← Product Code [OWB](/productcode/OWB) · K190064

# nView system 1 (K190064)

_Nview Medical, Inc. · OWB · Jul 8, 2019 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K190064

## Device Facts

- **Applicant:** Nview Medical, Inc.
- **Product Code:** [OWB](/productcode/OWB.md)
- **Decision Date:** Jul 8, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

nView system 1 is intended to provide both 2D and 3D imaging of adult and pediatric populations over 6 years of age. The device is intended to provide fluoroscopic and tomographic imaging of patients during orthopedic surgical procedures where the clinician benefits from 3D visualization of complex anatomical structures, such as high contrast objects, bones, joints, cervical, thoracic, and lumbar regions of the spine, and joint fractures of the upper and lower extremities. The device is indicated to image human anatomy up to 30 cm thickness. The device is not indicated for mammographic or lung nodule applications.

## Device Story

Mobile fluoroscopic C-arm and monitor cart system; provides 2D/3D X-ray imaging for orthopedic surgery. Input: X-ray projections captured via stationary gantry with rotating internal X-ray tube. Processing: Iterative reconstruction algorithm generates 3D tomographic volumes from multi-arc scan data. Output: 2D fluoroscopic projections and 3D anatomical reconstructions displayed on monitor cart. Used in hospital/clinic settings by clinicians. Benefits: Enables intraoperative visualization of complex bony anatomy; faster acquisition/reconstruction times compared to predicate; lower radiation dose for thin patients.

## Clinical Evidence

Bench testing only. Performance validated via engineering bench tests, imaging performance evaluations, and human factors/usability testing in simulated clinical environments. Conformance to IEC 60601 series and 21 CFR 1020.30/32 standards demonstrated. No clinical trial data presented.

## Technological Characteristics

Mobile C-arm/monitor cart; stationary anode X-ray tube (75kV/12mA/350W); 30cm x 30cm CMOS digital detector (1952x1952 resolution). Multi-arc source trajectory (117-degree). Iterative reconstruction algorithm. Connectivity via USB/HDMI. Power: 120 VAC. Software: Moderate level of concern.

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Ziehm Vision RFD 3D ([K142740](/device/K142740.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

July 8, 2019

Image /page/0/Picture/1 description: The image contains the logos of the U.S. Food and Drug Administration (FDA) and the Department of Health & Human Services. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The Department of Health & Human Services logo is a symbol of three figures in profile, representing health, services, and people.

nView medical, Inc. % Ms. Lisa Last Regulatory Affairs Leader 2681 E. Parleys Way, Suite 107 SALT LAKE CITY UT 84109

Re: K190064

Trade/Device Name: nView system 1 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO, JAA, JAK Dated: June 7, 2019 Received: June 12, 2019

Dear Ms. Last:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K190064

Device Name nView system 1

#### Indications for Use (Describe)

n View system 1 is intended to provide both 2D and 3D imaging of adult and pediations over 6 years of age. The device is intended to provide fluoroscopic imaging of patients during orthopedic surgical procedures where the clinician benefits from 3D visualization of complex anatomical structures, such as high contrast objects, bones, joints, cervical, thoracic, and lumbar regions of the spine, and joint fractures of the upper and lower extremities.

The device is indicated to image human anatomy up to 30 cm thickness.

The device is not indicated for mammographic or lung nodule applications.

Type of Use (Select one or both, as applicable)

| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>             |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for nView medical. The "n" in nView is blue, while the rest of the letters are dark gray. There is a blue dot above the "i" in nView. Below nView, the word "medical" is written in a smaller font.

# 510(k) Summary

The following statement is being submitted in accordance with the requirements of 21 CFR 807.92.

# Submitter Information

- Submitter: nView medical, Inc. 2681 E. Parleys Way, Suite 107 Salt Lake City, Utah 84109
- Contact: Ms. Lisa Last Chief Operating Officer/ Regulatory Affairs lisa.last@nviewmed.com T 617.283.7053

Preparation Date: June 7, 2019

# Subject Device Information

| Device Name:          | nView system 1                              |
|-----------------------|---------------------------------------------|
| Common/Usual Name:    | Mobile Fluoroscopic C-Arm                   |
| Regulation Number:    | CFR 892.1650                                |
| Regulation Name:      | Image-intensified fluoroscopic x-ray system |
| Regulatory Class:     | II                                          |
| Primary Product Code: | OWB                                         |

