K183621 · Time Medical Limited · LNH · Feb 22, 2019 · Radiology
Device Facts
Record ID
K183621
Device Name
EMMA 1.5T MRI System
Applicant
Time Medical Limited
Product Code
LNH · Radiology
Decision Date
Feb 22, 2019
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The EMMA 1.5T MRI System is indicated for use as a magnetic resonance diagnostic device (MRDD) which produces transverse, sagittal, coronal and oblique cross-sectional images, and those display the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images when interpreted by trained physician yield information that may assist medical diagnosis.
Device Story
The EMMA 1.5T MRI System is a 1.5T superconducting magnet MRI scanner used in clinical settings to produce cross-sectional images of the human body. It consists of a magnet, gradient coils, RF transmission/receive coils, patient table, and a Windows-based workstation running Prodiva software. The system operates by placing the patient in a strong magnetic field, exciting nuclei with RF signals, and acquiring relaxation signals via receiver coils. The spectrometer controls pulse sequences, while the workstation reconstructs raw data into images. Operators (technicians/radiologists) use the system to configure scans and manage data. The resulting images are interpreted by physicians to assist in medical diagnosis, providing anatomical and functional information to benefit patient care.
Clinical Evidence
No formal clinical trials were conducted. Evidence consists of bench testing (biocompatibility per ISO 10993-1, electrical safety/EMC per IEC 60601-1/2-33, acoustic noise per NEMA MS 4-2010, and performance per NEMA MS-1, 2, 3, 5, 6, 8) and a comparison of sample clinical images between the subject and predicate devices to demonstrate consistent diagnostic image quality.
Technological Characteristics
1.5T superconducting magnet system; materials include FRP and ABS. Components: gradient coils, RF transmission/receive coils, spectrometer, and multi-core PC workstation. Connectivity: DICOM compliant. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33, NEMA MS series. Software: Prodiva (Windows-based).
Indications for Use
Indicated for use as a magnetic resonance diagnostic device (MRDD) to produce transverse, sagittal, coronal, and oblique cross-sectional images of the head, body, or extremities for diagnostic purposes. May be used with contrast agents. For use by trained physicians.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is blue.
Time Medical Limited % Mr. Nick Tse Compliance Manager G/F, Bio-Informatics Center No., 2 Science Park West Avenue Shatin, N.T., Hong Kong CHINA
February 22, 2019
Re: K183621
Trade/Device Name: EMMA 1.5T MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: December 12, 2018 Received: December 26, 2018
Dear Mr. Tse:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 2.2 INDICATIONS FOR USE
FDA Indications for Use Form (FDA Form 3881) is shown below and at Annex K.
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | |
| K183621 | |
| Device Name | |
EMMA 1.5T MRI System
#### Indications for Use (Describe)
The EMMA 1.5T MR System is a magnetic resonance diagnostic device (MRDD) which produces transverse, sagittal, coronal and oblique cross-sectional images, and those display the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images when interpreted by trained physician yield information that may assist medical diagnosis.
