K183678 · Abbott Point of Care, Inc. · CGA · Feb 7, 2020 · Clinical Chemistry
Device Facts
Record ID
K183678
Device Name
i-STAT CHEM8+ cartridge with the i-STAT 1 System
Applicant
Abbott Point of Care, Inc.
Product Code
CGA · Clinical Chemistry
Decision Date
Feb 7, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The i-STAT CHEM8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose and creatinine in arterial or venous whole blood in point of care or clinical laboratory settings. Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Device Story
The i-STAT CHEM8+ is a single-use, disposable cartridge used with the handheld i-STAT 1 Analyzer. It accepts 2-3 drops of arterial or venous whole blood. The system performs amperometric measurement of oxidized hydrogen peroxide produced by glucose oxidase (for glucose) and creatinine amidinohydrolase (for creatinine) activity. The analyzer automatically moves fluid across sensors, performs internal quality checks, and provides quantitative results in approximately 2 minutes. Used by trained medical professionals at the point of care or in clinical labs. The output assists clinicians in diagnosing and monitoring metabolic and renal conditions. The system includes accessories like a downloader/recharger, electronic simulator, and printer.
Clinical Evidence
No clinical trials were performed. Evidence consists of analytical bench testing, including precision/reproducibility studies (n=80-320 per analyte), linearity (polynomial regression), interference testing (CLSI EP07), and method comparison against predicate devices (n=180-185 samples). Method comparison showed high correlation (r=0.99-1.00) for both analytes across the reportable ranges.
Technological Characteristics
Single-use disposable cartridge containing sensors, fluid pouch, and waste chamber. Amperometric sensing principle using enzymatic reactions (glucose oxidase for glucose; creatinine amidohydrolase/creatinine amidinohydrolase/sarcosine oxidase for creatinine). Dimensions/materials not specified. Standalone operation with i-STAT 1 Analyzer. Traceable to NIST SRM 967 (creatinine) and SRM 965 (glucose).
Indications for Use
Indicated for in vitro quantification of glucose and creatinine in arterial or venous whole blood for patients requiring diagnosis, monitoring, or treatment of carbohydrate metabolism disorders (e.g., diabetes, hypoglycemia) or renal diseases/dialysis monitoring. For use in point-of-care or clinical laboratory settings by trained professionals.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
SYNCHRON Systems Glucose Reagent on UniCel DxC 600/800 SYNCHRON Clinical System (k042291)
SYNCHRON Systems Creatinine Reagent on UniCel DxC 600/800 SYNCHRON Clinical System (k042291)
Submission Summary (Full Text)
{0}
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
k183678
B Applicant
Abbott Point of Care Inc.
C Proprietary and Established Names
i-STAT CHEM8+ cartridge with the i-STAT 1 System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CGA | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry |
| CGL | Class II | 21 CFR 862.1225 - Creatinine test system | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
Modification of a previously cleared device — modification to the i-STAT CHEM8+ (blue) cartridge run on the i-STAT 1 Analyzer
B Measurand:
Glucose and creatinine
C Type of Test:
Quantitative amperometric assays
K183678 - Page 1 of 15
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The i-STAT CHEM8+ cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of glucose and creatinine in arterial or venous whole blood in point of care or clinical laboratory settings.
Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma.
Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
For point-of-care or clinical laboratory setting
D Special Instrument Requirements:
i-STAT 1 Analyzer
IV Device/System Characteristics:
A Device Description:
The i-STAT 1 System consists of the i-STAT 1 Analyzer and the i-STAT CHEM8+ (blue) cartridges. The system is designed for use by trained medical professionals at the patient point of care or in the clinical laboratory and is for prescription use only.
The i-STAT 1 Analyzer (previously cleared under k103195 as the i-STAT 1 Wireless Analyzer) is a handheld analytical device designed to run i-STAT test cartridges. The instrument interacts with the cartridge to move fluid across the sensors and generate a quantitative result.
The single-use, disposable i-STAT CHEM8+ (blue) cartridge contains test reagents to analyze whole blood at the point of care or in the clinical laboratory for glucose, creatinine, and other analytes. The cartridge format allows all the tests in the cartridge to be performed simultaneously. The cartridges contain the required sensors, a fluid pouch, a sample entry well and closure, fluid channels, waste chamber, and the necessary mechanical features for controlled fluid movement within the cartridge. Cartridges require two to three drops of whole blood which are typically applied to the cartridge using a transfer device, by the trained user before the cartridge is placed within the analyzer.
