← Product Code [FOZ](/productcode/FOZ) · K190001

# Supercath 5 (K190001)

_Togo Medikit Co., Ltd. · FOZ · Sep 20, 2019 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K190001

## Device Facts

- **Applicant:** Togo Medikit Co., Ltd.
- **Product Code:** [FOZ](/productcode/FOZ.md)
- **Decision Date:** Sep 20, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Supercath 5 is intended to access a vein or artery and to administer fluids and/or blood. The Supercath 5 is designed for single use, and is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18 - 22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi.

## Device Story

Supercath 5 is a single-use intravascular catheter for venous or arterial access; used to administer fluids and blood. Features include a metallic introducer needle with a side-notch for visual confirmation of vessel entry; a built-in check valve to reduce blood flashback during needle withdrawal; and a spring-activated needle retraction mechanism to prevent needlestick injuries. Operated by healthcare professionals in clinical settings. The device is inserted into the patient; blood flashback confirms placement; the needle is retracted via a button on the hub. 18-22G models support power injection. Output is the established vascular access path for fluid/blood delivery. Benefits include reduced risk of accidental needlesticks and rapid confirmation of vessel entry.

## Clinical Evidence

Bench testing only. No clinical data. Testing included biocompatibility (leveraged from predicates), hemocompatibility (leveraged from reference device K160592), power injection pressure testing (ISO 10555-1), vent fitting (ISO 10555-5), flow rate, assembly performance, and tensile strength. Sterilization validated per ISO 11135:2014.

## Technological Characteristics

Polyurethane catheter; radiopaque; ethylene oxide sterilized. Features: side-notch introducer needle, spring-retraction needlestick prevention, check valve, and needle hub packing. Dimensions: 14G-26G. Standards: ISO 10993-1 (biocompatibility), ISO 10555-1/5 (catheter performance), ISO 594-2 (plunger assembly).

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- Supercath 5 ([K140419](/device/K140419.md))
- Supercath 5 ([K172496](/device/K172496.md))

## Reference Devices

- Supercath 6 ([K160592](/device/K160592.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 20, 2019

TOGO MEDIKIT Co., Ltd. % Izumi Maruo Senior Consultant MIC International 4-1-17 Hongo Bunkyo-ku, Tokyo 113-0033 JAPAN

Re: K190001

Trade/Device Name: Supercath 5 Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: August 15, 2019 Received: August 19, 2019

Dear Izumi Maruo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Nikhil Thakur Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190001

Device Name Supercath 5

Indications for Use (Describe)

The Supercath 5 is intended to access a vein or attery and to administer fluids and/or blood. The Supercath 5 is designed for single use, and is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18 - 22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi.

Type of Use (Select one or both, as applicable)

| <span style="font-size:12pt">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:12pt">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo with a globe-like grid pattern in the background. In the foreground, there is a stylized human figure with arms outstretched, forming a 'P' shape with its body. The figure is positioned in front of the globe, creating a sense of global reach or connection.

TOGO MEDIKIT CO., LTD.
17148-6, Aza Kamekawa, Oaza Hichiya, Hyuga City
Miyazaki Prefecture 883-0062, Japan
TELEPHONE: +81-982-53-8027
FACSIMILE: +81-982-53-8008

Page 1 of 8

# 510(k) Summary - K190001

#### Owner/Company name, address a.

TOGO MEDIKIT CO., LTD. 17148-6 Aza Kamekawa, Oaza Hichiya Hyuga City, Miyazaki, 883-0062, Japan

Daisuke Nagamizu Manager, Regulatory Affairs Section, Quality Assurance Department Phone: 011-81-982-53-8000 Fax: 011-81-982-53-8008 qc@togomedikit.co.jp Email:

### b. Contact/Application Correspondent

Izumi Maruo Senior Consultant MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, 113-0033, Japan

