← Product Code [GCJ](/productcode/GCJ) · K190029

# Disposable Bladeless Trocar, Disposable Optical Trocar, Disposable Blunt-Tip Trocar, Disposable Spiral Trocar, Disposable Bladed Trocar (K190029)

_Changzhou Xin Neng Yuan Medical Stapler Co.,Ltd · GCJ · Sep 26, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K190029

## Device Facts

- **Applicant:** Changzhou Xin Neng Yuan Medical Stapler Co.,Ltd
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Sep 26, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The proposed devices, Disposable Bladeless Trocar, Disposable Optical Trocar and Disposable Bladed Trocar, have applications in abdominal, thoracic, and gynecologic minimally invasive surgical procedures to establish a path of entry for endoscopic instruments.

## Device Story

Disposable trocars (bladeless, optical, bladed) used in minimally invasive surgery; provide access channel for endoscopic instruments. Components include puncture needle/knife, puncture sleeve, injection valve, cap, and safety lock. Operated manually by surgeons in clinical settings. Puncture needle/knife introduced into sleeve to penetrate abdominal/thoracic wall; needle/knife removed post-penetration; sleeve remains as conduit for endoscopes/instruments. Benefits include established surgical access path. No electronic or software components.

## Clinical Evidence

No clinical study was done. Performance evaluated via bench testing (instrument insertion/removal force, leak resistance, snap feature retention) and in vivo porcine study (penetration force, fixation force, visualization performance, tip integrity). Biocompatibility testing per ISO 10993 (cytotoxicity, irritation, sensitization, systemic toxicity) and sterility/packaging testing (EO residuals, seal strength, leak detection, endotoxin limits) performed.

## Technological Characteristics

Materials: ABS, PC, PE, Stainless Steel, Rubber, Colorant. Principle: Manual mechanical penetration. Form factor: Puncture needle/sleeve assembly in various diameters (3-15mm) and lengths (75-100mm). Sterilization: Ethylene Oxide (EO) to SAL 10^-6. No software or electronic components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- ENDOPATH® XCEL® Bladeless Trocar with OPTIVIEWTM Technology ([K122511](/device/K122511.md))
- Autosuture Modified Versaport Trocar with Fixation Sleeve ([K062362](/device/K062362.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 26, 2019

Changzhou Xin Neng Yuan Medical Stapler Co., Ltd. % Ms. Diana Hong Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China

Re: K190029

Trade/Device Name: Disposable Bladeless Trocar, Disposable Optical Trocar, Disposable Blunt-Tip Trocar, Disposable Spiral Trocar, and Disposable Bladed Trocar Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: August 26, 2019 Received: August 28, 2019

Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190029

#### Device Name

Disposable Bladeless Trocar, Disposable Optical Trocar and Disposable Bladed Trocar

Indications for Use (Describe)

The devices, Disposable Bladeless Trocar, Disposable Optical Trocar and Disposable Bladed Trocar, have applications in abdominal, thoracic, and gynecologic minimally invasive surgical procedures to establish a path of entry for endoscopic instruments.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span style="text-decoration: underline;"><b></b></span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)                                                                      |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: K190029

- 1. Date of Preparation: 09/23/2019
- 2. Sponsor Identification Changzhou Xin Neng Yuan Medical Stapler Co., Ltd. No.51 Shuishan Road, Zhonglou Economic Development Zone, Changzhou, 213023, China. Establishment Registration Number: 3013534295

Contact Person: Boping Ma Position: General Manager Tel: +86-519-88830515 Fax: +86-519-86865852 Email: mbp129@126.com

- 3. Designated Submission Contact Persons Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person) Mr. Lee Fu (Alternative Contact Person)

Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net

- 4. Identification of Proposed Device Device Trade Name: Disposable Bladeless Trocar, Disposable Optical Trocar and Disposable Bladed Trocar Device Common Name: Disposable Trocar/Cannula
Regulatory Information: Classification Name: Laparoscope, General & Plastic Surgery; Classification: II; Product Code: GCJ; Regulation Number: 876.1500; Review Panel: General & Plastic Surgery;

{4}------------------------------------------------

510(k) Number: K190029

#### Intended Use Statement:

The proposed devices, Disposable Bladeless Trocar, Disposable Optical Trocar and Disposable Bladed Trocar, have applications in abdominal, thoracic, and gynecologic minimally invasive surgical procedures to establish a path of entry for endoscopic instruments.

