Nalu Neurostimulation Kit (Integrated, 40 cm: Single 8/Dual 8), Nalu Neurostimulation Kit (Ported, 2 cm: Single 8/Dual 8), Dual 8 Ported Nalu Implantable Pulse Generator with 40 cm Kit, 40 cm/ 60 cm Trial/Extension Lead Kits, Patient Kits and miscellaneous replacement kits
K183579 · Boston Scientific Neuromodulation · GZF · Mar 29, 2019 · Neurology
Device Facts
Record ID
K183579
Device Name
Nalu Neurostimulation Kit (Integrated, 40 cm: Single 8/Dual 8), Nalu Neurostimulation Kit (Ported, 2 cm: Single 8/Dual 8), Dual 8 Ported Nalu Implantable Pulse Generator with 40 cm Kit, 40 cm/ 60 cm Trial/Extension Lead Kits, Patient Kits and miscellaneous replacement kits
Applicant
Boston Scientific Neuromodulation
Product Code
GZF · Neurology
Decision Date
Mar 29, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5870
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This system is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. The system is not intended to treat pain in the craniofacial region. The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device.
Device Story
Nalu Neurostimulation System provides peripheral nerve stimulation for chronic, intractable pain. System includes miniature implantable pulse generator (IPG), leads, surgical tools, externally worn Therapy Disc, and clinician/patient software. IPG (ceramic/polyurethane/silicone) delivers pulsed electrical current to peripheral nerves via 8-electrode leads to inhibit pain signals. Therapy Disc (rechargeable Li-ion) powers IPG wirelessly via 40.68 MHz RF; worn by patient via adhesive clip or belt. Clinician Programmer (Android tablet) configures therapy parameters via secure Bluetooth Low Energy (BLE). Patient Remote Control (iOS/Android app) allows basic control (on/off, program selection) via BLE. Used in hospital/home settings by physicians and patients. System benefits patients by providing therapeutic pain relief through adjustable, patient-controlled stimulation.
Clinical Evidence
No clinical data. Evidence consists of bench testing (electrical, mechanical, software verification/validation), biocompatibility testing (ISO 10993-1), and a 90-day porcine animal study evaluating surgical usability, RF communication, device stability, and tissue response. All devices performed as expected.
Technological Characteristics
Implantable neurostimulator (IPG) with ceramic/polyurethane/silicone housing. Leads: 8-electrode Pt/Ir array, polyurethane insulation. Power: 40.68 MHz RF wireless coupling. Connectivity: BLE for clinician/patient apps. Sterilization: Ethylene Oxide. Standards: ISO 14708-1/3, IEC 60601-1, IEC 60601-1-2, IEC 62304. Software: Moderate level of concern.
Indications for Use
Indicated for pain management in adults with severe intractable chronic pain of peripheral nerve origin, as sole therapy or adjunct in multidisciplinary approach. Not for craniofacial pain. Trial devices limited to 30 days for efficacy determination.
Regulatory Classification
Identification
An implanted peripheral nerve stimulator for pain relief is a device that is used to stimulate electrically a peripheral nerve in a patient to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed around a peripheral nerve and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.
Predicate Devices
StimQ Peripheral Nerve Stimulator (PNS) System (K171366)
Reference Devices
StimQ Peripheral Nerve Stimulator (PNS) System (K152178)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo, which features a stylized human figure. The logo on the right is the FDA U.S. Food & Drug Administration logo, with the FDA acronym in a blue square and the full name of the administration written in blue text.
March 29, 2019
Nalu Medical, Inc Michele Chin-Purcell, Ph.D. Vice President Regulatory Affairs and Quality Assurance 2320 Faraday Ave.. Suite 100 Carlsbad, CA 92008
Re: K183579
Trade/Device Name: Nalu Neurostimulation System for Peripheral Nerve Stimulation Regulation Number: 21 CFR 882.5870 Regulation Name: Implanted Peripheral Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZF Dated: February 26, 2019 Received: February 27, 2019
Dear Michele Chin-Purcell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdr/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Pamela D. Scott-S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183579
Device Name
Nalu Neurostimulation System for Peripheral Nerve Stimulation
Indications for Use (Describe)
This system is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent or as an adjunct to other modes of therapy used in a multidisciplinary approach. The system is not intended to treat pain in the craniofacial region.