K183678 - Page 2 of 15
{2}
K183678 - Page 3 of 15
B Principle of Operation:
Creatinine on the i-STAT CHEM8+ (blue) cartridge is measured amperometrically. Creatinine (in the sample) is hydrolyzed to creatine by the enzyme creatinine amidohydrolase. Creatine is then hydrolyzed to sarcosine by the enzyme creatine amidinohydrolase. The oxidation of sarcosine, catalyzed by the enzyme sarcosine oxidase, produces hydrogen peroxide. A voltage is applied to the cartridge electrodes with current generated by the reduction of hydrogen peroxide. The current is directly proportional to the concentration of the creatinine.
Glucose on the i-STAT CHEM8+ (blue) cartridge is measured amperometrically. A voltage is applied to the cartridge electrodes with current generated by the reduction of hydrogen peroxide that was produced by catalytic oxidation of glucose present in the sample by glucose oxidase. The current is directly proportional to the concentration of the glucose.
V Substantial Equivalence Information:
A Predicate Device Name(s):
SYNCHRON Systems Glucose Reagent on UniCel DxC 600/800 SYNCHRON Clinical System; SYNCHRON Systems Creatinine Reagent on UniCel DxC 600/800 SYNCHRON Clinical System
B Predicate 510(k) Number(s):
k042291
C Comparison with Predicate(s):
| Device & Predicate Device(s): | k183678 | k042291 |
| --- | --- | --- |
| Device Trade Name | i-STAT CHEM8+ Cartridge with the i-STAT 1 System (Glucose) | SYNCHRON Systems Glucose Reagent on UniCel DxC 600/800 SYNCHRON Clinical System |
| General Device Characteristic Similarities | k183678 | k042291 |
| Intended Use/Indications For Use | Intended for in vitro quantification of glucose.
Glucose measurements are used in the diagnosis, monitoring, and treatment of carbohydrate metabolism disorders including, but not limited to, diabetes | Same |
{3}
K183678 - Page 4 of 15
| | mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and pancreatic islet cell carcinoma. | |
| --- | --- | --- |
| General Device Characteristic Differences | k183678 | k042291 |
| Reportable Range | 20 - 700 mg/dL | 0 - 600 mg/dL; (serum, plasma) up to 1200 mg/dL by sample dilution |
| Sample Type | Arterial or venous whole blood | Serum, plasma, urine, CSF |
| Sample Volume | 95 μL | 0.5 mL |
| Principle of Measurement | Amperometric | Colorimetric |
| Reagent Format | Cartridge | Reagent handling system, stored within analyzer |
| Device & Predicate Device(s): | k183678 | k042291 |
| --- | --- | --- |
| Device Trade Name | i-STAT CHEM8+ Cartridge with the i-STAT 1 System (Creatinine) | SYNCHRON Systems Creatinine Reagent on UniCel DxC 600/800 SYNCHRON Clinical System |
| General Device Characteristic Similarities | k183678 | k042291 |
| Intended Use/Indications For Use | Intended for in vitro quantification of creatinine.
Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. | Same |
{4}
| General Device Characteristic Differences | k183678 | k042291 |
| --- | --- | --- |
| Reportable Range | 0.2 - 20.0 mg/dL | 0.1 - 25.0 mg/dL (Serum) |
| Sample Type | Arterial or venous whole blood | Serum, urine |
| Sample Volume | 95 μL | 0.5 mL |
| Principle of Measurement | Amperometric | Colorimetric |
| Reagent Format | Cartridge | Reagent handling system, stored within analyzer |
VI Standards/Guidance Documents Referenced:
CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline
CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
CLSI EP07 Interference Testing in Clinical Chemistry, 3rd Edition
CLSI EP 17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Internal site precision
A single site precision study for the creatinine and glucose assays was conducted following the recommendations in CLSI EP05. Five concentration levels of commercially available calibration verification samples were tested using one lot of i-STAT CHEM8+ (blue) cartridges and twelve i-STAT 1 Analyzers. Each sample was measured in duplicates per run, with two runs per day for 20-days resulting in a total of 80 test results per level. The results are summarized below.