011-81-3-3818-8577 Phone: Fax: 011-81-3-3818-8573 Email: maruo@mici.co.jp

# c. Date prepared

September 18, 2019

#### d. Name of device

| Trade Name:           | Supercath 5                                                           |
|-----------------------|-----------------------------------------------------------------------|
| Common Name:          | Catheter, Intravascular, Therapeutic, Short-term less than 30<br>days |
| Regulation Name:      | Intravascular Catheter                                                |
| Regulation Number:    | 21 CFR 880.5200                                                       |
| Regulation Class:     | II                                                                    |
| Product Code:         | FOZ                                                                   |
| Classification Panel: | General Hospital                                                      |

Hereinafter, the Proposed Device is called as "Supercath 5 (Proposed)" in this submission because predicate devices have the same trade name.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo with a globe-like structure made of intersecting lines, resembling a grid. Inside the globe, there is a stylized human figure with arms outstretched, as if embracing the world. The logo is presented in grayscale, giving it a simple and clean appearance.

TOGO MEDIKIT CO., LTD.
17148-6, Aza Kamekawa, Oaza Hichiya, Hyuga City
Miyazaki Prefecture 883-0062, Japan
TELEPHONE: +81-982-53-8027
FACSIMILE: +81-982-53-8008

# e. Predicate and reference devices

## Predicate devices

The Supercath 5 (Proposed) has two predicate devices because those two devices altogether cover the sizes which the Supercath 5 (Proposed) includes.

# Predicate Device for gauges 14G, 16G, 18G, 20G, 22G, and 24G

| Trade Name:           | Supercath 5                                                        |
|-----------------------|--------------------------------------------------------------------|
| 510(k) Number:        | K140419                                                            |
| Common Name:          | Catheter, Intravascular, Therapeutic, Short-term less than 30 days |
| Regulation Name:      | Intravascular Catheter                                             |
| Regulation Number:    | 21 CFR 880.5200                                                    |
| Regulation Class:     | II                                                                 |
| Product Code:         | FOZ                                                                |
| Classification Panel: | General Hospital                                                   |

Hereinafter "Supercath 5 (K140419)" in this submission. This predicate device is primary predicate.

| Predicate Device for 26G |                                                                       |
|--------------------------|-----------------------------------------------------------------------|
| Trade Name:              | Supercath 5                                                           |
| 510(k) Number:           | K172496                                                               |
| Common Name:             | Catheter, Intravascular, Therapeutic, Short-term less than<br>30 days |
| Regulation Name:         | Intravascular Catheter                                                |
| Regulation Number:       | 21 CFR 880.5200                                                       |
| Regulation Class:        | II                                                                    |
| Product Code:            | FOZ                                                                   |
| Classification Panel:    | General Hospital                                                      |

Hereinafter "Supercath 5 (K172496)" in this submission.

| Reference Devices     |                                                                    |
|-----------------------|--------------------------------------------------------------------|
| Trade Name:           | Supercath 6                                                        |
| 510(k) Number:        | K160592                                                            |
| Common Name:          | Catheter, Intravascular, Therapeutic, Short-term less than 30 days |
| Regulation Name:      | Intravascular Catheter                                             |
| Regulation Number:    | 21 CFR 880.5200                                                    |
| Regulation Class:     | II                                                                 |
| Product Code:         | FOZ                                                                |
| Classification Panel: | General Hospital                                                   |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo with a globe-like structure in the background. In the foreground, there is a stylized human figure with arms raised, seemingly integrated into the globe design. The logo has a circular shape and appears to be in grayscale.

# f. Description of the device

The Supercath 5 (Proposed) is intended to access a vein or artery and to administer fluids and/or blood. The Supercath 5 (Proposed) is designed for single use, and is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18 - 22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi.

