VITEK® 2 AST-Gram Negative Delafloxacin is designed for antimicrobial susceptibility testing of Gram Negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.
Device Story
VITEK 2 AST-GN Delafloxacin is an automated, miniaturized, broth-based dilution system for determining minimum inhibitory concentrations (MIC) of delafloxacin against Gram-negative bacilli. Input: standardized saline suspension of clinical isolates. Operation: system fills, seals, and incubates cards containing premeasured antimicrobial concentrations; optical scanner monitors growth via light attenuation in 64 micro-wells over time. Output: MIC values and interpretive category results (S/I/R) generated automatically. Used in clinical laboratories by technicians/microbiologists. Results assist clinicians in selecting appropriate antimicrobial therapy for patients with infections caused by susceptible organisms.
Clinical Evidence
Bench testing only. Performance evaluated using 404 clinical and 206 challenge isolates (total 609 results). Overall performance: 98.0% Essential Agreement (EA) and 95.9% Category Agreement (CA). Major error rate 0.7% (2/291 susceptible isolates); minor error rate 3.7%. Growth failure rate 0.25%. Reproducibility >99%. Trending analysis showed higher MIC values for Enterobacteriaceae compared to reference method, addressed in labeling.
Technological Characteristics
Miniaturized broth microdilution card; automated optical growth monitoring; incubation up to 18 hours. Compatible with VITEK 2 and VITEK 2 Compact platforms. Quantitative measurement of MIC. No specific materials or software algorithm class (e.g., ML) specified beyond standard automated incubation/reading.
Indications for Use
Indicated for antimicrobial susceptibility testing of isolated colonies of clinically significant aerobic gram-negative bacilli, including Enterobacter cloacae, Escherichia coli, Klebsiella pneumoniae, and Pseudomonas aeruginosa, to determine in vitro susceptibility to Delafloxacin.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
{0}
Page 1 of 13
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K183524
B. Purpose for Submission:
To obtain a substantial equivalence determination for Delafloxacin for testing of Gram-negative bacilli on the VITEK 2 and VITEK 2 Compact Antimicrobial Susceptibility Test (AST) Systems
C. Measurand:
Delafloxacin ≤ 0.06 – ≥ 4 µg/mL for Enterobacteriaceae and ≤ 0.125 – ≥ 4 µg/mL for Pseudomonas aeruginosa.
D. Type of Test:
Automated quantitative or qualitative antimicrobial susceptibility test for Delafloxacin
E. Applicant:
bioMérieux, Inc.
F. Proprietary and Established Names:
VITEK 2 AST-Gram Negative Delafloxacin (≤ 0.06 – ≥ 4 µg/mL)
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1645: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System
{1}
Page 2 of 13
2. Classification:
Class II
3. Product code(s):
LON – Fully automated short-term incubation cycle antimicrobial susceptibility system
LTW – Susceptibility Test Cards, Antimicrobial
LTT – Panels, Test, Susceptibility, Antimicrobial
4. Panel:
Microbiology, 83
H. Intended Use/Indications for Use:
1. Intended Use (s):
The VITEK 2 Gram-negative Susceptibility Card is intended to for use with the VITEK 2 Systems in clinical laboratories as an *in vitro* test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.
2. Indications for Use:
VITEK 2 AST-Gram Negative Delafloxacin is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of *in vitro* susceptibility to antimicrobial agents. VITEK 2 AST-Gram Negative Delafloxacin is a quantitative test. Delafloxacin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.
Active both *in vitro* and in clinical infections
- Escherichia coli
- Enterobacter cloacae
- Klebsiella pneumoniae
- Pseudomonas aeruginosa
The VITEK 2 Gram-negative Susceptibility Card is intended to for use with the VITEK 2 Systems in clinical laboratories as an *in vitro* test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.
