Direct anterior and posterior restorations (including occlusal surfaces) Core build-ups Splinting Indirect restorations including inlays, onlays, and veneers
Device Story
Visible-light activated restorative composite; used for anterior/posterior restorations, core build-ups, splinting, and indirect restorations. Input: restorative material dispensed via syringe (extraoral) or single-dose capsule (intraoral). Operation: material placed in tooth cavity; light-induced polymerization via photoinitiator system creates hard composite bonded to tooth structure using methacrylate-based dental adhesive. Used in dental clinics by dentists. Output: cured dental restoration. Benefits: provides strength, low wear, and esthetic tooth-colored restoration; pink opaque shade masks discolored tooth structure. Capsules may be warmed to 70°C/158°F prior to use.
Clinical Evidence
Bench testing only. Performance evaluated via physical property testing including Compressive Strength, Diametral Tensile Strength, Flexural Strength, Flexural Modulus, Surface Hardness, Radiopacity, Water Solubility, Volumetric Shrinkage, Wear, Depth of Cure, Cusp Deflection, and Polish Retention. Biocompatibility testing performed per ISO 10993 series and ISO 7405.
Technological Characteristics
Visible-light activated composite. Fillers: 20nm silica, 4-11nm zirconia, aggregated zirconia/silica clusters, and 100nm ytterbium trifluoride. Inorganic filler loading ~76.5% wt (58.4% vol). Resin matrix: AUDMA, AFM, diurethane-DMA, 1,12-dodecane-DMA. Form factor: syringes or single-dose capsules. Curing: light-induced polymerization. Standards: ISO 4049:2009, ISO 14971:2007, ISO 10993 series, ISO 7405:2008/Amd1 2013.
Indications for Use
Indicated for direct anterior and posterior restorations (including occlusal surfaces), core build-ups, splinting, and indirect restorations (inlays, onlays, veneers) in patients requiring dental restorative procedures.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
January 2, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
3M ESPE Dental Products % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, Nw Buffalo, Minnesota 55313
Re: K183476
Trade/Device Name: 3M Filtek Universal Restorative Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin material Regulatory Class: Class II Product Code: EBF Dated: December 13, 2018 Received: December 17, 2018
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Mary S. Mary S. Bunner -S3 Runner -S3 Date: 2019.01.02
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The logo is simple and recognizable, and it is often used on 3M products and marketing materials. The logo is a registered trademark of 3M Company.
#### Indications for Use Statement বঁ
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
Device Name
3MTM Filtek™ Universal Restorative
Indications for Use (Describe)
· Direct anterior and posterior restorations (including occlusal surfaces)
· Core build-ups
· Splinting
· Indirect restorations including inlays, onlays, and veneers
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Survices (301) 443-6740 BB
{3}------------------------------------------------
ЗМ ЕЅРЕ Dental Products 2510 Conway Avenue St. Paul, MN 55144-1000
Image /page/3/Picture/2 description: The image shows the 3M logo. The logo is in red and features the number 3 followed by the letter M. The logo is simple and recognizable.
### 510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.
| 510(k) Submitter | 3M ESPE Dental Products<br>2510 Conway Ave.<br>St. Paul, MN 55144 USA |
|----------------------------|-------------------------------------------------------------------------------------------------------------|
| Contact person | Lam Duong<br>Regulatory Affairs Associate<br>Phone: (651) 733-5945<br>Fax: (651) 736-1599<br>lduong@mmm.com |
| Date Summary was Prepared | 12Dec2018 |
| Trade Name | 3M™ Filtek™ Universal Restorative |
| Common Name(s) | Tooth shade resin material<br>Restorative |
| Recommended Classification | 21 CFR 872.3690<br>Tooth shade resin material<br>Product Code: EBF |
Predicate Device: Filtek™ Bulk Fill Posterior Restorative (K141081)
{4}------------------------------------------------
### Description of Device:
3MTM FiltekTM Universal Restorative, is a visible-light activated, restorative composite optimized to create esthetic anterior and posterior restorations. This material provides excellent strength and low wear for durability and improved esthetics with higher level of visual opacity. The shades are body like opacity enabling up to a 2mm depth-of-cure. The pink opaquer can be placed in 1mm thick increments. All shades are radiopaque. Filtek Universal Restorative is offered in the following shades: A1. A2, A3. A3.5. A4, B1. B2, D3. XW, and PO. The pink opaque shade option can be used to mask discolored or stained tooth structure.
Filtek™ Universal Restorative is packaged in traditional syringes, for dispensing restorative on a pad outside the mouth, and single-dose capsules for dispensing restorative intraorally. The capsules are dispensed using the 3M ESPE Restorative Dispenser. Capsules may be warmed up to 70°C/158°F for one hour in a commercial warmer prior to use.
## Indications for Use:
- · Direct anterior and posterior restorations (including occlusal surfaces)
- · Core build-ups
- Splinting
- · Indirect restorations including inlays, onlays and veneers
The proposed intended use of Filtek™ Universal Restorative is the same to that of the predicate device, Filtek™ Bulk Fill Posterior Restorative. The proposed indications for Filtek™ Universal Restorative are a subset to Filtek™ Bulk Fill Posterior Restorative indications, all of which are still covered within the intended use of a dental restorative. Filtek™ Universal Restorative is a modification of predicate device with no new ingredients introduced, thus it is expected that the performance of Filtek™ Univeral Restorative and its predicate will be equivalent. The differences do not affect the safety and effectiveness of the device. Therefore, the intended use and indications of Filtek™ Universal Restorative are not significantly different to those of FiltekTM Bulk Fill Posterior Restorative.
