K183473 · Amsino International, Inc. · NGT · Jul 12, 2019 · General Hospital
Device Facts
Record ID
K183473
Device Name
AMSafe(R) Pre-Filled Normal Saline Flush Syringe
Applicant
Amsino International, Inc.
Product Code
NGT · General Hospital
Decision Date
Jul 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AMSafe® Pre-Filled Normal Saline Flush Syringe This device is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. It is intended for single use only.
Device Story
AMSafe® Pre-Filled Normal Saline Flush Syringe is a single-use, pre-filled polypropylene syringe containing 0.9% sodium chloride for injection, USP. The device is used by clinicians to flush compatible intravenous administration sets and indwelling intravenous access devices. It features a Luer lock connection and is terminally sterilized via gamma radiation. The device is available in 12mL (3mL, 5mL, 10mL fill) and 20mL (20mL fill) sizes. It is supplied in either a sterile blister package for use in a sterile field or a polypropylene wrapper for non-sterile field use. The device facilitates maintenance of intravenous access lines by providing a sterile saline flush, reducing the risk of occlusion and ensuring line patency.
Clinical Evidence
Bench testing only. Testing included ISO 7886-2017 syringe performance, USP 40 chemical/physical testing (pH, extractable metals, saline content, particulate matter), and biocompatibility (ISO 10993-4, 5, 10, 11). Packaging integrity was validated via ASTM F88/F88M-15 and ASTM F1929-15. All tests met acceptance criteria.
Technological Characteristics
Materials: Polypropylene (barrel, tip cap), butyl rubber (plunger). Design: Piston syringe with Luer lock fitting. Sterilization: Gamma radiation (10^-6 SAL). Shelf life: 2 years. Standards: ISO 7886-2017, USP <791>, <233>, <232>, <191>, <788>, <85>, ISO 10993-11, -10, -5, -4 (ASTM F756), ASTM F88/F88M-15, ASTM F1929-15, ISO 11607-1.
Indications for Use
Indicated for flushing compatible intravenous administration sets and indwelling intravenous access devices. For use by clinicians per manufacturer recommendations.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
AM USA 0.9% Sodium Chloride Flush Syringe (K133685)
Reference Devices
AM USA 0.9% Sodium Chloride Flush Syringe (K111034)
AM USA 0.9% Sodium Chloride Flush Syringe (K120836)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 12, 2019
Amsino International, Inc. Jim Barley RA/QA Consultant 708 Corporate Center Drive Pomona, California 91768
Re: K183473
Trade/Device Name: AMSafe® Pre-Filled Normal Saline Flush Syringe Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: NGT Dated: June 6, 2019 Received: June 13, 2019
Dear Jim Barley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183473
Device Name
AMSafe® Pre-Filled Normal Saline Flush Syringe
Indications for Use (Describe)
The AMSafe® 0.9% Sodium Chloride Pre-Filled Normal Saline Flush Syringe, is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. Use according to the recommendations of the manufacturer for the appropriate device.
Type of Use (Select one or both, as applicable)
| <span style="font-size:12px">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|------------------------------------------------------------------------------------|
| <span style="font-size:12px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for AMSINO, a healthcare company. The logo is in blue and white, with the company name in large, bold letters. Below the name is the tagline "Advancing Healthcare Worldwide." The logo is simple and professional, and it conveys a sense of trust and reliability.