{2}
3. Special conditions for use statement(s):
For Prescription Use Only
4. Special instrument requirements:
VITEK 2 and VITEK 2 Compact Systems using VITEK 2 Systems 9.02 software
I. Device Description:
The VITEK 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK 2 AST card contains 64 wells. A control well(s) which contain only nutrient medium is resident on all cards. The remaining wells contain premeasured portions of antimicrobials combined with the nutrient media. The isolate to be tested is diluted to a standardized concentration with 0.45% to 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System will automatically dilute the bacterial suspension to prepare an inoculum for susceptibility cards. Then the VITEK 2 will fill, seal and place the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing, and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time (up to 24 hours for Streptococcus species). The analysis program determines when a well demonstrates growth based on attenuation of light measured by an optical scanner. This data is used to determine the minimum inhibitory concentration or "MIC" values for the anti-microbial agent. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.
VITEK 2 AST-GN Delafloxacin has the following concentrations in the card: 0.06, 0.25, 0.5, and 2 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The Delafloxacin MIC result ranges for the VITEK 2 are:
- ≤ 0.06 – ≥ 4 µg/mL for Enterobacteriaceae
- ≤ 0.125 – ≥ 4 µg/mL for Pseudomonas aeruginosa
J. Substantial Equivalence Information:
1. Predicate device name(s):
VITEK 2 AST-GN Amikacin
2. Predicate 510(k) number(s):
K172731
Page 3 of 13
{3}
# 3. Comparison with predicate:
Table 1: Comparison with Predicate Device
| Similarities | | |
| --- | --- | --- |
| Item | Device: VITEK 2 AST-GN Delafloxacin (K183524) | Predicate Device: VITEK 2 AST-GN Amikacin (K172731) |
| Intended Use | The VITEK 2 Gram-negative Susceptibility Card is intended to for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed. | The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast. |
| Test Method | Automated quantitative antimicrobial susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems to determine the in vitro susceptibility of Gram negative bacilli | Same |
| Inoculum | Standardized saline suspension of test organism | Same |
| Test Card | VITEK 2 Gram Negative Susceptibility Test Card | Same |
| Instrument | VITEK 2 and VITEK 2 Compact Systems | Same |
| Analysis Algorithm | Growth pattern analysis | Same |
| Differences | | |
| Antimicrobial Agent | Delafloxacin | Amikacin |
| Antimicrobial Concentrations | 0.06, 0.25, 0.5, and 2 μg/mL | 2, 4, 16, and 48 μg/mL |
| Reporting Range | Enterobacteriaceae: ≤ 0.06 – ≥ 4 μg/mL P. aeruginosa: ≤ 0.125 - ≥ 4 μg/mL | Enterobacteriaceae and P. aeruginosa: ≤ 1 – ≥ 64 μg/mL Acinetobacter spp.: ≤ 2 – ≥ 64 μg/mL |
# K. Standard/Guidance Documents Referenced (if applicable)
FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)
- CLSI M07-A10, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard-Ninth Edition" Vol. 35 No. 2 (January 2015)
{4}
- CLSI M100, "Performance Standards for Antimicrobial Susceptibility Testing"; Twenty-eighth Edition (January 2018)
# L. Test Principle:
The VITEK 2 and VITEK 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics in the systems uses visible light to directly measure organism growth within each of the 64 micro-wells. Transmittance optics is based on an initial light reading of a well before significant growth has begun. Every 15 minutes throughout the incubation cycle (defined period of time based on the VITEK 2 card), light transmittance readings of each well measures organism growth by the amount of light that is prevented from passing through the well. At the completion of the incubation period, the MIC values and their associated interpretive category results for each antimicrobial on the test card are displayed in an automatically generated report.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
A reproducibility study for the VITEK 2 AST-GN card with Delafloxacin was conducted at three external clinical sites using a panel of ten isolates of Gram-negative bacilli consistent with the indications for use. Testing was performed on three separate days and in triplicate for a total of 270 data points. The isolates tested in the reproducibility study included five Klebsiella pneumoniae pneumoniae, two $E.$ cloacae, and three $P.$ aeruginosa isolates. Inocula were prepared using both the auto-dilution and manual dilution methods for testing in the VITEK 2 System. In addition, inocula were prepared by the manual dilution method only for use with the VITEK 2 Compact. The mode MIC value was determined and the reproducibility was calculated based on MIC values that fell within $+/-$ one doubling dilution from the mode MIC value. The majority of data points were on-scale and within $\pm$ one doubling dilution agreement as compared to the mode MIC. Two data points were off-scale when tested by the VITEK 2 and the VITEK 2 Compact. The data was analyzed taking into consideration best case and worst case scenarios as described in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems. The reproducibility performance is shown in Table 2.