Creatinine:
| Level | Mean (mg/dL) | Within-run | | Between Run | | Between Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| L5 | 0.18 | 0.026 | 14.4 | 0.007 | 3.9 | 0.007 | 0.2 | 0.028 | 15.6 |
| L4 | 0.51 | 0.027 | 5.3 | 0.008 | 1.6 | 0.006 | 1.2 | 0.029 | 5.7 |
| L3 | 1.69 | 0.033 | 2.0 | 0.008 | 0.5 | 0.007 | 0.4 | 0.035 | 2.1 |
K183678 - Page 5 of 15
{5}
| Level | Mean (mg/dL) | Within-run | | Between Run | | Between Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| L2 | 4.23 | 0.093 | 2.2 | 0.029 | 0.7 | 0.027 | 0.6 | 0.101 | 2.4 |
| L1 | 15.90 | 0.321 | 2.0 | 0.096 | 0.6 | 0.030 | 0.2 | 0.337 | 2.1 |
Glucose:
| Level | Mean (mg/dL) | Within-run | | Between Run | | Between Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| L1 | 26.7 | 0.47 | 1.8 | 0.13 | 0.5 | 0.10 | 0.4 | 0.49 | 1.8 |
| L2 | 40.9 | 0.46 | 1.1 | 0.16 | 0.4 | 0.18 | 0.4 | 0.52 | 1.3 |
| L3 | 123.0 | 0.43 | 0.3 | 0.13 | 0.1 | 0.12 | 0.1 | 0.47 | 0.4 |
| L4 | 286.5 | 1.25 | 0.4 | 0.55 | 0.2 | 0.33 | 0.1 | 1.40 | 0.5 |
| L5 | 608.1 | 5.21 | 0.9 | 1.46 | 0.2 | 1.31 | 0.2 | 5.56 | 0.9 |
## Point of Care precision – aqueous control material
A three site precision study was performed using a panel of five aqueous control solutions containing different levels of creatinine and glucose. Each sample was assayed at each site once per day for five days across each of six i-STAT 1 Analyzers for a total of 90 measurements. At each site, all testing was conducted by a different operator per site using six lots of i-STAT CHEM8+ (blue) cartridges. The results were analyzed for within-day (between-analyzer), within-site, and overall (summation of between-site, between-day and between-analyzer) variance components and provided in the tables below:
Creatinine:
| Sample | Mean (mg/dL) | Within-day | | Within-site | | Overall | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV |
| Level 1 | 16.16 | 0.362 | 2.2 | 0.390 | 2.4 | 0.422 | 2.6 |
| Level 2 | 4.21 | 0.085 | 2.0 | 0.088 | 2.1 | 0.091 | 2.2 |
| Level 3 | 1.61 | 0.039 | 2.4 | 0.040 | 2.5 | 0.041 | 2.5 |
| Level 4 | 0.60 | 0.020 | 3.3 | 0.021 | 3.5 | 0.021 | 3.5 |
| Level 5 | 0.20 | 0.024 | 12.0 | 0.024 | 12.0 | 0.024 | 12.0 |
Glucose:
| Sample | Mean (mg/dL) | Within-day | | Within-site | | Overall | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV |
| Level 1 | 27.0 | 0.55 | 2.0 | 0.81 | 3.0 | 1.01 | 3.7 |
| Level 2 | 40.5 | 0.68 | 1.7 | 0.82 | 2.0 | 0.91 | 2.2 |
| Level 3 | 124.4 | 0.58 | 0.5 | 0.64 | 0.5 | 0.64 | 0.5 |
| Level 4 | 291.0 | 1.34 | 0.5 | 1.91 | 0.7 | 2.19 | 0.8 |
| Level 5 | 609.6 | 11.30 | 1.9 | 11.85 | 1.9 | 11.85 | 1.9 |
## Point of Care precision – whole blood
A three site precision study was performed using lithium heparin venous whole blood samples targeted to four different concentration levels of creatinine or glucose. For creatinine, a total of 14 native and nine contrived samples were tested using one cartridge lot common to all sites. For glucose, a total of 14 native and four contrived samples were tested using one cartridge lot common to all sites. Each sample was tested three times on each of
K183678 - Page 6 of 15
{6}
seven analyzers on one day for a total of 21 results. Certain samples had fewer replicates due to non-reportable errors, identified by the test system as quality check codes (QCCs) or star outs. A total of 14 i-STAT 1 Analyzers were used per site, and all testing was completed by the same operator at each site for a total of three operators. The results were analyzed for variance from within-analyzer (repeatability) and total (combined within-analyzer and between-analyzer variance components). Results are summarized in the tables below.
