SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software (DACS)
K190069 · Arrowhead De, LLC · KTT · May 14, 2019 · Orthopedic
Device Facts
Record ID
K190069
Device Name
SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software (DACS)
Applicant
Arrowhead De, LLC
Product Code
KTT · Orthopedic
Decision Date
May 14, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SixFix™ Hexapod and Deformity Analysis and Correction Software (DACS) are intended to be used for post-traumatic joint contracture which has resulted in loss of range of motion; fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction; open and closed fracture fixation; pseudo-arthrosis of long bones; limb lengthening by epiphyseal or metaphyseal distraction; correction of bony or soft tissue deformities; correction of bony or soft tissue defects; joint arthrodesis; infected fractures or nonunions.
Device Story
SixFix™ Hexapod Fixator is a multilateral circular external fixation system comprising rings, footplates, arches, struts, threaded rods, and wires. System treats bone deformities and fractures by adjusting strut lengths to alter ring orientation, thereby manipulating bone segments connected via half-pins and wires. DACS is an optional software component used by physicians to calculate required strut lengths for desired bone translation/rotation. Physician inputs parameters into software; software computes necessary adjustments and visualizes bone position. Used in clinical settings to assist surgeons in planning and implementing orthopedic corrections. Benefits include precise, gradual bone manipulation to achieve treatment goals for complex orthopedic conditions.
Clinical Evidence
No clinical data. Evidence consists of static and dynamic mechanical bench testing and software verification/validation testing. Software documentation provided for Moderate Level of Concern, including hazard analysis, requirements, design specifications, and cybersecurity risk analysis.
Technological Characteristics
Multilateral circular external fixation system. Components: rings, footplates, arches, struts, threaded rods, wires. Materials: metallic. Software: DACS for deformity analysis and strut length calculation. Connectivity: not specified. Sterilization: provided non-sterile, single use. Software class: Moderate Level of Concern.
Indications for Use
Indicated for patients requiring treatment for post-traumatic joint contracture, fractures, pseudo-arthrosis of long bones, limb lengthening, bony or soft tissue deformities/defects, joint arthrodesis, or infected fractures/nonunions.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA part includes the acronym in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 5, 2022
Arrowhead DE, LLC Patrick Mullaney President/CEO 328 Poplar View Lane East, Suite 2 Collierville, Tennessee 38017
Re: K190069
Trade/Device Name: SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software (DACS) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT, HTY, OSN
Dear Patrick Mullaney:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated May 14, 2019. Specifically, FDA is updating this SE Letter to address the inconsistent indications for use in the FDA Form 3881 as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Ting Song, Ph.D., OHT6: Office of Orthopedic Devices, 301-796-7677, Ting.Song@fda.hhs.gov.
Sincerely,
Ting Song -S
Ting Song, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 14, 2019
Arrowhead DE, LLC Patrick Mullaney President/CEO 328 Poplar View Lane East, Suite 2 Collierville, Tennessee 38017
Re: K190069
Trade/Device Name: SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software (DACS) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories Regulatory Class: Class II Product Code: KTT, HTY, OSN Dated: April 17, 2019 Received: April 18, 2019
Dear Patrick Mullaney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Digitally signed by Peter G. Peter G. Allen -S Date: 2019.05.14 Allen -S 10:08:37 -04'00'
FOR CAPT Raquel Peat, PhD, MPH, USPHS Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190069
#### Device Name
SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software (DACS)
#### Indications for Use (Describe)
The SixFix™ Hexapod and Deformity Analysis and Correction Software (DACS) are intended to be used for posttraumatic joint contracture which has resulted in loss of range of motion; fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction; open and closed fracture fixation; pseudo-arthrosis of long bones; limb lengthening by epiphyseal distraction; correction of bony or soft tissue deformities; correction of bony or soft tissue defects; joint arthrodesis; infected fractures or nonunions.
