Embrella Endoscopic Distal Attachment

K190030 · Micro-Tech (Nanjing) Co., Ltd. · FED · Mar 5, 2019 · Gastroenterology, Urology

Device Facts

Record IDK190030
Device NameEmbrella Endoscopic Distal Attachment
ApplicantMicro-Tech (Nanjing) Co., Ltd.
Product CodeFED · Gastroenterology, Urology
Decision DateMar 5, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

Embrella™ Endoscopic Distal Attachment is intended to the distal end of the endoscope to facilitate endoscopic therapy and maintain an appropriate endoscopic field of view.

Device Story

Embrella™ Endoscopic Distal Attachment is a single-use, umbrella-shaped medical-grade silicone rubber device; attaches to the distal end of an endoscope. Features six flexible wings that fold backwards and forwards to maintain an appropriate endoscopic field of view and facilitate endoscopic therapy. Available in two models (EM-S-01, EM-S-02) compatible with endoscope distal diameters of 11.5-13.0mm and 12.8-14.5mm. Used by clinicians in endoscopic settings. The device provides a mechanical aid to improve visualization during procedures; no electronic, software, or algorithmic components.

Clinical Evidence

No clinical study included. Evidence consists of bench testing (dimension, operating performance, matching performance, tension performance) and animal testing (endoscopy view testing). Device complies with ISO 10993-1 (biocompatibility), ISO 11135 (EO sterilization), and ISO 10993-7 (EO residuals).

Technological Characteristics

Material: Medical grade silicone rubber. Configuration: Short tube with 6 flexible wings. Dimensions: Compatible with endoscope distal diameters 11.5-13.0mm (EM-S-02) and 12.8-14.5mm (EM-S-01). Sterilization: Ethylene Oxide (EO). Single-use. No energy source, software, or connectivity.