The trial devices are solely used for trial stimulation (no longer thank 30 days) to determine efficacy before recommendation for a permanent (long term) device.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div></div> |
|----------------------------------------------|-------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div></div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 5.1. Submission Sponsor
Nalu Medical, Incorporated 2320 Faraday Ave., Suite 100 Carlsbad, CA 92008 Phone: (760) 448-2360 Fax: (760) 448-2377 Contact: Michele Chin-Purcell, Vice President of Regulatory Affairs and Quality Assurance
## 5.2. Date Prepared
March 22, 2019
### 5.3. Device Identification
Trade/Proprietary Name: Nalu Neurostimulation System for Peripheral Nerve Stimulation Common/Usual Name: Peripheral Nerve Stimulator Product Code: GZF Regulation number: 21 CFR 882.5870: Stimulator, peripheral nerve, implanted (Pain Relief) Class: Class II Device Classification Panel: Neurology
## 5.4. Legally Marketed Predicate Device(s)
StimQ Peripheral Nerve Stimulator (PNS) System (K171366) by StimQ LLC Hereafter, also referred to as the StimQ PNS System or StimQ
For areas where slight differences occur between the Nalu Neurostimulation system and the primary predicate (K171366), substantial equivalence to other reference devices in this same product code is demonstrated. These reference devices were used as part of the predicate history to the primary predicate in this submission. The history of the predicates is summarized in Table 5-1:
| Device | 510(k) | Predicate(s) used for clearance |
|--------------------------------------------------------------------------|---------|-----------------------------------------------------------------------------------------------------------------------------------|
| StimQ Peripheral Nerve<br>Stimulator (PNS) System | K152178 | Stimwave Freedom SCS (K150517)<br>Medtronic Mattrix 3271/3272 (K934065)<br>Medtronic Xtrel, 3425 (K883780)<br>ANS Renew (K000852) |
| StimQ Peripheral Nerve<br>Stimulator (PNS) System<br>(Primary Predicate) | K171366 | K152178 |
{4}------------------------------------------------
The 510(k) history of the StimQ system includes design changes over time. The original Medtronic and ANS devices are part of the predicate history of the StimQ device and are also used as references devices in this 510(k).
## 5.5. Device Description
The Nalu Neurostimulation System (also referred to as the "Nalu System") is used for peripheral nerve stimulation to provide therapeutic relief for chronic, intractable pain of peripheral nerve origin. The Nalu Neurostimulation system incorporates a miniature implanted neurostimulator, powered by an externally worn Therapy Disc device. Similar to the predicate StimQ, the Nalu Neurostimulation therapy utilizes pulsed electrical current to create an energy field that acts on the peripheral nerves to inhibit the transmission of pain signals to the brain. The Nalu System may be implanted following a successful trial period using the Nalu Neurostimulation trial system.
The Nalu Neurostimulation System is comprised of 5 elements:
| 1. | Nalu<br>Implantable<br>Pulse<br>Generator | The implantable pulse generator (IPG) provides electrical<br>stimulation pulses that are transmitted through the leads to the<br>desired peripheral nerve. The IPG is available in two different<br>implant architectures: an “integrated" system with pre-attached leads<br>and a "ported" system where leads may be attached, via connector<br>ports. In addition, both of these versions are available in single or<br>dual lead configurations. The hermetic IPG housing includes a<br>ceramic enclosure and a feedthrough connected internally to a<br>printed circuit board assembly. Wires leaving the IPG are<br>encapsulated in polyurethane and a silicone over mold forms the<br>final biocompatible surface of the IPG for direct patient tissue<br>contact. |
|----|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Leads | Leads are implantable and are designed to deliver electrical pulses to<br>the peripheral nerve via an array of eight cylindrical electrodes at the<br>distal end. Leads may be integrated with or connected to the IPG.<br>Both Trial and Permanent Implant leads are available for use. The<br>leads use polyurethane insulation with Pt/Ir electrodes. The leads<br>may be secured in place with the Nalu Lead Anchor. |