| Type of Use (Select one or both, as applicable) | | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|--|--|--|
| 2 Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | | |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .* | | | | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | | |
| Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov | | | | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number." | | | | |
FORM FDA 3881 (7/17)
Page 1 of 1
POC Fultidaing Services (301) 443-4345 EF
{3}------------------------------------------------
K183621
#### 2.1 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| Date of Submission: | February, 9, 2018 |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Time Medical Limited<br>G/F Bio-Informatics Centre, No. 2 Science Park West Avenue,<br>Hong Kong Science Park, Shatin, New Territories, Hong Kong, China<br>Contact Person: Nick Tse - Compliance Manager<br>Telephone number: +852 21561711 Fax.: +852 2156 0908 |
| Correspondent: | Time Medical Limited<br>G/F Bio-Informatics Centre, No. 2 Science Park West Avenue,<br>Hong Kong Science Park, Shatin, New Territories, Hong Kong, China<br>Contact Person: Nick Tse - Compliance Manager<br>Telephone number: +852 21561711 Fax.: +852 2156 0908 |
| Filing Device: | EMMA 1.5T MRI System |
| Classification Name: | System. Nuclear Magnetic Resonance Imaging |
| Regulatory Description: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Class | Class II / LNH / 21 CFR 892. 1000 |
| Intended Use: | The EMMA 1.5T MRI System is indicated for use as a magnetic resonance<br>diagnostic device (MRDD) which produces transverse, sagittal, coronal and<br>oblique cross-sectional images, and those display the internal structure<br>and/or function of the head, body, or extremities. Depending on the region of<br>interest, contrast agents may be used. These images when interpreted by<br>trained physician yield information that may assist medical diagnosis. |
| Device Description: | The EMMA 1.5T MRI System is a 1.5T superconducting magnet MRI<br>system which produces transverse, sagittal, coronal and oblique<br>cross-sectional images, and those display the internal structure and/or<br>function of the head, body, or extremities. It is composed of Magnet, Magnet<br>Enclosure, Patient Table, Gradient Coil, Transmission Coil, Receiver Coil,<br>Client PC, and Imaging Cabinet. The system software, Prodiva, a<br>Windows-based software, is an interactive program with user friendly<br>interface. The device is conformed to IEC and DICOM standards. |
#### General Comparison to Predicate Device:
| Comparison Element | Filing Device: | Predicate Device: |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | N/A | K161567 |
| Model | 15000-01 | SIGNA Voyager 1.5T |
| Device Name | EMMA 1.5T MRI System | SIGNA Voyager |
| Applicant | Time Medical Limited | GE Healthcare |
| Applicant Address | Unit G01-G03, 05 & 08, G/F,<br>Bio-Informatics Centre, Hong Kong<br>Science Park, Shatin, N.T | 3200 N. Grandview Blvd.<br>Waukesha, WI 53188, USA |
| Classification Name | System, Nuclear Magnetic<br>Resonance Imaging | Magnetic Resonance Diagnostic<br>Device |
| Product Code | LNH | LNH, MOS |
| Regulation Number | 21 CFR 892.1000 | |
| Panel | Radiology | |
| Class | Class II | |
| Indications for Use | The EMMA 1.5T MRI System is<br>indicated for use as a magnetic<br>resonance diagnostic device<br>(MRDD) which produces<br>transverse, sagittal, coronal and<br>oblique cross-sectional images,<br>and those display the internal<br>structure and/or function of the | The SIGNA Voyager is a whole body<br>magnetic resonance scanner<br>designed to support high resolution,<br>high signal-to-noise ratio, and short<br>scan time imaging. The SIGNA<br>Voyager is indicated for use as a<br>diagnostic imaging device to produce<br>axial, sagittal, coronal, and oblique |
{4}------------------------------------------------
The filing device has the same US FDA classification including classification name, product code, regulation number, panel, and class with the predicate device.
The filing device has the same major indications for use as the predicate device produces transverse, sagittal, coronal and oblique cross-sectional images, and those display the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images when interpreted by trained physician yield information that may assist medical diagnosis.
The minor differences in indications for use do not constitute any safety and effectiveness issue, as indicated in performance data provided.
