← Product Code [PQK](/productcode/PQK) · K183615

# Ellipsys Vascular Access System (Ellipsys System), Power Controller, Ellipsys Vascular Access System (Ellipsys System), Catheter, Ellipsys Vascular Access System (Ellipsys System), Crossing Needle (K183615)

_Avenu Medical, Inc. · PQK · Jan 25, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K183615

## Device Facts

- **Applicant:** Avenu Medical, Inc.
- **Product Code:** [PQK](/productcode/PQK.md)
- **Decision Date:** Jan 25, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1252
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Ellipsys System is indicated for the creation of a proximal radial artery to perforating vein anastomosis via a retrograde venous access approach in patients with a minimum vessel diameter of 2.0 mm and less than 1.5 mm of separation between the artery and vein at the fistula creation site who have chronic kidney disease requiring dialysis.

## Device Story

Catheter-based system for percutaneous creation of arteriovenous (AV) fistula; used in surgical/radiological suites by physicians. System comprises Power Controller (AMI-1001), Crossing Needle (AMI-3000), and Catheter (AMI-6005). Procedure involves retrograde venous access; needle crosses from vein to artery; catheter approximates vessel walls. Power Controller provides DC thermal energy to catheter tip to fuse vessels. Controller uses visual prompts on LCD to guide user and monitors tip closure. System enables minimally invasive AV fistula creation for hemodialysis, potentially reducing surgical trauma compared to traditional open access.

## Clinical Evidence

No clinical data provided; substantial equivalence established via predicate device data. Bench testing, animal testing, EMC/electrical safety, software validation, biocompatibility, and sterilization data from the predicate were adopted.

## Technological Characteristics

Endovascular catheter system; DC thermal energy source. Materials: ABS, Stainless Steel, Polyimide, Parylene. Connectivity: Standalone Power Controller with LCD GUI. Sterilization: EtO (SAL 10^-6). Software: C++ based, PID control algorithm. Dimensions: Compatible with .014" guidewires and 6 French introducer sheaths.

## Regulatory Identification

This device is a single use percutaneous catheter system that creates an arteriovenous fistula in the arm of patients with chronic kidney disease who need hemodialysis.

## Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must evaluate:
(i) The ability to safely deliver, deploy, and remove the device;
(ii) The ability of the device to create an arteriovenous fistula;
(iii) The ability of the arteriovenous fistula to attain a blood flow rate and diameter suitable for hemodialysis;
(iv) The ability of the fistula to be used for vascular access for hemodialysis;
(v) The patency of the fistula; and
(vi) The rates and types of all adverse events.
(2) Animal testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be assessed:
(i) Delivery, deployment, and retrieval of the device;
(ii) Compatibility with other devices labeled for use with the device;
(iii) Patency of the fistula;
(iv) Characterization of blood flow at the time of the fistula creation procedure and at chronic followup; and
(v) Gross pathology and histopathology assessing vascular injury and downstream embolization.
(3) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Simulated-use testing in a clinically relevant bench anatomic model to assess the delivery, deployment, activation, and retrieval of the device;
(ii) Tensile strengths of joints and components;
(iii) Accurate positioning and alignment of the device to achieve fistula creation; and
(iv) Characterization and verification of all dimensions.
(4) Electrical performance, electrical safety, and electromagnetic compatibility (EMC) testing must be performed for devices with electrical components.
(5) Software verification, validation, and hazard analysis must be performed for devices that use software.
(6) All patient-contacting components of the device must be demonstrated to be biocompatible.
(7) Performance data must demonstrate the sterility of the device components intended to be provided sterile.
(8) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(9) Labeling for the device must include:
(i) Instructions for use;
(ii) Identification of system components and compatible devices;
(iii) Expertise needed for the safe use of the device;
(iv) A detailed summary of the clinical testing conducted and the patient population studied; and
(v) A shelf life and storage conditions.

## Predicate Devices

- Ellipsys Vascular Access System ([K181725](/device/K181725.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.