| 3. | Surgical and<br>Trial Tools | Implantation of the Nalu IPG and lead components for Peripheral<br>Nerve Stimulation (PNS) is performed via standard PNS surgical<br>techniques. The desired implant location is accessed via needle<br>placement, followed by lead placement through an introducer. The<br>leads are anchored and the IPG is placed in a subcutaneous pocket.<br>Patient contacting materials include medical grade stainless steel,<br>thermoplastic elastomers, ABS, silicone, and Urethane. |
| 4. | Externally<br>worn | Two types of Therapy Disc are available. One is to be used during<br>the trial phase (Trial Therapy Disc), and one is to be used after<br>permanent IPG implantation (Therapy Disc). Both devices are worn<br>by the patient using one of the Nalu-provided options The Therapy |
{5}------------------------------------------------
| Therapy<br>Discs | Discs house a rechargeable lithium ion battery, and electronics<br>including a microcontroller running software for therapy control,<br>patient interaction and communication with Nalu's Clinician<br>Programmer and Remote Control devices. The Therapy Disc used to<br>power and command the implant does so wirelessly using Radio<br>Frequency (RF) and is held in place by an adhesive clip applied to<br>the skin or a belt/cuff worn over clothing. |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 5. Clinician<br>Programmer<br>and Remote<br>Control | A Clinician Programmer Application is provided to configure the<br>Trial Therapy Disc and Therapy Disc devices during surgery and<br>programming. A Patient Remote Control Application is available to<br>provide the patient with a convenient secondary option to control<br>their system in addition to built-in controls on the Therapy Disc.<br><br>The Clinician Programming Application runs on an Android tablet<br>and communicates over a secure Bluetooth Low Energy link with<br>the Trial Therapy Disc and Therapy Disc devices. The programmer<br>is responsible for configuring the devices to deliver therapy<br>according to clinician defined levels and patient preferences, and for<br>managing patient and session records.<br><br>The Patient Remote Control Application runs on iOS and Android<br>platforms and offers basic control of the Trial Therapy Disc and<br>Therapy Disc through a secure Bluetooth Low Energy link. The<br>controls include selecting between clinician-defined therapy options<br>(programs), turning stimulation on and off, and managing alerts. |
#### Indications for Use Statement 5.6.
"This system is indicated for pain management in adults who have severe intractable chronic pain of peripheral nerve origin, as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach. The system is not intended to treat pain in the craniofacial region.
The trial devices are solely used for trial stimulation (no longer than 30 days) to determine efficacy before recommendation for a permanent (long term) device. "
The Indications for Use statement for the Nalu Neurostimulation System is not identical to the predicate device; however, the differences do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate devices have the same intended use for the stimulation of peripheral nerves for treatment of chronic, intractable pain.
## 5.7. Substantial Equivalence Discussion
The following tables compare the Nalu Neurostimulation System to the predicate device with respect to intended use, technological characteristics and principles of operation,
{6}------------------------------------------------
providing more detailed information regarding the basis for the determination of substantial equivalence.
{7}------------------------------------------------
## Table 5-2: Substantial Equivalence Table – General and Implanted Components
| | Nalu Neurostimulation System<br>(Subject Device) | StimQ PNS<br>System<br>(Primary<br>Predicate) | StimQ PNS<br>System<br>(Reference<br>Device) | Medtronic<br>Mattrix<br>3271/3272<br>(Reference<br>Device) | Medtronic<br>Xtrel 3425<br>(Reference<br>Device) | ANS Renew<br>(Reference Device) | Analysis of<br>Technological<br>Differences from<br>Primary Predicate |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| 510(k) | K183579 | K171366 | K152178 | K934065 | K883780 | K000852 | NA |
