K183503 · Inrad, Inc. · NEU · Oct 17, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183503
Device Name
EasyMark Tissue Marker
Applicant
Inrad, Inc.
Product Code
NEU · General, Plastic Surgery
Decision Date
Oct 17, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Indications for Use
The EasyMark Tissue Marker is intended for use to attach to soft breast tissue at the surgical site during an open surgical breast biopsy or a percutaneous breast biopsy to radiographically mark the biopsy procedure.
Device Story
EasyMark Tissue Marker is a sterile, single-use device for marking breast biopsy sites. It consists of a disposable delivery system (plastic handle, 17 ga. stainless steel cannula, push rod) preloaded with a radiopaque marker. Available in two configurations: 316 LVM stainless steel (anchor-shaped) or Titanium 6Al-4V ELI (ribbon-shaped). Used by clinicians during breast biopsy procedures. The clinician inserts the cannula into the biopsy site and uses the push rod to deploy the marker into the soft tissue. The marker remains in the tissue to provide a radiographic landmark for future imaging (X-ray, MRI), aiding in the localization of the biopsy site for clinical follow-up or subsequent surgical planning.
Clinical Evidence
Bench testing only. Performance evaluations included marker deployment accuracy, deployment force, usability, imaging assessment (X-ray/MRI), MR safety/compatibility, migration potential, and tensile testing of the cannula and push rod. Biocompatibility, sterilization, packaging, and shelf-life testing were also conducted.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.
October 17, 2019
INRAD Inc. Steve Field CEO 4375 Donker Court SE Kentwood, Michigan 49512
Re: K183503
Trade/Device Name: EasyMark Tissue Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: September 16, 2019 Received: September 18, 2019
Dear Mr. Field:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K183503
Device Name EasyMark Tissue Marker
Indications for Use (Describe)
The EasyMark Tissue Marker is intended for use to attach to soft breast tissue at the surgical site during an open surgical breast biopsy or a percutaneous breast biopsy to radiographically mark the biopsy procedure.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the INRAD company logo. The logo consists of a blue square with a white diagonal line in the upper left corner, followed by the word "INRAD" in bold, black letters. A registered trademark symbol is located in the upper right corner of the logo.
# 510(k) Summary
| Date Prepared: | October 14, 2019 |
|-------------------------|------------------------------------------------------|
| Applicant: | INRAD Inc. |
| | 4375 Donker Court SE, |
| | Kentwood, MI 49512, USA |
| Official Correspondent: | Steve Field |
| | Tel. (616) 301-7800 / Fax. (616) 301-7799 |
| | sfield@inradinc.com |
| Trade Name: | EasyMark™ Tissue Marker |
| Common Name: | Implantable Radiographic Marker |
| Classification Name: | Implantable Clip (21 CFR 878.4300, Product Code NEU) |
| Device Classification: | Class II |
| Predicate Device: | UltraClip® Tissue Marker, K993785 |
## Device Description:
The EasyMark Tissue Marker is a sterile, single use device comprised of a disposable delivery device preloaded with a tissue marker. The disposable delivery device includes an introducer needle comprised of a plastic molded deployment handle, a thumb slide, a 17 ga. cannula with 1 cm depth marks and a push rod. The tissue marker is preloaded in the distal end of the cannula. This device offers the user the choice of two unique tissue markers (i) 316 LVM Stainless Steel Anchor shaped Marker (ii) Titanium 6AI-4V ELI Ribbon shaped Marker.
{4}------------------------------------------------
#### Indications for Use:
The EasyMark Tissue Marker is intended for use to attach to soft breast tissue at the surgical site during an open surgical breast biopsy or a percutaneous breast biopsy to radiographically mark the location of the biopsy procedure.
#### Technological Comparison to Predicate Device:
The EasyMark Tissue Marker has the same indications for use, principles of operation, fundamental scientific technology, and performance as that of predicate device UltraClip. In addition, SelectCore was chosen as the reference device to support the biocompatibility of the EasyMark's delivery device. The patient contact materials of the EasyMark (cannula and push rod) and SelectCore (cannula and stylet) are made of same material 304 Stainless Steel and have the same manufacturing processes. Therefore, a biological evaluation was conducted for the EasyMark delivery device in reference to SelectCore, to establish the biocompatibility of the delivery device.
| | Subject Device | Predicate Device | Reference Device |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description | EasyMark Tissue Marker | UltraClip Tissue Marker<br>(K993785) | SelectCore<br>(K093256) |
| Indications<br>for Use | The EasyMark Tissue<br>Marker is intended for use<br>to attach to soft breast<br>tissue at the surgical site<br>during an open surgical<br>breast biopsy or a<br>percutaneous breast<br>biopsy to radiographically<br>mark the location of the<br>biopsy procedure. | The UltraClip® Tissue<br>Marker is intended for use<br>to attach to soft breast<br>tissue at the surgical site<br>during an open surgical<br>breast biopsy or a<br>percutaneous breast biopsy<br>to radiographically mark the<br>location of the biopsy<br>procedure. | The SelectCore device is<br>intended for use in obtaining<br>biopsies from soft tissues<br>such as liver, breast, kidney,<br>prostate, spleen, lymph<br>nodes and various soft<br>tissue tumors. For breast<br>biopsy this product is for<br>diagnosis only - not for<br>therapeutic use. |
| Device | The EasyMark Tissue<br>Marker is a sterile, single<br>use device comprised of a<br>disposable delivery device<br>preloaded with a tissue<br>marker. | Same | The SelectCore is a sterile,<br>single use device comprised<br>of a disposable delivery<br>device. |
| Marker | Titanium 6Al-4V ELI - 3D<br>Ribbon shaped<br>316 LVM Stainless Steel -<br>Anchor shaped | Titanium 6Al-4V ELI -<br>Ribbon shaped<br>316 Stainless Steel -<br>Ribbon shaped | N/A |
| Markers -<br>Visualization<br>Techniques | X-ray, MRI | Same | X-ray |
{5}------------------------------------------------
| Description | Subject Device<br>EasyMark Tissue Marker | Predicate Device<br>UltraClip Tissue Marker<br>(K993785) | Reference Device<br>SelectCore<br>(K093256) |
|----------------------------|-----------------------------------------------|----------------------------------------------------------|--------------------------------------------------|
| Delivery<br>Device | Plastic Handle and<br>Stainless Steel Cannula | Same | Plastic Handle and Stainless<br>Steel Cannula |
| Needle<br>Dimensions | 17ga x 7cm<br>17ga x 10cm | 17ga x 10cm | 16ga x 10cm |
| Packaging | Poly-tyvek pouch | Same | Thermoformed tray (material<br>PETG) - Tyvek lid |
| Method of<br>Sterilization | Ethylene Oxide | Same | Same |
{6}------------------------------------------------
## Performance Data:
Non-clinical testing and evaluation was conducted to demonstrate that the EasyMark Tissue Marker is substantially equivalent to the predicate device. The following nonclinical testing and evaluations were performed:
- Performance Testing Bench ●
- Accuracy of Marker Deployment O
- Marker Deployment Force O
- Usability of the Device O
- Imaging Assessment (X-ray and MRI) O
- Safety and Compatibility in Magnetic Resonance (MR) Environment O
- Tissue Marker Migration Potential O
- Delivery Device Cannula Tensile Test o
- Delivery Device Push Rod Tensile Test
- Biocompatibility ●
- Sterilization ●
- Packaging ●
- Shelf Life ●
## Conclusion:
The results of non-clinical testing and evaluation demonstrates that the EasyMark Tissue Marker does not raise new concerns of safety or effectiveness and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.