## Device Description

The proposed devices, Disposable Bladeless Trocar, Disposable Optical Trocar and Disposable Bladed Trocar, are basic equipment used during laparoscopic surgical, which consist of Puncture Needle, Puncture Sleeve, Injection Valve, Cap and Safety lock. They are available in multiple configurations, including bladed type, bladeless type and optical type. In order to obtain access to the surgical site during laparoscopic surgery, the Puncture Needle is introduced into Puncture Sleeve to accomplish cannula penetration of the abdominal wall. The sleeve is connected to the Injection Valve at its proximal end and once the abdominal/thoracic wall is punctured, the puncture needle is removed. The sleeve acts as a channel for the introduction of the endoscopes and instruments. Generically, puncture needle and sleeve are available in a range of lengths and diameters to accommodate different sizes surgical instrument

- 5. Identification of Predicate Devices
Predicate Device 1 510(k) Number: K122511 Product Name: ENDOPATH® XCEL® Bladeless Trocar with OPTIVIEWTM Technology

Predicate Device 2 510(k) Number: K062362 Product Name: Autosuture Modified Versaport Trocar with Fixation Sleeve

#### 6. Non-Clinical Test Conclusion

Following non clinical tests were conducted. Performance Tests, both bench and in vivo study, were performed on both the proposed devices and predicate devices. The results demonstrated that the performances of the proposed device are substantial equivalent to those of the predicate device. Results of biocompatibility, sterility and packaging tests demonstrated that the proposed devices met their specifications and requirements.

#### Performance Tests - Bench

- Instrument Insertion and Removal Force Test; A
- A Leak Resistance Test;
- > Snap Feature Retention Force Test.

## Performance Tests - in vivo Study

An in vivo study was conducted on porcine model to evaluate the penetration force, fixation force and visualization performance (for optical type only), in addition, tip integrity was also evaluated after each insertion.

#### Biocompatibility

- Cytotoxicity per ISO 10993-5:2009 Biological Evaluation of Medical Devices-Part 5: Tests for In A Vitro Cytotoxicity;
- > Irritation and Sensitization per ISO 10993-10:2010 Biological Evaluation of Medical Devices-Part

{5}------------------------------------------------

10: Tests for Irritation and Skin Sensitization;

- A Pyrogen Study per ISO 10993-11:2006 Biological Evaluation of Medical Devices-Part 11: Tests for Systemic Toxicity;
Sterility and Packaging

- > ISO 10993-7:2008 Biological Evaluation Of Medical Devices-Part 7: Ethylene Oxide Sterilization Residuals;
- > ASTM F88/F88M-15 Standard test method for seal strength of flexible barrier materials;
- > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration;
- > USP <85> Endotoxin Limit Test
- 7. Clinical Testing No clinical study was done.
- 8. Substantially Equivalent (SE) Comparison

| Table 1 Comparison for Disposable Bladeless Trocar and Disposable Optical Trocar with predicate device |
|--------------------------------------------------------------------------------------------------------|
| (k122511)                                                                                              |

| ITEM            | Proposed Device<br>Disposable Bladeless Trocar                                                                                                                                     | Proposed Device<br>Disposable Optical Trocar                                                                                                                                       | Predicate Device 1<br>K122511                                                                                                                                                                                                                                                                                                |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification  | II                                                                                                                                                                                 | II                                                                                                                                                                                 | II                                                                                                                                                                                                                                                                                                                           |
| Regulation No.  | 876.1500                                                                                                                                                                           | 876.1500                                                                                                                                                                           | 876.1500                                                                                                                                                                                                                                                                                                                     |
| Product Code    | GCJ                                                                                                                                                                                | GCJ                                                                                                                                                                                | GCJ                                                                                                                                                                                                                                                                                                                          |
| Intended Use    | The device has applications in<br>abdominal, thoracic, and<br>gynecologic minimally<br>invasive surgical procedures to<br>establish a path of entry for<br>endoscopic instruments. | The device has applications in<br>abdominal, thoracic, and<br>gynecologic minimally<br>invasive surgical procedures<br>to establish a path of entry for<br>endoscopic instruments. | The ENDOPATH®<br>XCEL® Bladeless Trocar<br>has applications in<br>abdominal, thoracic, and<br>gynecologic minimally<br>invasive surgical<br>procedures to establish a<br>path of entry for<br>endoscopic instruments.<br>The trocar may be used<br>with or without<br>visualization for primary<br>and secondary insertions. |
| Configuration   | Puncture Needle<br>Puncture Sleeve<br>Reducer Cap                                                                                                                                  | Puncture Needle<br>Puncture Sleeve<br>Reducer Cap                                                                                                                                  | Puncture Needle<br>Puncture Sleeve<br>Reducer Cap                                                                                                                                                                                                                                                                            |
| Single Use      | Single Use                                                                                                                                                                         | Single Use                                                                                                                                                                         | Single Use                                                                                                                                                                                                                                                                                                                   |
| Operation Mode  | Manually                                                                                                                                                                           | Manually                                                                                                                                                                           | Manually                                                                                                                                                                                                                                                                                                                     |
| Accessory       | Puncture sleeve, reducer cap                                                                                                                                                       | Puncture sleeve, reducer cap                                                                                                                                                       | No accessory                                                                                                                                                                                                                                                                                                                 |
| Safety features | No safety feature                                                                                                                                                                  | No safety feature                                                                                                                                                                  | No safety feature                                                                                                                                                                                                                                                                                                            |