| Product Code<br>and class | GZF, Class II | Same | Same | GZF and GZB | GZB | GZF and GZB | Same |
| Regulation<br>number | 21 CFR §882.5870 | Same | Same | Same, plus 21<br>CFR 882.5880 | Same | Same, plus 21 CFR<br>§882.5880 | Same |
| Classification<br>name | Stimulator, Peripheral Nerve,<br>Implanted (pain relief) | Same | Same | Same plus<br>Stimulator,<br>Spinal Cord,<br>Implanted (Pain<br>Relief) | Same | Same plus<br>Stimulator, Spinal<br>Cord, Implanted<br>(Pain Relief) | Same |
| Intended Use | Stimulation of peripheral nerves<br>for chronic, intractable pain | Same | Same | Same, plus<br>Stimulation of<br>spinal cord for<br>chronic,<br>intractable pain | Same | Same, plus<br>Stimulation of spinal<br>cord for chronic,<br>intractable pain | Same |
| Indications for<br>Use | This system is indicated for pain<br>management in adults who have<br>severe intractable chronic pain of<br>peripheral nerve origin, as the<br>sole mitigating agent, or as an<br>adjunct to other modes of<br>therapy used in a<br>multidisciplinary approach. The<br>system is not intended to treat<br>pain in the craniofacial region. | The StimQ Peripheral Nerve<br>Stimulator (PNS) System is<br>indicated for pain management<br>in adults who have severe<br>intractable chronic pain of<br>peripheral nerve origin, as the<br>sole mitigating agent, or as an<br>adjunct to other modes of<br>therapy used in a<br>multidisciplinary approach. The<br>StimQ PNS System is not | | Indicated as an aide in the<br>management of chronic,<br>intractable pain of the trunk or<br>limbs | | Indicated for the<br>treatment of chronic<br>pain of trunk and<br>limbs, either as the<br>sole mitigating<br>agent, or as an<br>adjunct to other<br>modes of therapy<br>used in a | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| | Nalu Neurostimulation System<br>(Subject Device) | StimQ PNS<br>System<br>(Primary<br>Predicate) | StimQ PNS<br>System<br>(Reference<br>Device) | Medtronic<br>Mattrix<br>3271/3272<br>(Reference<br>Device) | Medtronic<br>Xtrel 3425<br>(Reference<br>Device) | ANS Renew<br>(Reference Device) | Analysis of<br>Technological<br>Differences from<br>Primary Predicate |
| | The trial devices are solely used<br>for trial stimulation (no longer<br>than 30 days) to determine<br>efficacy before recommendation<br>for a permanent (long term)<br>device. for a permanent (long<br>term) device. | intended to treat pain in the<br>craniofacial region.<br>The StimQ Trial Lead Kit is only<br>used in conjunction with the<br>StimQ Stimulator Receiver Kit.<br>The trial devices are solely used<br>for trial stimulation (no longer<br>than 30 days) to determine<br>efficacy before<br>recommendation for a<br>permanent (long term) device. | | | | multidisciplinary<br>approach. | |
| Prescription<br>Use? | Yes | Same | Same | Same | Same | Same | Same |
| Implant site | Peripheral nerves, excluding<br>craniofacial region | Same | Same | Same | Same | Same | Same |
| Environmental<br>Use | Hospital, Home | Same | Same | Same | Same | Same | Same |
| Intended<br>Clinician | Orthopedic, Neurosurgeon,<br>Anesthesiologist | Same | Same | Same | Same | Same | Same |
| Intended User | Physician, Layperson | Same | Same | Same | Same | Same | Same |
| Mode of Action | Radio Frequency (RF) wireless<br>transmission of energy to<br>produce stimulation at stimulator<br>electrodes. | Same | Same | Same | Same | Same | Same |
| Software Level<br>of Concern | Moderate | Moderate | Moderate | Unreported | Unreported | Moderate | Same |
| | Nalu Neurostimulation<br>System<br>(Subject Device) | StimQ PNS System<br>(K171366)<br>(Primary Predicate) | StimQ PNS System<br>(K152178)<br>(Reference<br>Device) | Medtronic<br>Mattrix<br>3271/3272<br>(K934065)<br>(Reference<br>Device) | Medtronic<br>Xtrel 3425<br>(K883780)<br>(Reference<br>Device) | ANS Renew<br>(K000852)<br>(Reference<br>Device) | Analysis of<br>Technological<br>Differences from<br>Primary<br>Predicate |
| IPG | | | | | | | |
