# VITEK 2 AST-Gram Negative Imipenem (</=0.25->= 16 ug/mL) (K183415)

_bioMerieux, Inc. · LON · Feb 20, 2019 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K183415

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [LON](/productcode/LON.md)
- **Decision Date:** Feb 20, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1645
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

VITEK® 2 AST-Gram Negative Imipenem is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Negative Imipenem is a quantitative test. Imipenem has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Acinetobacter spp. Citrobacter spp. E.cloacae / E.cloacae complex Escherichia coli Klebsiella spp. Pseudomonas aeruginosa The VITEK® 2 Gram-negative Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.

## Device Story

VITEK 2 AST-Gram Negative Imipenem is an automated, quantitative in vitro diagnostic test for determining the minimum inhibitory concentration (MIC) of imipenem against aerobic Gram-negative bacilli. The system uses miniaturized test cards containing premeasured antimicrobial concentrations and nutrient media. Clinical laboratories prepare a standardized saline suspension of the bacterial isolate; the VITEK 2 system performs automated dilution, filling, and sealing, while the VITEK 2 Compact requires manual preparation. Cards are incubated within the instrument, which uses an optical scanner to monitor bacterial growth via light attenuation. The system's algorithm analyzes growth patterns to calculate MIC values and interpretive categories (Susceptible, Intermediate, Resistant). Results are reported to clinicians to aid in antimicrobial therapy selection. The device provides rapid susceptibility data, potentially improving patient outcomes by guiding targeted antibiotic treatment.

## Clinical Evidence

Bench testing only. Performance evaluated using clinical isolates (fresh and stock) and challenge strains compared to CLSI broth microdilution reference method. Overall performance: 95.9% Essential Agreement, 95.4% Category Agreement. Subgroup performance: Enterobacteriaceae (96.2% EA, 97.0% CA), Acinetobacter spp. (98.3% EA, 98.3% CA), and P. aeruginosa (94.4% EA, 91.0% CA). Reproducibility and Quality Control results were acceptable.

## Technological Characteristics

Miniaturized broth microdilution card containing Imipenem concentrations (≤0.25 - ≥16 μg/mL). Automated incubation and optical growth monitoring system. Compatible with VITEK 2 and VITEK 2 Compact platforms. Quantitative MIC determination.

## Regulatory Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

## Predicate Devices

- VITEK® 2 AST-GN Amikacin ([K172731](/device/K172731.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
K183415

B. Purpose for Submission:
To obtain a substantial equivalence determination for imipenem at concentrations of 0.25 – 16 µg/mL for susceptibility testing of gram-negative aerobic organisms on the VITEK 2 and VITEK 2 Compact Antimicrobial Susceptibility Test (AST) Systems.

C. Measurand:
Imipenem 0.25 – 16 µg/mL

D. Type of Test:
Automated quantitative antimicrobial susceptibility (AST)

E. Applicant:
bioMérieux, Inc.

F. Proprietary and Established Names:
VITEK 2 AST-Gram Negative Imipenem (≤0.25 - ≥16 µg/mL)

G. Regulatory Information:

1. Regulation section:
21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System

2. Classification:
Class II

3. Product code:
LON – Fully automated short-term incubation cycle antimicrobial susceptibility system

{1}

LTW – Susceptibility Test Cards, Automated

LTT – Panels, Test, Susceptibility, Automated

4. Panel:
83 Microbiology

H. Intended Use:

1. Intended use(s):
The VITEK 2 Gram-negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.

2. Indication(s) for use:
VITEK 2 AST-Gram Negative Imipenem is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 AST Gram Negative Imipenem is a quantitative test. Imipenem has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.

Active in vitro and in clinical infections:
Acinetobacter spp.
Citrobacter spp.
Enterobacter cloacae/E. cloacae complex
Escherichia coli
Klebsiella spp.
Pseudomonas aeruginosa

The VITEK 2 Gram-negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.

3. Special conditions for use statement(s):
Prescription use only

{2}

Limitations

Perform an alternative method of testing prior to reporting of results for the following antibiotic/organism combinations:

Imipenem (ipm05n): Klebsiella (Enterobacter) aerogenes, Proteus species, Providencia species, Morganella species and Serratia species.