January 25, 2019

Avenu Medical, Inc. Dave Campbell VP of Quality Assurance & Regulatory Affairs 27123 Calle Arroyo. Suite 2101 San Juan Capistrano, California 92675

#### Re: K183615

Trade/Device Name: Ellipsys Vascular Access System (Ellipsys System), (Power Controller Model No. AMI-1001, Catheter Model No. AMI-6005, and Crossing Needle Model No. AMI-3000) Regulation Number: 21 CFR 870.1252 Regulation Name: Percutaneous Catheter for Creation of an Arteriovenous Fistula for Hemodialysis Access Regulatory Class: Class II Product Code: POK Dated: December 19, 2018 Received: December 26, 2018

Dear Dave Campbell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brian D. Pullin -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K183615

#### Device Name

The Ellipsys Vascular Access System (Ellipsys System)

(Power Controller, Model No. AMI-1001; Catheter, Model No. AMI-6005, & Crossing Needle, Model No. AMI-3000)

#### Indications for Use (Describe)

The Ellipsys System is indicated for the creation of a proximal radial artery to perforating vein anastomosis via a retrograde venous access approach in patients with a minimum vessel diameter of 2.0 mm and less than 1.5 mm of separation between the artery and vein at the fistula creation site who have chronic kidney disease requiring dialysis.

Type of Use (Select one or both, as applicable):

|  | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
|  | Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

# 510(k) Summary

#### I. SUBMITTER

Avenu Medical, Inc. 27123 Calle Arroyo, Suite 2101 San Juan Capistrano, CA 92675

Phone: 949-878-0886 FAX: 949-276-2493

Contact Person: Dave Campbell Date Prepared: January 21, 2019

#### DEVICE II.

Name of Device: Ellipsys Vascular Access System (Ellipsys® System), Model AMI-1001, Model AMI-6005 and AMI-3000

Common or Usual Name: Percutaneous catheter for creation of an arteriovenous fistula for hemodialysis access

Regulatory Class: II Product Code: PQK Regulation Number: 21 CFR 870.1252

#### III. PREDICATE DEVICE

Name of Device: Ellipsys Vascular Access System (Ellipsys® System), Model AMI-1001, Model AMI-6005 510(k) Number: K181725 Regulation Number: 21 CFR 870.1252

This predicate has not been subject to a recall.

## IV. DEVICE DESCRIPTION

The device that is the subject of this 510(k) is a modified Ellipsys Vascular Access System. The Ellipsys System recently gained market clearance through the 510(k) pathway (K181725). This submission describes the modification to the Ellipsys Vascular Access Kit (AMI-6005) only, which consists of the Ellipsys Crossing Needle and Catheter. Specifically, the Crossing Needle and Catheter are being separated and individually packaged using the same packaging materials and processes. The catheter and crossing needle are supplied sterile (EtO).

{4}------------------------------------------------

The modified Ellipsys System remains a catheter-based system that is used to percutaneously create a vascular anastomosis of the proximal radial artery and adjacent vein using direct current (DC) thermal heating. The Ellipsys Power Controller (AMI-1001) is software driven and guides the user through the procedure using visual prompts via an LCD display. The Controller monitors the closure of the catheter tip, and supplies controlled DC energy to the catheter's heating element. The Ellipsys Crossing Needle (AMI-3000) and Ellipsys Catheter (AMI-6005) are sterile single-use disposable devices that are responsible for approximating (bringing together) the arterial and venous vessel walls and applying thermal energy to create an anastomosis and join the two target vessels. Both the crossing needle and catheter are designed to be compatible with .014" guidewires and 6 French introducer sheaths commonly used with vascular interventional procedures. The system is designed to be used in a surgical or radiological suite or office based procedure room.

#### INDICATIONS FOR USE V.

The Ellipsys® System is indicated for the creation of a proximal radial artery to perforating vein anastomosis via a retrograde venous access approach in patients with a minimum vessel diameter of 2.0mm and less than 1.5mm of separation between the artery and vein at the fistula creation site who have chronic kidney disease requiring dialysis.

## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Ellipsys Vascular Access System described and cleared by FDA in 510(k) K181725, serves as the predicate for the modified device that is the subject of this 510(k). Both the modified device and the predicate device are identical in design and intended use. The technological principle for both the modified Ellipsys System and the predicate device is endovascular creation of an AV fistula. Both systems utilize ultrasonically guided endovascular techniques and instrumentation for approximating (bringing together) the arterial and venous vessel walls and applying DC thermal energy to join the target vessels creating a side by side anastomosis and thereby creating an AV fistula for dialysis access.

