K183416 · Olympus Medical Systems Corp. · EOB · Sep 6, 2019 · Ear, Nose, Throat
Device Facts
Record ID
K183416
Device Name
Rhino-Laryngo Videoscope Olympus ENF-VT3
Applicant
Olympus Medical Systems Corp.
Product Code
EOB · Ear, Nose, Throat
Decision Date
Sep 6, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment. This instrument is indicated for use within the nasal lumens and airway anatomy (including nasopharynx and trachea). This instrument is intended for use in a clinical or ambulatory medical environment.
Device Story
Rhino-Laryngo Videoscope ENF-VT3 is a flexible endoscope for visualization of nasal lumens and airway anatomy. Device inputs include reflected light from internal anatomy captured via distal objective lens; internal CCD converts light to electrical signals transmitted to external Olympus video system center for display on monitors. Operator (physician) controls distal end angulation (up/down 130°, right/left 70°) via control levers to navigate anatomy. Device includes instrument channel for EndoTherapy accessories and suction valve. Used in clinical/ambulatory environments for diagnosis and treatment. Output is real-time video feed assisting physician in anatomical assessment and procedural guidance. Benefits include improved visualization via higher resolution and 4-direction angulation, allowing navigation without rotating the entire device body, potentially reducing tissue trauma.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including reprocessing validation, biocompatibility (ISO 10993-1), software verification/validation, electrical safety (ANSI/AAMI ES 60601-1), EMC (IEC 60601-1-2), and performance bench testing (thermal, mechanical, resolution, phototoxicity).
Technological Characteristics
Flexible endoscope with 4-direction angulation (up/down 130°, right/left 70°). Features distal CCD sensor for image capture. Materials biocompatibility per ISO 10993-1. Electrical safety per ANSI/AAMI ES 60601-1 and IEC 60601-2-18; EMC per IEC 60601-1-2. Software includes pixel defect correction. Reprocessing validated per FDA guidance.
Indications for Use
Indicated for endoscopic diagnosis and treatment within nasal lumens and airway anatomy, including nasopharynx and trachea, in clinical or ambulatory settings.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
VISERA RHINO-LARYNGO VIDEOSCOPE OLYMPUS XENF TYPE VTY1 (K061313)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
September 6, 2019
Olympus Medical Systems Corp. % Daphney Germain-Kolawole Project Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, Pennsylvania 18034-0610
Re: K183416
Trade/Device Name: Rhino-Laryngo Videoscope Olympus ENF-VT3 Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOB, NWB Dated: August 7, 2019 Received: August 7, 2019
Dear Daphney Germain-Kolawole:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas Nandkumar, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K183416
#### Device Name RHINO-LARYNGO VIDEOSCOPE OLYMPUS ENF-VT3
Indications for Use (Describe)
RHINO-LARYNGO VIDEOSCOPE OLYMPUS ENF-VT3:
This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment is indicated for use within the nasal lumens and airway anatomy (including nasopharynx and trachea). This instrument is intended for use in a clinical or ambulatory medical environment.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10pt">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10pt">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are sans-serif and appear to be a custom font. A thin, horizontal gold line is located beneath the word. The registered trademark symbol is located to the right of the word.
# 510(k) Summary
# 1. GENERAL INFORMATION
- 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
- Contact Person: Daphney Germain-Kolawole Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610 Phone: 484-896-5691 Fax: 484-896-7128 Email: daphney.germain-kolawole@olympus.com
- Preparation Date: 9/6/2019
## 2. DEVICE IDENTIFICATION
- Device Name RHINO-LARYNGO VIDEOSCOPE OLYMPUS ENF-VT3
- Common Name RHINO-LARYNGO VIDEOSCOPE
II
- Regulation Number 874.4760, 876.1500
- Regulation Name Nasopharyngoscope (flexible or rigid) and accessories Endoscope and accessories
- · Regulatory Class
- Product Code EOB, NWB
- Classification Panel Ear Nose & Throat, Gastroenterology/Urology
{4}------------------------------------------------
## 3. PREDICATE DEVICE
#### ■ Predicate device
| Device name | 510(k) Submitter | 510(k) No. |
|--------------------------------------------------------------|----------------------------------|------------|
| VISERA RHINO-LARYNGO<br>VIDEOSCOPE OLYMPUS XENF<br>TYPE VTY1 | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K061313 |
#### 4. DEVICE DESCRIPTION
## ■ General Description of the subject device
This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment. This instrument is indicated for use within the nasal lumens and airway anatomy (including nasopharynx and trachea).
It is mainly consisted of four functional parts:
- video connection section including video connector and video cable that connect the endoscope to the video system center for displaying images on compatible video monitors
- light guide connector section that links the endoscope with a light source and transmits light to the distal end of the endoscope
- control section that operates the endoscope such as controlling angulation to guide insertion and observation, activating suction by single-use suction valve and performing treatment by means of allowing EndoTherapy accessories to enter into the instrument channel.
- insertion section that goes through and contacts with the nasal lumens and airway anatomy (including nasopharynx and trachea) under the light guide.
## ■ Principle of Operation
- a. Basic principle
- -Illumination light supply
The endoscope receives the illumination light from the light source by the light guide connector connected to the light source device. The illumination light is transferred to the distal end through the optical fiber bundle inside of the endoscope and illuminates the inside of the patient body through the illumination lens at the distal end.