# Predicate Device Information

| Device Name (510(k) number): | Ziehm Vision RFD 3D (K142740)               |
|------------------------------|---------------------------------------------|
| Common/Usual Name:           | Mobile Fluoroscopic C-Arm                   |
| Regulation Number:           | CFR 892.1650                                |
| Regulation Name:             | Image-intensified fluoroscopic x-ray system |
| Regulatory Class:            | II                                          |
| Primary Product Code:        | OWB                                         |

# Device Description:

Device Identification: The nView system 1 mobile fluoroscopic system is cone beam computed tomography X-ray system and a fluoroscopic X-ray imaging system consisting of two mobile units: a mobile C-arm and a monitor cart. The mobile C-arm is comprised of a fixed anode X-ray tube, a high voltage generator, X-ray controls, and a mechanical "C" shaped structure which supports the X-ray tube 001 S001 to K190064 nViewSvstem1

{4}------------------------------------------------

S001 to K190064 - Attachment 1 - Updated 510(k) summary and generator and the image receptor flat panel detector.

The monitor cart is a mobile platform that connects to the mobile C-arm by USB and HDMI cables, and which integrates the flat panel display monitors and user controls.

Device Characteristics: The device contains software. It does not contain biologics, drugs, patient-contacting materials, coatings, additives, single-use, or sterile components.

Environment of Use: The device is intended to be used in a hospital facility during surgery.

Brief Written Description of the Device: The nView system 1 employs X-rays as its imaging technology for visualizing human anatomy in both 2D and 3D. The X-ray tube powered by a generator produces X-rays, which image the patient under control of the user, at the direction of a physician. The images from the system assist the physicians in visualizing the patient's anatomy during surgical procedures. The device provides both real-time image capture visualization suitable for use both during surgery or immediately pre or post-surgery.

Key Performance Specifications/Characteristics of the Device performs both 2D and 3D medical imaging generated by means of an iterative algorithm. The system uses the images of a scan captured with relation to a predefined scan reference frame to compute the three-dimensional representation of the imaged object. The images are displayed on the screen of the monitor cart. It is possible to display projection views as well as orthogonal tomographic views.

## Indications for Use:

nView system 1 is intended to provide both 2D and 3D imaging of adult and pediatric populations over 6 years of age. The device is intended to provide fluoroscopic and tomographic imaging of patients during orthopedic surgical procedures where the clinician benefits from 3D visualization of complex anatomical structures, such as high contrast objects, bones, joints, cervical, thoracic, and lumbar regions of the spine, and joint fractures of the upper and lower extremities.

The device is indicated to image human anatomy up to 30 cm thickness.

The device is not indicated for mammographic or lung nodule applications.

### Comparison of Technology with Predicate:

The technological principle of operation for both the predicate devices is limited angle cone beam CT X-ray imaging. It is based on the collection of a series of projection images of the same anatomy, and applying iterative reconstruction techniques to generate a 3D tomographic reconstruction. The following tables compare the intended use and technological characteristics of the subject and predicate device.

{5}------------------------------------------------

| SUBJECT<br>DEVICE<br>nView system<br>1                     | nView system 1 is intended to provide both 2D and 3D imaging of adult and<br>pediatric populations over 6 years of age. The device is intended to<br>provide fluoroscopic and tomographic imaging of patients during orthopedic<br>surgical procedures where the clinician benefits from 3D visualization of<br>complex anatomical structures, such as high contrast objects, bones, joints,<br>cervical, thoracic, and lumbar regions of the spine, and joint fractures of the<br>upper and lower extremities.<br>The device is indicated to image human anatomy up to 30 cm thickness.<br>The device is not indicated for mammographic or lung nodule applications.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PREDICATE<br>DEVICE<br>Ziehm Vision<br>RFD 3D<br>(K142740) | The Ziehm Vision RFD 3D system is intended for use in providing both 2D<br>and 3D medical imaging for all adult and pediatric populations, using pulsed<br>and continuous fluoroscopic imaging.<br><br>The device provides 2D medical imaging for fluoroscopy, digital subtraction,<br>and acquisition of cine loops during diagnostic interventional and surgical<br>procedures where intraoperative imaging and visualization of complex<br>anatomical structures of both lower and higher contrast density are<br>required. Such procedures may include but are not limited to those of<br>interventional cardiology, heart surgery, hybrid procedures, interventional<br>radiology, interventional angiography, electrophysiology, pediatrics,<br>endoscopic, urological, gastroenterology, orthopedic, maxillofacial surgery,<br>neurology, neurosurgery, critical care, emergency room procedures, and<br>those procedures visualizing structures of the cervical, thoracic, and lumbar<br>regions of the spine and joint fractures of the upper and lower extremities,<br>and where digital image data is required for Computer-Assisted Surgery<br>procedures.<br>The device is also intended to provide 3D medical imaging of patients<br>during orthopedic, neurological, intraoperative surgical procedures and<br>where the clinician benefits from 3D visualization of complex anatomical<br>structures, such as but not limited to those of high contrast objects, bones,<br>joints, maxillofacial, cervical, thoracic, and lumbar regions of the spine,<br>pelvis, acetabulum and joint fractures of the upper and lower extremities,<br>and where digital image and C-arm positioning data is required for<br>Computer-Assisted Surgery procedures.<br>The visualization of such anatomical structures assists the clinician in the<br>clinical outcome. At the discretion of a physician, the device may be used<br>for other imaging applications. This device does not support direct<br>radiographic film exposures and is not intended for use in performing<br>mammography. The system is not intended for use near MRI systems. |