Type of Use (Select one or both, as applicable):
| <span> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------------------|
| <span> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary SixFix™ Hexapod Fixator and DACS K190069
## l. Submitter:
Arrowhead Medical Device Technologies, LLC 328 Poplar View Lane East, Suite 2 Collierville, TN 38017
Contact Person: Thomas J. Twardzik Vice President, Marketing and Operations Office: (901) 853-4366 Fax: (206) 222-9173 Email: INFO@ArrowheadDevices.com
Date of Summary: May 7, 2019
## II: Device
Proprietary Name: SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software (DACS)
Common Name: Orthopaedic Software for Treatment and Instrumentation Regulatory Class: Class II
Regulation:
21 CFR 888.3030 Single/multiple component metallic bone fixation appliances and accessories
21 CFR 888.3040 Smooth or threaded metallic bone fixation fastener Device Product Codes: KTT, HTY, and OSN Panel: Orthopedic
| III. Predicate Devices | |
|------------------------|--|
|------------------------|--|
| Device | Manufacturer | 510(k) No. | Clearance<br>Date |
|--------------------------------------------------------------------------------|--------------------------------------------------|------------|-------------------|
| Primary Predicate | | | |
| Taylor Spatial Frame | Smith & Nephew | K093047 | 09/27/2010 |
| | | K970748 | 05/09/1997 |
| Secondary Predicates | | | |
| Deformity Analysis and<br>Correction Software<br>(DASC) and<br>Instrumentation | Arrowhead Medical<br>Device<br>Technologies, LLC | K180539 | 08/13/2018 |
| Ilizarov External<br>Fixation Wires | Smith & Nephew | K870961 | 03/19/1987 |
| | | K962808 | 08/19/1996 |
{5}------------------------------------------------
## 510(k) Summary SixFix™ Hexapod Fixator and DACS K190069
## IV. Device Description
The SixFix™ Hexapod Fixator is a multilateral circular external fixation system. The system includes the following external fixator elements: rings, footplates, arches, struts, threaded rods, wires, external fixation accessories, and software. All of the elements are provided non-sterile and are for single use only.
The system is designed such that gradually adjusting the lengths of the struts in relation to one another alters the orientation of the rings and, consequently, the bone segments connected to the rings by half-pins and wires during the treatment period in order to achieve the patient's treatment goals.
The Deformity Analysis and Correction Software is an optional software component and is used to assist the physician in calculating the lengths of the struts connecting the rings to manipulate the bone fragments. The software receives inputs from the physician and allows the physician to visualize the moving bone position. The program computes the strut lengths necessary to implement any desired translation and/or rotation required by the surgeon.
## V. Intended Use
The SixFix™ Hexapod and Deformity Analysis and Correction Software (DACS) are intended to be used for post-traumatic joint contracture which has resulted in loss of range of motion; fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction; open and closed fracture fixation; pseudo-arthrosis of long bones; limb lengthening by epiphyseal or metaphyseal distraction; correction of bony or soft tissue deformities; correction of bony or soft tissue defects; joint arthrodesis; infected fractures or nonunions.
## VI. Comparison of Technological Characteristics with the Predicate Devices
The SixFix™ Hexapod Fixator is technologically substantially equivalent to predicate devices in terms of intended use, material, design, mechanical performance and safety. Testing and analyses confirmed that the Hexapod™ Fixator is substantially equivalent when compared to the predicate device. The design characteristics of the subject system do not raise any new types of questions of safety or effectiveness. From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate device.
The subject device and predicate devices are optional software programs that are used with a spatial frame external fixation system. The functions performed by the SixFix™ DACS match the functions performed by the predicate DACS (K180539). Software validation confirmed that the DACS should perform as intended.
{6}------------------------------------------------
## 510(k) Summary SixFix™ Hexapod Fixator and DACS K190069
### VII. Performance Data
Static and dynamic mechanical testing were performed in support of the submission. The testing confirmed that the subject SixFix™ Hexapod Fixator is substantially equivalent to the predicate device.
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a Moderate Level of Concern. Documentation for Moderate Level of Concern was provided as follows.
- Device Hazard Analysis
- Software Requirements Specification ●
- Architectural Design Charts
- Software Design Specification Traceability Analysis
- Software Development Environment Description
- Software Verification Plan and Protocol
- Variety Testing Software Verification Protocol
- Cybersecurity Risk Analysis Report
- Off-the-Shelf Software Analysis
- Software Release Record
- Unresolved Anomaly Report
- Verification and Validation Documentation
From the evidence submitted in this 510(k) for the software, the Deformity Analysis and Correction Software (DACS) demonstrates that the device is as safe, as effective, and performs as well as or better than the legally marketed device predicates. The design characteristics of the subject system software do not raise any new types of questions of safety or effectiveness.
## VIII. Conclusions
A review of the device indications, material composition, external element design, and technological characteristics confirmed that the SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software are substantially equivalent to the predicate device. While the SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software are not identical to the predicate device, comparisons of the subject and predicate device confirmed that any differences between the subject device and predicate do not render the device NSE as there is not a new intended use; and any differences in technological characteristics do not raise different questions of safety and effectiveness than the predicate device. Therefore, it is concluded that the SixFix™ Hexapod Fixator and Deformity Analysis and Correction Software is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.