K190007 · Vydence Medical Industria E Comercio Ltda · GEX · Mar 1, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190007
Device Name
Vydence Laser Family
Applicant
Vydence Medical Industria E Comercio Ltda
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 1, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vydence Laser Family is indicated for a wide range of dermatological and surgical procedures depending on the specific handpiece and wavelength used. IPL filters (400-1200nm, 515-1200nm, 540-1200nm, 580-1200nm, 640-1200nm, 695-1200nm) are indicated for the treatment of moderate inflammatory acne vulgaris, benign pigmented epidermal lesions (dyschromia, hyperpigmentation, melasma, ephelides, lentigines, nevi, cafe-au-lait macules), benign cutaneous lesions (warts, scars, striae), benign cutaneous vascular lesions (port wine stains, hemangiomas, telangiectasias, rosacea, angiomas, Poikiloderma of Civatte, leg veins, venous malformations), tattoo removal, and stable long-term or permanent hair reduction on various Fitzpatrick skin types (I-VI). The Intense IR Handpiece is indicated for topical heating, temporary relief of minor muscle/joint pain and stiffness, temporary increase in local circulation, and muscle relaxation. The 1064nm Long Pulse Handpiece is indicated for hair reduction, PFB, benign vascular lesions, benign pigmented lesions, benign cutaneous lesions, non-ablative treatment of facial wrinkles, laser skin resurfacing, reduction of red pigmentation in scars, and podiatry applications including onychomycosis. The Acroma Handpiece (1064nm) is indicated for benign vascular lesions, benign cutaneous lesions, benign pigmented lesions, tattoo removal, non-ablative facial wrinkle treatment, and laser skin resurfacing. The 532nm wavelength is indicated for soft tissue incision/excision/ablation/vaporization, tattoo removal, benign vascular lesions, and benign pigmented lesions. The 1540 GoSmooth Handpiece is indicated for coagulation of soft tissue, skin resurfacing, and treatment of melasma, striae, and scars. The 2940 DualMode Handpiece is indicated for skin resurfacing, wrinkle treatment, and various dermatological/podiatric lesions. The 1340 ProDeep Handpiece is indicated for skin resurfacing. The 755nm Fiber Delivered Handpiece is indicated for hair reduction, benign pigmented lesions, wrinkles, and vascular lesions. The 1064nm Fiber Delivered Handpiece is indicated for vascular lesions, non-ablative wrinkle treatment, hair reduction, PFB, and onychomycosis.
Device Story
Vydence Laser Family consists of multi-wavelength laser and IPL system with interchangeable handpieces; delivers light energy (IPL, 532nm, 755nm, 1064nm, 1340nm, 1540nm, 2940nm) to target tissue; used by physicians in clinical settings for dermatological, plastic surgery, and podiatric applications; system transforms electrical energy into specific optical wavelengths; output used for coagulation, ablation, vaporization, or photothermolysis of target lesions; clinical decision-making guided by visual assessment of skin/lesion type; benefits include non-invasive or minimally invasive treatment of aesthetic and medical skin conditions; includes specialized modes like DYNAMICS for onychomycosis and fractional modes for resurfacing.
Clinical Evidence
Bench testing only. No clinical data provided in the summary.
Technological Characteristics
Multi-platform laser and IPL system; wavelengths include 532nm, 755nm, 1064nm, 1340nm, 1540nm, 2940nm and IPL (400-1200nm range); energy delivery via fiber-delivered or direct-contact handpieces; includes fractional and long-pulse modes; cooling systems integrated for skin protection; system is standalone; sterilization via standard clinical protocols for reusable handpieces.
Indications for Use
Indicated for patients requiring dermatological, plastic surgery, or podiatric procedures including hair reduction, vascular/pigmented lesion treatment, skin resurfacing, wrinkle reduction, scar/striae treatment, acne vulgaris management, tattoo removal, and temporary pain relief. Applicable to Fitzpatrick skin types I-VI depending on the specific wavelength/handpiece. Contraindications not explicitly listed beyond standard laser safety precautions.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Vydence Medical Industria E Comercio Ltda % Kathy Maynor Vydence Regulatory Consultant Vydence Medical 26 Rebecca Ct. Homosassa, Florida 34446
March 1, 2019
# Re: K190007
Trade/Device Name: Vydence Laser Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX, PDZ, ONF Dated: December 30, 2018 Received: January 2, 2019
Dear Kathy Maynor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Neil R Neil R Ogden -S Date: 2019.03.01 Oaden -S 15:35:30 -05'00'
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
# K190007
Device Name Vydence Laser Family
Indications for Use (Describe)
· IPL 400nm-1200nm Filter
· The treatment of moderate inflammatory acne vulgaris;
· The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, and ephelides (freckles):
· the treatment of benign cutaneous lesions including warts, scars and striae;
· The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas, Poikiloderma of Civatte, leg veins and venous malformations.
