Lung Vision System

K183593 · Bodyvision Medical , Ltd. · LLZ · Apr 18, 2019 · Radiology

Device Facts

Record IDK183593
Device NameLung Vision System
ApplicantBodyvision Medical , Ltd.
Product CodeLLZ · Radiology
Decision DateApr 18, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

Indications for Use

The LungVision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.

Device Story

LungVision System assists physicians in guiding endobronchial tools to pulmonary targets. Input: pre-operative 3D CT datasets and live fluoroscopic X-ray images. Operation: system segments CT data; registers and overlays 3D segmented data onto live fluoroscopic images to provide 3D visualization of tool position relative to target. Modifications include virtual bronchoscopy, C-arm based tomography (stacking images for alternative visualization), multi-view registration (using multiple C-arm angles), and real-time compensation for tool movement. Used in clinical settings by physicians. Output: visual overlay on monitor/tablet to support navigation. Benefits: improved accuracy in navigating to target areas in deformable lung tissue. Tablet used for planning; not for diagnostic purposes.

Clinical Evidence

No clinical testing performed. Evidence consists of bench testing, including verification of requirements, hazard mitigation testing, and performance testing in pig lungs to assess accuracy in deformable tissue.

Technological Characteristics

System includes main unit, tablet, router, and software. Hardware complies with ANSI/AAMI/ES 60601-1 and IEC 60601-1-2. Features 2D/3D medical image acquisition, DICOM CT import, segmentation, and real-time fluoroscopic registration. Software-based image processing includes contrast/brightness adjustment, zoom, pan, and multi-planar reformatting. Operates on Windows 10. No patient-contacting parts.