EQUIVALENT - nView's indications are a subset of the predicate, fully encompassed by the predicate indications. nView system 1 intended use is limited to high contrast imaging only and applicable to a more restricted patient population.

Table 2 – Comparison of attributes for the subject device and predicate device

{6}------------------------------------------------

## S001 to K190064 - Attachment 1 - Updated 510(k) summary

| ATTRIBUTE                          | SUBJECT DEVICE<br>nView system 1                                                                          | PREDICATE<br>DEVICE<br>Ziehm Vision RFD<br>3D (K142740)                                              | SUBSTANTIAL EQUIVALENCE<br>DISCUSSION                                                                                                                                                                                                                                                               |
|------------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| target population                  | adult and pediatric<br>populations over 6<br>years of age with<br>imaged anatomy up<br>to 30 cm thickness | all adult and pediatric<br>populations                                                               | SUBSTANTIALLY<br>EQUIVALENT<br>nView's target population is a<br>subset, fully encompassed by<br>the predicate.                                                                                                                                                                                     |
| anatomical site                    | 2D and 3D imaging<br>of high contrast<br>bony anatomy                                                     | 2D imaging of high<br>and lower contrast<br>anatomy, 3D imaging<br>of high contrast bony<br>anatomy. | SUBSTANTIALLY<br>EQUIVALENT<br>nView's target population is a<br>subset, fully encompassed by<br>the predicate.                                                                                                                                                                                     |
| where used                         | Hospital/ clinic                                                                                          | Hospital/ clinic                                                                                     | IDENTICAL                                                                                                                                                                                                                                                                                           |
| Input Power<br>(VAC)               | 120                                                                                                       | 120                                                                                                  | IDENTICAL                                                                                                                                                                                                                                                                                           |
| X-Ray Tube Max<br>kV/ mA/ W        | 75/12/350                                                                                                 | 120/250/25k                                                                                          | SUBSTANTIALLY<br>EQUIVALENT<br>The subject device outputs<br>lower power X-ray to optimize<br>image quality of thin patients at<br>lower dose                                                                                                                                                       |
| Mobile Platform                    | Yes                                                                                                       | Yes                                                                                                  | IDENTICAL                                                                                                                                                                                                                                                                                           |
| C-arm Gantry                       | Yes                                                                                                       | Yes                                                                                                  | IDENTICAL                                                                                                                                                                                                                                                                                           |
| # of Axes                          | 6 axes of motion, 1<br>motorized                                                                          | 4 axes of motion, all<br>motorized                                                                   | SUBSTANTIALLY<br>EQUIVALENT<br>The subject device does not<br>move during acquisition so does<br>not need motorized axes.                                                                                                                                                                           |
| User Interface                     | Touch control                                                                                             | Touch control                                                                                        | IDENTICAL                                                                                                                                                                                                                                                                                           |
| Fluoroscopic                       | Yes - via digital<br>projections                                                                          | Yes - via physical<br>projections                                                                    | SUBSTANTIALLY<br>EQUIVALENT<br>Both systems provide 2D<br>projection visualization.                                                                                                                                                                                                                 |
| Temporal<br>resolution             | 0.5 fps                                                                                                   | 25/30 fps standard                                                                                   | SUBSTANTIALLY<br>EQUIVALENT<br>The predicate has a higher<br>temporal resolution to allow for<br>broader clinical applications<br>such as interventional<br>cardiology, hybrid procedures,<br>and interventional angiography.<br>For the clinical applications<br>indicated for the subject device, |
|                                    |                                                                                                           |                                                                                                      | S001 to K190064 - Attachment 1 - Updated<br>"fluoro shots" are used in both<br>the predicate and subject<br>device. A temporal resolution<br>of 0.5 fps is consistent with this<br>clinical use.                                                                                                    |
| Tomographic                        | Yes                                                                                                       | Yes                                                                                                  | IDENTICAL                                                                                                                                                                                                                                                                                           |
| Tube Type                          | Stationary Anode                                                                                          | Rotating Anode                                                                                       | SUBSTANTIALLY<br>EQUIVALENT<br>Both systems utilize similar<br>x-ray technology. The predicate<br>has a rotating anode to account<br>for higher power output.                                                                                                                                       |