4. Special instrument requirements:

VITEK 2 and VITEK 2 Compact Systems

I. Device Description:

The VITEK® 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK® 2 AST card contains 64 wells. A control well(s) which contain only nutrient medium is resident on all cards. The remaining wells contain premeasured portions of antimicrobials combined with the nutrient media. The isolate to be tested is diluted to a standardized concentration with 0.45% to 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System will automatically dilute the bacterial suspension to prepare an inoculum for susceptibility cards. Then the VITEK® 2 will fill, seal and place the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time (up to 24 hours for Streptococcus species). The analysis program determines when a well demonstrates growth based on attenuation of light measured by an optical scanner. This data is used to determine the minimum inhibitory concentration or MIC values for the anti-microbial agent. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.

VITEK 2 AST-GN Imipenem has the following concentrations in the card: 0.5, 2, 8, and 16 µg/mL (equivalent standard method concentration by efficacy in µg/mL).

The MIC result range for VITEK 2 AST-GN Imipenem on the VITEK 2 card is ≤ 0.25 - ≥ 16 µg/mL for Enterobacteriaceae and ≤ 0.5 - ≥ 16 µg/mL for Acinetobacter spp. and P. aeruginosa.

J. Substantial Equivalence Information:

1. Predicate device name(s):

VITEK 2 AST-GN Amikacin

2. Predicate 510(k) number(s):

K172731

{3}

# 3. Comparison with predicate:

Table 1. Comparison with the Predicate Device

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device K183425 VITEK 2 AST-GN Imipenem | Predicate K172731 VITEK 2 AST-GN Amikacin  |
|  Intended Use | VITEK 2 AST-Gram Negative Imipenem is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 AST Gram Negative Imipenem is a quantitative test. Imipenem has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Acinetobacter spp. Citrobacter spp. Enterobacter cloacae/E. cloacae complex Escherichia coli Klebsiella spp. Pseudomonas aeruginosa | VITEK 2 AST Gram Negative Amikacin is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 AST Gram Negative Amikacin is a quantitative test. Amikacin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Pseudomonas spp. Escherichia coli Proteus mirabilis Klebsiella spp. Enterobacter spp. Serratia spp. Acinetobacter species (excluding A. baumannii Complex) In vitro data available but clinical significance is unknown: Citrobacter freundii  |
|  Test Methodology | Automated quantitative | Same  |

{4}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device K183425 VITEK 2 AST-GN Imipenem | Predicate K172731 VITEK 2 AST-GN Amikacin  |
|   | antimicrobial susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems to determine the in vitro susceptibility of Gram negative bacilli |   |
|  Inoculum | Saline suspension of organism | Same  |
|  Test Card | VITEK 2 Gram Negative Susceptibility Test Card | Same  |
|  Instrument | VITEK 2 and VITEK 2 Compact | Same  |
|  Analysis Algorithm | Growth Pattern Analysis | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device K183425 VITEK 2 AST-GN Imipenem | Predicate K172731 VITEK 2 AST-GN Amikacin  |
|  Antimicrobial Agent | Imipenem | Amikacin  |
|  Antimicrobial Concentrations | 0.5, 2, 8, 16 | 2, 4, 16, 48  |

# K. Standard/Guidance Document Referenced (if applicable):

CLSI Document M07-A10, Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard - Tenth Edition, Vol. 35, No. 2; January, 2015.

Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems Guidance for Industry and FDA, August 2009.

# L. Test Principle:

The VITEK® 2 and VITEK® 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics used in the systems use visible light to directly measure organism growth. Transmittance optics are based on an initial light reading of a well before significant growth has begun. Periodic light transmittance samplings of the same well measure organism growth by how much light is prevented from going through the well. The VITEK 2 System monitors the growth of each well in the card over a defined period of time. An interpretive call is made between 4 and 16

{5}

hours for a “rapid” read but may be extended to 18 hours in some instances. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic on the card.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

A reproducibility study was performed at three external sites using a panel comprised of 10 isolates from indicated species: *P. aeruginosa* (three isolates) and *K. pneumoniae* (seven isolates). All isolates were tested in triplicate at each of the testing sites with each inoculation method and with the VITEK 2 and the VITEK 2 Compact. The mode MIC was determined and the reproducibility was calculated based on MIC values falling within ± 1 dilution of the mode MIC value. Reproducibility was greater than 95% for all inoculation and read methods and was considered to be acceptable.