Table 2: Reproducibility Performance
| | VITEK 2 | | VITEK 2 Compact |
| --- | --- | --- | --- |
| | Auto-Dilution | Manual Dilution | Manual Dilution |
| Best Case | 100% | 99.63% | 100% |
| Worst Case | 100% | 99.63% | 99.63% |
{5}
The combined reproducibility results for all three sites were acceptable.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
# Quality Control (QC) Testing
The CLSI recommended QC organisms (E. coli ATCC 25922 and Pseudomonas aeruginosa ATCC 27853) were tested using both the VITEK 2 card and the reference method at each site. Both the automatic dilution and manual dilution methods were used for the VITEK 2 and the manual dilution method was used for the VITEK 2 Compact.
As shown in Table 3, both the auto-dilution and the manual dilution methods for VITEK 2 and the manual dilution for VITEK 2 Compact were within the expected range $>95\%$ of the time.
Table 3: Quality Control Summary Results for VITEK 2 (Auto-Dilution and Manual Dilution Methods) and VITEK 2 Compact (Manual Dilution Method)
| Delafloxacin | | VITEK 2 | | | | VITEK 2 Compact Manual Dilution | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Auto-Dilution | | Manual Dilution | | | |
| Organism | Conc. (μg/mL) | Reference | Test | Reference | Test | Reference | Test |
| E. coliATCC 25922FDA/CLSIExpected Range0.008 – 0.03μg/mL(VITEK 2:≤0.06 μg/mL) | ≤0.004 | | | | | | |
| | 0.008 | | | | | | |
| | 0.015 | 35 | | 30 | | 30 | |
| | 0.03 | 80 | | 60 | | 60 | |
| | 0.06 | | | | | | |
| | ≤0.06* | | 115 | | 90 | | 90 |
| | 0.12 | | | | | | |
| | 0.25 | | | | | | |
| | ≥0.5 | | | | | | |
| P. aeruginosaATCC 27853FDA/CLSIExpected Range0.12 - 0.5μg/mL(VITEK 2:0.12 - 0.5μg/mL) | ≤0.015 | | | | | | |
| | 0.03 | | | | | | |
| | 0.06 | | | | | | |
| | 0.12 | 10 | | 7 | | 7 | |
| | ≤0.12** | | 0 | | 0 | | 0 |
| | 0.25 | 102 | 100 | 82 | 87 | 82 | 89 |
| | 0.5 | 4 | 16 | 3 | 4 | 3 | 3 |
| | 1 | | | | | | |
| | 2 | | | | | | |
| | ≥4 | | | | 1 | | |
* The lowest dilution of the VITEK 2 Delafloxacin MIC range is $\leq 0.06\mu \mathrm{g / mL}$ for Enterobacteriaceae. Obtaining this value was considered an indicator that the quality control test results were acceptable.
**The lowest dilution of the VITEK 2 Delafloxacin MC range is $\leq 0.12\mu \mathrm{g / mL}$ for $P_{\cdot}$ aeruginosa. Obtaining this value was considered an indicator that the quality control results were acceptable.
{6}
The CLSI Delafloxacin expected ranges for *E. coli* ATCC 25922 and *P. aeruginosa* ATCC 27853 are 0.008 – 0.03 and 0.12 – 0.5 μg/mL respectively. However, the VITEK 2 MIC reporting range is ≤0.06 – ≥4 μg/mL (MIC results: ≤0.06, 0.12, 0.25, 0.5, 1, 2, ≥4 μg/mL) for *Enterobacteriaceae* and ≤0.12 – ≥4 μg/mL (MIC results: ≤0.12, 0.25, 0.5, 1, 2, ≥4 μg/mL) for *P. aeruginosa*. The two recommended QC strains tested provided off-scale with the VITEK 2 card as the acceptable ranges are lower than the lowest concentration on the card (0.06 μg/mL for *Enterobacteriaceae* and 0.12 μg/mL for *P. aeruginosa*). The quality control test results were considered acceptable for both QC strains when the MIC value of ≤0.06 μg/mL was obtained for *E. coli* ATCC 25922 and ≤0.12 μg/mL for *P. aeruginosa* ATCC 27853. Given this, bioMérieux included the following footnote indicating that QC for delafloxacin is off-scale:
> VITEK 2 AST-GN Delafloxacin does not include the full CLSI/FDA recommended dilution range for QC testing with this organism.
