------------------------------|-----------------------------------------------|
| Cisplatin 1.0 mg/ml                 | ≥ 240                                         |
| Cyclophosphamide (Cytoxan) 20 mg/ml | ≥ 240                                         |
| Dacarbazine (DTIC) 10.0 mg/ml       | ≥ 240                                         |
| Doxorubicin Hydrochloride 2.0 mg/ml | ≥ 240                                         |
| Etoposide (Toposar) 20.0 mg/ml      | ≥ 240                                         |
| Fluorouracil 50.0 mg/ml             | ≥ 240                                         |
| Paclitaxel (Taxol) 6.0 mg/ml        | ≥ 240                                         |

Please note that the following drugs have extremely low permeation time for

* Carmustine (BCNU) 3.3mg.ml - 54.1 (mins)

*Thiotepa 10.0 mg/ml - 16.0 (mins)

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

# 510(k) SUMMARY K190080

# POWDER FREE NITRILE EXAMINATION GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUGS LABELING CLAIM (BLACK)

#### 1.0 Submitter:

| Name           | Comfort Rubber Gloves Industries Sdn. Bhd.                            |
|----------------|-----------------------------------------------------------------------|
| Address        | Lot 821, Jalan Matang,<br>34750 Matang, Perak, Malaysia.<br>Malaysia. |
| Phone No.      | 605-847 2777                                                          |
| Fax No.        | 605-847 9108                                                          |
| Contact Person | Chan Yew Men (Mr.)                                                    |

Date of Preparation: June 27, 2019

#### 2.0 Name of the Device

Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black)

Common Name: Patient Examination Gloves

Classification Name: Patient Examination Gloves (21 CFR 880.6250 product code LZA) Patient Examination Gloves Specialty (21 CFR 880.6250 product code LZC

510(K) Number: K190080

#### 3.0 Predicate Device

Device Name: Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Green) Company: Comfort Rubber Gloves Industries Sdn. Bhd. 510(K)

No.: K180476

#### 4.0 Description of the Device:

Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black) meets all the requirements of ASTM D6319 - 10(2015) Standard Specification for Nitrile Examination Gloves for Medical Application.

#### 5.0 Indication for Use of the Device

The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black) are a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition, these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use with chemotherapy drugs.

#### 6.0 Summary of the Technological Characteristics of the Device:

The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim (Black) are summarized with the following technological characteristics compared to ASTM D6319 - 10(2015) Standard Specification for Nitrile Examination Gloves for Medical Application or equivalent standards as shown in Table 1.

{4}------------------------------------------------

Chemotherapy claim is similar to Predicate, which has a gloves thickness comply with the ASTM Standards.

