Doctor Tecar

K183693 · Mectronic Medicale S.R.L. · PBX · Feb 22, 2019 · General, Plastic Surgery

Device Facts

Record IDK183693
Device NameDoctor Tecar
ApplicantMectronic Medicale S.R.L.
Product CodePBX · General, Plastic Surgery
Decision DateFeb 22, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Doctor Tecar device is intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation.

Device Story

Doctor Tecar is a radiofrequency (RF) therapy device; generates high-frequency sinusoidal current (460-600 KHz) in monopolar mode. System consists of power console on trolley, LCD monitor, and capacitive/resistive/multipolar electrodes. Therapist operates device by placing fixed electrode on patient and manipulating handheld electrode; electrical circuit closes upon contact. Device provides topical heating to elevate tissue temperature. Used in clinical settings by therapists; intensity adjustable 0-100%. Benefits include pain relief, reduction of muscle spasms, and improved local circulation.

Clinical Evidence

Bench testing only. Studies conducted on human volunteers of varying skin colors confirmed the device's capacity to safely elevate tissue temperature using various electrode sizes. Biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity), electrical safety (IEC 60601-1, IEC 60601-2-2), and electromagnetic compatibility (IEC 60601-1-2) testing yielded satisfactory results.

Technological Characteristics

Cart-mounted console; RF frequency 460-600 KHz; max power 120 W. Monopolar operation using capacitive/resistive/multipolar electrodes. Standards: ISO 10993 (biocompatibility), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-2-2:2017 (electrosurgical safety). Intensity adjustment 0-100%.