| Tube Focal Spot<br>Size            | 0.6mm                                                                                                     | 0.6 mm                                                                                               | IDENTICAL                                                                                                                                                                                                                                                                                           |
| Detector Type                      | CMOS Digital<br>Detector                                                                                  | CMOS Digital<br>Detector                                                                             | IDENTICAL                                                                                                                                                                                                                                                                                           |
| Detector Shape                     | Square                                                                                                    | Square                                                                                               | IDENTICAL                                                                                                                                                                                                                                                                                           |
| Detector Size<br>(cm)              | 30 cm x 30 cm                                                                                             | 20 cm x 20 cm<br>30 cm x 30 cm                                                                       | IDENTICAL (Predicate has<br>multiple configurations)                                                                                                                                                                                                                                                |
| Detector<br>Resolution             | 1952 x 1952                                                                                               | 1024 x 1024<br>1536 x 1536                                                                           | SUBSTANTIALLY<br>EQUIVALENT<br>nView utilizes a detector with<br>higher resolution than the<br>predicate                                                                                                                                                                                            |
| Distortion Free<br>Imaging         | Yes                                                                                                       | Yes                                                                                                  | IDENTICAL                                                                                                                                                                                                                                                                                           |
| Collimator/Beam<br>Limiter         | Yes                                                                                                       | Yes                                                                                                  | IDENTICAL                                                                                                                                                                                                                                                                                           |
| Reconstruction<br>Geometry         | Multi Arc Source<br>Trajectory                                                                            | Multi Arc SmartScan<br>trajectory                                                                    | SUBSTANTIALLY<br>EQUIVALENT<br>Both systems utilize a multi<br>positional scan to acquire<br>images                                                                                                                                                                                                 |
| Dataset<br>Capabilities            | 2D and 3D                                                                                                 | 2D and 3D                                                                                            | IDENTICAL                                                                                                                                                                                                                                                                                           |
| Acquisition time<br>(s)            | 2 or 4                                                                                                    | 48                                                                                                   | SUBSTANTIALLY<br>EQUIVALENT<br>nView utilizes a shorter<br>acquisition time. This<br>characteristic reduces radiation<br>and motion artifacts.                                                                                                                                                      |
| Reconstruction<br>Time (s)         | 1 to 30                                                                                                   | 8 to 18                                                                                              | SUBSTANTIALLY<br>EQUIVALENT<br>Both systems provide a range of<br>reconstruction time based on<br>imaging parameters.                                                                                                                                                                               |
| 3D resolution                      | 512 x 512 x 512                                                                                           | 512 x 512 x 512                                                                                      | IDENTICAL                                                                                                                                                                                                                                                                                           |
| 3D<br>Reconstruction<br>Type (deg) | 117 multi arc                                                                                             | 180 via SmartScan<br>predicated on 165                                                               | SUBSTANTIALLY<br>EQUIVALENT<br>Both systems predicated on a<br>limited angle reconstruction                                                                                                                                                                                                         |
| 3D<br>reconstruction<br>algorithm  | Iterative                                                                                                 | Iterative                                                                                            | IDENTICAL                                                                                                                                                                                                                                                                                           |

{7}------------------------------------------------

### S001 to K190064 - Attachment 1 - Updated 510(k) summary

{8}------------------------------------------------

S001 to K190064 - Attachment 1 - Updated 510(k) summary

### Discussion of similarities and differences vs predicate

The subject device has an intended use that is a subset of the predicate device, Ziehm Vision RFD 3D (K142740). Both systems are intended to be used to generate fluoroscopic and tomographic images of human anatomy. The subject device's indications are a reduction of claims from the predicate (less clinical applications and smaller patient population) but are fully encompassed in the predicate indications. There are no new indications or claims for the subject device.