## Table 1

| CHARACTERISTICS                      | STANDARDS                                                                                                                                                 | PREDICATE DEVICE<br>K180476                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | SUBJECT DEVICE<br>K190080                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | COMPARISON |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer(s)                      |                                                                                                                                                           | Comfort Rubber<br>Gloves Industries<br>Sdn. Bhd                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Comfort Rubber Gloves<br>Industries Sdn. Bhd                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same       |
| Device Name                          |                                                                                                                                                           | Powder Free Nitrile<br>Examination Gloves Tested<br>for Use with Chemotherapy<br>Drugs Labeling Claim<br>(Green)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Powder<br>Free<br>Nitrile<br>Examination Gloves Tested<br>for Use with Chemotherapy<br>Drugs Labeling Claim<br>(Black)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Similar    |
| Dimension                            | ASTM 6319-10<br>(2015)                                                                                                                                    | Length-Min 240mm<br>Thickness palm and finger-<br>Min 0.05mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Length-Min 230mm<br>Thickness palm and finger-<br>Min 0.05mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Similar    |
| Physical Properties                  | ASTM 6319-10<br>(2015)                                                                                                                                    | Min - 20.23<br>MPa Min - 445<br>%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Min - 22.31 MPa<br>Min - 579%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Similar    |
| Thickness - Finger<br>- Palm         | ASTM 6319-10<br>(2015)                                                                                                                                    | 0.09 mm- 0.10<br>mm 0.12mm –<br>0.14 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 0.09 mm- 0.10 mm<br>0.11 mm – 0.13 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Similar    |
| Powder Content                       | ASTM 6124-06<br>(2011)<br>(≤ 2 mg/glove)                                                                                                                  | Max - 2 mg/gloves<br>0.20 mg/glove                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Max - 2 mg/gloves<br>0.60 mg/glove                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same       |
| Chemotherapy Drug<br>Permeation Test | ASTM D6978-<br>05                                                                                                                                         | Below                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Below                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Similar    |
| Test Chemotherapy<br>Drug            | Concentration                                                                                                                                             | Minimum Breakthrough Detection Time (min)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |            |
| *Carmustine (BCNU)                   | 3.3 mg/ml                                                                                                                                                 | 23.4                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 54.1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |            |
| Cisplatin                            | 1.0 mg/ml                                                                                                                                                 | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |            |
| Cyclophosphami<br>de (Cytoxan)       | 20 mg/ml                                                                                                                                                  | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |            |
| Dacarbazine (DTIC)                   | 10.0 mg/ml                                                                                                                                                | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |            |
| Doxorubicin Hydrochloride            | 2.0 mg/ml                                                                                                                                                 | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |            |
| Etoposide (Toposar)                  | 20.0 mg/ml                                                                                                                                                | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |            |
| Fluorouracil                         | 50.0 mg/ml                                                                                                                                                | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |            |
| Paclitaxel (Taxol)                   | 6.0 mg/ml                                                                                                                                                 | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | >240                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |            |
| *Thiotepa                            | 10.0 mg/ml                                                                                                                                                | 16.2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 16.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |            |
| Warning Statement                    |                                                                                                                                                           | * WARNING :<br>Please note that the<br>following drugs have<br>extremely low permeation<br>times Carmustine (BCNU):<br>23.4 minutes and Thiotepa:<br>16.0 minutes.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | * WARNING :<br>Please note that the<br>following drugs have<br>extremely low permeation<br>times Carmustine (BCNU):<br>54.1 minutes and Thiotepa:<br>16.0 minutes.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |            |
| CHARACTERISTICS                      | STANDARDS                                                                                                                                                 | PREDICATE DEVICE<br>K180476                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | SUBJECT DEVICE<br>K190080                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | COMPARISON |
| Biocompatibility                     | Primary Skin Irritation ISO<br>10993-10:2010<br>Biological evaluation of<br>medical devices -- Part<br>10: Tests for irritation and<br>skin sensitization | Under the conditions of the<br>study, the subject device<br>extracts are not irritating to<br>the animal model.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Under the conditions of the<br>study, the subject device<br>extracts are not irritating to<br>the animal model.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same       |
|                                      | ISO 10993-10:2010<br>Biological evaluation of<br>medical devices -- Part<br>10: Tests for irritation and<br>skin sensitization                            | Under the conditions of the<br>study, the subject device<br>extracts are not sensitizing<br>to the animal model.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Under the conditions of the<br>study, the subject device<br>extracts are not sensitizing<br>to the animal model.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same       |
|                                      | Cytotoxicity ISO 10993-<br>5:2009 - Biological<br>evaluation of medical<br>devices -- Part 5: Tests for<br>in vitro cytotoxicity                          | Under the conditions of<br>the study, the subject<br>device extract exhibits<br>mild cytotoxicity reactivity<br>result (score of 2) with the<br>neat extract (100%).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Under the conditions of<br>the study, the subject<br>device extract exhibits<br>mild cytotoxicity reactivity<br>result (score of 2) with the<br>neat extract (100%).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same       |
| Watertight (1000ml)                  | 21 CFR 800.20<br>ASTM D5151                                                                                                                               | Passes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Passes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same       |
| Indication for Use                   |                                                                                                                                                           | The Powder Free Nitrile<br>Examination Gloves Tested<br>for Use with Chemotherapy<br>Drugs Labeling Claim<br>(Green) is a specialty<br>medical glove which is a<br>disposable device intended<br>for medical purpose that is<br>worn on the examiner's hand<br>or finger to prevent<br>contamination between<br>examiner and patient. In<br>addition, these gloves are<br>worn to protect the wearer<br>against exposure to<br>chemotherapy drugs.<br>Tested chemotherapy drugs<br>are as follows:<br>Average Breakthrough<br>Detection Time (minutes<br>Cisplatin, 1.0 mg/ml - ≥ 240<br>Cyclophosphamide<br>(Cytoxan), 20.0 mg/ml - ≥<br>240<br>Dacarbazine (DTIC), 10.0<br>mg/ml - ≥ 240<br>Doxorubicin Hydrochloride,<br>2.0 mg/ml - ≥ 240<br>Etoposide (Toposar), 20.0<br>mg/ml - ≥ 240<br>Fluorouracil, 50.0 mg/ml - ≥<br>240<br>Paclitaxel (Taxol), 6.0 mg/ml<br>- ≥ 240<br>Please note that the | The Powder Free Nitrile<br>Examination Gloves Tested<br>for Use with Chemotherapy<br>Drugs Labeling Claim<br>(Black) is a specialty medical<br>glove which is a disposable<br>device intended for medical<br>purpose that is worn on the<br>examiner's hand or finger to<br>prevent contamination<br>between examiner and<br>patient. In addition, these<br>gloves are worn to protect<br>the wearer against exposure<br>to chemotherapy drugs.<br>Tested chemotherapy drugs<br>are as follows: Average<br>Breakthrough<br>Detection Time (minutes)<br>Cisplatin, 1.0 mg/ml - ≥ 240<br>Cyclophosphamide<br>(Cytoxan), 20.0 mg/ml - ≥<br>240<br>Dacarbazine (DTIC), 10.0<br>mg/ml - ≥ 240<br>Doxorubicin Hydrochloride,<br>2.0 mg/ml - ≥ 240<br>Etoposide (Toposar), 20.0<br>mg/ml - ≥ 240<br>Fluorouracil, 50.0 mg/ml - ≥<br>240<br>Paclitaxel (Taxol), 6.0 mg/ml<br>- ≥ 240<br>Please note that the | Same       |
|                                      |                                                                                                                                                           | following drugs have<br>extremely low permeation<br>time for: Carmustine (BCNU)<br>3.3mg/ml - 23.4 (mins)<br>Thiotepa 10.0 mg/ml -<br>16.2 (mins                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | following drugs have<br>extremely low permeation<br>time for: Carmustine (BCNU)<br>3.3mg/ml - 54.1 (mins)<br>Thiotepa 10.0 mg/ml -<br>16.0 (mins                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |            |
| Material                             | ASTM D6319 - 10(2015)                                                                                                                                     | Nitrile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Nitrile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same       |
| Color                                |                                                                                                                                                           | Green                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Black                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Different  |
| Size                                 | Medical Glove<br>Guidance Manual<br>Labeling                                                                                                              | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same       |
| Single Use                           | Medical Glove Guidance<br>Manual -<br>Labeling                                                                                                            | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same       |