{6}------------------------------------------------

| Label/Labeling   | Comply with 21, CFR Section 801        | Comply with 21, CFR Section 801                | Comply with 21, CFR Section 801 |
|------------------|----------------------------------------|------------------------------------------------|---------------------------------|
| Shaft Diameter   | Available in 3, 5, 8, 10, 12, and 15mm | Available in 5, 8, 10 and 12mm                 | Available in 5 and 12mm         |
| Shaft Length     | Available in 75 and 100mm              | Available in 75 and 100mm                      | Available in 75, 100 and 150mm  |
| Material         | ABS, PC, PE, Colorant, Rubber          | ABS, Stainless Steel, PC, PE, Rubber, Colorant | Unknown                         |
| Trocar Width     | 52.4mm, 61.4m                          | 52.4mm, 61.4mm                                 | Unknown                         |
| Total Length     | 158mm, 184mm, 172mm, 197mm             | 158mm, 172mm, 184mm, 197mm                     | Unknown                         |
| Needle Tip Shape | Fin                                    | Fin                                            | Fin                             |
| Sterilization    | EO sterilized                          | EO sterilized                                  | Irradiation sterilized          |
| SAL              | 10-6                                   | 10-6                                           | 10-6                            |
| Endotoxin Limit  | 20 EU per device                       | 20 EU per device                               | 20 EU per device                |
| Shelf life       | 2 years                                | 2 years                                        | 3 years                         |
| Packaging method | Sealing method                         | Sealing method                                 | Sealing method                  |
| Cytotoxicity     | No cytotoxicity                        | No cytotoxicity                                |                                 |
| Skin Irritation  | No irritation                          | No irritation                                  | Comply with ISO 10993           |
| Sensitization    | No sensitization                       | No sensitization                               |                                 |
| Pyrogen          | No pyrogen                             | No pyrogen                                     |                                 |

| ITEM            | Proposed Device                                                                                                                                                              | Predicate Device 2                                                                                                                                                                                                         |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                 | Disposable Bladed Trocar                                                                                                                                                     | K062326                                                                                                                                                                                                                    |
| Classification  | II                                                                                                                                                                           | II                                                                                                                                                                                                                         |
| Regulation No.  | 876.1500                                                                                                                                                                     | 876.1500                                                                                                                                                                                                                   |
| Product Code    | GCJ                                                                                                                                                                          | GCJ                                                                                                                                                                                                                        |
| Intended Use    | The device has applications in abdominal,<br>thoracic, and gynecologic minimally<br>invasive surgical procedures to establish a<br>path of entry for endoscopic instruments. | The autosutureTM Modified VERSAPORTTM<br>trocar with fixation sleeve is intended for use<br>in a variety of gynecologic, general, thoracic<br>and urologic endoscopic procedures to create<br>and maintain a port of entry |
| Configuration   | Puncture Knife<br>Puncture Sleeve<br>Reducer Cap<br>Puncture Needle                                                                                                          | Puncture Knife<br>Puncture Sleeve<br>Reducer Cap<br>Puncture Needle                                                                                                                                                        |
| Single Use      | Single Use                                                                                                                                                                   | Single Use                                                                                                                                                                                                                 |
| Operation Mode  | Manually                                                                                                                                                                     | Manually                                                                                                                                                                                                                   |
| Accessory       | Puncture sleeve, reducer cap                                                                                                                                                 | No accessory                                                                                                                                                                                                               |
| Safety features | Yes                                                                                                                                                                          | Yes                                                                                                                                                                                                                        |

{7}------------------------------------------------

| Label/Labeling   | Comply with 21, CFR Section 801                | Comply with 21, CFR Section 801 |
|------------------|------------------------------------------------|---------------------------------|
| Shaft Diameter   | Available in 5, 8, 10 and 12mm                 | Available in 5, 11 and 12mm     |
| Shaft Length     | Available in 75 and 100mm                      | Short, regular, long            |
| Material         | ABS, Stainless Steel, PC, PE, Rubber, Colorant | Unknown                         |
| Trocar Width     | 52.4mm, 61.4mm                                 | Unknown                         |
| Total Length     | 158mm, 172mm, 184mm, 198mm                     | Unknown                         |
| Needle Tip Shape | Blunt                                          | Blunt                           |
| Sterilization    | EO sterilized                                  | EO sterilized                   |
| SAL              | 10-6                                           | 10-6                            |
| Endotoxin Limit  | 20 EU per device                               | 20 EU per device                |
| Shelf life       | 2 years                                        | 3 years                         |
| Packaging method | Sealing method                                 | Sealing method                  |
| Cytotoxicity     | No cytotoxicity                                | Comply with ISO 10993           |
| Skin Irritation  | No irritation                                  |                                 |
| Sensitization    | No sensitization                               |                                 |
| Pyrogen          | No pyrogen                                     |                                 |

Any technical differences have been justified, both scientifically and using performance testing. These do not affect the safety or effectiveness of the proposed device.

# CONCLUSION

Based on testing and comparison with the predicate devices shows no adverse indications or results. It is our determination that the proposed device is safe, effective and performs within its design specifications and is substantially equivalent to the predicate devices.

Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K190029](https://fda-staging.innolitics.com/device/K190029)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