### Intended Use

The intended use of the subject device and the predicate device are the same within the guidance of "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] Guidance for Industry and Food and Drug Administration Staff. Document issued on: July 28, 2014". Both systems are intended for 2D and 3D imaging of adult and pediatric populations. The subject device limits use in pediatric populations to patients over 6 years of age and imaged human anatomy up to 30 cm thickness. A reduction in patient population does not result in a new intended use. Both systems are indicated for orthopedic surgical procedures where the clinician benefits from 3D visualization of complex anatomical structures, such as those of high contrast objects, bones, joints, cervical, thoracic, and lumbar regions of the spine, and joint fractures of the upper and lower extremities. The predicate includes additional claims in the clinical procedures of cardiology, angiography, electrophysiology, urology, gastroenterology, etc. A reduction in the diseases a device is intended to treat do not result in a new intended use as they don't affect the safety and/or effectiveness of the new device.

### Technological Characteristics

The technological characteristics governing how the device produces images of human anatomy are the equivalent in the subject device and the cleared Ziehm Vision RFD 3D (K142740) system. Both systems use X-ray to image the anatomy. Both systems are used for fluoroscopic and tomographic imaging.

The predicate displays 2D fluoroscopic views via real-time physical projection images. The subject device displays 2D fluoroscopic views via a digital projection through the 3D tomographic volume. Both systems result in 2D fluoroscopic projection views displayed to the user.

The predicate reconstructs 3D images from a series of X-ray images collected while the gantry is 001 S001 to K190064 nViewSystem1 Page 30

{9}------------------------------------------------

S001 to K190064 - Attachment 1 - Updated 510(k) summary moving around the patient. The subject device reconstructs tomographic images from a series of X-ray images collected while the gantry stays in a fixed position with respect to the patient, but while the X-ray tube rotates within the system. Therefore, motorized axes are not required as the gantry is stationary during acquisition. The predicate device reconstructs tomographic images from a series of X-ray images collected while the gantry rotates and shifts around the patient. Therefore, motorized axes were implemented to automatically move the gantry while acquisitions are occurring.

The Ziehm Vision RFD 3D (K142740) is based on a predicate limited angle device employing 135 degrees (+90 degrees and -45 degrees) of orbital rotation. The Ziehm Vision RFD 3D (K142740) employs a shift-scan-shift (SmartScan) motion of this predicate to increase relative coverage to 180 degrees. 3D images are generated from this shift-scan-shift dataset. The nView system 1 is a limited angle device utilizing a 117-degree multi-arc scan for 3D reconstruction. The images are acquired in two steps, first the gantry is rotated to +30 degrees, and images are collected as the tube rotates 360 degrees. Next, the gantry is rotated to -30 degrees, and images are collected as the tube rotates 360 degrees.

Both the subject device and the predicate device have a fluoroscopic mode. The predicate has a higher temporal resolution to allow for broader clinical applications such as interventional cardiology, hybrid procedures, and interventional angiography. For the clinical applications indicated for the subject device, "fluoro shots" are used in both the predicate and subject device. A temporal resolution of 0.5 fps is consistent with this clinical use.

Both the subject device and the predicate use a 30 cm CMOS detector. The subject device CMOS detector has a higher resolution but is used in a binned mode to be equivalent to the resolution of the predicate device detector. The subject device has a lower power x-ray tube with a fixed anode. A higher power x-ray tube with rotating anode is not required for the subject device with limitations on imaged anatomy thickness. The subject device is designed to operate at a lower radiation dose than the predicate.

The acquisition and reconstruction times for the subject device are faster than the predicate device. This allows the subject device to perform real-time fluoroscopic imaging via digital projections and faster 3D tomographic imaging and does not impact safety or efficacy of the device.

# Summary of non-clinical test data

The demonstration of substantial equivalence is based on a comparison of features to the predicate device and on an assessment of non-clinical performance data.