Indications for Use

Indicated for patients undergoing pulmonary procedures requiring endobronchial tool navigation to a target area within the lungs.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | | 510(k) Summary<br>April 18, 2019 | |----------------------------|-------------------------------------------------------------------------------------------------------------------| | Official Contact: | Dorian Averbuch<br>Body Vision Medical Ltd.<br>34 Sokolov St.<br>Ramat Hasharon Israel<br>Tel: 646- 863-7848 (US) | | Submission Correspondent: | Paul Dryden<br>ProMedic, LLC<br>131 Bay Point Dr. NE<br>St. Petersburg, FL 33704 | | Proprietary or Trade Name: | LungVision System | | Common/Usual Name: | System, image processing, radiological | | Classification Name: | Picture archiving and communications system<br>LLZ, Class II, CFR 892.2050 | | Predicate Device: | BodyVision – LungVision – K163622 | | Reference Device: | Covidien - superDimension™ Navigation System V7.2<br>K173244 | #### Device Modification - 1. Virtual Bronchoscopy - a. A standard technology in the field of pulmonology. This is a visualization technique showing the same information already available in the CT scan. - 2. Software upgrades - a. C-Arm based Tomography - visual aid. Collecting a number of successive images produced by a standard C-Arm which are digitally "stacked" together to form an additional alternative visualization of imaging information sourcing from standard C-Arm. - b. Multi view Setup - an algorithm improvement. Same airway based registration algorithm, extended to multiple C-Arm angles. previous was limited to one angle (AP, LAO, RAO or other C-Arm position) - 3. Real time compensation - Predicate displays a static target overlay on the fluoroscopic image, whereas the proposed a. modification performs small adjustments related to the tool movement on the fluoroscopic screen to the target overlay. - 4. 3D Guidance - Predicate calculates the tool position with respect to pre-operative CT. The proposed a. modification provides 3D visualization for those calculations in addition to the existing screen. - 5. Use of a Tablet in addition to PC - The Tablet image displayed is not for diagnostic purposes. a. None of these modifications changes the indications for use, environment of use or patient population. #### Description: The Lung Vision System (K163622) is designed to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures. {1}------------------------------------------------ ### 510(k) Summary April 18, 2019 The Lung Vision System is designed to assist the physician in guiding endobronchial tools towards the target area of interest inside the patient lungs. Prior to the endoscopic procedure the system allows planning the target location and the path to the target area on the CT scan. During the endoscopic procedure the system overlays planned data over fluoroscopic images to support endobronchial tool navigation towards the area of interest. The system does not include the Fluoroscope, Bronchoscope or the external monitor. Lung Vision system includes a main unit and a tablet vs. the previous PC based hardware platform. Image processing algorithms are executed on the main unit and the tablet is used as a primary method of interacting with the system. The Tablet is for planning but is not for diagnostic purposes. Both can perform the following - functions: segment previously acquired DICOM 3D CT image data, . - . register DICOM 3D CT image data with live fluoroscopic X-ray image - . overlay the segmented 3D CT dataset over a live fluoroscopic X-ray image of the same anatomy, obtained on a Fluoroscopic system. # System Components Overview The following is a list of the LungVision System's main components: LungVision Main Unit LungVision Tablet LungVision Router LungVision Board (passive device) LungVision Software Tablet is off-the-shelf information technology equipment powered from the medical grade power supply. The system has been evaluated to AAIM ANSI ES 60601-1 2005+ A1:2012 and IEC 60601-1-2: 2014. #### Indications for Use: The Lung Vision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures. #### Contraindications: None # Device Comparison Table 1 compares the subject device to the predicate | Feature | Proposed Device<br>LungVision | Primary Predicate<br>K163622<br>Body Vision-LungVision | |------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | The LungVision System is intended to<br>enable users to segment previously<br>acquired 3D CT datasets and overlay and<br>register these 3D segmented data sets<br>with fluoroscopic live X-ray images of<br>the same anatomy in order to support<br>catheter/device navigation during<br>pulmonary procedures. | The LungVision System is intended to<br>enable users to segment previously<br>acquired 3D CT datasets and overlay and<br>register these 3D segmented data sets<br>with fluoroscopic live X-ray images of<br>the same anatomy in order to support<br>catheter/device navigation during<br>pulmonary procedures. | | Classification | Picture archiving and communications<br>system<br>21CFR 892.2050 | Picture archiving and communications<br>system<br>21CFR 892.2050 | | Feature | April 18, 2019<br>Proposed Device<br>LungVision | Primary Predicate<br>K163622<br>Body Vision-LungVision | | | Product Code: LLZ | Product Code: LLZ | | Target anatomy | Lungs | Lungs | | Anatomy access | Bronchial airways | Bronchial airways | | Windows OS | Windows 10 | Windows 10 | | Medical imaging<br>software | Yes | Yes | | General Image 2D/3D<br>review | Yes | Yes | | 3D rendering view | Yes | Yes | | Multi-modality<br>Support | Yes | Yes | | Image registration | Yes | Yes | | Multi-planar<br>reformatting (MPR) | Yes | Yes | | DICOM import | Yes | Yes | | Fluoroscopic video | Yes | Yes | | Standard Image<br>viewing tools | Yes | Yes | | Segmentation tool | Yes | Yes | | Video capture | Yes | Yes | | Live Image overlays | Yes | Yes | | Import prior plans | Yes | Yes | | Point<br>marking/ tagging | Yes | Yes | | Navigation type | Visual | Visual | | Modifications | | | | Virtual bronchoscopy | Yes | No<br>Reference Covidien -<br>superDimension™ Navigation<br>System V7.2 - K173244 | | C-Arm based | Yes | No | | Tomography | | | | Multi-view set-up | Yes | No | | Real-time | Yes - software upgrade | No | | compensation | | | {2}------------------------------------------------ #### 0(k) Summar 000 0010 # Substantial Equivalence Discussion We will discuss the table above. Indications for Use / Patient Population / Environment of Use: As in comparison of Indications For Use above, we can conclude that the indications for use for the LungVision and the predicate are substantially equivalent. Discussion: The differences in proposed indications for use are minor. The minor differences do not raise new risk or safety concerns, and the subject device can be found substantially equivalent. Prescriptive: Both the LungVision and predicate are prescription devices. {3}------------------------------------------------ ### 510(k) Summary April 18, 2019 ### Discussion: There are no differences. Design and Technology: The LungVision utilizes the same technological characteristics as the predicate and reference devices. Both include: - . software applications that provide 2D and 3D medical image acquisition including real-time video image acquisition and visualization of the anatomy - . allow co-registration of real-time images to previously created 3D image sets based on previously collected DICOM CT images - . include image enhancements such as contrast and brightness, zoom and pan capabilities - addition of a Tablet to the PC but not to be used for diagnostic purposes ● For the Virtual Bronchoscopy feature the reference device - Covidien - superDimension™ Navigation System V7.2 - K173244 has the same feature and intended use. Discussion: The software modifications including virtual bronchoscopy, C-arm based Tomography, 3D Guidance, and Real-time compensation are similar to the reference device. ### Performance and Specifications: The performance and specifications demonstrate that the Lung Vision and predicate devices perform the same functions using the same technologies thus can be found substantially equivalent. Discussion: There are no differences, thus the subject device can be found substantially equivalent. ### Compliance with Standards: LungVision includes hardware and complies with ANSI/AAMI/ES 60601-1:2005(2012) and IEC 60601-1-2:2014. Discussion: The proposed device complies with the latest standards and thus can be found substantially equivalent. # Performance Testing: Nonclinical / Bench: We have performed bench tests and found that the Lung Vision met all requirements specifications and was found to be equivalent in comparison to the predicate. Testing includes verification testing of the requirements, testing of hazards mitigations) and performance testing of the system. Testing has also been performed with pig lungs to test accuracy in deformable tissue. Biocompatibility: There are no patient contacting parts of the LungVision System Animal: No animal testing was performed Clinical: No clinical testing was performed # Substantial Equivalence Conclusion Based upon the foregoing performance testing and comparison to the legally marketed predicate devices for indications for use, technology, and performance we believe we have demonstrated that the Lung Vision System is substantially equivalent in safety and effectiveness to the predicate device. {4}------------------------------------------------ | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement on last page. | |-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------| |-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------| **Indications for Use** | 510(k) Number ( <i>if known</i> ) | K183593 | |----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | LungVision System | | Indications for Use ( <i>Describe</i> ) | The LungVision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures. | | Type of Use ( <i>Select one or both, as applicable</i> ) | | | XX Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION" in a smaller font. April 18, 2019 Body Vision Medical Ltd. % Mr. Paul Dryden Consultant ProMedic, LLC 131 Bay Point Dr., NE ST. PETERSBURG FL 33704 Re: K183593 Trade/Device Name: LungVision System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: April 1, 2019 Received: April 2, 2019 Dear Mr. Dryden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {6}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Thalia T. Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...