Indications for Use

Indicated for patients requiring topical heating for relief of pain, muscle spasms, and increase in local circulation.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. February 22, 2019 Mectronic Medicale S.r.l. Ms. Gloria Aloisini Quality Manager Via Orio al Serio, 15 24050 Grassobbio (BG), ITALY Re: K183693 Trade/Device Name: Doctor Tecar Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: PBX Dated: December 19, 2018 Received: December 31, 2018 Dear Ms. Aloisini: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, David Krause - Digitally signed by David David Krause -s Date: 2019.02.22 08:29:41 ട -05'00' for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (ifknown) -K183693 Device Name Doctor Tecar Indications for Use (Describe) The Doctor Tecar device is intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation. Type of Use (Select one or both, as applicable } [X] Prescription Use (Part 21 CFR 801 Subpart D) [ ] Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRA Staff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valtd 0 MB number. " {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows a logo for a company called "ELECTRONIC MEDICALE". The logo features a stylized star on the left side, with the company name written in a stylized font to the right of the star. The word "MEDICALE" is written in smaller letters below the word "ELECTRONIC". # 510(k) Summary Pursuant to 21 CFR 807.92 Date: February 18, 2019 K183693 | 1. | Submitted By: | Mectronic Medicale S.r.l.<br>Via Orio al Serio, 15<br>24050 Grassobbio (BG)<br>ITALY<br>ennioaloisini@mectronicmedicale.it<br>Phone: +39 035 656080, FAX: +39<br>035 65736 | |----|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 2. | Contact: | Mrs. Gloria Aloisini<br>Quality Manager<br>Mectronic Medicale S.r.l.<br>gloriaaloisini@mectronicmedicale.it<br>Phone: +39 035 656080, FAX: +39<br>035 65736 | | 3. | US Agent: | Mr. Guilliermo Laureiro<br>President<br>Luna America LLC<br>751 Willard Ave<br>Jefferson, PA, 18436<br>Cell 570-840-2225<br>Fax 570-955-3783<br>e-mail gil@lunaamerica.com | | 4. | Product: | Doctor Tecar<br>(21CFR§878.4400) Class II<br>Doctor Tecar<br>Product code PBX | | 5. | Common/Trade Name: | Radiofrequency Induced Heat | Doctor Tecar ## Description: The Doctor Tecar generates a high frequency sinusoidal current with a monopolar mode of application using two electrodes. A fixed electrode is placed in contact with the patient and a hand held electrode is manipulated by a therapist. When both electrodes are in contact with a patient the electrical circuit is closed and RF therapy can be provided. The device can be operated in a capacitor resistive monopolar mode . The product consists of a power console on a moveable trolley, LCD monitor, and accessories including capacitive resistive electrodes and multipolar electrodes. The unit can be adjusted to provide various levels of treatment frequency ranging from 460 KHz to 600 KHz. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows a logo for a company called "Electronic Medicale". The logo features a blue star above the word "Electronic", with the word "Medicale" written in a smaller font size below "Electronic". The text and star are all in a blue color. ## Intended Use: The Doctor Tecar device is intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation. ### Substantial Equivalence/Technological Characteristics: The Doctor Tecar device is substantially equivalent to the following device: - •Predicate device clearance: K162828 - •Trade Name: Winback 3SE - •Company: WINBACK USA Corp Both devices are cart mounted consoles with electrode accessories capable of operation in monopolar modes in the range of 460 kHz to 600 KHz radiofrequency. Both devices operate in the same treatment range and voltage and feature intensity adjustments from 0 to 100%. Electrical safety and biocompatibility have been established for both devices. No direct comparison was made since there are no significant differences in operation and test results indicate identical safety. The table below summarizes the equivalence of the devices. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows a logo for a company called "Mectronic Medicale". The logo features a blue star above the word "Mectronic", which is also in blue. Below "Mectronic" is the word "Medicale" in a smaller font size and also in blue. The logo has a slightly blurred or shadowed effect, giving it a three-dimensional appearance. # Predicate Device Comparison Table | Element of Comparison | 510(k) Device: | Predicate Device: | Explanation of Differences | |-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------| | | Doctor Tecar | Winback 3SE K162828 | | | Regulation and Product<br>Classification Code | 21 CFR 878.4400<br>PBX | 21 CFR 878.4400<br>PBX | None | | Indications for Use | The Doctor Tecar device is<br>intended to provide topical<br>heating for the purpose of<br>elevating tissue temperature for<br>the treatment of selected<br>medical conditions such as<br>relief of pain, muscle spasms,<br>and increase in local<br>circulation. | The Winback Back 3SE devise is<br>intended to provide topical heating<br>for the purpose of elevating tissue<br>temperature for the treatment of<br>selected medical conditions such as<br>relief of pain, muscle spasms, and<br>increase in local circulation.<br>The Winback 3SE massage device<br>is intended to provide a temporary<br>reduction in the appearance of<br>cellulite. | No significant<br>difference for<br>radiofrequency<br>device. | | Electrode Shapes | Square and circular | Square and circular | Electrode shape is<br>not a significant<br>difference | | Infrared Light | No | No | Identical | | Vacuum (suction) | No | No | Identical | | Treatment Activation | Finger selection on console | Finger selection on console | No significant<br>difference | | RF Type | Unipolar | Multipolar/Unipolar | No significant<br>difference | | RF Frequency | 460 KHz - 600 KHz | 300kHz - 1 MHz | No significant<br>difference | | Max RF Power | 120 W | 300 W | Max RF Power NO<br>significant difference | | Intensity Adjustment | 0-100% | 0-100% | Identical | | Configuration | Cart mounted console with<br>accessories | Cart mounted console with<br>accessories | Identical | | Patient Safety Switch | Yes | Yes | Identical | | Standards Compliance | ISO10993, IEC60601-1,<br>IEC60601-1-2 Compliant & IEC<br>60601-2-2: 2017 (6.0 Ed.) for use<br>in conjunction with<br>IEC 60601-1: 2005 AM1:2012 | ISO10993, IEC60601-1 &<br>IEC60601-1-2 Compliant | Identical | {6}------------------------------------------------ Image /page/6/Picture/2 description: The image shows a logo for Electronic Medicale. The logo features a blue star-like shape on the left, followed by the text "ELECTRONIC" in a stylized, italicized font. Below "ELECTRONIC" is the word "MEDICALE" in a smaller, less stylized font. The overall design has a slightly blurred or shadowed effect, giving it a three-dimensional appearance. ### Summary of Testing: The technological characteristics of the Doctor Tecar System has been verified based on assessments of electrical safety, performance, biocompatibility, software and usability. In addition, an assessment was done to demonstrate that the device effectively raises tissue temperature in a safe and efficient manner. The following testing has been conducted with satisfactory results: - Doctor Tecar Usability & Risk Management: Usability and Risk Management . assessments were done using worse-case assumptions to verify user interface, safety features and satisfactory performance. - Tissue Temperature Elevation Assessment: Studies were done in Italy using volunteers of varying skin colors to assess the capacity of the device to elevate tissue temperature in the treatment areas. The smallest and the largest electrodes were tested as well as the facial applicator electrode. Results indicated satisfactory safe therapeutic increases in tissue temperature. - Biocompatibility: Samples of the tissue contacting probes were tested for cytotoxicity, sensitization and intracutaneous reactivity. - Software Assessment: Software features were assessed in accordance with FDA software validation guidelines. Levels of Concern, User & System Requirements, Hazard Analysis, Software Requirements, Architectural Design, Software Validation & Testing were all addressed. - . Electromagnetic compatibility: EMC testing was done to evaluate emissions and immunity to electromagnetic fields in accordance with IEC 60601-1-2. - Electrical safety: Full electrical safety testing was done in compliance with IEC 60601-1. - Additional Electrical Testing: Additional electrical testing was done in . compliance with IEC 60601-2-2: 2017 (6.0 Ed.) for use in conjunction with IEC 60601-1: 2005 AM1:2012 ### Substantial Equivalent Conclusion: All testing has determined that Doctor Tecar is substantially equivalent to the predicate Winback 3SE device clearance: K162828. Both devices operate in the same treatment range and voltage and feature intensity adjustments from 0 to 100%. Electrical safety and biocompatibility have been established for both devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...