| Dimensions | Lead = 1.30 mm diameter<br>IPG = 28 x 11 x 4.9 mm | Integrated with lead body,<br>1.35 mm diameter | Integrated with<br>lead body, 1.35<br>mm diameter | Details unavailable. See Section 12.2.1 below | | | Differences do<br>not affect safety<br>and effectiveness<br>of intended use |
| Housing<br>material | Silicone and Pellethane<br>2363-55D | Pellethane 2363-55D | Pellethane 2363-<br>55D | Details unavailable. See Section 12.2.1 below | | | Differences do<br>not affect safety<br>and effectiveness<br>of intended use |
| Implant site | Peripheral nerves,<br>excluding craniofacial<br>region | Same | Same | Same | Same | Same | Same |
| Electrical<br>components | Embedded receiver,<br>flexible circuit board | Same | Same | Sealed<br>electronic<br>circuits | Sealed<br>electronic<br>circuits | Sealed<br>electronic<br>circuits | Same |
| Power Delivery | Coupled receiver radio<br>frequency transmission | Same | Same | Same | Same | Coupled<br>receiver,<br>hardwired<br>with<br>connector | Same |
| Lead | | | | | | | |
| Electrode<br>Material | Platinum-iridium 90:10 | Same | Same | Same | Same | Same | Same |
| Insulation Body<br>Material | Pellethane 2363-55D | Same | Same | Same | Same | Same | Same |
| Cable features | Multilumen tube | Same | Same | Coiled Wires | Coiled Wires | Braided Wire | Same |
| | Nalu Neurostimulation<br>System<br>(Subject Device) | StimQ PNS System<br>(K171366)<br>(Primary Predicate) | StimQ PNS System<br>(K152178)<br>(Reference<br>Device) | Medtronic<br>Mattrix<br>3271/3272<br>(K934065)<br>(Reference<br>Device) | Medtronic<br>Xtrel 3425<br>(K883780)<br>(Reference<br>Device) | ANS Renew<br>(K000852)<br>(Reference<br>Device) | Analysis of<br>Technological<br>Differences from<br>Primary<br>Predicate |
| Lead length | 40 cm, 60 cm | 44 cm | 45 cm | 30 to 110 cm | 30 to 110 cm | 30 and 60 cm | Differences do<br>not affect safety<br>and effectiveness<br>of intended use |
| Diameter | 1.30 mm | 1.35 mm | 1.35 mm | 1.3 mm | 1.3 mm | 1.37 mm | Differences do<br>not affect safety<br>and effectiveness<br>of intended use |
| Electrode Array<br>length | 52 mm | 24 mm (FRE-4)<br>52 mm (FRE-8) | 24 mm | 24 mm | 24 mm | 24 mm | Differences do<br>not affect safety<br>and effectiveness<br>of intended use |
| No. of<br>Electrodes, per<br>lead | 8 | 4 (FRE-4)<br>8 (FRE-8) | 4 | Same | Same | 4 or 8 | Differences do<br>not affect safety<br>and effectiveness<br>of intended use |
| Individual<br>Electrode<br>length | 3.0 mm | Same | Same | Same | Same | Same | Same |
| Electrode<br>spacing | 4.0 mm | Same | Same | Same | Same | Same | Same |
| Electrode<br>surface area | 12.25 mm² | 12.72 mm² | 12.72 mm² | 12.25 mm² | 12.25 mm² | ~13 mm² | Differences do<br>not affect safety<br>and effectiveness<br>of intended use |
| Lead extension | Lead extension available | NA | NA | Lead extension<br>available | Lead extension<br>available | Lead<br>extension<br>available | Differences do<br>not affect safety<br>and effectiveness<br>of intended use |
| | Nalu Neurostimulation<br>System<br>(Subject Device) | StimQ PNS System<br>(K171366)<br>(Primary Predicate) | StimQ PNS System<br>(K152178)<br>(Reference<br>Device) | Medtronic<br>Mattrix<br>3271/3272<br>(K934065)<br>(Reference<br>Device) | Medtronic<br>Xtrel 3425<br>(K883780)<br>(Reference<br>Device) | ANS Renew<br>(K000852)<br>(Reference<br>Device) | Analysis of<br>Technological<br>Differences from<br>Primary<br>Predicate |
| Lead Anchor | Molded silicone anchor<br>with Ti locking mechanism | Suture Sleeve Cap,<br>Pellethane 55-D, placed<br>over proximal end of<br>stimulator | Suture Sleeve Cap,<br>Pellethane 55-D,<br>placed over<br>proximal end of<br>stimulator | Molded silicone<br>anchor | Molded<br>silicone anchor | Molded<br>silicone<br>anchor | Differences do<br>not affect safety<br>and effectiveness<br>of intended use |
| Configurations | Integrated and with Ports | Integrated | Integrated | With Ports | With Ports | With Ports | Differences do<br>not affect safety<br>and effectiveness<br>of intended use |
| Sterilization | Ethylene Oxide | Same | Same | Same | Same | Same | Same |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
Table 5-3: Substantial Equivalence Table – Therapy
{12}------------------------------------------------
