Renuvion/J-Plasma Precise Handpiece, Renuvion/J-Plasma Precise Open handpieces, Renuvion/J-Plasma Precise Open handpieces
K183610 · Bovie Medical Corporation · GEI · Mar 12, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183610
Device Name
Renuvion/J-Plasma Precise Handpiece, Renuvion/J-Plasma Precise Open handpieces, Renuvion/J-Plasma Precise Open handpieces
Applicant
Bovie Medical Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Renuvion®/J-Plasma® Precise Handpiece is intended to be used with compatible electrosurgical generators for the delivery of helium gas plasma for cutting, coagulation and ablation of soft tissue during open and laparoscopic surgical procedures. The Renuvion®/J-Plasma® Precise Open Handpiece is intended to be used with compatible electrosurgical generators for the delivery of radiofrequency energy and/or helium plasma for cutting, coagulation and ablation of soft tissue during open surgical procedures. The Renuvion®/J-Plasma® Precise Handpiece is compatible with the Electrosurgical Generators BVX-200H and BVX-200P.
Device Story
Sterile, single-use electrosurgical handpieces; used with compatible generators (BVX-200H/BVX-200P). Input: helium gas and electrical power (up to 40W). Transformation: generator ionizes helium gas into plasma via electrode tip; delivers RF energy and/or helium plasma to tissue. Output: plasma/RF energy for cutting, coagulation, and ablation. Used in OR by surgeons for open or laparoscopic procedures. Features retractable cutting blade tip. Benefits: precise tissue management. Design modifications from predicates include improved plug connector with pneumatic seal before electrical connection and reduced insertion/extraction force.
Clinical Evidence
Bench testing only. Includes mechanical/electrical verification (pneumatic testing, electrical continuity, cable flexure/tension, dielectric strength, mold stress, drop test, insertion/extraction integrity), usability evaluation, and biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, material-mediated pyrogen).
Technological Characteristics
Electrosurgical accessory; helium plasma and RF energy delivery. Materials: biocompatible components (flow tube coupler, O-ring). Dimensions: 15cm, 27cm, 33cm (Precise); 44mm, 150mm (Precise Open). Power: up to 40W. Gas flow: 1-5 LPM. Connectivity: pneumatic/electrical connector to compatible generators. Sterilization: sterile, single-use.
Indications for Use
Indicated for soft tissue cutting, coagulation, and ablation during open and laparoscopic surgical procedures using helium plasma and/or radiofrequency energy.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 12, 2019
Bovie Medical Corporation Topaz Kirlew Executive Director of OA & RA 5115 Ulmerton Road Clearwater, Florida 33760
Re: K183610
Trade/Device Name: Renuvion/J-Plasma Precise Handpiece and Renuvion/J-Plasma Precise Open Handpieces Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 11, 2019 Received: January 15, 2019
Dear Dr. Topaz Kirlew:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Long Long H. H. Chen -S Date: 2019.03.12 Chen -S 13:26:07 -04'00' for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183610
#### Device Name
The Renuvion®/J-Plasma® Precise Handpiece and The Renuvion®/J-Plasma® Precise Open Handpiece
Indications for Use (Describe)
The Renuvion®/J-Plasma® Precise Handpiece is intended to be used with compatible J-Plasma electrosurgical generators for the delivery of helium plasma for cutting, coagulation of soft tissue during open and laparoscopic surgical procedures.
The Renuvion®/J-Plasma® Precise Open Handpiece is intended to be used with compatible J-Plasma electrosurgical generators for the delivery of radiofrequency energy and/or helium plasma for cutting, coagulation and ablation of soft tissue during open surgical procedures. The Renuvion®/J-Plasma® Precise Open® Handpieces are compatible with electrosurgical Generators BVX-200H, and BVX-200P.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Bovie.
# 510(k) Summary
| Submitted by: | Bovie Medical Corporation<br>5115 Ulmerton Road Clearwater, Florida 33760 -4004<br>United States of America |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Rubiela Maldonado<br>Regulatory Affairs Manager<br>Phone: (727) 803-8508<br>Fax: (727) 803-4491 |
| Submission Correspondent: | Topaz Kirlew<br>VP, Quality Assurance & Regulatory Affairs<br>Phone: (727) 803-8617<br>Email: topaz.kirlew@apyxmedical.com |
| Date Prepared: | December 21, 2018 |
| Trade Name: | Renuvion®/J-Plasma® Precise Handpiece and<br>Precise Open Handpiece |
| Common Name: | Electrosurgical Handpiece |
| Classification: | Class II per 21CFR 878.4400 -Electrosurgical<br>Cutting and Coagulation Device and Accessories<br>Product Code GEI |
| Predicate Devices: | Bovie J-Plasma Handpiece K151325 &<br>Bovie J-Plasma Precise Open K170188 |
### Intended Use / Indications for Use:
The Renuvion®/J-Plasma® Precise Handpiece is intended to be used with compatible electrosurgical generators for the delivery of helium gas plasma for cutting, coagulation and ablation of soft tissue during open and laparoscopic surgical procedures.
