← Product Code [EIH](/productcode/EIH) · K183569

# DD cube ONE ML (K183569)

_Dental Direkt GmbH · EIH · Feb 14, 2019 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K183569

## Device Facts

- **Applicant:** Dental Direkt GmbH
- **Product Code:** [EIH](/productcode/EIH.md)
- **Decision Date:** Feb 14, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 872.6660
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

DD cube ONE ML dental zirconia blanks are indicated for crowns, multi-unit bridges and inlay bridges. Applications include both, anterior and posterior bridges.

## Device Story

DD cube ONE ML is a dental milling blank composed of yttria-stabilized pre-sintered zirconium dioxide. It is used by dental professionals to fabricate dental restorations, specifically crowns and bridges. The device is a multilayer ceramic blank, where layers contain varying concentrations of coloring oxides to achieve a natural tooth shade post-sintering. It is processed via milling and sintering. The output is a finished dental restoration that provides high translucency, strength, and fracture toughness. The device is intended for use in dental clinics or laboratories by trained personnel to restore patient dentition.

## Clinical Evidence

No clinical data was conducted or relied upon. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (EN ISO 10993-1, -5) and physical/chemical property testing per ISO 6872:2015 (flexural strength of 1100 MPa vs 800 MPa requirement; chemical solubility of 15.3 μg/cm² vs <100 μg/cm² requirement).

## Technological Characteristics

Yttria-stabilized zirconium dioxide ceramic (Type II, Class 5 per DIN EN ISO 6872). Multilayer composition with coloring oxides. Tested per ISO 6872:2015 and ISO 10993. No software or electronic components.

## Regulatory Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

## Predicate Devices

- DD cube ONE (former DD cubeX2 HS) ([K170885](/device/K170885.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.

Dental Direkt GmbH Vera Escher Regulatory Affairs Manager International Industriezentrum 106-108 Spenge, 32139 GERMANY

February 14, 2019

## Re: K183569

Trade/Device Name: DD cube ONE ML Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: Class II Product Code: EIH Dated: January 9, 2019 Received: January 15, 2019

## Dear Vera Escher:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely. Digitally signed by Mary S. Mary S. Runner -S3 Runner -S3 Date: 2019.02.14

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known): K183569

Device Name: DD cube ONE ML

#### Indications for Use:

DD cube ONE ML dental zirconia blanks are indicated for crowns, multi-unit bridges and inlay bridges. Applications include both, anterior and posterior bridges.

| Prescription Use            | X                      |
|-----------------------------|------------------------|
| (Part 21 CFR 801 Subpart D) | AND/OR                 |
|                             | Over-The-Counter Use   |
|                             | (21 CFR 801 Subpart C) |

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{3}------------------------------------------------

| Page 1 of 2 |  | 510(k) Number: | K183569 |
|-------------|--|----------------|---------|
|             |  | Date:          |         |

**510(k) Summary**

| Submitter of 510(k)            | Dental Direkt GmbH<br>Industriezentrum 106-108<br>32139 Spenge / Germany                                                |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Contact Person                 | Uwe Greitens, CEO<br>Phone: +49 5225 86319-0<br>Fax: +49 5225 86319-99<br>u.greitens@dentaldirekt.de                    |
| Regulatory Correspondent:      | Vera Escher, Manager Regulatory Affairs<br>Phone:+49 5225 86319-0<br>Fax: +49 5225 86319-99<br>v.escher@dentaldirekt.de |
| Establishment Registration No. | 3008347275                                                                                                              |
| Date Prepared                  | 2018/12/18                                                                                                              |
| Trade Name of Device           | DD cube ONE ML                                                                                                          |
| Common Name                    | Powder, Porcelain                                                                                                       |
| Classification Name            | Porcelain Powder for clinical use                                                                                       |
| Regulation Number              | 21 CFR 872.6660                                                                                                         |
| Classification Product Code    | EIH                                                                                                                     |
| Panel                          | Dental                                                                                                                  |
| Classification                 | Class II                                                                                                                |
| Predicate Device               | K170885:<br>Dental Direkt GmbH<br>DD cube ONE (former DD cubeX2 HS)                                                     |