#### b. Linearity/assay reportable range:

N/A

#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Quality control strains recommended by the CLSI were tested with imipenem at four sites. The QC organisms tested were *E. coli* ATCC 25922 and *P. aeruginosa* ATCC 27853. The QC strains were tested a minimum of 20 time per site and inoculated using both the automated dilution and manual dilution for VITEK 2 and using manual dilution for the VITEK 2 Compact. Testing with *E. coli* ATCC 25922 provided off scale results with the VITEK 2 card as the acceptable range for this strain is at or lower than the lowest dilution on the card. The sponsor included the following footnote to the *E. coli* ATCC 25922 expected range in the device labeling QC table:

&gt; FDA/CLSI Broth Microdilution expected QC range = 0.06 – 0.25 μg/mL. Does not include the full recommended dilution range for QC testing with this organism.

As an additional check of the reference method, two gram positive organisms were tested throughout the study at each clinical site. Isolates tested were *E. faecalis* ATCC 29212 and *S. aureus* ATCC 29213. All results were within the expected range.

The results demonstrate that the VITEK 2 can provide quality control results within the recommended ranges for all inoculation methods &gt; 95% of the time (Table 2).

6

{6}

Table 2. Quality Control Results for Imipenem with VITEK 2 (Automated and Manual Dilution) and VITEK 2 Compact (Manual Dilution)

|  Organism | Conc. (μg/mL) | VITEK 2 Automatic Dilution |   | VITEK 2 Manual Dilution |   | VITEK 2 Compact Manual Dilution  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  Test | Ref.ª | Test | Ref. | Test | Ref.  |
|  E. coli
ATCC 25922
0.06 – 0.25 | 0.125 | - | 46 | - | 42 | - | 42  |
|   |  0.25ᵇ | 115 | 68 | 90 | 48 | 90 | 48  |
|   |  0.5 | 0 | 1 |  |  |  |   |
|   |  1 |  |  |  |  |  |   |
|   |  2 |  |  |  |  |  |   |
|   |  4 |  |  |  |  |  |   |
|   |  8 |  |  |  |  |  |   |
|   |  ≥16 |  |  |  |  |  |   |
|  |   |   |   |   |   |   |   |
|  P. aeruginosa
ATCC 27853
1-4 μg/mL | ≤0.25 |  |  |  |  |  |   |
|   |  0.5 |  |  |  |  |  |   |
|   |  1 |  | 5 |  | 3 |  | 3  |
|   |  2 | 116 | 96 | 92 | 79 | 92 | 79  |
|   |  4 |  | 15 |  | 10 |  | 10  |
|   |  8 |  |  |  |  |  |   |
|   |  ≥16 |  |  |  |  |  |   |

ª Reference panel
ᵇ Lowest dilution of imipenem on the VITEK 2 GN card

Inoculum Density Check. The inoculum density was monitored using the DensiCHEK Plus instrument. The DensiCHEK Plus was standardized weekly with all results recorded and in expected range.

Purity Check: A purity check of all organisms was performed at the time of VITEK2 card inoculation. Only results obtained with pure cultures were evaluated.

Growth Failure Rate: One clinical isolate of *P. aeruginosa* failed to grow in the VITEK 2 AST card.

d. Detection limit:
N/A

e. Analytical specificity:
N/A

{7}

f. Assay cut-off:

N/A

2. Comparison studies:

a. Method comparison with predicate device:

Testing of imipenem on the VITEK AST gram-negative card was performed at three external sites and one internal site. Results obtained with the VITEK 2 AST-Gram Negative card were compared to results obtained with the CLSI broth microdilution reference panel using direct colony suspension inoculation. Reference panels were incubated at 35°C for 16 to 20 hours for Enterobacteriaceae and P. aeruginosa and 20 to 24 hours for Acinetobacter spp.