{5}------------------------------------------------

{6}------------------------------------------------

#### 7.0 Summary of Non-Clinical Performance Data

Non-clinical tests were conducted to demonstrate that the proposed device met all design specifications. The test results demonstrated that the proposed device met the performance criteria with the following standards:

- ASTM D412-2016 Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers— ● Tension
- ASTM D573-2004 (Reapproved 2010) Standard Test Method for Rubber-Deterioration in an Air Oven .
- ASTM D3767-03 Standard Practice for Rubber Measurement of Dimensions ●
- ASTM D5151-2006 (Reapproved 2015) Standard Test Method for Detection of Holes in Medical ● Gloves
- . ASTM D6124-2006 (Reapproved 2001) Standard Tested Method for Residual Powder on Medical Gloves
- ASTM D6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application ●
- ASTM D6978-2005(Reapproved 2019) Standard Practice for Assessment of Resistance of Medical ● Gloves to Permeation by Chemotherapy Drugs
- . ISO 2859 Sampling Procedures and Tables for Inspection by Attributes
- ISO 10993-5 Biological evaluation of medical devices-Part5 Tests for in vivo cytotoxicity ●
- ISO 10993-10 Biological evaluation of medical devices-Part 10 Test for irritation and delayed-type . hypersensitivity

## 8.0 Clinical Performance Data

Clinical data is not needed.

#### 9.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K190080](https://fda-staging.innolitics.com/device/K190080)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).