| General | Filing Device | Predicate Device |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| Design | Cylindrical | Cylindrical |
| Materials | FRP, ABS | |
| Power Source | Isolated power supply | |
| Performance Data | Comply with:<br>- NEMA MS standards<br>- IEC standards: IEC 60601-1,<br>IEC60601-1-2,<br>IEC60601-2-33 | |
| Magnet | Filing Device | Predicate Device |
| Type | Superconducting | |
| Field Strength | 1.5T | |
| Weight | 3,200 kg | 3,900 kg |
| Overall Dimensions | 1497 x 1880 x 2421 mm<br>(L x W x H) | 1730 x 2070 x 2360 mm<br>(L x W x H) |
| Cryogen Type | Liquid Helium | |
| Spatial Homogeneity | 1.45 ppm at 40 cm DSV | 0.40 ppm at 40 cm DSV |
| Fringe Field | 5 gauss line < 2.5m radial x 4m<br>axial | 2.5 m radial x 4.0 m axial |
| Magnet Room Lighting | For sufficient and even lighting, 16<br>overhead lamps (100 lumen each)<br>recommended | Available |
| Room Ventilation | HVAC is installed for ventilation. | |
| Gradient System | Filing Device | Predicate Device |
| Amplitude | 43.6 mT/m | 36 mT/m |
| Rise time | 0.17 ms | 0.24 ms |
| Slew rate | 257 T/m/s | 150 T/m/s |
| Patient Table | Filing Device | Predicate Device |
| Dimensions (L x W x H) | 92.1 x 24.6 x 34.0 inch<br>(2340 x 624 x 864 mm) | 96.5 x 27.6 x 36.7 inch<br>(2450 x 700 x 932 mm) |
| Patient positioning | Laser Localizer assisted automatic<br>patient table positioning.<br>Manual Position mode | Laser Localizer Available |
| Max Patient Weight | 200kg | 250 kg |
| Patient-control room<br>communication | Microphones and speaker for<br>two-way communication between<br>patient and radiologist | Magnacoustic Communication &<br>Music System |
| Workstation | Filing Device | Predicate Device |
| Model | Multi-Processing, Multi-Core PC<br>with 24-inch monitor | Quad-Core Intel Xeon with 24-inch<br>monitor |
| Memory and Capacity | 64GB of RAM, 250GB SSD Storage | 32GB of RAM, 2 x 400 GB SSD<br>Storage |
| Operating System | Microsoft Windows | Linux OS |
| Receive Coils | Filing Device | Predicate Device |
| Multi-Channel Head Coil | ✓ | ✓ |
| Multi-Channel Head and<br>neck Coil | ✓ | ✓ |
| Multi-Channel Shoulder<br>Coil | ✓ | ✓ |
| Multi-Channel Spine Coil | ✓ | ✓ |
| Multi-Channel Body Coil | ✓ | ✓ |
| Multi-Channel Closed<br>Breast Coil | ✓ | ✓ |
| Multi-Channel Knee Coil | ✓ | ✓ |
| Multi-Channel Ankle Coil | ✓ | ✓ |
| Multi-Channel Wrist Coil | ✓ | ✓ |
| Multi-Channel<br>Multi-purpose Coils | ✓ | ✓ |
| Pulse Sequences | Filing Device | Predicate Device |
| T1 Spin Echo | ✓ | ✓ |
| T1, T2 Fast Spin Echo,<br>HASTE | ✓ | ✓ |
| T1, T2* Gradient Echo | ✓ | ✓ |
| FLAIR | ✓ | ✓ |
| TOF Angiography | ✓ | ✓ |
| Inversion Recovery | ✓ | ✓ |
| Fat-Water Separation | ✓ | ✓ |
| Chemical Fat Sat | ✓ | ✓ |
| MRCP | ✓ | ✓ |
| DWI EPI | ✓ | ✓ |
| CMGE | ✓ | ✓ |
| SWI | ✓ | ✓ |
| 2D Image Processing<br>Features | Filing Device | Predicate Device |
| Image panning, zooming,<br>windowing | ✓ | ✓ |
| Image flipping, inverting<br>and rotation | ✓ | ✓ |
| Statistical Analysis | ✓ | ✓ |
| Geometric Measurement | ✓ | ✓ |
| Phase Image | ✓ | ✓ |
| CINE | ✓ | ✓ |
| Shutter | ✓ | ✓ |
| Image Filtering | ✓ | ✓ |
| Image Subtraction | ✓ | ✓ |
| Slice Profile | ✓ | ✓ |
| Histogram | ✓ | ✓ |
| 3D Image Processing<br>(MPR/Max.IP) | Filing Device | Predicate Device |
| Image panning, zooming,<br>windowing | ✓ | ✓ |
| Cutting Operation Output<br>Stack | ✓ | ✓ |
| Image export | ✓ | ✓ |
| The filing device basically utilizes same technologies as predicate device, and can give consistent<br>results when using the images in medical diagnosis. The minor differences in technological<br>characteristics do not constitute any safety and effectiveness issue, as indicated in performance data<br>provided. | | |
#### Comparison of Technological Characteristics with the Predicate Device:
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