The maximum flow rate and maximum pressure setting for each catheter size are listed in the table shown as follows. The maximum flow rate may vary due to differences in injector characteristics, injectate viscosity and temperature, and other variables.

| Gauge | Length |      | Maximum<br>Flow Rate<br>(mL/sec) | Maximum<br>Pressure<br>(psi) |
|-------|--------|------|----------------------------------|------------------------------|
|       | (in)   | (mm) |                                  |                              |
| 22G   | 1      | 25   | 7                                | 300                          |
| 22G   | 1 1/4  | 31   | 7                                | 300                          |
| 20G   | 1      | 25   | 9                                | 300                          |
| 20G   | 1 1/4  | 31   | 9                                | 300                          |
| 20G   | 2      | 50   | 9                                | 300                          |
| 18G   | 1 1/4  | 31   | 12                               | 300                          |
| 18G   | 2      | 50   | 12                               | 300                          |

Table 1. The maximum flow rate and the maximum pressure setting

The Supercath 5 (Proposed) catheter hub has a built-in check valve which together with the healthcare professional's finger pressure on the blood vessel, assists to reduce blood flashback when the metallic introducer needle is withdrawn following blood vessel puncture. The built-in check valve is not intended to stop bleeding completely. Pressing the button on the needle hub activates the coiled spring in the hub, retracting the metallic introducer needle into the needle hub.

The Supercath 5 (Proposed) introducer needle has a side-notch to provide rapid visual confirmation of vessel entry. When the introducer needle is inserted into the vein, blood flows up and through the side-notch and returns down along the inside of the catheter tube.

The Supercath 5 (Proposed) is available in 14G, 16G, 18G, 20G, 22G, 24G and 26G.

# g. Indications for Use Statement

The Supercath 5 is intended to access a vein or artery and to administer fluids and/or blood. The Supercath 5 is designed for single use and is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18 - 22-gauge catheters may be used with power injectors at a maximum pressure of 300 psi.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo with a globe-like background. In the center of the globe is a stylized human figure with arms raised. The figure is white, and the globe is made up of a grid pattern.

TOGO MEDIKIT CO., LTD.
17148-6, Aza Kamekawa, Oaza Hichiya, Hyuga City
Miyazaki Prefecture 883-0062, Japan
TELEPHONE: +81-982-53-8027
FACSIMILE: +81-982-53-8008

### Page 4 of 8

## h. Comparative Information

The Supercath 5 (Proposed) was modified from the Supercath 5 (K140419) and the Supercath 5 (K172496). The primary difference in the Supercath 5 (proposed) is the indications for use statement includes administration of blood and 18-22 gauge catheters may be used with power injectors. These changes do not affect the intended use of the device compared to the predicates.

## Technological characteristics

The similarities of the subject device to the predicate devices are:

- Same catheter material (Polyurethane)
- Radiopaque
- Flashback Visualization
- Side-Notch Needle ●
- Same Needlestick Injury Prevention Feature ●
- Check Valve
- Ethylene Oxide Sterilized ●
- Single Sterile Wrapped ●
- Multiple Gauge Sizes and Needle Lengths ●

The Supercath 5 (Proposed) contains following modifications as compared to the predicate devices:

- Change of material formulation of the plug, the wing, the plunger, the cap, and the protector
- The adjustable attachment on the tip of the grip is not used for all models
- Addition of a packing part for blood on the needle hub for all models ●
- Addition of power injection to 18G, 20G, and 22G ●
- Addition of administration of blood in the Indications for Use ●

Performance testing was done to support the modifications of the subject device

Followings are comprehensive comparison table for the proposed, predicate, and reference devices:

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a logo featuring a stylized human figure within a globe. The figure is composed of simple lines and shapes, suggesting a person with arms raised. The globe is represented by a grid pattern, giving it a three-dimensional appearance. The logo is in grayscale and appears to be a symbol for an organization or concept related to global humanity.