· Use on all skin types (Fitzpatrick I-VI).
IPL 515-1200nm Filter
· The treatment of moderate inflammatory acne (acne vulgaris).
· The treatment of tattoos and benign pigmented epidermal and benign cutaneous lesions including warts, scars, striae, dyschromia, hyperpigmentation, melasma, ephelides (freckles), lentigines, nevi, melasma, and cafè-aulait macules. · The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas, poikiloderma of Civate, leg veins and venous malformations.
· The removal of unwanted hair to effect stable long-term or permanent hair reduction in hair growth is defined as the long-term , stable reduction in the number of hairs regrowing when measured at 6,9 an 12 months after the completion of a treatment regime.
· Use on Fitzpatrick skin types I-V.
### IPL 540nm-1200nm Filter
· benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, and ephelides (freckles),
· lentigines, nevi, and cafe-au-lait macules;
· the treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas, poikiloderma of Civate, leg veins and venous malformations.
· The treatment of benign cutaneous lesions including warts, scars and striae.
· The removal of unwanted hair to effect stable long-term or permanent hair reduction in hair growth is defined as the long-term , stable reduction in the number of hairs regrowing when measured at 6,9 and 12 months after the completion of a treatment regime.
· Use on all skin types (Fitzpatrick I-VI).
#### IPL 540nm-1200nm Filter
· benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, and ephelides (freckles),
· lentigines, nevi, and cafe-au-lait macules;
· the treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas, poikiloderma of Civatte, leg veins and
{3}------------------------------------------------
venous malformations.
· The treatment of benign cutaneous lesions including warts, scars and striae.
· The removal of unwanted hair to effect stable long-term or permanent hair reduction in hair growth is defined as the long-term , stable reduction in the number of hairs regrowing when measured at 6,9 and 12 months after the completion of a treatment regime.
· Use on all skin types (Fitzpatrick I-VI).
IPL 580nm-1200nm Filter
· The treatment of moderate inflammatory acne vulgaris.
· The treatment of benign pigmented epidermal lesions, including dyschromia, hyperpigmentation, melasma and ephelides (freckles).
· The treatment of face and body benign vascular and benign pigmented lesions.
· The treatment of benign cutaneous lesions, including scars and striae.
· The treatment of benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, melasma, and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations.
· The removal of unwanted hair to effect stable long-term or permanent hair reduction in hair growth is defined as the long-term , stable reduction in the number of hairs regrowing when measured at 6,9 an 12 months after the completion of a treatment regime.
· Use on all skin types (Fitzpatrick I-VI).
#### IPL 640nm-1200nm Filter
· the treatment of tattoos;
· The treatment of mild to moderate inflammatory and pustular inflammatory acne vulgaris;
· The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, and ephelides (freckles), lentigines, nevi, melasma, and cafe-au-lait;
· the treatment of benign cutaneous lesions including warts, scars and striae;
· the treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, erythema of rosacea, angiomas, poikiloderma of Civate, leg veins and venous malformations;
· removal of unwanted hair, for stable long term or permanent hair reduction is defined as the long-term, stable reduction in the number if hair regrowing when measured at 6, 9 and 12 months after the completion of a treament regimen).
· Use on all skin types (Fitzpatrick I-VI), including tanned skin.
#### IPL 695nm-1200nm Filter
· removal of unwanted hair, for stable long term or permanent hair reduction is defined as the long-term, stable reduction in the number if hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regimen).
• Use on all skin types (Fitzpatrick I-VI), including tanned skin
#### Intense IR Handpiece
· intended to emit energy in the infrared spectrum to provide topical heating for the tissue temperature. It's also indicated for the temporary relief of minor muscle pain and joint pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and the relaxation of muscles; may also help muscle spasms, minor strains, and minor muscular back pain.
1064nm Long Pulse Handpiece
· removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The laser is indicated for all skin tvpes, Fitzpatrick I-VI, including tanned skin. Permanent hair reduction is defined as the long-term. stable reduction in the number if hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime:
· Treatment of pseudofolliculitis barbae (PFB)
{4}------------------------------------------------
· Benign vascular lesions such as, but not limited to treatment of: Port wine stains; Hemangiomas; Warts; Superficial and deep telangiectasias (venulectasias); Reticular veins (0.1-4.0 mm dia.) of the leg: Rosacea; Venus Spider veins; Poikiloderma of Civatte; Angiomas
· Benign pigmented lesions such as, but not limited to: Lentigos (age spots); Solar lentigos (sun spots); Cafe-au-lait macules; Sebortheic keratoses; Nevi and nevus of Ota; Chloasma; verrucae, skin tags, keratoses, the removal of black, blue, or green tattoos (significant reduction in the intensity of black and/or blue/black tattoos), plaques
· Benign cutaneous lesions, such as, but not limited to warts, scars, striae and psoriasis
- · Pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
- · The non-ablative treatment of facial wrinkles, such as, but not limited to: Periocular wrinkles
- · Laser skin resurfacing procedures for the treatment of: Acne scars, wrinkles
- · Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar;
# 1064nm Long Pulse Handpiece - Short Pulse Operation
· intended for the coagulation and hemostasis of benign vascular lesions such as but not limited to, rosaceal diffiuse redness, poikiloderma of civatte, scar reduction (including hypertropic and keloidscars), and warts.
