The Radio Frequecy Ablation System is intended for coagulation and ablation of soft tissue. It is not intended for use in cardiac procedures.
Device Story
Radio Frequency Ablation System delivers 470 kHz RF energy to induce coagulation necrosis in soft tissue. System components include RF generator with integrated cooling pump, RF electrode kits, temperature probe, and foot switch. Generator monitors tissue impedance and temperature; peristaltic pump provides internal electrode cooling. Operates in three modes: Auto (automatic power adjustment for 1-3 electrodes), Power (user-set maximum power, single electrode), and Temp (temperature-regulated power, single electrode). Used by clinicians in surgical settings to ablate lesions. Output is thermal energy generated by ion collision friction. System includes return pad to complete circuit. Benefits include controlled tissue destruction for tumor management.
Clinical Evidence
No clinical data provided. Substantial equivalence demonstrated through bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), performance testing (IEC 60601-2-2), thermal effects, temperature monitoring, leakage current, shelf life, packaging, sterilization (ISO 11135), and biocompatibility (ISO 10993).
Technological Characteristics
Monopolar RF ablation system; 470 kHz output; 200W max power. Components: RF generator, peristaltic pump, stainless steel 304 electrodes, hydrogel return pads. Connectivity: Foot switch. Sterilization: EO (SAL 10^-6). Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, ISO 11607-1, ISO 11135, ISO 10993.
Indications for Use
Indicated for coagulation and ablation of soft tissue. Not for use in cardiac procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
June 25, 2019
Surgnova Healthcare Technologies (Zhejiang) Co., Ltd. Grace Guofang Ma Quality Director No.1 XinXing Yilu Road, Emerging Industrial Cluster Area Zonghan Subdistrict, Cixi City Zhejiang, China 315300
Re: K190052
Trade/Device Name: Radio Frequency Ablation System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: December 31, 2018 Received: January 11, 2019
Dear Grace Guofang Ma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K190052
Device Name Radio Frequency Ablation System
Indications for Use (Describe)
The Radio Frequecy Ablation System is intended for coagulation and ablation of soft tissue. It is not intended for use in cardiac procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section III 510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.
There is no prior submission for the device.
### 3.1 Submitter Information
#### ● 510(k) Submitter/Holder:
Surgnova Healthcare Technologies (Zhejiang) Co., Ltd. No.1 Xinxing Yilu Road, Emerging Industrial Cluster Area, Zonghan Subdistrict, Cixi City, Zhejiang, China
#### ● Contact
Grace Ma Quality Director Telephone: +86-10-64116115 Fax: +86-10-64113261 Email: QARA@surgnova.com
● Date Prepared: November 12, 2018
### 3.2 Proposed Device Information
Device Common Name: Electrosurgical cutting and coagulation device and accessories Device Trade/Proprietary Name: Radio Frequency Ablation System Model: R150E Classification Name: Electrosurgical cutting and coagulation device and accessories Regulation Number: 878.4400 Product Code: GEI Class: II Panel: General & Plastic Surgery
### 3.3 Predicate Device
510(k) Number: K052796 Device Trade/Proprietary Name: Cool-tip™ RF Ablation System Common/Classification Name: Electrosurgical cutting and coagulation device and accessories Regulation Number: 878.4400 Product Code: GEI Class: II Panel: General & Plastic Surgery Manufacturer: VALLEYLAB
{4}------------------------------------------------
### 3.4 Device Description
The proposed device consists of Radio Frequency Generator, Radio Frequency Electrode Kits, Temperature Probe and Foot Switch. Wherein, the radiofrequency generator integrated with cooling pump is capable of delivering up to 200W while monitoring tissue impedance, tissue temperature and electrode tip temperature during the delivery of the RF energy. The Radio Frequency Electrode Kits consists of Radio Frequency Electrode, introducer, inflow - outflow tubing sets and Return Pad to be used for puncturing the patient's lesion position during operation and also for outputting Radio Frequency energy to ablate the tumor tissue through connection with Radio Frequency Generator. The Temperature Probe is applied to monitor the temperature of the target location and to protect important organs and tissues in the periphery of the lesion from unexpected damage by Radio Frequency energy.
The Radio Frequency generator supplies 470 kHz of radio frequency current, which enters into target tissue of patient, and builds a circuit loop with the Return Pad. The alternating current cause alternating motion of the positive ions in the tumor region. Because of the different size, mass and charge of each ion, the moving speeds of the ions are different, resulting in collision friction and thermal energy. When the temperature is above 50 °C, tumor tissue come into irreversible coagulation necrosis.
lt supports three working modes: Auto Mode, POWER Mode and TEMP Mode.
#### Auto Mode
The Auto Mode allows the use of 1-3 RF ablation electrodes for ablation. When multiple RF Ablation electrodes are connected, the RF generator will switch between the electrodes for energy output. In Auto Mode, the system automatically adjusts the power output according to the built-in algorithm.
#### Power mode
In POWER Mode, the RF generator outputs according to the maximum power set by users. In this mode, only single RF electrode is allowed to be used. The RF Electrode is kept below 40°C by means of internal cooling with cooling-water delivers from a peristaltic pump.
#### Temperature mode
In TEMP Mode, the RF generator detects the temperature of the RF electrode automatically, and adjusts the power output according to the detected temperature. In this mode, "SET POWER(W)" is forbade to set. Single RF electrode is permitted to use. Once the temperature of RF electrode reaches 85℃, a beep sounds. Then users can pull the RF electrode out of patient at a rate of 1centimetre every 2 seconds.
The Radio Frequency Generator is non-sterile, the return pad is non-sterile and for single use, while Radio Frequency Electrode and Temperature Probe are sterile and for single
{5}------------------------------------------------
use.