Creatinine:
| | | | | Within-analyzer | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Concentration (mg/dL) | Site | N | Mean (mg/dL) | SD | %CV | SD | %CV |
| < 1 | 01 | 21 | 0.82 | 0.038 | 4.6 | 0.044 | 5.4 |
| | 01 | 21 | 0.60 | 0.038 | 6.3 | 0.038 | 6.3 |
| | 01 | 21 | 0.52 | 0.049 | 9.4 | 0.049 | 9.4 |
| | 02 | 20 | 0.96 | 0.049 | 5.1 | 0.051 | 5.3 |
| | 02 | 21 | 0.96 | 0.053 | 5.5 | 0.053 | 5.5 |
| | 02 | 21 | 0.56 | 0.058 | 10.4 | 0.058 | 10.4 |
| | 02 | 21 | 0.87 | 0.044 | 5.1 | 0.049 | 5.6 |
| | 03 | 21 | 0.81 | 0.031 | 3.8 | 0.031 | 3.8 |
| | 03 | 21 | 0.70 | 0.031 | 4.4 | 0.032 | 4.6 |
| | 03 | 21 | 0.59 | 0.031 | 5.3 | 0.031 | 5.3 |
| 1-15 | 03 | 21 | 1.23 | 0.049 | 4.0 | 0.049 | 4.0 |
| | 03 | 21 | 1.17 | 0.049 | 4.2 | 0.049 | 4.2 |
| 1.5-2.0 | 01 | 21 | 1.53 | 0.049 | 3.2 | 0.049 | 3.2 |
| | 02 | 14 | 1.83 | 0.053 | 2.9 | 0.053 | 2.9 |
| | 02 | 21 | 1.97 | 0.062 | 3.1 | 0.062 | 3.1 |
| | 03 | 21 | 1.70 | 0.058 | 3.4 | 0.058 | 3.4 |
| 5.0-7.0 | 01 | 21 | 5.62 | 0.172 | 3.1 | 0.172 | 3.1 |
| | 02 | 21 | 6.31 | 0.246 | 3.9 | 0.246 | 3.9 |
| | 03 | 21 | 5.30 | 0.072 | 1.4 | 0.072 | 1.4 |
| 7.0-12 | 02 | 21 | 9.47 | 0.127 | 1.3 | 0.155 | 1.6 |
| >12 | 01 | 21 | 14.37 | 0.388 | 2.7 | 0.388 | 2.7 |
| | 02 | 21 | 14.90 | 0.515 | 3.5 | 0.515 | 3.5 |
| | 03 | 21 | 14.30 | 0.558 | 3.9 | 0.558 | 3.9 |
Glucose
| | | | | Within-analyzer | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Concentration (mg/dL) | Site | N | Mean (mg/dL) | SD | %CV | SD | %CV |
| 30-110 | 01 | 21 | 95.3 | 0.98 | 1.0 | 0.98 | 1.0 |
| | 01 | 21 | 72.3 | 1.23 | 1.7 | 1.23 | 1.7 |
| | 02 | 21 | 95.1 | 0.90 | 0.9 | 0.90 | 0.9 |
| | 02 | 21 | 95.5 | 0.69 | 0.7 | 0.69 | 0.7 |
| | 02 | 21 | 80.0 | 0.38 | 0.5 | 0.38 | 0.5 |
| | 03 | 21 | 101.3 | 0.76 | 0.8 | 0.76 | 0.8 |
| | 03 | 21 | 87.8 | 0.58 | 0.7 | 0.58 | 0.7 |
K183678 - Page 7 of 15
{7}
| | | | | Within-analyzer | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Concentration (mg/dL) | Site | N | Mean (mg/dL) | SD | %CV | SD | %CV |
| | 03 | 21 | 98.9 | 0.58 | 0.6 | 0.63 | 0.6 |
| 111-150 | 01 | 21 | 148.2 | 0.62 | 0.4 | 0.62 | 0.4 |
| | 02 | 20 | 143.0 | 0.85 | 0.6 | 0.85 | 0.6 |
| | 02 | 14 | 143.3 | 1.25 | 0.9 | 1.25 | 0.9 |
| | 03 | 21 | 142.2 | 0.79 | 0.6 | 0.79 | 0.6 |
| 151-400 | 01 | 21 | 385.7 | 2.38 | 0.6 | 2.98 | 0.8 |
| | 02 | 21 | 318.0 | 3.25 | 1.0 | 3.25 | 1.0 |
| | 03 | 21 | 151.8 | 1.02 | 0.7 | 1.02 | 0.7 |
| 401-700 | 01 | 21 | 618.4 | 7.95 | 1.3 | 7.95 | 1.3 |
| | 02 | 21 | 444.2 | 2.23 | 0.5 | 2.23 | 0.5 |
| | 03 | 21 | 582.0 | 2.82 | 0.5 | 2.93 | 0.5 |
## 2. Linearity:
The linearity of the creatinine and glucose assays on the i-STAT CHEM8+ (blue) cartridge was evaluated following the recommendations in CLSI EP06-A. Lithium heparin venous whole blood was obtained from a healthy subject and altered to produce a high sample pool and a low sample pool. Samples of intermediate concentrations were prepared by intermixing the high and low pools. Each sample was measured in replicates of 15 using five lots of i-STAT CHEM8+ (blue) cartridges. An assessment of linearity was performed using polynomial regression analysis.