TOGO MEDIKIT CO., LTD
17148-6, Aza Kamekawa, Oaza Hichiya, Hyuga City
Miyazaki Prefecture 883-0062, Japan
TELEPHONE: +81-982-53-8027

Table 2 Comparison table between the Supercath 5 (Proposed), the predicate and the reference devises

| Item                                     | Supercath<br>(Proposed)                                                                                                                                                                                                                                                                                                                                                                 | Predicate Devices                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                           | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                        | Substantial<br>Equivalence<br>Analysis                                                                                                                   |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                          | 5                                                                                                                                                                                                                                                                                                                                                                                       | Supercath<br>(K140419)                                                                                                                                                                                                                                                                                    | 5<br>Supercath<br>(K172496)                                                                                                                                                                                                                               | 5<br>Supercath<br>(K160592)                                                                                                                                                                                                                                                                                                                                                                                             | 6                                                                                                                                                        |
| Indications for Use<br>Statement         | The Supercath 5 is<br>intended to access a<br>vein or artery and to<br>administer fluids<br>and/or blood. The<br>Supercath 5 is<br>designed for single<br>use, and is designed<br>to minimize<br>inadvertent<br>needlesticks and to<br>reduce accidental<br>needlesticks. The 18<br>- 22 gauge catheters<br>may be used with<br>power injectors at a<br>maximum pressure<br>of 300 psi. | The Supercath 5 is<br>intended to access a<br>vein or artery and to<br>administer fluids.<br>The Supercath 5 is<br>designed for single<br>use and for<br>short-term use (less<br>than 30 days), is<br>designed to<br>minimize<br>inadvertent<br>needlesticks and to<br>reduce accidental<br>needlesticks. | The Supercath 5 is<br>intended to access a<br>vein or artery and to<br>administer fluids.<br>The Supercath 5 is<br>designed for single<br>use, and is designed<br>to minimize<br>inadvertent<br>needlesticks and to<br>reduce accidental<br>needlesticks. | The Supercath 6 is<br>intended to access a<br>vein or artery and to<br>administer fluids.<br>The Supercath 6 is<br>designed for single<br>use and for<br>short-term use (less<br>than 30 days), is<br>designed to<br>minimize<br>inadvertent<br>needlesticks and to<br>reduce accidental<br>needlesticks. The 18<br>- 22 gauge catheters<br>may be used with<br>power injectors at a<br>maximum pressure<br>of 300 psi. | Substantially<br>equivalent.<br>For detail about<br>substantial<br>equivalence<br>analysis for<br>Indications for Use<br>statement, please<br>see above. |
| Needlestick injury<br>prevention feature | The metallic introducer needle is retracted into the needle hub when the activation button<br>on the needle hub is pressed.                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                     |
| Flashback<br>visualization               | Yes                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                         | Same.                                                                                                                                                    |
| Side-notch needle*1                      | Yes                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                     |

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows a logo that appears to be for an organization or company. The logo features a stylized human figure standing within a globe-like structure. The figure is abstract, with a large head and arms raised, giving the impression of someone reaching upwards or outwards. The globe is represented by a grid of lines, suggesting a network or global connection.

TOGO MEDIKIT CO., LTD

17148-6, Aza Kamekawa, Oaza Hichiya, Hyuga City
Miyazaki Prefecture 883-0062, Japan
TELEPHONE: +81-982-53-8027
FACSIMILE: +81-982-53-8008