Amsino International. Inc. 708 Corporate Center Dr. Pomona, CA91768 USA
Tel:909-626-5888 Fax:909-626-3888 Toll Free: 1-800-MD-AMSINO http://www.amsino.com email: amsino@amsino.com
# Traditional 510(k) Summary (As required by 21 CFR 807.92(a)) For K183473
## a) Submitter Information:
| Submitter: | Richard Lee, CEO of Amsino<br>Amsino International Inc.<br>708 Corporate Center Drive Pomona,<br>CA 91768<br>Phone: +1 (909)626-5888<br>Fax: +1 (909)626-3888 |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jim Barley, RA/QA Consultant<br>Cell phone: 949-4333058<br>jimbarley@aol.com |
| Date of Summary: | July 9, 2019 |
## b) Device Information:
| Trade or Proprietary Name: | AMSafe® Pre-Filled Normal Saline Flush Syringe |
|----------------------------|------------------------------------------------|
| Common or Usual Name: | Pre-Filled Normal Saline Flush Syringe |
| Regulation Number: | 21 CFR 880.5200 |
| Classification: | II |
| Product Code: | NGT |
### c) Identification of Leqally Marketed Device(s):
Predicate Device: AM USA 0.9% Sodium Chloride Flush Syringe, 510(k) number K133685, 20 cc Syringe with 20 cc fill volume
### Reference Devices:
- For Fluid Path Only Sterile: AM USA 0.9% Sodium Chloride Flush Syringe, 510(k) ● number K111034, 12 cc Syringe with 3, 5, 10 cc fill volumes
- For Device provided Sterile: AM USA 0.9% Sodium Chloride Flush Syringe, 510(k) ● number K120836, 12 cc Syringe with 3, 5, 10 cc fill volumes
### d) Device Description:
AMSafe® Pre-Filled Normal Saline Flush Syringe is a polypropylene plastic syringe filled with 0.9% sodium chloride for injection, USP, and capped with a polypropylene cap. The device will be terminally sterilized by gamma radiation sterilization. The device will be marketed as a 12mL syringe with a 3mL, 5mL, or 10mL fill volume, and a 20mL syringe with 20mL fill volume. The products are in two different packages, one is poly blister package and the entire packaged device are gamma radiation sterilized for sterile delivery to a sterile field; another is PP wrapper as a dust cover for non-sterile field.
The solution is sterile normal saline for injection and meets the requirements of
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for AMSINO. The logo is in blue and white, with the word "AMSINO" in large, bold letters. Below the word "AMSINO" is the phrase "Advancing Healthcare Worldwide." The logo is simple and professional, and it conveys a sense of trust and reliability.
Advancing Healthcare Worldwi
Amsino International. Inc. 708 Corporate Center Dr. Pomona, CA91768 USA
Tel:909-626-5888 Fax:909-626-3888 Toll Free: 1-800-MD-AMSINO http://www.amsino.com email: amsino@amsino.com
USP<40>.
#### e) Intended Use:
AMSafe® Pre-Filled Normal Saline Flush Syringe
This device is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices. It is intended for single use only.
### f) Technological Characteristics
Shown below is a side by side comparison of the subject device with the predicate device.
#### Table 5-1
| Device<br>Characteristic | Proposed device | Primary<br>Predicate<br>device<br>(K133685) | Reference<br>device 1<br>(K120836) | Reference<br>device 2<br>(K111034) | Results |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Indications for<br>Use | The AMSafe 0.9%<br>sodium chloride<br>pre- filled normal<br>saline flush<br>syringe, is<br>intended for use<br>in flushing<br>compatible<br>intravenous<br>administration<br>sets and<br>indwelling<br>intravenous<br>access devices.<br>Use according to<br>the recommend<br>at ions of the<br>manufacturer for<br>the appropriate<br>device. | Intended<br>use: 0.9%<br>sodium<br>chloride flush<br>syringes are<br>intended for<br>use in<br>flushing<br>compatible<br>intravenous<br>administratio<br>n sets and<br>indwelling<br>intravenous<br>access<br>devices. Use<br>according to<br>the<br>recommend<br>at ions of the<br>manufacturer<br>for the<br>appropriate<br>device | Intended<br>use: 0.9%<br>sodium<br>chloride flush<br>syringes are<br>intended for<br>use in<br>flushing<br>compatible<br>intravenous<br>administratio<br>n sets and<br>indwelling<br>intravenous<br>access<br>devices. Use<br>according to<br>the<br>recommend<br>at ions of the<br>manufacturer<br>for the<br>appropriate<br>device. | Intended<br>use: 0.9%<br>sodium<br>chloride<br>flush<br>syringes are<br>intended for<br>use in<br>flushing<br>compatible<br>intravenous<br>administrati<br>on sets and<br>indwelling<br>intravenous<br>access<br>devices.<br>Use<br>according to<br>the<br>recommend<br>at ions of<br>the<br>manufacturer<br>for the<br>appropriate<br>device. | Same |
| Design | Prefilled plastic<br>piston syringe with<br>Luer lock<br>connection | Prefilled<br>plastic piston<br>syringe with<br>Luer lock<br>connection | Prefilled<br>plastic piston<br>syringe with<br>Luer lock<br>connection | Prefilled<br>plastic piston<br>syringe with<br>Luer lock<br>connection | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "AMSINO" in large, bold, sans-serif font. The letters are white and are set against a blue background. There are two horizontal blue lines, one above and one below the word, that appear to be part of a logo.