However, during this study, all VITEK QC results for *P. aeruginosa* ATCC 27853 were either at 0.5 or 0.25 μg/mL and none gave a value of ≤0.12μg/mL (See Table 3). In addition, bioMérieux includes the CLSI broth microdilution expected range in the labeling when the VITEK 2 system range is not aligned with that of the FDA/CLSI range. The additional information is acceptable.
## Inoculum Density Control
The DensiCHEK Plus was used to standardize the inoculum to a 0.5 McFarland standard. The instrument was standardized daily with all results recorded at each site. Calibration values were within the expected range.
## Purity Check
A purity check of all organisms was performed on the dilution tube used to prepare the VITEK 2 card inoculum. Only those cultures that were pure were evaluated in the study.
## Growth Failure Rate
There was one isolate that failed to grow in the VITEK 2 card in the clinical study. Complete test results were available for 403 isolates from a total of 404 clinical isolates. The growth failure rate was 0.25% and was acceptable.
A challenge set of 206 isolates was evaluated at one site. All 206 challenge organisms grew in the VITEK 2 GN card with Delafloxacin using both the auto-dilution and manual dilution methods for the VITEK 2 and manual inoculation for the VITEK 2 Compact System.
A total of 609 VITEK 2 AST results were available.
Page 7 of 13
{7}
d. Detection limit:
Not applicable
e. Analytical Specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Testing of delafloxacin on the VITEK 2 AST-Gram Negative card was performed at three external sites and on internal site using VITEK 2 Systems (PC) version 9.02. Results obtained with the bioMérieux VITEK 2 AST-Gram Negative card with Delafloxacin were compared to results obtained with the CLSI broth microdilution reference panel. The following concentrations of Delafloxacin are contained in the VITEK 2 AST-GN test card: 0.06, 0.25, 0.5, and 2 µg/mL. However, the reporting range is ≤ 0.06 – ≥4 µg/mL for Enterobacteriaceae and ≤ 0.12 – ≥4 µg/mL for Pseudomonas aeruginosa. The reference panel contained two-fold serial dilutions with a range of ≤ 0.0039 to ≥16 µg/mL. The testing conditions for the reference method consisted of the following:
- Medium – Mueller Hinton broth with appropriate dilutions of antimicrobial solution added
- Inoculum – Direct colony suspension
- Incubation – 35°C ambient air incubator; 16-20 hours
The VITEK 2 AST cards were inoculated with test organisms using the auto-dilution method (VITEK 2 System) and using the manual dilution method (VITEK 2 System and VITEK 2 Compact). All test inocula used for the VITEK 2 AST cards and the reference method were standardized using the DensiCHEK Plus instrument.
A total of 404 clinical isolates were evaluated at three sites with VITEK 2 AST – Gram Negative cards inoculated by automatic dilution and interpreted using the VITEK 2 instrument. There was one isolate that did not grow in the studies and was excluded from the study. The no growth rate was 0.25% (1/404). Most isolates were recently isolated from clinical specimen (270 isolates, 66.8%). The remainder were stock isolates (134 isolates, 33.2%). The clinical study included several species within Enterobacteriaceae that were not indicated in the FDA drug label for delafloxacin.
A total of 206 challenge organisms (147 Enterobacteriaceae and 59 P. aeruginosa)
Page 8 of 13
{8}
were evaluated at one site. The challenge set was tested with the auto-dilution and manual dilution options of the VITEK 2 System and with the manual dilution method on the VITEK 2 Compact System. Challenge isolates included E. cloacae (18 isolates), E. coli (12 isolates), K. pneumoniae pneumoniae (14 isolates), K. pneumoniae (103 isolates), and P. aeruginosa (59 isolates).