{4}------------------------------------------------

| Indications for Use                                     | DD cube ONE ML dental zirconia blanks are<br>indicated for crowns, multi-unit bridges and inlay<br>bridges. Applications include both, anterior and<br>posterior bridges.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description                                      | DD cube ONE ML are dental milling blanks made of<br>yttria-stabilized pre-sintered zirconium dioxide,<br>showing a super high translucency and a high<br>strength and fracture toughness at the same time.<br>The ceramics are of type II (not powder), Class 5<br>according to DIN EN ISO 6872 (FDA Recognition<br>Number 4-223). The material is suited ideally for<br>crowns and bridges for both, the anterior and<br>posterior region.                                                                                                                                                                                                                                                                                                                                                                             |
| Technological Characteristics                           | The technological characteristics of the modified<br>device are the same as those of the legally<br>marketed unmodified device. Both devices are<br>made of yttria-stabilized zirconia ceramics of high<br>quality and with outstanding material properties. The<br>only difference to the unmodified device is the<br>shade. The unmodified device DD cube ONE is a<br>white blank and the new device is a multilayer-<br>blank. The base material is identical, and the new<br>device only contains small amounts of coloring<br>oxides to achieve a natural tooth shade directly after<br>sintering. The blank has different layers: The upper<br>part contains less coloring oxides and therefore is<br>brighter than the medium dentine area and the<br>darker bottom layer which contain more coloring<br>oxides. |
| Comparison of<br>Required Technology<br>Characteristics | The following table shows a summary of the<br>technological characteristics of DD cube ONE<br>ML (modified product) compared to the predicate<br>device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

{5}------------------------------------------------

| Comparison of Required Technology Characteristics |                                                                                                                                                                                        |                                                                                                                                                                                           |                                                                                                                         |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Feature                                           | Unmodified<br>product                                                                                                                                                                  | Modified product                                                                                                                                                                          | Comment                                                                                                                 |
| Trade name                                        | DD cube ONE                                                                                                                                                                            | DD cube ONE ML                                                                                                                                                                            | n.a.                                                                                                                    |
| 510(k)                                            | K170885                                                                                                                                                                                | this submission                                                                                                                                                                           | n.a.                                                                                                                    |
| Product code                                      | EIH                                                                                                                                                                                    | EIH                                                                                                                                                                                       | Same                                                                                                                    |
| Regulatory class                                  | Class II                                                                                                                                                                               | Class II                                                                                                                                                                                  | Same                                                                                                                    |
| Manufacturer                                      | Dental Direkt GmbH                                                                                                                                                                     | Dental Direkt GmbH                                                                                                                                                                        | Same                                                                                                                    |
| Intended use                                      | DD cube ONE<br>dental zirconia<br>blanks are indicated<br>for crowns, multi-<br>unit bridges and<br>inlay bridges.<br>Applications include<br>both, anterior and<br>posterior bridges. | DD cube ONE ML<br>dental zirconia<br>blanks are indicated<br>for crowns, multi-<br>unit bridges and<br>inlay bridges.<br>Applications include<br>both, anterior and<br>posterior bridges. | Same                                                                                                                    |
| Multi-unit bridges                                | Yes                                                                                                                                                                                    | Yes                                                                                                                                                                                       | Same.                                                                                                                   |
| Accessories                                       | Yes, coloring liquids                                                                                                                                                                  | None.                                                                                                                                                                                     | The application of<br>coloring liquids with<br>the new product is<br>not reasonable,<br>since it is already<br>colored. |
| Feature                                           | Unmodified<br>product                                                                                                                                                                  | Modified product                                                                                                                                                                          | Comment                                                                                                                 |
| Chemical composition                              |                                                                                                                                                                                        |                                                                                                                                                                                           |                                                                                                                         |
| ZrO2 + HfO2 + Y2O3 [wt%]                          | > 99.0                                                                                                                                                                                 | > 99.0                                                                                                                                                                                    | Same                                                                                                                    |
| Y2O3 [wt%]                                        | < 8                                                                                                                                                                                    | < 8                                                                                                                                                                                       | Same                                                                                                                    |
| Al2O3 [wt%]                                       | < 0.1                                                                                                                                                                                  | < 0.1                                                                                                                                                                                     | Same                                                                                                                    |
| Other oxides                                      | < 0.05                                                                                                                                                                                 | ≤ 0.8                                                                                                                                                                                     | Minor deviation due<br>to color oxides but<br>still below 1%                                                            |
| Physical characteristics [Units]                  |                                                                                                                                                                                        |                                                                                                                                                                                           |                                                                                                                         |
| Performance<br>Testing                            | Tested according to<br>DIN EN ISO 6872                                                                                                                                                 | Tested according to<br>DIN EN ISO 6872                                                                                                                                                    | Same                                                                                                                    |
| Biocompatibility                                  | EN ISO 10993-1, -5                                                                                                                                                                     | EN ISO 10993-1, -5                                                                                                                                                                        | Same                                                                                                                    |