Test inocula for VITEK 2 AST were standardized using the DensiCHEK Plus instrument. The DensiCHEK instrument was calibrated within specified intervals; all recorded calibration values were within acceptable parameters. VITEK 2 AST-Gram Negative cards were inoculated using automatic dilution (for reading on the VITEK 2 instrument) or using a manual dilution method (for reading on the VITEK 2 instrument or on the VITEK 2 COMPACT instrument).

A total of 385 clinical isolates were evaluated, 379 of these were from indicated species. A total of 239 (61.9%) isolates were fresh or recent isolates (tested within one year of isolation from clinical specimens); 146 (37.8%) were stock. Clinical isolates included: Acinetobacter spp. (36 isolates), P. aeruginosa (122 isolates), E. coli (97 isolates), Citrobacter spp. (13 isolates), Enterobacter cloacae/E. cloacae complex (32 isolates), Klebsiella spp. (79 isolates). An additional six isolates of non-indicated species were tested. One clinical isolate of P. aeruginosa failed to grow in the VITEK 2 AST-GN card.

A total of 218 challenge isolates were tested at one clinical site using both automatic and manual dilution methods for the VITEK 2 System and manual dilution for the VITEK 2 Compact System. Challenge isolates tested included: Acinetobacter spp. (23 isolates), P. aeruginosa (55 isolates), E. coli (6 isolates), Citrobacter spp. (1 isolate), Enterobacter cloacae/E. cloacae complex (18 isolates), Klebsiella spp. (115 isolates).

For the VITEK 2 with automatic dilution, the MIC results obtained with clinical and challenge isolates of Enterobacteriaceae demonstrated an EA of 96.2% and CA of 97.0%, EA of 98.3% and CA of 98.3% for Acinetobacter spp. and EA of 94.4% and CA of 91.0% for P. aeruginosa. For all organism groups combined the MIC results demonstrated an EA of 95.9% and CA of 95.4%. The majority of errors for all organisms were minor errors. There were 100 resistant P. aeruginosa isolates by the reference method; the VITEK 2 with automatic dilution showed a very major error rate of 2.0% for this organism, which is considered acceptable. The EA of evaluable results was low for all organism groups (88.1%, 83.3% and 88.9% for

8

{8}

Enterobacteriaceae, Acinetobacter spp. and  $P$  aeruginosa, respectively); the low EA of evaluable results for Acinetobacter spp. was due to the large number of isolates showing MICs  $\geq 16$  for the reference method and the test device (Table 3).

For VITEK 2 with manual dilution the MIC results obtained with challenge isolates demonstrated an EA of  $98.6\%$  and CA of  $97.9\%$  for Enterobacteriaceae, EA and CA of  $95.7\%$  for Acinetobacter spp., and EA of  $96.4\%$  and CA of  $94.5\%$  for  $P.$  aeruginosa. The EA of evaluable results for Acinetobacter spp. was  $80.0\%$  due to the majority of isolates (20/23) having MICs  $\geq 16$  for both the reference method and test device resulting in a low number of evaluable results. (Table 4).

For VITEK 2 Compact with manual dilution, the MIC results obtained with challenge isolates demonstrated an EA of  $99.3\%$  and CA of  $97.9\%$  for Enterobacteriaceae, EA and CA of  $95.7\%$  for Acinetobacter spp., and EA of  $96.4\%$  and CA of  $94.5\%$  for  $P.$  aeruginosa. The EA of evaluable results for Acinetobacter spp. was  $80.0\%$  due to the majority of isolates (20/23) having MICs  $\geq 16$  for both the reference method and test device (Table 5).

The sponsor included the following limitation in the device labeling to address testing for species for which no performance data was provided for FDA review in this submission:

Perform an alternative method of testing prior to reporting of results for the following antibiotic/organism combinations: Imipenem (ipm05n): Klebsiella (Enterobacter) aerogenes, Proteus species, Providencia species, Morganella species and Serratia species.