Advancing Healthcare Worldw
Amsino International. Inc. 708 Corporate Center Dr. Pomona, CA91768 USA
Te1:909-626-5888 Fax:909-626-3888 Toll Free: 1-800-MD-AMSINO http://www.amsino.com email: amsino@amsino.com
| | fitting and<br>non-vented,<br>female Luer<br>lock tip cap | fitting and<br>non-vented,<br>female Luer<br>lock tip cap | fitting and<br>non-vented,<br>female Luer<br>lock tip cap | fitting and<br>non-vented,<br>female Luer<br>lock tip cap | |
|---------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------|---------|
| Syringe Size<br>and Fill<br>Volumes | 3ml, 5ml,<br>10ml in 12cc<br>syringe<br>20ml in 20cc<br>syringe | 20ml in 20cc<br>syringe | 3ml, 5ml,<br>10ml in 12cc<br>syringe | 3ml, 5ml,<br>10ml in 12cc<br>syringe | Similar |
| Fill Volume<br>Graduations | On syringe<br>label | On syringe<br>label | On syringe<br>label | On syringe<br>label | Same |
| Syringe<br>Content | 0.9% Sodium<br>chloride<br>injection,<br>USP | 0.9% Sodium<br>chloride<br>injection,<br>USP | 0.9% Sodium<br>chloride<br>injection,<br>USP | 0.9% Sodium<br>chloride<br>injection,<br>USP | Same |
| Labeled non-<br>pyrogenic | Yes | Yes | Yes | Yes | Same |
| Single<br>use only | Yes | Yes | Yes | Yes | Same |
| Sterile | Yes | Yes | Yes | Yes | Same |
| Devices<br>with Fluid<br>Path<br>Only Sterile<br>Or<br>Devices<br>provided<br>sterile | Devices with<br>Fluid Path<br>Only Sterile<br>Or Devices<br>provided<br>sterile | Devices<br>provided<br>sterile | Devices<br>provided<br>sterile | Devices with<br>Fluid Path<br>Only Sterile | Same |
| Shelf Life | 2 years | 2 years | 2 years | 2 years | Same |
| Sterilization<br>method | Terminally<br>sterilized by<br>Gamma<br>radiation, 10-6<br>SAL | Terminally<br>sterilized by<br>Gamma<br>radiation, 10-6<br>SAL | Terminally<br>sterilized by<br>Gamma<br>radiation, 10-6<br>SAL | Terminally<br>sterilized by<br>Gamma<br>radiation, 10-6<br>SAL | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for AMSINO. The logo is in blue and white. The word "AMSINO" is written in large, bold letters. There are two horizontal lines underneath the word.
Advancing Healthcare Worldy
Amsino International. Inc. 708 Corporate Center Dr. Pomona, CA91768 USA
Tel:909-626-5888 Fax:909-626-3888 Toll Free: 1-800-MD-AMSINO http://www.amsino.com email: amsino@amsino.com
| | USA | | | | |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|---------|
| | email: amsino@amsino.com | | | | |
| | Barrel and<br>plunger:<br>polypropyle ne | Barrel and<br>plunger:<br>polypropylene | Barrel and<br>plunger:<br>polypropylene | Barrel and<br>plunger:<br>polypropylene | |
| Syringe<br>material | Plunger: Butyl<br>rubber (not<br>made with<br>natural rubber<br>latex)<br>Tip cap:<br>polypropylene<br>with white<br>colorant | Plunger: Butyl<br>rubber (not<br>made with<br>natural rubber<br>latex)<br>Tip cap:<br>polypropylene<br>and TPE | Plunger: Butyl<br>rubber (not<br>made with<br>natural rubber<br>latex)<br>Tip cap:<br>polypropylene<br>and TPE | Plunger: Butyl<br>rubber (not<br>made with<br>natural rubber<br>latex)<br>Tip cap:<br>polypropyle<br>ne<br>and TPE | Similar |
| Syringe<br>packaging | PP wrap or<br>Sterile<br>barrier<br>Plastic peel<br>pouch | Sterile barrier<br>Plastic peel<br>pouch | Sterile barrier<br>Plastic peel<br>pouch | PP wrap | Same |
| Content of<br>Syringe<br>Package | One syringe per<br>pouch | One syringe<br>per pouch | One syringe<br>per pouch | One syringe<br>per pouch | Same |
# g) Summary of Non-clinical Testing(Bench):
The non-clinical testing for AMSafe® Pre-Filled Normal Saline Flush Syringe was performed to demonstrate verification and validation testing in conformance with the acceptance criteria of test methods and recognized consensus standards shown below. Table 5-2