{6}------------------------------------------------

| Discussion of<br>Tests Performed | Clinical Tests                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                  | Dental Direkt GmbH did not conduct, nor rely upon, clinical<br>tests to determine substantial equivalence as dental<br>ceramics that fall under FDA product code EIH have a long<br>history in the US.                                                                                                                                                                                                                                                                         |
|                                  | Non Clinical Tests                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                  | Non-clinical testing was performed in order to validate the<br>product against the company's specified design require-<br>ments according to the following standards:                                                                                                                                                                                                                                                                                                          |
|                                  | EN ISO 10993-1 (biological compatibility) and<br>EN ISO 10993-5 (cytotoxicity)                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                  | DD cube ONE ML products were tested by an accredited<br>testing laboratory with respect to biocompatibility. The<br>laboratory certified that "the insolubility of the material is in<br>compliance with the requirements of the standard. There is<br>no evidence that hazardous effects will arise by release of<br>leachable ingredients/residues". Based on the test results<br>DD cube ONE ML was classified as eminently suitable for<br>use in the dental applications. |

- ISO 6872:2015, Dentistry Ceramic materials . (FDA Recognition # 4-223)

{7}------------------------------------------------

| Requirements acc. to ISO 6872:2015:                           |                                                                                          |                   |                   |                    |
|---------------------------------------------------------------|------------------------------------------------------------------------------------------|-------------------|-------------------|--------------------|
| Requirements<br>ISO 6872<br>(Clause)                          | Short<br>description                                                                     | Required<br>value | Achieved<br>value | Passed /<br>Failed |
| Uniformity<br>(Clause 5.1)                                    | Checked by<br>visual inspection                                                          | N/A               | -                 | Passed             |
| Freedom from<br>extraneous materials<br>(Clause 5.2.1)        | Checked by<br>visual inspection                                                          | N/A               | -                 | Passed             |
| Physical and chemical<br>properties<br>(Clause 5.2.2)         | Activity<br>concentration of<br>uranium 238                                              | < 1.0 Bq/g        | < 0.03 Bq/g       | Passed             |
| Physical and chemical<br>properties<br>(Clause 4)             | Flexural<br>strength                                                                     | 800 MPa           | 1100 MPa          | Passed             |
| Physical and chemical<br>properties<br>(Clause 4)             | Chemical<br>solubility                                                                   | < 100 μg/cm²      | 15.3 μg/cm²       | Passed             |
| Information supplied by<br>the manufacturer<br>(Clause 8.1.3) | Detailed<br>information<br>regarding the<br>handling and<br>treatment of the<br>material | N/A               | -                 | Passed             |

# Substantial Equivalence Conclusion

Dental Direkt GmbH believes that DD cube ONE ML is as safe and effective as the predicate device when used as instructed by knowledgeable and trained dental personnel. There is no change of the intended use of the modified device and no difference in fundamental scientific technology. The modified device is made from the same materials as the predicate device. The new device was verified in accordance with the above FDA recognized standard. Dental Direkt GmbH therefore believes that DD cube ONE ML (modified device) is substantially equivalent to the legally marketed unmodified device.

---

**Source:** [https://fda-staging.innolitics.com/device/K183569](https://fda-staging.innolitics.com/device/K183569)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).