The modified and predicate devices are based on the same technological elements as described below:

## Power Controller, Model #AMI-1001

- Software controlled device that guides the user through the procedure using visual ● prompts via a graphical user interface, monitors the closure of the catheter tip, and supplies controlled DC energy to the catheter's heating element.
## Ellipsvs Catheter, Model # AMI-6005

- The Catheter is inserted over the guidewire through the introducer sheath so the distal tip ● is inside the artery and the proximal portion of the tip remains in the vein. The catheter mechanically captures and approximates the vessel walls.
- The Catheter seals the walls of the proximal radial artery and the adjacent vein creating an arteriovenous fistula utilizing DC thermal energy delivered by the Power Controller.

{5}------------------------------------------------

#### Ellipsys Crossing Needle, Model # AMI-3000

- The Crossing Needle is inserted into the vasculature over a guidewire through an ● introducer sheath to cross from the vein into the adjacent artery.
The following provides a technological comparison between the modified and predicate devices: The Power Controller (AMI-1001) remains unchanged from the predicate. No changes have been made to the hardware, software, power source, control algorithm or thermal profile.

The Vascular Access Kit containing the Crossing Needle and Catheter has been modified to separate the Crossing Needle and Catheter and provide each device in individual sterile packages. The labeling will change to reflect the new model number for the Crossing Needle. There are no other changes to the Crossing Needle or Catheter and both are identical to the predicate. There are no changes to the basic design, patient contact materials, packaging materials or function of the modified Crossing Needle and Catheter aside from separating them into individual packages.

A comparison of the characteristics of the modified Ellipsys System versus the predicate is provided in TABLE 6.1 below.

| Characteristic                   | Modified Device<br>(AMI-1001/AMI-<br>6005/AMI-3000) | Predicate Device<br>(AMI-1000/AMI-6050)                                                                                                                                                                                                                                                                                                                                      | Comparison |
|----------------------------------|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indication for Use               | No change                                           | The Ellipsys System is indicated for the<br>creation of a proximal radial artery to<br>perforating vein anastomosis via a retrograde<br>venous access approach in patients with a<br>minimum vessel diameter of 2.0mm and less<br>than 1.5mm of separation between the artery<br>and vein at the fistula creation site who have<br>chronic kidney disease requiring dialysis | Identical  |
| Power Controller                 |                                                     |                                                                                                                                                                                                                                                                                                                                                                              |            |
| Design:<br>Hardware/Electronics  | No change                                           | Existing                                                                                                                                                                                                                                                                                                                                                                     | Identical  |
| Energy Source & Type             | No change                                           | DC thermal energy                                                                                                                                                                                                                                                                                                                                                            | Identical  |
| System Software                  | No change                                           | System software (C++ code)                                                                                                                                                                                                                                                                                                                                                   | Identical  |
| Temperature Control<br>Algorithm | No change                                           | PID control                                                                                                                                                                                                                                                                                                                                                                  | Identical  |
| Temperature Profile              | No change                                           | Maximum of 7 cycles (2 sec @ 700°F followed<br>by 8 sec cooling phase)                                                                                                                                                                                                                                                                                                       | Identical  |
| Catheter Compatibility           | No change                                           | AMI-6005                                                                                                                                                                                                                                                                                                                                                                     | Identical  |
| Crossing Needle                  |                                                     |                                                                                                                                                                                                                                                                                                                                                                              |            |
| Design                           | No change                                           | Existing                                                                                                                                                                                                                                                                                                                                                                     | Identical  |
| Materials - Patient Contact      | No change                                           | ABS/Stainless Steel (SS)                                                                                                                                                                                                                                                                                                                                                     | Identical  |