| Technological<br>Characteristic | EMMA 1.5T MRI system is based on the principle that certain atomic nuclei<br>present in the human body, when placed in a strong magnetic field and<br>excited by a radio frequency signal at the precession frequency, will emit a<br>relaxation signal. The emitted relaxation signals are acquired by signal<br>receiving coils and are analyzed by the system. Human-interpretable<br>images are reconstructed by and stored in the computer and displayed on<br>the screen. |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The principle components of EMMA 1.5T MRI system includes |
| | Superconducting magnet with corresponding cooling devices Gradient system RF transmission and receive system Cooling system for different components of the system Patient table and support components Spectrometer Server/Client PC and application software |
| | The superconducting magnet is made of coils of superconducting wires<br>and is cooled to and maintained at cryogenic temperature by an enclosed<br>dedicated cooling system. When energized, current flows continuously<br>inside the superconducting wires as a closed loop; thus producing a<br>constant magnetic field. This has the same major technological<br>characteristics as the predicate device. |
| | The gradient system consists of a gradient coil and the gradient amplifier.<br>The gradient coil produces varying linear gradient fields in the three<br>orthogonal spatial axes with currents given by the gradient amplifier. The<br>gradient coil can also produce various non-linear gradients fields to<br>improve the homogeneity of the main magnetic field. This has the same<br>major technological characteristics as the predicate device. |
| | The RF transmission and receive system consists of the RF amplifier, the<br>RF transmission coil and various RF receiver coils for receiving the<br>relaxation signals emitted from the subject after RF transmission. They are<br>coils made of copper wires tuned to transmit and receive around the<br>designed frequency corresponding to the magnetic field strength. This has<br>the same major technological characteristics as the predicate device. |
| | The cooling system maintains suitable operating temperatures for<br>amplifiers and other system devices. This has the same major<br>technological characteristics as the predicate device. |
| | The patient table and support components provide a comfortable platform<br>for the patient and allows the operator to easily position the patient to the<br>location for imaging. This has the same major technological characteristics<br>as the predicate device. |
| | The spectrometer controls the execution of the pulse sequence for RF<br>transmit and receive and gradient output. This has the same major<br>technological characteristics as the predicate device. |
| | The server/client PC and the application software provides the operator a<br>platform to configure scans and manage scanned data. Raw data received<br>from the spectrometer is also reconstructed here. This has the same major<br>technological characteristics as the predicate device. |
| | technological characteristics as the predicate device. |
| | Therefore, the same scientific technology theory is applied to both the<br>EMMA 1.5T MRI system and predicate device. |
| | Any minor differences in physical attributes do not constitute any safety and<br>effectiveness issue, as indicated in performance data provided. |
{8}------------------------------------------------
{9}------------------------------------------------
| Nonclinical Testing<br>Summary: | Like the predicate device, the following performance data were provided in<br>support of the substantial equivalence determination. |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Biocompatibility testing<br>The biocompatibility evaluation for the filing device was conducted in<br>accordance with ISO 10993-1 "Biological Evaluation of Medical Devices –<br>Part 1: Evaluation and Testing Within a Risk Management Process" as<br>recognized by FDA. |