The combined data set of clinical and challenge isolates included 609 AST results (428 Enterobacteriaceae and 181 P. aeruginosa). A total of 289 (47.4%) resistant isolates were tested among the clinical (n=130) and challenge (n=159) isolates. A sufficient number of resistant isolates from the indicated species were included.
The overall performance using the auto-dilution method of the VITEK 2 System demonstrated an EA of 98.0% and a CA of 95.9%. There were two major errors (2/291 susceptible isolates, 0.7% error rate) and 23 minor errors (23/609 total isolates, 3.7% error rate). The major error and minor error rates were acceptable. With Enterobacteriaceae isolates, performance was acceptable with an EA of 98.6% and a CA of 96.0%. With P. aeruginosa isolates, performance was acceptable with an EA of 96.7% and a CA of 95.6%. The performance based on combined clinical and challenge data was acceptable. The overall performance is shown in Table 4. In addition, there was no significant difference in performance between indicated organisms only when compared to performance of all Enterobacteriaceae isolates tested (Tables 4.1 and 4.2).
To address testing of non-indicated species the sponsor included the following statement in the Precautions section of the device labeling:
Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.
Page 9 of 13
{9}
Table 4. Performance of Clinical and Challenge Isolates, VITEK 2 Auto-Dilution Method
| Organism Type | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #R | Min | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Enterobacteriaceae | | | | | | | | | | | | |
| Clinical | 281 | 275 | 97.9 | 136 | 130 | 95.6 | 267 | 95.0 | 72 | 12 | 2 | 0 |
| Challenge | 147 | 147 | 100 | 31 | 31 | 100 | 144 | 98.0 | 129 | 3 | 0 | 0 |
| Combined | 428 | 422 | 98.6 | 167 | 161 | 96.4 | 411 | 96.0 | 201 | 15 | 2 | 0 |
| P. aeruginosa | | | | | | | | | | | | |
| Clinical | 122 | 117 | 95.9 | 68 | 63 | 92.6 | 115 | 94.3 | 58 | 7 | 0 | 0 |
| Challenge | 59 | 58 | 98.3 | 30 | 29 | 96.7 | 58 | 98.3 | 30 | 1 | 0 | 0 |
| Combined | 181 | 175 | 96.7 | 98 | 92 | 93.9 | 173 | 95.6 | 88 | 8 | 0 | 0 |
| All species | | | | | | | | | | | | |
| Clinical | 403 | 392 | 97.3 | 204 | 193 | 94.6 | 382 | 94.8 | 130 | 19 | 2 | 0 |
| Challenge | 206 | 205 | 99.5 | 61 | 60 | 98.4 | 202 | 98.1 | 159 | 4 | 0 | 0 |
| Combined | 609 | 597 | 98.0 | 265 | 253 | 95.5 | 584 | 95.9 | 289 | 23 | 2 | 0 |
EA - Essential Agreement (± 1 dilution)
Min - minor discrepancies
CA - Category Agreement
Maj - major discrepancies
Eval - Evaluable isolates
Vmj - very major discrepancies
R - Resistant isolates
Essential agreement (EA) occurs when the result of the reference method and that of the VITEK card are within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the reference method and the VITEK card. Category agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation provided by the VITEK card.