| Comparator | Nalu<br>Neurostimulation<br>System<br>(Subject Device) | StimQ PNS System<br>(K171366)<br>(Primary Predicate) | StimQ PNS System<br>(K152178)<br>(Reference Device) | Medtronic Mattrix<br>3271/3272<br>(K934065)<br>(Reference Device) | Medtronic<br>Xtrel 3425<br>(K883780)<br>(Reference<br>Device) | ANS Renew<br>(K000852)<br>(Reference<br>Device) | Analysis of<br>Technological<br>Differences from<br>Primary Predicate |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|-----------------------------------------------------|-------------------------------------------------------------------|---------------------------------------------------------------|-------------------------------------------------|-----------------------------------------------------------------------------|
| Pulse Frequency | 2 Hz to 1500 Hz1 | 5 to 1500 Hz | 5 Hz to 1500 Hz | 5 to 240 Hz | 5 to 1400 Hz | 10 to 1500 Hz | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Pulse Width | 12 to 1000 µs | 50 to 500 µs | 50 to 500 µs | 50 to 500 µs | 50 to 1000 µs | 50 to 500 µs | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Current/Voltage<br>Regulated | Current | Current | Current | Voltage | Voltage | Current | Same |
| Output Voltage<br>(300 Ohms) | 0 to 3.1 V | 0 to 4.1 V | 0 to 6.3 V | 0 to 7 V | 0 to 5.4 V | 0 to 5.7 V | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Output Voltage<br>(500 Ohms) | 0 to 5.1 V | 0 to 6.4 V | 0 to 7.2 V | 0 to 10.8 V | 0 to 7.1 V | 0 to 7.6 V | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Output Voltage<br>(800 Ohms) | 0 to 8.2 V | 0 to 7.5 V | 0 to 8.0 V | 0 to 11.6 V | 0 to 8.4 V | 0 to 9.6 V | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Output Current<br>(300 Ohms) | 0 to 10.2 mA | 0 to 13.5 mA | 0 to 21 mA | 0 to 23.3 mA | 0 to 18.0 mA | 0 to 19.0 mA | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Comparator | Nalu<br>Neurostimulation<br>System<br>(Subject Device) | StimQ PNS System<br>(K171366)<br>(Primary Predicate) | StimQ PNS System<br>(K152178)<br>(Reference Device) | Medtronic Mattrix<br>3271/3272<br>(K934065)<br>(Reference Device) | Medtronic<br>Xtrel 3425<br>(K883780)<br>(Reference<br>Device) | ANS Renew<br>(K000852)<br>(Reference<br>Device) | Analysis of<br>Technological<br>Differences from<br>Primary Predicate |
| Output Current<br>(500 Ohms) | 0 to 10.2 mA | 0 to 12.8 mA | 0 to 15 mA | 0 to 21.6 mA | 0 to 14.2 mA | 0 to 15.2 mA | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Output Current<br>(800 Ohms) | 0 to 10.2 mA | 0 to 9.4 mA | 0 to 10 mA | 0 to 14.5 mA | 0 to 10.5 mA | 0 to 12.0 mA | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Waveform | charge balanced<br>(delayed) biphasic<br>asymmetrical | Same | Same | Same | Same | Same | Same |
| Pulse Shape | Decaying<br>Exponential | Decaying<br>Exponential | Decaying<br>Exponential | Decaying<br>Exponential | Decaying<br>Exponential | Decaying<br>Exponential | Same |
| Maximum phase<br>charge (300 Ohms) | 10.2 µC/pulse | 6.8 µC/pulse | 10.5 µC/pulse | 11.7 µC/pulse | 18.0 µC/pulse | 9.5 µC/pulse | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Maximum phase<br>charge (500 Ohms) | 10.2 µC/pulse | 6.4 µC/pulse | 7.2 µC/pulse | 10.8 µC/pulse | 14.2 µC/pulse | 7.6 µC/pulse | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Maximum phase<br>charge (800 Ohms) | 10.2 µC/pulse | 4.7 µC/pulse | 5.0 µC/pulse | 7.3 µC/pulse | 10.5 µC/pulse | 6.0 µC/pulse | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Maximum charge<br>density (300 Ohm) | 83.3 µC/cm² | 53.1 µC/cm² | 82.5 µC/cm² | 97.2 µC/cm² | 150.0 µC/cm² | 73.1 µC/cm² | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Comparator | Nalu<br>Neurostimulation<br>System<br>(Subject Device) | StimQ PNS System<br>(K171366)<br>(Primary Predicate) | StimQ PNS System<br>(K152178)<br>(Reference Device) | Medtronic Mattrix<br>3271/3272<br>(K934065)<br>(Reference Device) | Medtronic<br>Xtrel 3425<br>(K883780)<br>(Reference<br>Device) | ANS Renew<br>(K000852)<br>(Reference<br>Device) | Analysis of<br>Technological<br>Differences from<br>Primary Predicate |
| Maximum charge<br>density (500 Ohm) | 83.3 µC/cm² | 50.3 µC/cm² | 56.6 µC/cm² | 90.0 µC/cm² | 118.3 µC/cm² | 58.5 µC/cm² | Differences do not<br>affect safety and<br>effectiveness of<br>intended use |
| Maximum charge<br>density (800 Ohm)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.