| Item                                                        | Supercath<br>(Proposed)                           | <span style="text-align:center;">Predicate Devices</span> | Supercath<br>(K140419)                    | Supercath<br>(K172496) | Reference Device | Supercath<br>(K160592) | Substantial<br>Equivalence<br>Analysis                                                                       |
|-------------------------------------------------------------|---------------------------------------------------|-----------------------------------------------------------|-------------------------------------------|------------------------|------------------|------------------------|--------------------------------------------------------------------------------------------------------------|
| Radiopaque                                                  | Yes                                               |                                                           | Yes                                       | Yes                    |                  |                        | Same                                                                                                         |
| Adjustable<br>attachment*2                                  | No                                                |                                                           | Yes                                       | No                     |                  |                        | Please see *2                                                                                                |
| Additional packing<br>part for blood on the<br>needle hub*3 | Yes                                               |                                                           | No                                        | No                     |                  |                        | Performance data                                                                                             |
| Sterilization                                               | Ethylene oxide                                    |                                                           | Ethylene oxide                            | Ethylene oxide         |                  |                        | Same                                                                                                         |
| Packaging                                                   | Blister                                           |                                                           | Blister                                   | Blister                |                  |                        | Same.                                                                                                        |
| Use with Power<br>Injectors (18G, 20G,<br>22G)              | Yes                                               |                                                           | No                                        | No                     |                  | Yes                    | Performance Data                                                                                             |
| Multiple gauge<br>sizes and needle<br>length                | Yes (14G, 16G, 18G,<br>20G, 22G, 24G, and<br>26G) |                                                           | Yes (14G, 16G, 18G,<br>20G, 22G, and 24G) | 26G                    |                  |                        | The two predicate<br>devices altogether<br>cover the sizes<br>which the Supercath<br>(Proposed)<br>includes. |
| Material of catheter<br>tube                                | Polyurethane                                      |                                                           | Polyurethane                              | Polyurethane           |                  |                        | Same.                                                                                                        |

*! Side-notch needle is the introducer needle with a side-noth, to provide rapid visual confirmation of vessel entry. When the introducer needle is inserted into the vein, blood flows upward into the side-notch alown along the inside of the catheter tube. This does not affect the intended use of the device.

*2: The adjustable attachment is used just when manufacturing the Supercath 5 (K140419). The attachment the tip of the introducer needle and the tip of the catheter. This does not affect the intended use of the device.

*3: The packing part on the needle hub improves the prevention of the device. This does not affect the intended use of the device.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows a logo with a stylized human figure standing inside a globe. The human figure is white, and the globe is made up of intersecting lines, giving it a grid-like appearance. The logo is simple and abstract, focusing on the concept of humanity within a global context.

#### i. Performance data

Evaluations of the modifications in support of the substantial equivalence determination is shown below;

## 1. Biocompatibility testing

The Supercath 5 (Proposed) contacts patient's skin and blood and is categorized as an External communicating. Circulating blood contact device with a Prolonged contact duration in accordance with ISO 10993-1:2009 and referencing FDA guidance entitled "Use of International Standard ISO 10993- 1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The nature of contact with the body and the category of the Supercath 5 (Proposed) are identical to those of the predicate devices.

Biocompatibility testing was leveraged from the sponsor's own predicate devices (K140419 and K172496).

The addition of administration of blood in the Indications for Use is supported by the Hemocompatibility testing (hemolysis, complement system, coagulation, platelets, and thrombosis) performed on the sponsor's own reference device (K160592).

2. Power injection test

Catheter burst pressure exceeds the peak pressure present in the catheter at maximum flow conditions when tested according to ISO 10555-1:2013.

3. Vent fitting test for the needle hub

The needle hub with the packing part is impervious to liquid infiltration when subjected to positive pressure in accordance with ISO 10555-5:2013.

- 4. Flow rate test
The flow rate through catheter meets allowable limits when tested according to ISO 10555-1:2013.

5. Ease of assembly performance testing for the plunger

The plunger passes through the check valve when tested according to ISO 594-2:1998 section 5.6.

6. Tensile strength testing for the cap

The tensile strength of the cap meets acceptable minimum force until breakage when tested according to in-house standard.

7. Sterilization validation in accordance with ISO 11135:2014 and shelf-life testing were conducted.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows a logo with a globe-like structure in the background. In the foreground, there is a stylized human figure with arms outstretched, superimposed over the globe. The figure appears to be made of interconnected lines or segments, giving it a modern and abstract look. The overall design is simple and monochromatic, with a focus on the interconnectedness of the human figure and the global sphere.

The results of testing above did not raise any new questions of safety or effectiveness.

# j. Conclusion

Based on the Indications for Use, technological characteristics, and performance testing, Togo Medikit concludes that the Supercath 5 (proposed) is substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K190001](https://fda-staging.innolitics.com/device/K190001)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).