Table 4.1: Performance of Enterobacteriaceae Species (Indicated Species Only), Auto-dilution Method
| Delafloxacin | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | Min | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Enterobacter cloacae | | | | | | | | | | | | |
| Clinical | 10 | 10 | 100 | 9 | 9 | 100 | 9 | 100 | 0 | 0 | 0 | 0 |
| Challenge | 18 | 18 | 100 | 9 | 9 | 100 | 17 | 94.4 | 13 | 1 | 0 | 0 |
| Combined | 28 | 28 | 100 | 18 | 18 | 100 | 27 | 96.4 | 13 | 1 | 0 | 0 |
| Escherichia coli | | | | | | | | | | | | |
| Clinical | 97 | 97 | 100 | 15 | 15 | 100 | 92 | 94.8 | 38 | 5 | 0 | 0 |
| Challenge | 12 | 12 | 100 | 3 | 3 | 100 | 12 | 100 | 10 | 0 | 0 | 0 |
| Combined | 109 | 109 | 100 | 18 | 18 | 100 | 104 | 95.4 | 48 | 5 | 0 | 0 |
| Klebsiella pneumoniae pneumoniae | | | | | | | | | | | | |
| Clinical | 4 | 4 | 100 | 2 | 2 | 100 | 4 | 100 | 1 | 0 | 0 | 0 |
| Challenge | 14 | 14 | 100 | 7 | 7 | 100 | 13 | 92.9 | 9 | 1 | 0 | 0 |
| Combined | 18 | 18 | 100 | 9 | 9 | 100 | 17 | 94.4 | 10 | 1 | 0 | 0 |
| Klebsiella pneumoniae | | | | | | | | | | | | |
| Clinical | 52 | 47 | 90.4 | 31 | 26 | 83.9 | 50 | 96.2 | 14 | 1 | 1 | 0 |
| Challenge | 103 | 103 | 100 | 12 | 12 | 100 | 102 | 99.0 | 97 | 1 | 0 | 0 |
| Combined | 155 | 150 | 96.8 | 43 | 38 | 88.4 | 152 | 98.1 | 111 | 2 | 1 | 0 |
| Enterobacteriaceae (all) | | | | | | | | | | | | |
| Clinical | 163 | 158 | 96.9 | 57 | 52 | 91.2 | 156 | 95.7 | 53 | 6 | 1 | 0 |
| Challenge | 147 | 147 | 100 | 31 | 31 | 100 | 144 | 98.0 | 129 | 3 | 0 | 0 |
| Combined | 310 | 305 | 98.4 | 88 | 83 | 94.3 | 300 | 96.8 | 182 | 9 | 1 | 0 |
{10}
Table 4.2: Performance of Delafloxacin, Comparison Between Indicated Species and All Species within Enterobacteriaceae
| | EA Total | EA# | EA% | Eval EA Tot | Eval EA# | Eval EA% | CA# | CA % | #R | Min | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Indicated Species | 310 | 305 | 98.4 | 88 | 83 | 94.3 | 300 | 96.8 | 182 | 9 | 1 | 0 |
| All Species within Enterobacteriaceae | 428 | 422 | 98.6 | 167 | 161 | 96.4 | 411 | 96.0 | 201 | 15 | 2 | 0 |
Challenge Data – Auto and Manual Dilution
The challenge data set was evaluated at one site with both the auto-dilution and manual dilution options of the VITEK 2 System and with the manual dilution method on the VITEK 2 Compact System. The performance was acceptable. Overall performance is shown in Table 5, and performances with $P$ aeruginosa and Enterobacteriaceae isolates are shown in Table 6 and Table 7, respectively.
Table 5. Performance of Challenge Isolates, VITEK 2 and VITEK 2 Compact – All Organisms
| Dilution | EA Total | EA % | EA % | Eval EA Total | Eval EA % | Eval EA % | CA % | CA % | #R | Min | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| VITEK 2 System | | | | | | | | | | | | |
| Auto | 206 | 205 | 99.5 | 61 | 60 | 98.4 | 202 | 98.1 | 159 | 4 | 0 | 0 |
| Manual | 206 | 203 | 98.5 | 61 | 58 | 95.1 | 200 | 97.1 | 159 | 5 | 1 | 0 |
| VITEK 2 Compact | | | | | | | | | | | | |
| Manual | 206 | 206 | 100 | 61 | 61 | 100 | 202 | 98.1 | 159 | 4 | 0 | 0 |
Table 6. Performance of Challenge Isolates, VITEK 2 and VITEK 2 Compact – $P$ aeruginosa
| Dilution | EA Total | EA % | EA % | Eval EA Total | Eval EA % | Eval EA % | CA % | CA % | #R | Min | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| VITEK 2 System | | | | | | | | | | | | |
| Auto | 59 | 58 | 98.3 | 30 | 29 | 96.7 | 58 | 98.3 | 30 | 1 | 0 | 0 |
| Manual | 59 | 57 | 96.6 | 31 | 29 | 93.5 | 57 | 96.6 | 30 | 2 | 0 | 0 |
| VITEK 2 Compact | | | | | | | | | | | | |
