K190051 · Zephyr Sleep Technologies, Inc. · LRK · Jun 7, 2019 · Dental
Device Facts
Record ID
K190051
Device Name
TD Clip
Applicant
Zephyr Sleep Technologies, Inc.
Product Code
LRK · Dental
Decision Date
Jun 7, 2019
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TD Clip is intended to be used with MATRx or MATRx plus Titration Trays to form a mandibular repositioning device intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in patients 18 years and older. The device is intended to be fitted with assistance from a healthcare professional and used during sleep for a total of less than 30 nights.
Device Story
TD Clip is a temporary mandibular repositioning device used to alleviate snoring and mild to moderate OSA while awaiting a custom oral appliance. The device consists of upper and lower dental titration trays custom-fitted with polyvinyl siloxane impression material; trays are fastened at a target therapeutic position using a clip assembly and pins. Used during sleep for a maximum of 30 nights. The workflow follows an OA Assessment study using the MATRx or MATRx plus system; the healthcare provider fastens the trays at the target protrusive position identified during the study. The patient inserts the device each evening. The device maintains mandibular protrusion to keep the airway open, potentially reducing apnea events and snoring. It is a prescription device fitted by a healthcare professional.
Clinical Evidence
No clinical data. Bench testing only. Testing included simulated aging, environmental conditions, and worst-case loading schemes to ensure structural integrity over 30 nights. Tray removal simulation confirmed 100% of users could remove the device safely. Biocompatibility testing performed per ISO 10993-1.
Technological Characteristics
Mandibular repositioning device; separate upper/lower trays; polyvinyl siloxane impression material; pin-based fastening system with serrated interface. Non-sterile. Biocompatibility per ISO 10993-1. Mechanical design allows for lateral/vertical jaw movement. Adjustable arch width.
Indications for Use
Indicated for adult patients (18+) with mild to moderate obstructive sleep apnea (OSA) and snoring. Contraindicated for patients with central sleep apnea, loose teeth, advanced periodontal disease, full dentures, dental implants, TMJ dysfunction syndrome, or severe respiratory disorders.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 7, 2019
Zephyr Sleep Technologies, Inc. Sabina Bruehlmann Director, Technology 102, 701 64 Ave SE Calgary, T2H 2C3 Ca
Re: K190051
Trade/Device Name: TD Clip Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: April 30, 2019 Received: May 1, 2019
Dear Sabina Bruehlmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
542 of the Act); 21 CFR 1000-1050.
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Srinivas Nandkumar, Ph.D. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190051
Device Name TD Clip
Indications for Use (Describe)
The TD Clip is intended to be used with MATRx or MATRx plus Titration Trays to form a mandibular repositioning device intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in patients 18 years and older. The device is intended to be fitted with assistance from a healthcare professional and used during sleep for a total of less than 30 nights.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 20px;"><b> </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image features the logo for Zephyr Sleep Technologies. The logo is in blue and includes the company name in a stylized font. Below the name is the company's address: #102, 701-64th Avenue SE, Calgary, AB, T2H 2C3, Canada.
K190051
## Section 5 - 510k Summary - TD Clip
| Date: | January 9, 2019 |
|------------------------------------|-----------------------------------------------------------------------------|
| Manufacturer Name: | Zephyr Sleep Technologies, Inc. |
| Contact Name: | Sabina Bruehlmann, PhD |
| Title: | Director, Technology |
| Postal Address: | #102, 701 64th Ave SE<br>Calgary, Alberta, Canada<br>T2H-2C3 |
| Phone Number: | 587-332-0285 |
| Fax Number: | 587-332-0208 |
| Establishment Registration Number: | 3008960597 |
| Device Proprietary Name: | TD Clip |
| Classification Name: | CFR 872.5570 |
| Classification Code: | Class II |
| Product Code: | LRK (anti-snoring device) |
| Predicate Device: | Apnea Guard (K111110) |
| Reference Devices: | MATRx plus (DEN170090)<br>MATRx (K103704)<br>Brux-TMD QuickSplint (K111066) |
### Device Description:
The TD Clip is a temporary oral appliance. The patient uses the temporary oral appliance to alleviate mild to moderate obstructive sleep apnea and/or snoring while waiting for a custom oral appliance to be manufactured.