For creatinine, regression analysis found that at each level, the deviation from linearity was less than 0.15 mg/dL or 7.5%.
For glucose, regression analysis found that at each level, the deviation from linearity was less than 3 mg/dL or 5%.
Linear regression analysis results for all five lots combined are presented in the tables below.
Creatinine:
| Range tested (mg/dL) | Slope | Intercept | R² |
| --- | --- | --- | --- |
| 0.13 – 21.32 | 0.97 | 0.11 | 1.00 |
Glucose:
| Range tested (mg/dL) | Slope | Intercept | R² |
| --- | --- | --- | --- |
| 16.5 – 821 | 1.00 | -2.14 | 1.00 |
The observed proportional response supports the claim that the creatinine and glucose assays on the i-STAT CHEM8+ (blue) are linear across the measurement ranges of:
Creatinine: 0.2 - 20.0 mg/dL
Glucose: 20 - 700 mg/dL
K183678 - Page 8 of 15
{8}
K183678 - Page 9 of 15
# 3. Analytical Specificity/Interference:
The analytical specificity of the creatinine and glucose assays on the i-STAT CHEM8+ (blue) cartridge was established by conducting interference testing following the recommendations in CLSI EP07 ED3. Interference from certain exogenous and endogenous substances was assessed using lithium heparin venous whole blood samples spiked to two concentrations of low and high: creatinine (0.6 mg/dL, 2.0 mg/dL) and glucose (40 mg/dL, 220 mg/dL). Each low and high sample was further divided into two aliquots: control (with no added interferent) and test (with added interferent). Each sample was measured in replicates of 10 using one lot of the i-STAT CHEM8+ (blue) cartridges. A substance was identified as an interferent if the difference in the mean between the control and test sample was outside of the predefined allowable error:
For Creatinine: ± 0.3 mg/dL up to 2 mg/dL and ± 15% above 2 mg/dL
For Glucose: greater of ± 6 mg/dL or ± 10%
For any substances identified as an interferent, a dose response analysis was performed to assess the highest concentration without significant error, as defined above.
The following table lists the concentrations of each substance at which no significant interference was found.
Creatinine:
| Substance | Highest concentration at which no interference was observed |
| --- | --- |
| Acetaldehyde | 0.2 mg/dL |
| Acetaminophen | 15.6 mg/dL |
| N-Acetyl-L-Cysteine | 15.0 mg/dL |
| L-Ascorbic Acid | 5.25 mg/dL |
| β-Hydroxybutyric Acid | 62.5 mg/dL |
| Bicarbonate | 294 mg/dL |
| Bilirubin | 40 mg/dL |
| Calcium Chloride | 20 mg/dL |
| Creatine | 5.01 mg/dL |
| Dopamine Hydrochloride | 0.0621 mg/dL |
| Formaldehyde | 0.399 mg/dL |
| Glycolic Acid | 76.0 mg/dL |
| Hemoglobin | 1000 mg/dL |
| Lithium Lactate | 90 mg/dL |
| Methyldopa | 2.25 mg/dL |
| pH | 8.0 pH units |
| Lithium Pyruvate | 5 mg/dL |
| Lithium Salicylate | 2.86 mg/dL |
| Sodium Thiosulfate | 264 mg/dL |
| Triglyceride | 1500 mg/dL |
| Uric Acid | 23.5 mg/dL |
{9}
For those substances that on initial screening where found to interfere, dose response testing was conducted to establish the concentration limit below which no significant interference is expected. The results are summarized in the table below:
| Substance | Concentration | Interference |
| --- | --- | --- |
| Lithium bromide | ≥ 159 mg/dL (18.3 mmol/L) | Increased creatinine results |
| Hydroxyurea | ≥ 0.23 mg/dL (0.03 mmol/L) | Increased creatinine results |
The sponsor includes the following statements in the labeling for the device:
Lithium Bromide at ≥ 18.3 mmol/L showed increased creatinine results. Bromide at 2.5 mmol/L is the peak plasma concentration associated with halothane anesthesia, in which bromide is released. Bromide may result in an increased rate of star-outs (***).