| | Electrical safety and electromagnetic compatibility (EMC)<br>Electrical safety and EMC testing were conducted in accordance with AAMI<br>/ ANSI ES60601-1:2005/(R)2012 And A1:2012 "Medical Electrical<br>Equipment - Part 1: General Requirements For Basic Safety And Essential<br>Performance", IEC 60601-2-33:2010+A1:2013 "Medical Electrical<br>Equipment - Part 2-33: Particular Requirements For The Basic Safety And<br>Essential Performance Of Magnetic Resonance Equipment For Medical<br>Diagnostic for safety" and IEC 60601-1-2:2014 Edition 4 "Medical Electrical<br>Equipment - Part 1-2: General Requirements For Basic Safety And<br>Essential Performance - Collateral Standard: Electromagnetic<br>Disturbances - Requirements And Tests" for EMC as recognized by FDA |
| | Surface Heating of RF Receive Coils<br>Surface heating was tested with both coil-plugged and coil-unplugged<br>configurations according to AAMI/ANSI ES60601-1. The measured<br>temperature of the surface of the coil never exceeds the maximum limit of<br>41°C in either configuration. |
| | Software Verification and Validation Testing<br>Software verification and validation testing were conducted and<br>documentation was provided as recommended by FDA's Guidance for<br>Industry and FDA Staff, "Guidance for the Content of Premarket<br>Submissions for Software Contained in Medical Devices." |
| | Acoustic testing<br>Acoustic testing were conducted in accordance with NEMA MS 4-2010<br>"Acoustic Noise Measurement Procedure for Diagnosing Magnetic<br>Resonance Imaging Devices" |
| | Performance testing - Bench<br>The filing device complies with the following NEMA MS standard as does the<br>predicate device:<br>• NEMA MS-1-2008 (R2014) - Determination of Signal-to-Noise Ratio<br>(SNR) in Diagnostic Magnetic Resonance Images<br>• NEMA MS 2-2008 (R2014) - Determination of Two-Dimensional<br>Geometric Distortion in Diagnostic Magnetic Resonance Images<br>• NEMA MS 3-2008 (R2014) - Determination of Image Uniformity in<br>Diagnostic Magnetic Resonance Images<br>• NEMA MS 5-2010 - Determination of Slice Thickness in Diagnostic<br>Magnetic Resonance Imaging<br>• NEMA MS 6-2008 (R2014) - Determination of Signal-to-Noise Ratio<br>and Image Uniformity for Single-Channel Non-Volume Coils in<br>Diagnostic MR Imaging<br>• NEMA MS 8-2016 - Characterization Of The Specific Absorption<br>Rate For Magnetic Resonance Imaging Systems<br>The performance testing results demonstrate the safety and performance<br>as expected. Therefore, the filing device is substantially equivalent to the<br>legally marketed predicate device. |
{10}------------------------------------------------
| Clinical Testing<br>Summary: | Sample clinical images are provided to verify the claim of filing device's<br>capability in generating images for diagnostic purposes. Sample clinical<br>image sets from filing device and predicate device on same pulse<br>sequences are provided to demonstrate the substantial equivalence of filing<br>device to legally marketed predicate device. |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusions Drawn<br>from Non-Clinical<br>Information and<br>Clinical Images: | The non-clinical information (safety and performance tests), combined with<br>clinical images (both from filing device and predicate device), demonstrate<br>that the filing device is as safe, as effective, and performs as well as the<br>predicate device. |
| SE Conclusion: | | The filing device has the same intended use as the predicate device. The<br>minor differences in indication for use do not constitute any safety and<br>effectiveness issue, as indicated in performance data provided. |
|----------------|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | The filing device utilizes same technologies as predicate device, and gives<br>consistent results (images) in medical diagnosis. Any minor differences in<br>physical attributes do not constitute any safety and effectiveness issue, as<br>indicated in performance data [non clinical/clinical] provided. |
| | | Therefore the filing device is demonstrated, for being substantially<br>equivalent to the legally marketed predicate device. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.