K183489 · 3D Systems, Inc. · LLZ · Aug 29, 2019 · Radiology
Device Facts
Record ID
K183489
Device Name
D2P
Applicant
3D Systems, Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 29, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Anatomic segmentation
Deep learning neural network
—
—
—
—
—
—
Indications for Use
The D2P software is intended for use as a software interface and image segmentation system for the transfer of DICOM imaging information from a medical scanner to an output file. It is also intended as pre-operative software for surgical planning. For this purpose, the output file may be used to produce a physical replical replica is intended for adjunctive use along with other diagnostic tools and expert clinical judgement for diagnosis, patient management, and/or treatment selection of cardiovascular, craniofacial, genitourinary, neurological, and/or musculoskeletal applications.
Device Story
D2P is a standalone software package for DICOM image visualization, analysis, and segmentation. It processes medical scanner data to create output files for surgical planning and 3D printing of physical anatomical replicas. Used by clinicians in clinical settings, the software features automated segmentation via a deep learning neural network, DICOM viewing (axial, coronal, sagittal), and VR visualization tools. Output files are used to produce physical models on qualified 3D printers. The device assists healthcare providers in diagnosis, patient management, and treatment selection by providing enhanced anatomical visualization and physical replicas, serving as an adjunct to standard diagnostic tools and expert clinical judgment.
Clinical Evidence
Bench testing only. Validation included measurement accuracy and calculation 3D studies, usability studies, decimation studies, and validation of 3D printing accuracy for anatomical replicas across cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and musculoskeletal applications. All tests met pre-established specifications.
Technological Characteristics
Standalone modular software. Features DICOM viewer, automated segmentation (deep learning neural network), and VR visualization. Supports 3D printing integration. Connectivity includes DICOM data import. No specific hardware materials or energy sources listed as it is software-only.
Indications for Use
Indicated for patients requiring surgical planning or diagnostic support in cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and musculoskeletal applications. Used as an adjunct to clinical judgment for diagnosis, patient management, and treatment selection.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
August 28, 2019
3D Systems, Inc. % Kim Torluemke VP Quality & Regulatory, Healthcare 5381 South Alkire Circle LITTLETON, CO 80127
Re: K183489
Trade/Device Name: D2P Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: July 16, 2019 Received: July 18, 2019
Dear Kim Torluemke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183489
Device Name D2P
#### Indications for Use (Describe)
The D2P software is intended for use as a software interface and image segmentation system for the transfer of DICOM imaging information from a medical scanner to an output file. It is also intended as pre-operative software for surgical planning. For this purpose, the output file may be used to produce a physical replical replica is intended for adjunctive use along with other diagnostic tools and expert clinical judgement for diagnosis, patient management, and/or treatment selection of cardiovascular, craniofacial, genitourinary, neurological, and/or musculoskeletal applications.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the 3D Systems logo. On the left is a 3D cube with the letters '3D' on its faces. To the right of the cube is the text '3D SYSTEMS' in a simple, sans-serif font. The logo has a modern and corporate feel.
K183489
# 510(K) SUMMARY
# 1. INTRODUCTION
This document contains the 510(k) summary for the D2P software. The content of this summary is based on the requirements of 21 CFR 807.92.
# 2. SUBMITTER
3.
| Name: | Simbionix LTD a 3D Systems Company |
|----------------------|------------------------------------------------------------------------------------------------------------------------------|
| Address: | Beit Golan<br>Corner of Golan and Hanegev St.<br>Airport City, 70151, Israel<br>Phone: +972-3-9114444<br>Fax: +972-3-9114455 |
| Official Contact: | Kim Torluemke<br>Vice President, Quality and Regulatory, Healthcare |
| Date Prepared: | July 16, 2019 |
| DEVICE | |
| Trade Name: | D2P |
| Common Name: | Image processing system |
| Classification Name: | System, Image Processing, Radiological |
| Classification: | Class II, 21 CFR 892.2050 |
| Product Code: | LLZ |
### 4. PREDICATE DEVICE
The D2P software is claimed to be substantially equivalent to the following legally marketed predicate device:
- D2P, Simbionix LTD a 3D Systems Company (K161841)
Reference device:
- InPrint, Materials N.V. (K173619)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the 3D Systems logo. On the left is a 3D cube with the letters "3D" on the sides. To the right of the cube is the word "3DSYSTEMS" in a sans-serif font. The logo is in a gray color scheme.
K183489
# 5. DEVICE DESCRIPTION
The D2P software is a stand-alone modular software package that provides advanced visualization of DICOM imaging data. This modular package includes, but is not limited to the following functions:
- DICOM viewer and analysis
- Automated segmentation
- · Editing and pre-printing
- Seamless integration with 3D Systems printers
- · Seamless integration with 3D Systems software packages
- Seamless integration with Virtual Reality visualization for non-diagnostic use.
#### INDICATIONS FOR USE 6.
The D2P software is intended for use as a software interface and image segmentation system for the transfer of DICOM imaging information from a medical scanner to an output file. It is also intended as pre-operative software for surgical planning. For this purpose, the output file may be used to produce a physical replica. The physical replica is intended for adjunctive use along with other diagnostic tools and expert clinical judgement for diagnosis, patient management, and/or treatment selection of cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and/or musculoskeletal applications.
The Indications for Use statement for the D2P software is nearly identical to the predicate device. The primary difference, confirmed through bench testing, includes the ability to utilize the electronic output to print physical anatomic models on qualified 3D Printing technologies and materials for diagnostic use. Both the subject and predicate devices have the same intended use for visualization, analysis and segmentation of medical images and rendering 3D objects.
### 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The D2P software employs the same fundamental technologies as the identified predicate devices, including:
- Viewing of medical imaging data in the axial, coronal and sagittal views
- Ability to process, review and analyze medical imaging data;
- Image transfer and manipulation via software used for the creation of a 3D object;
The following technological differences exist between the subject and predicate devices:
- Incorporation of a deep learning neural network used to create the prediction of the segmentation.
- The ability to remove images from a DICOM dataset.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the 3D Systems logo. On the left is a 3D cube with the letters "3D" on its faces. To the right of the cube is the text "3D SYSTEMS" in a sans-serif font. The logo is in a gray color.
K183489
- The addition of a measurement tool into VR visualization.
- The subject device's intended use explicitly reflects 3D printing of the output file which can be used for diagnostic purposes in cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and/or musculoskeletal applications.
#### 8. PERFORMANCE DATA
#### Non-clinical tests
The D2P application has been validated for its intended use to determine substantial equivalence to the predicate device. A measurement accuracy and calculation 3D study, usability study, and decimation study were performed and confirmed to be within specification. Validation of printing of physical replicas was performed and demonstrated that anatomic models for cardiovascular, craniofacial, gastrointestinal, genitourinary, neurological, and/or musculoskeletal cases can be printed accurately when using any of the compatible 3D printers and materials.
#### Summary
All performance testing, which were conducted as a result of risk analyses and design impact assessments, showed conformity to pre-established specifications and acceptance criteria. The acceptance criteria were established in order to demonstrate device performance and substantial equivalence of the software to the predicate device.
### 9. CONCLUSIONS
Based on a comparison of the intended use and technological characteristics, the D2P software is substantially equivalent to the identified predicate device. Minor differences in technological and performance characteristics did not raise new or different questions of safety and effectiveness. Additionally, the non-clinical testing supports that the system performs in accordance with its intended use and is as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.