Hydroxyurea at ≥ 0.03 mmol/L showed increased creatinine results. Hydroxyurea is a DNA synthesis inhibitor used in the treatment sickle cell anemia, HIV infection, and various types of cancer. The malignancies that it is used to treat include melanoma, metastatic ovarian cancer, and chronic myelogenous leukemia. It is also used in the treatment of polycythemia vera, thrombocythemia, and psoriasis. At typical doses ranging from 500 mg to 2 g/day, concentrations of hydroxyurea in a patient's blood may be sustained at approximately 100 to 500 μmol/L. Higher concentrations may be observed soon after dosing or at higher therapeutic doses.
## Glucose:
The following table lists the concentrations of each substance at which no significant interference was found.
| Substance | Highest concentration at which no interference was observed |
| --- | --- |
| Acetaldehyde | 0.2 mg/dL |
| Acetaminophen | 15.6 mg/dL |
| Lithium Acetoacetate | 20 mg/dL |
| N-Acetyl-L-Cysteine | 15.0 mg/dL |
| Ammonium Chloride | 10.7 mg/dL |
| L-Ascorbic Acid | 5.25 mg/dL |
| Bilirubin | 40 mg/dL |
| Cholesterol | 400 mg/dL |
| Creatinine | 15 mg/dL |
| Dopamine Hydrochloride | 0.0621 mg/dL |
| Ethanol | 600 mg/dL |
| Lithium Fluoride | 0.12 mg/dL |
| Formaldehyde | 0.399 mg/dL |
| Fructose | 18 mg/dL |
| Galactose | 60 mg/dL |
| Gentamicin Sulfate | 3 mg/dL |
| Glucosamine Hydrochloride | 0.647 mg/dL |
| Glutathione, reduced | 3 mEq/L |
K183678 - Page 10 of 15
{10}
| Substance | Highest concentration at which no interference was observed |
| --- | --- |
| Acetaldehyde | 0.2 mg/dL |
| Acetaminophen | 15.6 mg/dL |
| Lithium Acetoacetate | 20 mg/dL |
| N-Acetyl-L-Cysteine | 15.0 mg/dL |
| Glycolic Acid | 76.0 mg/dL |
| Guaifenesin | 0.45 mg/dL |
| Hemoglobin | 1000 mg/dL |
| Sodium Heparin | 330 U/dL |
| β-Hydroxybutyric Acid | 62.5 mg/dL |
| Ibuprofen | 21.9 mg/dL |
| Isoniazid | 6 mg/dL |
| Lithium Lactate | 90 mg/dL |
| Mannose | 18.0 mg/dL |
| Maltose | 360 mg/dL |
| pH | 8.0 pH units |
| Lithium Pyruvate | 5 mg/dL |
| Lithium Salicylate | 2.86 mg/dL |
| Lithium Thiocyanate | 5.22 mg/dL |
| Sodium Thiosulfate | 264 mg/dL |
| Triglyceride | 1500 mg/dL |
| Uric Acid | 23.5 mg/dL |
| Xylose | 45.0 mg/dL |
| Gentisic Acid | 1.50 mg/dL |
For those substances that on initial screening where found to interfere, dose response testing was conducted to establish the concentration limit below which no significant interference is expected. The results are given in the table below:
| Substance | Concentration | Interference |
| --- | --- | --- |
| Lithium bromide | ≥ 102 mg/dL (11.8 mmol/L) | Decreased glucose results |
| Hydroxyurea | ≥ 0.61 mg/dL (0.08 mmol/L) | Increased glucose results |
The sponsor includes the following statements in the labeling for the device:
Lithium Bromide at ≥ 11.8 mmol/L showed decreased results. Bromide at 2.5 mmol/L is the peak plasma concentration associated with halothane anesthesia, in which bromide is released. Bromide may result in an increased rate of star-outs (***).
Hydroxyurea at ≥ 0.08 mmol/L showed increased results. Hydroxyurea is a DNA synthesis inhibitor used in the treatment sickle cell anemia, HIV infection, and various types of cancer. The malignancies that it is used to treat include melanoma, metastatic ovarian cancer, and chronic myelogenous leukemia. It is also used in the treatment of polycythemia vera, thrombocythemia, and psoriasis. At typical doses ranging from 500 mg to 2 g/day, concentrations of hydroxyurea in a patient's blood may be sustained at approximately 100 to
K183678 - Page 11 of 15
{11}
500 μmol/L. Higher concentrations may be observed soon after dosing or at higher therapeutic doses.