The Renuvion®/J-Plasma® Precise Open Handpiece is intended to be used with compatible electrosurgical generators for the delivery of radiofrequency energy and/or helium plasma for cutting, coagulation and ablation of soft tissue during open surgical procedures. The Renuvion®/J-Plasma® Precise Handpiece is compatible with the Electrosurgical Generators BVX-200H and BVX-200P.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Bovie" in a bold, blue font. There is a curved line above the word, also in blue, with a dot at the end. The font is sans-serif and the word is slightly italicized.
### Device Description Renuvion/J-Plasma Handpieces (J-Plasma Precise and Precise Open):
The Renuvion/J-Plasma Handpieces are sterile, single use electrosurgical accessories intended to be used in conjunction with compatible electrosurgical generators for the delivery of helium gas plasma for cutting, coagulating soft tissue. The compatible Generators operate at an adjustable power of up to 40W (expressed as 0-100% is 40W) and provide an adjustable helium gas flow of 1-5 LPM. The Renuvion/J-Plasma Handpieces are available with a retractable cutting blade tip for excising tissue. The blade tip serves as an electrode to generate helium plasma.
The Renuvion/J-Plasma Precise devices are available in 15cm, 27cm and 33cm lengths for use in laparoscopic or open surgical procedures. The 27cm are standard lengths for laparoscopic instruments and are compatible with standard 5mm trocars.
The Renuvion/J-Plasma Precise Open devices are available in 44mm and 150mm for used in open surgical procedures.
#### Intended Use & Technological Characteristics
The handpieces have the same intended use as the predicate devices. There are no technological differences or changes to the principle of operation or to the method of application. There are no changes to sterilization methods.
The design of Handpieces compared to the predicate devices are identical except for the following differences:
- . The plug connector to the generator is dimensioned so that it will establish a pneumatic seal connection before an electrical connection (pins shortened, and O-ring added to replace silicone seal).
- Reduced insertion/extraction force to pluq/unplug the device to/from the generator.
### Performance Data
Performance testing to assure that the Renuvion/J-Plasma Precise and Precise Open Handpieces met performance requirements were performed and are summarized in the following table:
| Test | Objective |
|--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanical & Electrical<br>Verification and<br>Functionality | Verify the mechanical and electrical functionality of<br>the handpiece Pneumatic Testing Electrical Continuity test Perform Cable Flexure / Tension Test Connector Dielectric Strength Test Mold Stress Test Free-fall Drop Test Insertion/Extraction Integrity |
| Usability Evaluation | The purpose of this protocol is to verify and<br>validate the usability of the J-Plasma Pistol Grip<br>Handpiece as it relates to safety. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Bovie" in a bold, blue font. There is a curved, light blue line above the word, and a small, blue dot above the "i". The word appears to be a logo or brand name.
## Biocompatibility Testing
The biocompatibility evaluation for the Renuvion/J-Plasma Handpieces was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of evaluations included the following testing:
- l Cytotoxicity
- I Sensitization
- Irritation
- Systemic toxicity
- 트 Hemolysis
- . Material Mediated Pyrogen
The Renuvion/J-Plasma handpieces are an external communicating device with an indirect blood path contact for a duration of less than 24 hours; Biocompatibility has been established per ISO 10993 guidelines for this category. Due to the redesign of the plug connector, changes on materials were made to two patient contacting components, flow tube coupler and O-ring. Biocompatibility verification testing was satisfactorily conducted for these components.
## Conclusion
There is no difference between the predicates, Precise J-Plasma handpiece and Precise Open handpiece and the Subject devices in terms of intended use, principle of operation, and the technology used for device performance. There is no new technology and there are no differences that would raise new or different questions of safety or efficacy.
Verification and validation testing of the Renuvion/J-Plasma handpieces demonstrate that the devices perform as intended.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.