Indications for Use

Indicated for attachment to the distal end of an endoscope to facilitate endoscopic therapy and maintain an appropriate endoscopic field of view in patients undergoing endoscopic procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 5. 2019 Micro-Tech (Nanjing) Co., Ltd. Becky Li Quality and Regulatory Affairs Director NO. 10 Gaoke Third Road Nanjing, Jiangsu 210032 China Re: K190030 > Trade/Device Name: Embrella™ Endoscopic Distal Attachment Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FED Dated: December 27, 2018 Received: January 7, 2019 Dear Becky Li: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Mark J. Antonino -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) #### K190030 Device Name Embrella™ Endoscopic Distal Attachment Type of Use (Select one or both, as applicable) Indications for Use (Describe) Embrella™ Endoscopic Distal Attachment is intended to the distal end of the endoscope to facilitate endoscopic therapy and maintain an appropriate endoscopic field of view. | <span style="font-size: 10pt;"> <span style="text-decoration: underline;"></span> <span style="text-decoration: line-through;"></span> <span style="font-family: Symbol;"> <span style="color: rgb(0, 0, 0);">☑</span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> <span style="text-decoration: underline;"></span> <span style="text-decoration: line-through;"></span> <span style="font-family: Symbol;"> <span>☐</span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ 510k summary Image /page/3/Picture/1 description: The image shows a blue and white logo. The logo is a stylized, three-dimensional representation of the letters 'M' and 'T'. The 'M' is in white and appears to be in front of the blue 'T', creating a layered effect. The logo has a modern and geometric design. # Tab 7 # 510K Summary This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: K190030 # 1. Date of Preparation: 2019-03-01 # 2. Sponsor Identification # Micro-Tech (Nanjing) Co., Ltd. No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone, Nanjing, Jiangsu Province, PRC # Establishment Registration Number: 3004837686 Contact Person: Becky Li Position: Quality and Regulatory Affairs Director Tel: +86-25-58646378 Fax: +86-25-58744269 # Email: Beckyli mt@outlook.com # 3. Identification of Proposed Device Trade Name: Embrella™ Endoscopic Distal Attachment Common Name: Endoscopic Distal Attachment # Regulatory Information Classification Name: Endoscope and accessories Classification: 2 Product Code: FED Regulation Number: 21 CFR 876.1500 Review Panel: Gastroenterology/Urology {4}------------------------------------------------ Image /page/4/Picture/1 description: The image contains the logo for MICRO-TECH (Nanjing) Co., Ltd. The logo consists of the letters MT in a stylized, three-dimensional font on the left. To the right of the letters is the company name in English and Chinese. The English name is MICRO-TECH (Nanjing) Co., Ltd., and the Chinese name is 南京微创医学科技股份有限公司. ### 4. Identification of Predicate Device 510(k) Number: K133359 Product Name: Endoring™ Manufacturer: EndoAid, Ltd. # 5. Indications for Use Embrella™ Endoscopic Distal Attachment is intended to be attached to the distal end of the endoscope to facilitate endoscopic therapy and maintain an appropriate endoscopic field of view. # 6. Device Description The proposed device Embrella™ Endoscopic Distal Attachment is an additional device, made of medical grade silicone rubber designed to attach to the distal end of the endoscope. Its shape is umbrella shaped, with six flexible wings that easily to fold backwards and forwards. It can be used to facilitate endoscopic therapy and maintain an appropriate endoscopic field of view. Embrella™ Endoscopic Distal Attachment has two models, one is EM-S-01, which is compatible with endoscopy distal diameter 12.8-14.5mm; the other is EM-S-02, which is compatible with endoscopy distal diameter 11.5-13.0mm. # 7. Comparison of Technological Characteristics The Embrella™ Endoscopic Distal Attachment incorporates substantially equivalent device materials, design, configuration, packaging fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate device Endoring™. # Comparison to predicate Devices: {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows a blue logo with the letters "M" and "T" in a stylized, three-dimensional design. The "M" is positioned in front of the "T", creating a sense of depth. The logo has a modern and corporate feel due to its clean lines and simple color scheme. | | | 510k summary | | | |-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|--| | Item | Proposed Device<br>Embrella™ Endoscopic Distal<br>Attachment | Predicate Device<br>Endoring™<br>(K133359) | Substantial<br>Equivalenc<br>e | | | Product Code | FED | FED | Same | | | Regulation No. | 21 CFR 876.1500 | 21 CFR 876.1500 | Same | | | Class | 2 | 2 | Same | | | Supplied in<br>Sterile | Yes | Yes | Same | | | Material | Silicone Rubber | Silicone Rubber | Same | | | Configuration | One short tube with 6 flexible<br>wings | One short tube with 6 flexible<br>wings | Same | | | Compatible<br>endoscopy<br>working<br>channel | 11.5-13.0mm<br>12.8-14.5mm | 11.5-13.0mm<br>12.8-14.5mm | Same | | | Indications for<br>Use | Embrella™ Endoscopic Distal<br>Attachment is intended to be<br>attached to the distal end of the<br>endoscope to facilitate endoscopic<br>therapy and maintain an<br>appropriate endoscopic field of<br>view. | The EndoRings is intended to be<br>attached to the distal end of the<br>endoscope to facilitate endoscopic<br>therapy, to be used for the<br>following:<br>*Keeping the suitable depth of<br>endoscope's view field | Same | | | Single Use | Yes | Yes | Same | | | Packaging | Single-use EO sterilized pouch<br>with one device per pouch | Single-use EO sterilized pouch<br>with one device per pouch | Same | | | Shelf Life | One years | Three years | Different | | # 8. Performance Data The proposed device the Embrella™ Endoscopic Distal Attachment meets the requirements of ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within A Risk Management Process", ISO 11135 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for MICRO-TECH (Nanjing) Co., Ltd. The logo consists of a blue graphic on the left and the company name in blue text on the right. The company name is written in both English and Chinese. The English name is "MICRO-TECH (Nanjing) Co., Ltd." and the Chinese name is "南京微创医学科技股份有限公司". 510k summary Development, Validation, and Routine Control of Sterilization processes for Medical Devices", and ISO 10993-7 "Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals", The following bench tests were performed on the Embrella™ Endoscopic Distal ## Attachment - > Dimension; - > Operating Performance - > Matching Performance; - > Tension Performance; The testing performed demonstrated that the proposed device and predicate device are equivalent. # 9. Animal Testing Endoscopy View Testing were performed on the Embrella™ Endoscopic Distal Attachment. The testing performed demonstrated that the proposed device and predicate device are equivalent, #### 10. Clinical Test Conclusion No clinical study is included in this submission. #### 11. Substantially Equivalent (SE) Conclusion Based on the indications for use, technological characteristics, and safety and performance testing, the Embrella™ Endoscopic Distal Attachment has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared predicate device EndoRing™ (K133359).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...