### Technological Characteristics:
Filtek Universal Restorative is a modification of predicate device. Filtek™ Bulk Fill Posterior Restorative, also manufactured by 3M ESPE Dental Products. The formulation was modified to provide a higher level of visual opacity enhancing the final esthetics of restorations.
The fillers are a combination of a non-agglomerated 20 nm silica filler, a nonagglomerated/non-aggregated 4 to 11 nm zirconia filler, an aggregated zirconia/silica cluster filler (comprised of 20 nm silica and 4 to 11 nm zirconia particles), and a ytterbium trifluoride filler consisting of agglomerate 100 nm particles. The inorganic filler loading is about 76.5% by weight (58.4% by volume). Filtek Universal Restorative contains AUDMA, AFM. diurethane-DMA, and 1, 12-dodecane-DMA. Filtek™M Universal Restorative is applied to the tooth following use of a methacrylate-based dental adhesive, such as manufactured by 3M ESPE, which permanently bonds the restoration to the tooth structure.
{5}------------------------------------------------
When irradiated by light, the methacrylate functionalities of the resins and fillers undergo, in conjunction with the photoinitiator system, a light-induced polymerization to form a hard composite that is bonded to the tooth structure with a permanent dental adhesive.
{6}------------------------------------------------
# Substantial Equivalence:
| Technological property | FiltekTM<br>Universal<br>Restorative | FiltekTM Bulk<br>Fill Posterior<br>Restorative<br>(K141081) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Photoinitiator system | X | X |
| Methacrylate-based resin matrix | X | X |
| Compatible with methacrylate-based dental adhesives | X | X |
| Inorganic fillers | X | X |
| Oxide fillers are silane treated so that they bond to the<br>resin matrix when the restorative is cured | X | X |
| Single-surface light-cure (up to 2 mm depth of cure for<br>body shades and up to 1mm depth of cure for opaque<br>shade)1 | X | – |
| Single-surface light-cure (up to 4 mm depth of cure) 2 | – | X |
| Multi-surface light-cure (5 mm depth of cure, Class II) 2 | – | X |
| When irradiated by light, the methacrylate<br>functionalities of the resins and fillers undergo, in<br>conjunction with the photoinitiator system, a light-<br>induced polymerization to form a hard composite that is<br>bonded to the tooth structure with a permanent dental<br>adhesive. | X | X |
| Dispensing system:<br>single-use capsule (intraoral)3<br>reusable syringe (extraoral)4 | X | X |
| Recommended for load-bearing occlusal surfaces | X | X |
| FDA-recognized standards followed | Risk Management:<br>ISO 14971:2007<br>Biocomp stds:<br>ISO 10993-1:2009<br>ISO 10993-3:2014<br>ISO 10993-5:2009<br>ISO 10993-6:2007<br>ISO 10993-<br>10:2010<br>ISO 10993-<br>11:2006<br>ISO 10993-<br>12:2012<br>ISO 7405:2008/<br>Amd1 2013<br>Product stds:5<br>ISO 4049:2009 | Risk Management:<br>ISO 14971:2007<br>Biocomp stds:<br>ISO 10993-1:2009<br>ISO 10993-3:2014<br>ISO 10993-5:2009<br>ISO 10993-6:2007<br>ISO 10993-10:2010<br>ISO 10993-11:2006<br>ISO 10993-12:2012<br>ISO 7405:2008/<br>Amd1 2013<br>Product stds:5<br>ISO 4049:2009<br>ISO 6874:2005 |
{7}------------------------------------------------
- 1. Curing protocol for < 2mm depth, with reduced light-cure time versus 4mm depth, was not included for Filtek™ Bulk Fill Posterior Restorative 510(k) submission. This difference is not significant since test data shows both Filtek™ Universal Restorative and the predicate device pass ISO 4049 depth of cure requirement when using the Filtek™ Universal Restorative light-cure recommendations.
- 2. The increased visual opacity due to changes made in pigment levels from the predicate device, Filtek™ Bulk Fill Posterior Restorative, makes Filtek™ Universal Restorative a traditional restorative, requiring 2mm incremental placement. The slightly lower pigment loading used in Filtek™ Bulk Fill Posterior Restorative makes it a bulk fill restorative and can be cured in bulk up to 4mm depth from a single surface and up to 5mm depth from three surfaces (occlusal, buccal and lingual). This change in physical property does not affect the safety or efficacy of the device when used as labeled. This difference is not significant since test data shows both Filtek™ Universal Restorative and the predicate device pass ISO 4049 depth of cure requirement.
- 3. Restorative material is dispensed from a single-use capsule in the mouth.
- 4. Restorative material is dispensed from a reusable syringe outside the mouth (e.g., onto a pad).
- 5. ISO 6874:2005 was not used to evaluate Filtek™ Universal Restorative because the product does not have a sealant indication. The only test in ISO 6874 that is applicable for a lightcure material such as the predicate device. Filtek™ Bulk Fill Posterior Restorative (K141081), is Depth of Cure. The Depth of Cure test is also included in ISO 4049:2009. This submission includes data showing both Filtek™ Universal Restorative and predicate device, Filtek™ Bulk Fill Posterior Restorative, readily pass the ISO 4049 Depth of Cure requirement. Therefore, this difference is not significant and does not affect the safety or efficacy of the device.
Test results for stability testing, biocompatibility testing, and the following physical properties were included in this submission: Compressive Strength, Diametral Tensile Strength, Flexural Strength, Flexural Modulus, Surface Hardness, Radiopacity, Water Solubility, Volumetric Shrinkage, Wear, Depth of Cure, Cusp Deflection and Polish Retention.
### Conclusion:
3MTM Filtek™ Universal Restorative is substantially equivalent to predicate device Filtek™ Bulk Fill Posterior Restorative in terms of intended use, indications for use, formulation, physical properties and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.