#### TABLE 6.1 - Device Comparison

{6}------------------------------------------------

| Packaging Configuration     | Crossing Needle<br>individually packaged<br>in a separate Tyvek<br>pouch w/ backer card | Crossing Needle & Catheter packaged together<br>in a Tyvek pouch w/ backer card                                                                                                                                                          | Similar   |
|-----------------------------|-----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Catheter                    |                                                                                         |                                                                                                                                                                                                                                          |           |
| Function & Operation        | No change                                                                               | Function: Approximates (brings together) the<br>target arterial and venous vessel walls and<br>applies thermal energy to create an anastomosis.<br>Operation: Catheter Handle & Thumb Tab<br>control insertion and Actuation of the tip. | Identical |
| Materials - Patient Contact | No change                                                                               | ABS/Polyimide/Parylene/SS                                                                                                                                                                                                                | Identical |
| Catheter Design             | No change                                                                               | Existing                                                                                                                                                                                                                                 | Identical |
| Packaging Configuration     | Catheter individually<br>packaged in a<br>separate Tyvek Pouch<br>w/ backer card        | Crossing Needle & Catheter packaged together<br>in a Tyvek Pouch w/ backer card                                                                                                                                                          | Similar   |

The minor packaging configuration difference between the modified Ellipsys System and the predicate device does not raise different questions of safety and effectiveness and the performance data described in Section VII establishes that the modified device is substantially equivalent to the predicate device.

#### VII. PERFORMANCE DATA

The following performance data were provided in support of the special controls requirements of 21 CFR 870.1252 and the substantial equivalence determination.

#### 1) Clinical Performance Testing

Clinical data is not required in support of the special controls requirement and substantial equivalence since the intended use, technological characteristics, design and function of the Power Controller, Crossing Needle, and Catheter remain unchanged from the predicate.

#### 2) Animal Testing

The intended use, technological characteristics, design, function, and manufacturing processes of the Power Controller, Crossing Needle, and Catheter remain unchanged from the predicate. Therefore, the animal test data from the predicate has been adopted in support of the special controls requirement and substantial equivalence.

#### 3) Non-Clinical Performance Testing

The technological characteristics, design, function and the manufacturing processes of the Power Controller, Crossing Needle and Catheter remain unchanged from the predicate. Therefore, the previously validated non-clinical performance test data for the predicate has been adopted in support of the special control requirement and substantial equivalence.

{7}------------------------------------------------

## 4) Electromagnetic Compatibility (EMC) and Electrical Safety Testing

The technological characteristics, design, function and the manufacturing processes of the Power Controller, Crossing Needle and Catheter remain unchanged from the predicate. Therefore, the Electromagnetic Compatibility (EMC) and Electrical Safety test data from the predicate has been adopted in support of the special controls requirements and substantial equivalence.

## 5) Software Verification and Validation Testing

No new software was incorporated into the modified device; therefore, the software verification, validation and hazard analysis test data from the predicate has been adopted to support the software special control and substantial equivalence requirements.

## 6) Biocompatibility of Patient-Contacting Components

No new patient-contacting components are incorporated into the modified device; therefore, the biocompatibility test data from the predicate has been adopted to support the biocompatibility special control and substantial equivalence.

## 7) Sterilization Performance

The construction, packaging and configuration of the modified product was evaluated against the predicate and was determined to pose no additional challenge to the existing EO sterilization process (SAL 106) and therefore has been adopted for the modified device in support of the special control for sterility and substantial equivalence.

## 8) Shelf Life Performance

The construction and packaging of the modified product is identical to the predicate and does not pose any additional challenge to the packaging system; therefore, the existing package shelf life validation has been adopted in support of the special control for shelf life performance and substantial equivalence.

## 9) Device Labeling

Labeling for the device includes the following characteristics and is supplied as part of this application:

- (i) Instructions for use;
- (ii) Identification of system components and compatible devices;
- (iii) Expertise needed for the safe use of the device;
- (iv) A detailed summary of the clinical testing conducted and the patient population studied; and
- Shelf life and storage conditions. (v)

## VIII. CONCLUSIONS

Based on a comparison of the intended use and an assessment of the technological characteristics related to the minor packaging configuration change, it has been established that the modified Ellipsys System is substantially equivalent to the legally marketed predicate device. The minor difference in the packaging configuration of the Crossing Needle and Catheter does not raise new or different questions of safety and effectiveness. Furthermore, the information provided in this 510(k) demonstrates that the modified device is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K183615](https://fda-staging.innolitics.com/device/K183615)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
