O-Scan

K183685 · Esaote, S.p.A. · LNH · Jan 22, 2019 · Radiology

Device Facts

Record IDK183685
Device NameO-Scan
ApplicantEsaote, S.p.A.
Product CodeLNH · Radiology
Decision DateJan 22, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

O-scan is a Magnetic Resonance (MR) system that produces transversal, sagittal, coronal and oblique section images of limbs and joints. It is designed for imaging portions of the arm, including hand, wrist, forearm, and elbow, but excluding the upper part of the arm and portions of the leg, including foot, ankle, calf, and knee, but excluding the thigh. The M.R. images produced by O-scan correspond to the spatial distribution of the protons (hydrogen nuclei) which verify the properties of magnetic resonance and depend on the MR parameters spin-lattice relaxation time (T1), spin-spin relaxation time (T2), nucleus density, flow velocity and "chemical shift". When interpreted by a medical expert trained in use of MR equipment, the images can provide diagnostically useful information.

Device Story

O-scan is a Magnetic Resonance (MR) imaging system for limbs and joints. It acquires MR signals based on proton distribution, T1/T2 relaxation times, nucleus density, flow velocity, and chemical shift. The system processes these signals to generate transversal, sagittal, coronal, and oblique images. The device includes the 'EVO'18' software upgrade, featuring 'AGilExam' for automatic slice positioning (knee, ankle, wrist) and a '3D Viewer' for 3D series reconstruction and MPR curve generation. Operated by trained medical experts in clinical settings, the output images assist in diagnostic decision-making. The device benefits patients by providing non-invasive, detailed anatomical imaging of extremities.

Clinical Evidence

No clinical tests are included within this submission; substantial equivalence is supported by non-clinical testing, including software verification and validation, and conformance to recognized medical device safety standards.

Technological Characteristics

Magnetic Resonance Imaging system; utilizes proton density and relaxation times (T1, T2) for image generation. Software-based features include AGilExam (automated slice positioning) and 3D Viewer (MPR reconstruction). Conforms to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-33, ISO 14971, ISO 62304, IEC 62366, NEMA MS-1, and NEMA MS-3 standards.

Indications for Use

Indicated for imaging limbs and joints, specifically hand, wrist, forearm, elbow, foot, ankle, calf, and knee, for diagnostic purposes by trained medical experts.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 22, 2019 Esaote S.p.A % Allison Scott, RAC Associate Director Navigant Consulting, Inc 9100 Keystone Crossing, Suite 500 INDIANAPOLIS IN 46240 Re: K183685 Trade/Device Name: O-Scan Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: December 27, 2018 Received: December 28, 2018 Dear Ms. Scott: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael D. O'Hara For Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K183685 Device Name O-Scan #### Indications for Use (Describe) O-scan is a Magnetic Resonance (MR) system that produces transversal, sagittal, coronal and oblique section images of limbs and joints. It is designed for imaging portions of the arm, including hand, wrist, forearm, and elbow, but excluding the upper part of the arm and portions of the leg, including foot, ankle, calf, and knee, but excluding the thigh. The M.R. images produced by O-scan correspond to the spatial distribution of the protons (hydrogen nuclei) which verify the properties of magnetic resonance and depend on the MR parameters spinlattice relaxation time (T1), spin-spin relaxation time (T2), nucleus density, flow velocity and "chemical shift". When interpreted by a medical expert trained in use of MR equipment, the images can provide diagnostically useful information. **Type of Use (Select one or both, as applicable)** X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ 510(k) Summary O-scan Esaote S.p.A. #### 510(k) Summary The following 510(k) summary has been prepared pursuant to requirements specified in 21CFR 807.92(a). #### Submitter Information Esaote, S.p.A. Via E. Melen 77 Genoa 16152 Italy Allison Scott, RAC Contact Person: P: 317.272.5419 F: 317.217.5361 Allison.Scott@navigant.com Date: December 21, 2018 Trade Name: O-scan Classification Panel: Radiology Classification Name(s): Magnetic Resonance Diagnostic Device 90LNH Classification Number: # Predicate Device(s) | Trade Name | Common name | Class | Product<br>code | Manufacturer | K<br>number | |------------|-----------------------------------------------|-------|-----------------|---------------|-------------| | O-scan | System, nuclear magnetic<br>resonance imaging | II | LNH | ESAOTE S.P.A. | K092469 | #### Reference Device(s) | ACHIEVA<br>1.0T | System, nuclear magnetic<br>resonance Imaging | II | LNH | PHILIPS MEDICAL<br>SYSTEMS | K052078 | |-----------------|-----------------------------------------------|----|-----|----------------------------|---------| | G-scan Brio | System, nuclear magnetic<br>resonance imaging | II | LNH | ESAOTE S.P.A. | K180592 | {4}------------------------------------------------ 510(k) Summary O-scan Esaote S.p.A. #### Device Description The changes performed on the modified O-scan, with respect to the cleared version – O-scan K092469 are due to the upgrade of the software system. These modifications, which do not affect the intended use or alter the fundamental scientific technology of the device, are the following: - A new Software version (EVO'18) including the following features: - 1. AGilExam (automatic slice positioning for the examination of the anatomical regions of knee, ankle and wrist) - 2. 3D Viewer; this is an option that starting from the acquired data cube allows the running of a 3D viewer in which to reconstruct the desired 3D series, both using the three projections of the acquired package and generating a special MPR curve as reference. (Previously cleared by FDA via K180592 on G-Scan Brio) ### Intended Use(s) O-scan is a Magnetic Resonance (MR) system that produces transversal, sagittal, coronal and oblique section images of limbs and joints. It is designed for imaging portions of the arm, including hand, wrist, forearm, and elbow, but excluding the upper part of the arm and portions of the leg, including foot, ankle, calf, and knee, but excluding the thigh. The M.R. images produced by O-scan correspond to the spatial distribution of the protons (hydrogen nuclei) which verify the properties of magnetic resonance and depend on the MR parameters spin-lattice relaxation time (T1), spinspin relaxation time (T2), nucleus density, flow velocity and "chemical shift". When interpreted by a medical expert trained in use of MR equipment, the images can provide diagnostically useful information. #### Technological Characteristics The technological characteristics of the O-scan systems with the addition of AgilExam and 3D Viewer, reflected in this 510(k), do not alter the scientific technology of the O-Scan systems and are substantially equivalent to those of the predicate devices. The modification are implemented in order to improve display, analysis and comparison of Magnetic Resonance images. ## Summary of Non-Clinical Tests The O-Scan has been evaluated to demonstrate substantial equivalence related to medical electrical equipment, risk management, and software verification and has been found to conform to the following medical device safety standards: - IEC 60601-1 - IEC 60601-1-2 - IEC 60601-1-6 - IEC 60601-2-33 ● - ISO 14971 - ISO 62304 ● - IEC 62366 ● - NEMA MS-1 - NEMA MS-3 ● {5}------------------------------------------------ 510(k) Summary O-scan Esaote S.p.A. ### Summary of Clinical Tests No clinical tests are included within this submission. ### Conclusion The non-clinical testing demonstrates that the O-Scan is as safe, as effective, and performs as well as or better than the predicate. O-Scan is substantially equivalent to the legally marketed devices and conforms to applicable medical device safety and performance standards.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...