The dentist fabricates temporary titration trays consisting of upper and lower dental trays. The trays are custom fit to the patient using polyvinyl siloxane impression material.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Zephyr Sleep Technologies. The logo features the word "ZEPHYR" in a large, sans-serif font, with the words "Sleep Technologies" in a smaller font underneath. To the right of the text is a stylized image of a crescent moon and a star. Below the logo is the address "#102. 701 - 64th Avenue SE, Calgary, AB, T2H 2C3, Canada".
The healthcare provider may use the clip assembly to fasten the titration trays at the target therapeutic position to form the TD Clip temporary oral appliance. Patients may use the temporary oral appliance as treatment while waiting for their custom oral appliance. Adjustment of the position requires the setting of a new TD Clip by the healthcare professional.
In the expected workflow, the patient first completes an OA Assessment study with the MATRx or MATRx plus system. The healthcare provider then takes the titration trays used in the overnight OA Assessment study and fastens them at the intended target protrusive position for temporary treatment.
### Indications for Use:
The TD Clip is a mandibular repositioning device intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in patients 18 years and older. The device is intended to be fitted with assistance from a healthcare professional and used during sleep for less than 30 nights.
### Patient Population:
The TD Clip is intended to be used on adult patients upon referral from their healthcare provider.
# Contraindications:
The device is contraindicated for patients who:
- . have central sleep apnea
- . have loose teeth or advanced periodontal disease
- have full dentures or dental implants
- have temporomandibular joint (TMJ) dysfunction syndrome
- have severe respiratory disorders
- are under 18 years of age ●
#### Comparison to Predicate Device:
The TD Clip is substantially equivalent to the predicate, Apnea Guard. The following table provides a comparison of the intended use and technological features of the subject. predicate, and reference devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Zephyr Sleep Technologies. The logo features the word "ZEPHYR" in a large, bold, blue font, with the words "Sleep Technologies" in a smaller font underneath. To the right of the text is a stylized image of a crescent moon with a star.
| #102, 701 - 64th Avenue SE, Calgary, AB, T2H 2C3, Canada | | | | |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name<br>510(k) Number | Primary<br>Predicate | Reference Device | Subject Device | Substantial<br>Equivalence<br>Discussion |
| Manufacturer | | | | |
| | Apnea Guard<br>K111110 | MATRx plus<br>DEN170090 | TD Clip | |
| | Advanced Brain<br>Monitoring, Inc. | Zephyr Sleep<br>Technologies, Inc. | Zephyr Sleep<br>Technologies, Inc. | |
| Indications for Use | | | | |
| | The Apnea Guard<br>is a mandibular<br>repositioning<br>device intended to<br>reduce or alleviate<br>snoring and mild<br>to moderate<br>obstructive sleep<br>apnea (OSA) in<br>patients 18 years<br>and older. The<br>Apnea Guard is<br>intended to be<br>fitted with<br>assistance from a<br>healthcare<br>professional, and<br>used during sleep<br>for less than 30<br>nights. | The MATRx plus<br>device may also<br>be used with an<br>automated<br>mandibular<br>positioner that<br>uses feedback<br>control to record<br>changes in the<br>patient's<br>respiratory status<br>related to<br>repositioning of<br>the mandible<br>during an<br>overnight study.<br>MATRx plus uses<br>these recordings<br>to produce a<br>report for the HCP<br>that can be used to<br>prospectively<br>identify patients<br>with mild to<br>moderate<br>obstructive sleep<br>apnea who may be<br>suitable for<br>therapy with an<br>oral appliance and<br>to recommend a<br>target mandibular<br>position.