| Manual | 59 | 59 | 100 | 31 | 31 | 100 | 58 | 98.3 | 30 | 1 | 0 | 0 |
Table 7. Performance of Challenge Isolates, VITEK 2 and VITEK 2 Compact - Enterobacteriaceae
| Dilution | EA Total | EA % | EA % | Eval EA Total | Eval EA % | Eval EA % | CA % | CA % | #R | Min | Maj | Vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| VITEK 2 System | | | | | | | | | | | | |
| Auto | 147 | 147 | 100 | 31 | 31 | 100 | 144 | 98.0 | 129 | 3 | 0 | 0 |
| Manual | 147 | 146 | 99.3 | 30 | 29 | 96.7 | 143 | 97.3 | 129 | 3 | 1 | 0 |
| VITEK 2 Compact | | | | | | | | | | | | |
| Manual | 147 | 147 | 100 | 30 | 30 | 100 | 144 | 98.0 | 129 | 3 | 0 | 0 |
MIC Trends:
An analysis of trending was conducted using the combined clinical and challenge data for each organism group. This trending calculation considers MIC values that are determined to be one or more doubling dilutions lower or higher compared to the reference method regardless of whether the device MIC values are on-scale. Results
Page 11 of 13
{11}
that are not clearly at least one dilution lower, at least one dilution higher or in exact agreement with the CLSI reference method are not considered in the trending analysis.
Trending analysis results are shown in Table 8; results were stratified by species to assess species-related trends. Species for which the difference between the percentage of isolates with higher or lower readings was $\geq 30$ with a statistically significant confidence interval were considered to show evidence of trending. Trending that provides higher or lower MIC values compared to the reference method is addressed in labeling.
There was no trending noted for $P$ aeruginosa, however, when Enterobacteriaceae species were analyzed separately and combined, trends for higher MIC values compared to the reference were observed (Table 8). The following footnote was added to the performance table in the package insert to address trending:
MIC values tended to be in exact agreement or at least one doubling dilution higher when Enterobacteriaceae isolates compared to the reference micro-dilution.
Table 8. MIC Trending Analysis of All Organisms (Clinical and Challenge)
| Organism | Total Evaluable For Trending | ≥1 Dilution Lower No. (%) | Exact No. (%) | ≥1 Dilution Higher No. (%) | Percent Difference* (95% CI) | Trending Noted |
| --- | --- | --- | --- | --- | --- | --- |
| E. cloacae | 18 | 1 (5.56) | 7 (38.9) | 10 (55.6) | 50 (20.3 to 70.4) | Yes |
| E. coli | 23 | 0 (0) | 11 (47.8) | 12 (52.2) | 47.8 (24.3 to 67.0) | Yes |
| K. pneumoniae pneumoniae | 10 | 0 (0) | 7 (70) | 3 (30) | 30.0 (3.8 to 60.3) | Yes |
| K. pneumoniae | 42 | 3 (7.1) | 19 (45.2) | 20 (47.6) | (21.9 to 55.9) | Yes |
| All Enterobacteriaceae | 93 | 4 (4.3) | 44 (47.3) | 45 (48.4) | 44.1 (32.4 to 54.4) | Yes |
| P. aeruginosa | 102 | 11 (10.8) | 64 (62.8) | 27 (26.5) | 15.7 (5.0 to 26.1) | No |
* A positive percent difference indicates a higher MIC value compared to the reference method. A negative percent difference indicates a lower MIC value compared to the reference method.
b. Matrix comparison:
Not applicable
{12}
3. Clinical studies:
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Table 6: Interpretive Criteria for Delafloxacin (FDA STIC Webpage*)
| Organism | FDA Interpretive Criteria for Delafloxacin MIC (μg/mL) | | |
| --- | --- | --- | --- |
| | S | I | R |
| Enterobacteriaceae | ≤0.25 | 0.5 | ≥1 |
| Pseudomonas aeruginosa | ≤0.5 | 1 | ≥2 |
* FDA STIC Webpage
https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm
N. Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.