## Hematocrit (glucose only):
A study was performed with the i-STAT CHEM8+ (blue) cartridge glucose assay to assess the effect of hematocrit on the reported result. Lithium heparin venous whole blood samples were collected from 8 donors. Across the donors, eight samples were prepared with 4 glucose levels (32-40, 110-118, 255-268, 644-674 mg/dL). Each of samples was further altered to three hematocrit levels of low, mid, and high (15-17%, 43-45%, and 71-75%) for a total of 24 samples. Each sample was measured in replicates of 20 using one lot of i-STAT CHEM8+ (blue) cartridges and one i-STAT Analyzer. The hematocrit effect was assessed per donor by comparing the mean glucose at low hematocrit or high hematocrit to the reference glucose result - which was at the mid-level hematocrit. The results of the study demonstrated that there was no significant difference in glucose results at the extremes of the hematocrit range compared to a mid-level hematocrit.
## Altitude effects
A study was conducted to assess the effect of high altitude on the i-STAT CHEM8+ (blue) cartridge creatinine and glucose assays. The study was performed inside a barometric chamber with pressure set to simulate altitudes of:
2000 m (6562 ft.) for creatinine
3000 m (9842 ft.) for glucose
Six lithium heparin venous whole blood samples (one native and five contrived) were used for creatinine testing and seven heparin whole blood samples (one native and six contrived) were used for glucose testing. For creatinine each sample was run on seven analyzers using one lot cartridges for total of 42 measurements. For glucose each sample was run on six analyzers using one lot of i-STAT CHEM8+ (blue) cartridges for total of 42 measurements. The 42 individual creatinine and glucose measurements were compared to results obtained by running the samples on comparator methods established to be insensitive to effects of altitude. The impact on the performance of the tests at altitude was assessed by Passing-Bablok regression analysis:
| Analyte | Range (mg/dL) | Slope | Intercept | r |
| --- | --- | --- | --- | --- |
| Creatinine | 0.56 – 13.6 | 1.13 | -0.25 | 1.00 |
| Glucose | 21.5 – 683 | 1.00 | -2.26 | 1.00 |
## Oxygen sensitivity
A study was conducted to assess the effect of blood oxygen on the i-STAT CHEM8+ (blue) cartridge for the creatinine and glucose assays. Lithium heparin venous whole blood samples from two donors per analyte were altered to achieve four analyte levels (creatinine: 0.6, 2.5, 8.8, and 15 mg/dL and glucose: 30, 100, 260, and 550 mg/dL). The samples were further divided to two oxygen levels prepared by tonometry; high (>500 mmHg) and low (creatinine: 20-30 mmHg and glucose: 24-26 mmHg). For creatinine, each sample was measured in replicates of 12 on 12 i-STAT 1 Analyzers using one lot of i-STAT CHEM8+ (blue) cartridges. For glucose, each sample was measured in replicates of 12 on 12 i-STAT 1
K183678 - Page 12 of 15
{12}
Analyzers using one lot of i-STAT CHEM8+ (blue) cartridges. The results of the study demonstrated no significant differences in creatinine or glucose results between the high and low oxygen concentrations.
4. Assay Reportable Range:
See section A.2 Linearity
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The creatinine assay on the i-STAT CHEM8+ (blue) cartridges and creatinine values assigned to i-STAT controls and calibration verification materials are traceable to the U.S. National Institute of Standards and Technology (NIST) standard reference material NIST SRM 967.
The glucose assay on the i-STAT CHEM8+ (blue) cartridges and glucose values assigned to i-STAT controls and calibration verification materials are traceable to the U.S. National Institute of Standards and Technology (NIST) standard reference material SRM965.
6. Detection Limit:
Detection capability studies of limit of blank (LoB), limit f detection (LoD), and limit of quantitation (LoQ) for creatinine and glucose assays on the i-STAT CHEM8+ Cartridge with the i-STAT 1 Analyzer were conducted following the recommendations in CLSI EP17-A2.