<br>The use of the<br>device does not<br>replace the need<br>for follow-up<br>testing to<br>determine the<br>initial and<br>ongoing<br>effectiveness of<br>the therapy as<br>recommended by<br>clinical practice | The TD Clip is a<br>mandibular<br>repositioning<br>device intended to<br>reduce or alleviate<br>snoring and mild<br>to moderate<br>obstructive sleep<br>apnea (OSA) in<br>patients 18 years<br>and older. The<br>device is intended<br>to be fitted with<br>assistance from a<br>healthcare<br>professional and<br>used during sleep<br>for a total of less<br>than 30 nights. | The indications<br>for use of the<br>subject device are<br>identical to those<br>of the primary<br>predicate. |
| | #102, 701 - 64th Avenue SE, Calgary, AB, T2H 2C3, Canada | | | |
| | Primary<br>Predicate | Reference Device | Subject Device | |
| Proprietary Name | Apnea Guard | MATRx plus | TD Clip | Substantial |
| 510(k) Number | K111110 | DEN170090 | -- | Equivalence |
| Manufacturer | Advanced Brain<br>Monitoring, Inc. | Zephyr Sleep<br>Technologies,<br>Inc. | Zephyr Sleep<br>Technologies,<br>Inc. | Discussion |
| | | guidelines. | | |
| Use Environment | | | | |
| Intended for nighttime<br>use | | Yes | | Identical |
| Indicated for use at<br>home or in sleep<br>laboratories | | Yes | | |
| In-use claim | Less than 30 days | Duration of<br>Titration study<br>(max: 3 days) | Less than 30 days<br>(Including<br>titration study use) | Identical to<br>predicate - total<br>use < 30 days |
| Contraindications | | | | |
| | Missing, loose,<br>infected teeth,<br>temporary crowns<br>or fillings, TMJ<br>dysfunction<br>syndrome | Loose teeth, advanced periodontal<br>disease, full dentures/implants, TMJ<br>dysfunction syndrome, central sleep<br>apnea, severe respiratory disorders,<br>under 18 years of age | | Substantially<br>equivalent -<br>devices are<br>intended for use in<br>adults with mild<br>to moderate<br>obstructive sleep<br>apnea who have<br>sufficient<br>dentition and no<br>relevant<br>comorbidities |
| Technological Characteristics | | | | |
| Design | | | | |
| Operating Principle -<br>Mandibular<br>Advancement | | Yes | | |
| Separate upper and<br>lower tray pieces | | Yes | | |
| Placed in patient's<br>mouth each evening | | Yes | | |
| Mechanism of tray<br>fastening | | Same (Connected via brackets extending from the oral cavity<br>with directional guidance from rails) | | Identical |
| Permitted lateral and/or<br>vertical jaw movement | | Yes | | |
| Method of retention -<br>impression material | | Yes | | |
| Customized fit for each<br>patient | | Yes | | |
| | Primary<br>Predicate | Reference Device | Subject Device | Substantial<br>Equivalence<br>Discussion |
| Proprietary Name<br>510(k) Number | Apnea Guard<br>K111110 | MATRx plus<br>DEN170090 | TD Clip<br>-- | |
| Manufacturer | Advanced Brain<br>Monitoring, Inc. | Zephyr Sleep<br>Technologies,<br>Inc. | Zephyr Sleep<br>Technologies,<br>Inc. | |
| Method of fixing the<br>position | Pin connection<br>through one of<br>several slots | Pin connection to<br>a motorized<br>mandibular<br>positioner | Pin connection to<br>the TD clip<br>device; pin used<br>to fix in place | Substantially<br>equivalent - pin<br>used to secure<br>protrusive level |
| Dental tray shape | Same (Full arch, lingual and buccal walls to hold impression<br>material) | | | Identical |
| Dental tray sizing | One size fits all.<br>Adjustable arch<br>width. | Available in 2 sizes (medium and<br>large). Adjustable arch width. | | Substantially<br>equivalent –<br>multiple sizes in<br>subject device<br>accommodate full<br>range of arch<br>widths |
| Fastening resolution | 1.0 mm | 0.5 mm | | Substantially<br>equivalent –<br>ability to fasten<br>trays at a<br>clinically<br>meaningful<br>resolution |
| Can be adjusted or refit | Yes – with assistance from a healthcare professional | | | Identical |
| Materials | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.