The nView system 1 complies with the mandatory and voluntary standards listed in Table 3 below. The nView system 1 was developed in accordance with the FDA guidance documents listed in Table 4.

| Standards development<br>organization, reference<br>number, and date | Standard name                                                                                                                     |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| 21 CFR 1020.30, 32                                                   | Federal Performance Standard for Diagnostic X-ray Systems                                                                         |
| ES60601-1:2005/(R)2012<br>and A1:2012                                | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance (IEC 60601-1:2005, MOD) |

|  | Table 3 - Standards used in the development of nView system 1 |  |
|--|---------------------------------------------------------------|--|
|  |                                                               |  |

{10}------------------------------------------------

S001 to K190064 - Attachment 1 -- Updated 510(k),summary

| IEC 60601-1-2: 2014                | Medical Electrical Equipment, General requirements for basic safety<br>and essential performance - Collateral Standard: Electromagnetic<br>disturbances - Requirements and tests |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-3:2008 +<br>AMD1:2013  | Medical Electrical Equipment, General requirements for basic safety<br>and essential performance - Collateral Standard: Radiation<br>protection in diagnostic X-ray equipment    |
| IEC 60601-1-6:2010 +<br>AMD1:2013  | Medical electrical equipment, General requirements for basic safety<br>and essential performance - Collateral standard: Usability                                                |
| IEC 60601-2-43:2010                | Medical electrical equipment, Particular requirements for the basic<br>safety and essential performance of X-ray equipment for<br>interventional procedures                      |
| IEC 60601-2-54:2009 +<br>AMD1:2015 | Medical electrical equipment, Particular requirements for the basic<br>safety and essential performance of X-ray equipment for<br>radiography and radioscopy                     |
| ISO 14971: 2007                    | Application of risk management to medical devices                                                                                                                                |

|  | Table 4 - Guidance documents used in the development of nView system 1 |  |
|--|------------------------------------------------------------------------|--|
|--|------------------------------------------------------------------------|--|

| Guidance Document Name                                                                                                                                         | Issue Date        |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Guidance for Industry and Food and Drug Administration Staff: The 510(k)<br>Program: Evaluating Substantial Equivalence in Premarket Notifications<br>[510(k)] | July 28, 2014     |
| Guidance for Industry and FDA Staff: Guidance for the Submission Of<br>510(k)'s for Solid State X-ray Imaging (SSXI) Devices                                   | September 1, 2016 |
| Guidance for Industry and FDA Staff: Guidance for the Content of Premarket<br>Submission for Software in Medical Devices                                       | May 11, 2005      |
| Guidance for Industry and FDA Staff: Applying Human Factors and Usability<br>Engineering to Medical Devices                                                    | February 3, 2016  |
| Guidance for Industry and FDA Staff: Pediatric Information for X-ray Imaging<br>Device Premarket Notifications                                                 | November 28, 2017 |
| Guidance for Industry and FDA Staff: Content of Premarket Submissions for<br>Management of Cybersecurity in Medical devices                                    | October 18, 2018  |
| Guidance for Industry and FDA Staff: Information to Support a Claim of<br>Electromagnetic Compatibility (EMC) of Electrically-Powered Medical<br>Devices       | July 11, 2016     |
| Guidance for Industry and Food and Drug Administration Staff: Medical<br>X-Ray Imaging Devices Conformance with IEC Standards                                  | May 8, 2019       |
| Guidance for Industry and Food and Drug Administration Staff: Policy<br>Clarification for Certain Fluoroscopic Equipment Requirements                          | May 8, 2019       |

### Verification and Validation

Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
 Page 32

{11}------------------------------------------------

S001 to K190064 - Attachment 1 - Updated 510(k) summary issued on May 11, 2005 is also included as part of this submission. Non-clinical tests were conducted on the subject device during product development.

The risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results support that all requirements have met the acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence.

The subject device was tested and found to be safe and effective for intended users, uses and use environments through the design control verification and validation process. The human factor usability validation showed that human factors are addressed in the system in simulated clinical use tests.

The subject device conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or unauthorized use of information that is stored, accessed or transferred from a medical device to an external recipient.

Additional engineering bench testing was performed including the clinical and non-clinical testing identified in the guidance for submission of 510(k) s for Solid State X-Ray Imaging Devices (SSXI); demonstration of system performance; and an imaging performance evaluation.

# Conclusion as to substantial equivalence

In summary, the subject device has an intended use that is a subset of the predicate intended use and essentially the same technological characteristics as the predicate Ziehm Vision RFD 3D (K142740). Minor differences that do not impact the decision of substantial equivalency include: the subject device having lower temporal fluoroscopic views generated from tomographic reconstructions, a different acquisition geometry, a lower powered x-ray tube, and faster acquisition and reconstruction times. It has been demonstrated that the subject device has clinically acceptable performance per a qualified expert evaluation. The subject device does not have new functionalities when compared to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K190064](https://fda-staging.innolitics.com/device/K190064)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