- · is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.
- · For use on all skin types (Fitzpatrick I-VI), including tanned skin.
1064nm Long Pulse Handpiece - Onychomycosis
· podiatry (i.e. ablation, vaporization, excision, and coagulation of soft tissue) including matrixectomy, radical nail excision, periungual and subungual warts, plantar warts, neuromas,
· temporary increase of clear nail in patients with onychomycosis (e.g., dermatophyton rubrum and Trichophyton mentagrophytes, and/or yeasts Candida albicans, etc.).
### Acroma Handpiece
The 1064 nm wavelength is indicated for:
· treatment of benign vascular lesions such as (but not limited to): port wine stains; hemangiomas; warts; superficial and deep telangiectasias (venulectasias); reticular veins (0.1-4.0 mm diameter) of the leg; rosacea; venous lakes; leg veins; spider veins; Poikiloderma of Civatte; angiomas;
· benign cutaneous lesions, such as, but not limited to: warts; scars; striae; psoriasis;
· benign pigmented lesions such as, but not limited to: lentigos (sun lentigos (sun spots); cafe-au-lait macules; sebortheic keratoses; nevi and Nevus of Ota; chloasma; verrucae; skin tags; keratoses; the removal of black, blue or green tattoos (significant reduction in the intensity of black and/or blue/black tattoos); plaques;
· pigmented benign lesions to reduce lesions' size for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
- the non-ablative treatment of facial wrinkles, including but not limited to: periocular wrinkles;
- · laser skin resurfacing procedures for the treatment of acne scars; wrinkles,
- · reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
- · indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin.
# The 532 nm wavelength is indicated for:
- · incision, excision, ablation, vaporization of soft tissue;
- · tattoo removal: light blue, yellow, red; green;
· benign vascular lesions: Hemangiomas (port wine stains/birthmarks, cavernous, cherry, spider, hemangiomas); angiomas (cherry, spider); telangiectasias; spider nevi;
· benign pigmented lesions: cafe-au-lait macules: lentigines (senile and solar); freckles (ephelides); chloasma; nevi; nevus spillus: Nevus of Ota: Becker's Nevi:
· other benign pigmented cutaneous lesions: verrucae; skin tags; keratoses; plaques
# 1540 GoSmooth Handpiece
· The 1540 Fractional Non-ablative Laser Handed for use in the coagulation of soft tissue, skin resurfacing procedures as well as treatment of melasma, striae, acne scars and surgical scars.
{5}------------------------------------------------
# 2940 DualMode Handpiece · DERMATOLOGY AND PLASTIC SURGERY
Skin resurfacing; Treatment of wrinkles; Epidermal nevi; Telangiectasia; Spider veins;
Actinic chelitis; Keloids; Verrucae; Skin tags; Keratose; Scar revision (including acne scars); Debulking benign tumors; Debulking cysts; Superficial skin lesions; Diagnostic biopsies; Decubitus ulcers; · PODIATRY
Treatment of: Warts, plantar verrucae, large mosaic verrucae. Matrixectomy.
2940 DualMode - fractional - skin resurfacing
1340 ProDeep Handpiece · non-ablative laser handpiece is intended for use in skin resurfacing
755nm Fiber Delivered Handpiece (Zye and One)
755nm
• Intended for the temporary and permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin. Permanent hair reduction is defined as long-term, stable reduction in hair counts observed at 6, 9, and 12 months after the end of a treatment regime; treatment of benign pigmented lesions; treatment of wrinkles; photocoagulation of dermatological vascular lesions, such as, but not limited to, port wine stains, hemangiomas and telangiectasias.
# 1064nm Fiber Delivered Handpiece (Zye and One)
In long pulse mode, the ZYE YAG® applicator is designated for coagulation and hemostasis of vascular lessons and epidermal tissue, including the treatment of telangiectasia, superficial varicosities, angiomas and spider angiomas, hemangiomas, rosacea, and nevi. It is also designated for non-ablative treatment of facial wrinkles and for removal of unwanted hair, for stable long term or permanent hair reduction is defined as the long-term, stable reduction in the number if hair regrowing when measured at 6, 9 and 12 months after the completion of a treatment regimen); for the treatment of pseudofolliculitis barbae (PFB). Use on all skin types (Fitzpatrick I-V), including tanned skin.
· In the DYNAMICS® mode, for temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes Trichophyton rubrum and Trichophyton mentagrophytes, and/or yeasts Candida albicans, etc.).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{6}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.