The device is software-driven and the software validation is provided in Section of Software.
{6}------------------------------------------------
# 3.5 Comparison list of the technological characteristics
| Comparison Elements | Predicate Device<br>(K052796) | Proposed Device |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | Cool-tip™ RF Ablation System | Radio Frequency Ablation System |
| Regulation No. | 21 CFR 878.4400 | 21 CFR 878.4400 |
| Classification | II | II |
| Product Code | GEI | GEI |
| Indications for Use | The Valleylab Cool-tip RF System (generator and<br>accessories) is intended for the use in percutaneous,<br>laparoscopic,<br>intraoperative coagulation and<br>ablation of tissue, such as partial or complete<br>ablation of non-resectable liver lesions and osteoid<br>osteoma tumors within bone. | The Radiofrequency Ablation System is<br>intended for coagulation and ablation of soft<br>tissue. It is not intended for use in cardiac<br>procedures. |
| Component | RF Generator, peristaltic pump, electrodes, inflow<br>and outflow tubing for electrode cooling, and return<br>pads | Radio Frequency Generator, Radio Frequency<br>Electrode Kits, Temperature Probe and Foot<br>Switch |
| Electrosurgical unit | | |
| Output Parameters | $480KHZ±2%$ | $470KHZ±2%$ |
| Drive on time | Up to 30minutes | Up to 30minutes |
| Maximum power output | Up to 200 watts @50 ohm | Up to 200 watts @50 ohm |
| Voltage Supply | 100-240VAC 50-60 Hz | 100-240VAC 50-60 Hz |
| Active accessary | | |
| Monopolar/Bipolar | Monopolar | Monopolar |
| Antenna Length (mm) | 100, 150, 200, 250 | 70, 150, 200, 250 |
| Exposure Length (mm) | 7, 10, 20, 30 | 5, 7, 10, 15, 20, 30, 40 |
| Outer Diameter (mm) | 1.47 | 1.2, 1.47 |
| Material | Stainless steel 304 | Stainless steel 304 |
| Disposable<br>/Single-use<br>Device | The antennas are disposable and are to be used<br>within a single patient procedure only. | The antennas are disposable and are to be<br>used within a single patient procedure only. |
| Sterility | The accessories are sterilized with EO(SAL: 10-6) | The accessories are sterilized with EO(SAL:<br>10-6) |
| Biocompatibility | Patient-contacting materials are biocompatible. | Patient-contacting<br>materials<br>are<br>biocompatible. |
| Device<br>Temperature<br>Monitoring | Temperature monitoring features used to ensure<br>system safety | Temperature monitoring features used to<br>ensure system safety |
| Neutral electrodes | | |
| Conductive or Capacitive | Conductive | Conductive |
| Patient Contacted Material | Hydrogel | Hydrogel |
{7}------------------------------------------------
Premarket Notification 510(k) Submission—K190052
{8}------------------------------------------------
### 3.6 Indications for use
The Radio Frequency Ablation System is intended for coagulation and ablation of soft tissue. It is not intended for use in cardiac procedures.
## 3.7 Testing
### Non-Clinical Testing
The Radio Frequency Ablation System and Accessories and the predicate device are substantially equivalent in design concepts, technologies and materials. The Radiofrequency Ablation System and Accessories has been designed and tested and will be manufactured in accordance with both mandatory and voluntary standards, including:
• IEC 60601-1: 2005/A1:2012 Medical Electrical Equipment-Part 1: General requirements for safety.
• IEC 60601-2-2:2009 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.
• IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
•ISO 11607-1:2016 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems.
• ISO 11135: 2014 Sterilization of health care products -- Ethylene oxide -- Requirements for the development, validation and routine control of a sterilization process for medical devices
The Software Validation is in compliance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
The compatibility of the skin-contact component material in the finished product meets the requirement of Biocompatibility. The Biological Evaluation Tests are in compliance with the standards of ISO 10993, "Biological Evaluation of Medical Devices".
The shelf life of Radiofrequency Ablation Electrode Kits and Temperature Probe is 2 years.
The list of non-clinical test performed on the proposed device.
| No. | Test Name |
|-----|-------------------------------------------------|
| 1 | Electrical Safety Test According to IEC 60601-1 |
{9}------------------------------------------------
Premarket Notification 510(k) Submission—K190052
| 2 | Electromagnetic Compatibility Test According to IEC 60601-1-2 |
|----|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| 3 | Performance Test according to IEC 60601-2-2 |
| 4 | System Performance Test |
| 5 | Thermal Effects test according to FDA Guidance Premarket Notification (510(K)) Submissions for Electrosurgical Devices for General Surgery |
| 6 | Temperature Monitoring test according to FDA Guidance Premarket Notification (510(K)) Submissions for Electrosurgical Devices for General Surgery |
| 7 | High Frequency Leakage Current Test according to FDA Guidance Premarket Notification (510(K)) Submissions for Electrosurgical Devices for General Surgery |
| 8 | Shelf Life Test |
| 9 | Package Verification Test according to ISO 11607-1 |
| 10 | Sterilization validation according to ISO 11135 |
| 11 | Irritation, Sensitization, Cytotoxicity, Pyrogenicity, Acute Systemic Toxicity Test according to ISO 10993 |
### Clinical Testing
Clinical studies were not required to demonstrate the substantial equivalence of the microwave ablation system and the predicated device.
### 3.8 Determination of substantial equivalence
The proposed device is equivalent with respect to the basic system design and function to that of the predicate device. The proposed device isn't the implants and high-risk device. And it doesn't have new intended purposes, new medical, new target populations, and new users and so on. What's more, it can't use the medicinal substances or animal tissues. So differences between the predicate and proposed device do not raise new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.