LoB
The LoB for the i-STAT CHEM8+ (blue) cartridge for the creatinine and glucose assays was evaluated using lithium heparin venous whole blood samples that were collected from a healthy subject, and altered to create a blank sample. The sample was measured in 80 replicates across each of two i-STAT CHEM8+ (blue) cartridge lots for a total of 160 cartridges. The samples were tested over three days for creatinine and four days for glucose. The LoB was calculated using a parametric data analysis for each of the two lots. The higher of the two LoB values from each lot is reported in the labeling as the LoB.
LoD
The LoD for creatinine and glucose was evaluated using lithium heparin venous whole blood samples that were collected from four healthy subjects, and altered to two low concentrations of creatinine and glucose for each donor. Each of the eight samples was measured in 20 replicates per day for three days across each of two i-STAT CHEM8+ (blue) cartridge lots for a total of 320 cartridges. The samples were tested over three days for creatinine and four days for glucose. The LoD was calculated using a parametric data analysis for each of the two lots. The higher of the two LoD values from each lot is reported in the labeling as the LoD.
LoQ
The LoQ for creatinine and glucose was evaluated using lithium heparin venous whole blood samples that were collected on each day from a unique healthy subject. The whole blood sample was altered to four low concentrations for both creatinine and glucose. Each of the four samples was measured in 15 replicates per day for four days across each of two i-STAT
K183678 - Page 13 of 15
{13}
CHEM8+ (blue) cartridge lots for a total of 120 cartridges. The LoQ was calculated for each of the two lots. The sponsor defined LoQ as the greater of the two lots at which the lowest concentration met the pre-defined total error goals; given in the following table.
| Analyte | Total Error goals |
| --- | --- |
| Creatinine | < 0.30 mg/dL |
| Glucose | < 6.00 mg/dL |
The results from all studies are summarized in the table below.
| Analyte | Reportable Range | LoB | LoD | LoQ |
| --- | --- | --- | --- | --- |
| Creatinine | 0.2 - 20.0 mg/dL | 0.05 mg/dL | 0.10 mg/dL | 0.1 mg/dL |
| Glucose | 20 - 700 mg/dL | 1 mg/dL | 2 mg/dL | 12 mg/dL |
7. Assay Cut-Off:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
The accuracy of the creatinine and glucose assays on the i-STAT CHEM8+ (blue) cartridge on the i-STAT 1 Analyzer was evaluated by a method comparison study for agreement with the predicate devices. The study was conducted across two point of care sites.
Creatinine:
A total of 145 lithium heparin venous whole blood specimens and 32 lithium heparin arterial whole blood specimens were tested using one lot of i-STAT CHEM8+ (blue) cartridges. Eight specimens were contrived. The data were analyzed by Passing-Bablok regression analysis comparing the first replicate of the candidate device results to the singlicate result of the predicate device:
| Site | N | Sample Range Tested (mg/dL) | Regression Equation | r |
| --- | --- | --- | --- | --- |
| 1 | 138 | 0.50 - 16.10 | y = 1.050x - 0.060 | 1.000 |
| 2 | 42 | 0.36 - 4.64 | y = 1.000x - 0.055 | 0.990 |
| combined | 180 | 0.36 - 16.10 | y = 1.043x - 0.062 | 1.000 |
Glucose:
A total of 145 lithium heparin venous whole blood specimens and 32 lithium heparin arterial whole blood specimens were tested using one lot of i-STAT CHEM8+ (blue) cartridges. Eight specimens were contrived. Passing-Bablok regression analysis of the total of 185 samples comparing the first replicate of the candidate device results to the singlicate result of the predicate device:
K183678 - Page 14 of 15
{14}
| Site | N | Sample Range Tested (mg/dL) | Regression Equation | r |
| --- | --- | --- | --- | --- |
| 1 | 139 | 30 - 617 | y = 0.993x - 2.18 | 1.000 |
| 2 | 46 | 26 - 312 | y = 0.968 + 5.79 | 0.990 |
| combined | 185 | 26 - 617 | y = 0.98x + 0.00 | 1.000 |
2. Matrix Comparison:
Not applicable. Lithium heparin whole blood is the only acceptable sample type for this device.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
Expected values for the glucose and creatinine assays on the i-STAT CHEM8+ (blue) cartridge are cited from literature*:
| Analyte | Reference Range |
| --- | --- |
| Creatinine | 0.6 - 1.3 mg/dL |
| Glucose (fasting) | 70 - 105 mg/dL |
* Tietz Textbook of Clinical Chemistry and Molecular Diagnostics 4th Edition. CA Burtis, ER Ashwood, DE Bruns, ed., Elsevier Saunders Inc., 2006, page 2264.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K183678 - Page 15 of 15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.