{5}------------------------------------------------
#### Image construction -
The endoscope receives the reflected light from the inner lumen of a patient by the objective lens at the distal end. The built-in CCD at the distal end converts the light to the electrical signal, and the signal is sent to the video system center via the electrical cable and the video connector of the endoscope. The endoscope transfers the image signal and displays the observation image on the screen.
#### Bending operation -
UP/DOWN angulation control lever and RIGHT/LEFT angulation control lever and bending section are connected to each other by an independent wire. It is possible to bend in the up/down direction by operating the UP/DOWN angulation control lever, bend in theright/left direction by operating the RIGHT/LEFTLIGHT angulation control lever, respectively.
## b. Technological characteristics
- -4-direction angulation capability
Distal end is possible to bend in the UP/DOWN direction up to 130° and RIGHT/LEFT direction up to 70°, respectively.
- -User interface
In contrast to predicate device where the operations to combine the rotation of the device body and adjustment of angulation lever to reach an ideal observation site, this new subject device can alter the direction of the distal end by the angulation control levers solely to achieve the same observation effects and avoid potential damage caused by rotating the device body.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are closely spaced together, creating a solid block of text. The word appears to be a logo or brand name, and the overall impression is one of strength and reliability. There is a gold line underneath the word.
#### 5.INDICATIONS FOR USE
This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment. This instrument is indicated for use within the nasal lumens and airway anatomy (including nasopharynx and trachea). This instrument is intended for use in a clinical or ambulatory medical environment.
# 6. COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEIVCE
The ENF-VT3 has the same technological characteristics and design as the predicate device except for the following new features:
- 1. Changes on the optical system parameters (total number of pixels, resolution, etc.)
- 2. Material change of insertion section
- 3. Incorporated 4-direction angulation capability
- 4. Modify user interface to add RIGHT/LEFT operation
- 5. Software implementation to correct the pixel defect
#### Comments on the difference
1. The key parameter of observation is the image resolution. The subject device has been designed to improve the image resolution compared to the predicate device. This improvement does not affect the indication use. There is no problem in safety and effectiveness.
2. Biocompatibility test of the finished form was conducted, and the results support safety and effectiveness of the changed material.
3. The distal end is possible to bend in the UP/DOWN direction up to 130° and RIGHT/LEFT direction up to 70° respectively. We performed a risk assessment and determined that the residual risk is acceptable.
4. Although it seems the angulation range has increased in the subject device for the addition of the RIGHT/LEFT angulation, which is more likely to damage adjacent tissues around the observation site, we would like to emphasize that even for the original predicate device, operators are capable of not only changing the angulation by the UP/DOWN control lever inherently set in the device, but also adjusting the direction of the distal end by rotating the whole device by force outside the patient. However, in contrast to the predicate device where the operation to combine the rotation of the device body and adjustment of angulation lever to reach an ideal observation site, this new subject device can alter the direction of the distal end by angulation control levers solely to achieve the same observation effects and avoid the potential damage caused by rotating the device body. In addition, bench test results also demonstrate that the fundamental
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are spaced closely together, and the word is centered in the image. There is a registered trademark symbol to the right of the word. There is a gold line underneath the word.
function of the subject device has no changes and will not influence safety and effectiveness, and also risk management results considering human factors show that the residual risk is acceptable.
5. The function of the pixel defect correction was validated and it does not cause any new issue of safety and effectiveness.
All other technological characteristics of both the subject and predicate device are identical. Validation from non-clinical testing demonstrated that these technological features do not raise any new issues of safety or effectiveness of the subject device.
## 7. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
#### 1) Reprocessing validation testing
Reprocessing instruction and reprocessing method validation testing for the ENF-VT3 were conducted and documentation was provided as recommended by Guidance for Industry and Food and Drug Administration Staff, "Reprocessing Medical Devices in Health Care Setting: Validation Methods and Labeling".
#### 2) Biocompatibility testing
Biocompatibility testing for the ENF-VT3 were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:
- Cytotoxicity Study Using the Colony Assay ।
- । Intracutaneous Study in Rabbits
- Guinea Pig Maximization Sensitization Test
- Systemic Toxicity Study in Mice ।
## 3) Software verification and validation testing
Software verification and validation testing for the ENF-VT3 were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".
## 4) Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the ENF-VT3. The system complies with the ANSI/AAMI ES 60601-1:2005/(R)2012 and A1:2012 and IEC 60601-2-18:2009 standards for safety and the IEC 60601-1-2:2014 standards for
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are a sans-serif font and are evenly spaced. There is a thin yellow line underneath the word, which appears to be a design element. The registered trademark symbol is present to the right of the word.
EMC.
#### 5) Performance testing - Bench
Bench testing for the ENF-VT3 as listed below was conducted to ensure that the subject device performs as intended and meet design specifications.
- Thermal safety test
- Mechanical durability test
- Resolution test
- Phototoxicity test
## 6) Performance testing - Animal
No animal study was performed to demonstrate substantial equivalence.
## 7) Performance testing - Clinical
No clinical study was performed to demonstrate substantial equivalence.
#### 8) Risk analysis
Risk analysis for the ENF-VT3 was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2007 and the human factors validation was conducted in accordance with the FDA Guidance, "Applying Human Factors and Usability Engineering to Medical Devices". The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
## 8. CONCLUSION
Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate device, the ENF-VT3 raises no new issue of safety and effectiveness and is substantially equivalent to the predicate device in terms of safety, effectiveness and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.