| ID# | Test | Method | Acceptance criteria | Result/Conclusion |
|----------------------------------------|-------------------------------------------------------|-----------------|---------------------|-------------------|
| 1 | Plastic syringe | ISO7886-2017 | ISO7886-2017 | Conforms/Pass |
| Sodium Chloride Injection, USP Testing | | | | |
| 2 | pH value | USP40<791> | PH: 4.5-7.0 | Pass |
| | Limits of<br>extractable metals<br>of saline solution | USP<233>, <232> | USP<233>, <232> | Pass |
| | Chemical<br>Identification Tests | USP <191> | USP <191> | Pass |
| | 0.9% normal saline<br>content test | VP200 | 0.86% -- 0.94% | Pass |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "AMSINO" in large, white, sans-serif font. The word is set against a blue background. There are two horizontal white lines underneath the word, adding a sense of structure to the design.
Advancing Healthcare Worldwide
Amsino International. Inc. 708 Corporate Center Dr. Pomona, CA91768 USA
Tel:909-626-5888 Fax:909-626-3888 Toll Free: 1-800-MD-AMSINO http://www.amsino.com email: amsino@amsino.com
| | Oxidizable<br>substance test | VP200 | VP200 | Pass |
|---|----------------------------------------------|---------------------------|------------------------------------------------------------|------|
| | Iron test | USP40<241> | < 2ppm | Pass |
| | Ammonium | USP40<191> | USP40<191> | Pass |
| | Calcium | USP40<191> | USP40<191> | Pass |
| | Carbonate | USP40<191> | USP40<191> | Pass |
| | Sulfate | USP40<191> | USP40<191> | Pass |
| | Particulate matter | USP 40 <788> | ≥10um, ≤6000;<br>≥25um, ≤600. | Pass |
| 3 | Biocompatibility testing | | | |
| | Bacterial<br>endotoxins test | USP40<85> | Bacterial<br>endotoxins≤0.25<br>EU/mL | Pass |
| | Acute system<br>toxicity | ISO10993-11 | The device extracts did not<br>ellicit a systemic response | Pass |
| | Irritation /<br>Intracutaneous<br>reactivity | ISO10993-10 | Non-irritant | Pass |
| | Material-mediated<br>pyrogenicity | ISO10993-11 | Non-pyrogenic response | Pass |
| | Sensitization | ISO10993-10 | Non-sensitizer | Pass |
| | Cytotoxicity | ISO10993-5 | Non-cytotoxic | Pass |
| | Hemolysis | ISO10993-4<br>(ASTM F756) | Non-hemolytic | Pass |
| | Chemical<br>characterization | USP<232>,<br>USP<233> | Acceptable<br>extractable /<br>leachable profile | Pass |
| 4 | Blister package integrity | | | |
| | Seal strength test | ASTM F88/F88M-<br>15 | Should<br>not be<br>less than 2 N/ inch | Pass |
| | Dye integrity test | ASTM F1929-15 | ASTM F1929-15 | Pass |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for AMSINO. The logo is in blue and white, with the word "AMSINO" in large, bold letters. There are two horizontal lines above and below the word "AMSINO". The logo is simple and clean, and it is likely used to represent the company or organization.
Advancing Healthcare Worldwide®
Amsino International. Inc. 708 Corporate Center Dr. Pomona, CA91768 USA
Tel:909-626-5888 Fax:909-626-3888 Toll Free: 1-800-MD-AMSINO http://www.amsino.com email: amsino@amsino.com
The shelf life of the final finished sterilized device was evaluated using the recognized consensus standard on the requirements for materials, sterile barrier systems and packaging systems for terminally sterilized medical devices (ISO11607-1)
#### h) Conclusions:
The conclusions drawn from the nonclinical test that demonstrate that AMSafe® Pre-Filled Normal Saline Flush syringe is as safe, as effective, and performs as well as or better than the legally marketed predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.