Table 3. Performance of Imipenem, VITEK 2 Automatic Dilution, Clinical and Challenge Isolates

|  Organism Type | Tot | No. EA | EA % | Eval EA Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R | No. S | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Enterobacteriaceae (Citrobacter spp., E. cloacae/E. cloacae complex, E. coli, Klebsiella spp.)  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 227 | 215 | 94.7% | 54 | 42 | 77.8% | 219 | 96.5% | 8 | 212 | 8 | 0 | 0  |
|  Challenge | 140 | 138 | 98.6% | 55 | 54 | 98.2% | 137 | 97.9% | 120 | 24 | 3 | 0 | 0  |
|  Combined | 367 | 353 | 96.2% | 109 | 96 | 88.1% | 356 | 97.0% | 128 | 236 | 11 | 0 | 0  |
|  Acinetobacter spp.  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 36 | 36 | 100.0% | 1 | 1 | 100.0% | 36 | 100.0% | 19 | 17 | 0 | 0 | 0  |
|  Challenge | 23 | 22 | 95.7% | 5 | 4 | 80.0% | 22 | 95.7% | 20 | 3 | 1 | 0 | 0  |
|  Combined | 59 | 58 | 98.3% | 6 | 5 | 83.3% | 58 | 98.3% | 39 | 20 | 1 | 0 | 0  |
|  P. aeruginosa  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 122 | 114 | 93.4% | 68 | 60 | 88.2% | 109 | 89.3% | 61 | 51 | 11 | 0 | 2  |
|  Challenge | 55 | 53 | 96.4% | 22 | 20 | 90.9% | 52 | 94.5% | 39 | 15 | 3 | 0 | 0  |
|  Combined | 177 | 167 | 94.4% | 90 | 80 | 88.9% | 161 | 91.0% | 100 | 66 | 14 | 0 | 2  |
|  All organisms  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 385 | 365 | 94.8% | 123 | 103 | 83.7% | 364 | 94.5% | 88 | 280 | 19 | 0 | 2  |
|  Challenge | 218 | 213 | 97.7% | 82 | 78 | 95.1% | 211 | 96.8% | 179 | 42 | 7 | 0 | 0  |
|  Combined | 603 | 578 | 95.9% | 205 | 181 | 88.3% | 575 | 95.4% | 267 | 322 | 26 | 0 | 2  |

{9}

EA - Essential Agreement (± 1 dilution)

CA - Category Agreement

EVAL - Evaluable isolates

R - Resistant isolates

min - minor discrepancies

maj - major discrepancies

vmj - very major discrepancies

Essential agreement (EA) occurs when the result of the reference method and that of the VITEK card are within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the reference method and the VITEK card. Category agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation provided by the VITEK card.

Table 4. Performance of Imipenem, VITEK 2, Manual Dilution, Challenge Isolates

|  Organism Type | Tot | No. EA | EA % | Eval EA Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R | No. S | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Enterobacteriaceae (Citrobacter spp., E. cloacae/E. cloacae complex, E. coli, Klebsiella spp.)  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Challenge | 140 | 138 | 98.6% | 51 | 50 | 98.0% | 137 | 97.9% | 120 | 20 | 3 | 0 | 0  |
|  Acinetobacter spp.  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Challenge | 23 | 22 | 95.7% | 5 | 4 | 80.0% | 22 | 95.7% | 20 | 3 | 1 | 0 | 0  |
|  P. aeruginosa  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Challenge | 55 | 53 | 96.4% | 23 | 21 | 91.3% | 52 | 94.5% | 39 | 15 | 2 | 0 | 0  |

Table 5. Performance of Imipenem, VITEK 2, Compact, Challenge Isolates

|  Organism Type | Tot | No. EA | EA % | Eval EA Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R | No. S | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Enterobacteriaceae (Citrobacter spp., E. cloacae/E. cloacae complex, E. coli, Klebsiella spp.)  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Challenge | 140 | 139 | 99.3% | 64 | 63 | 98.4% | 137 | 97.9% | 120 | 19 | 3 | 0 | 0  |
|  Acinetobacter spp.  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Challenge | 23 | 22 | 95.7% | 5 | 4 | 80.0% | 22 | 95.7% | 20 | 3 | 1 | 0 | 0  |
|  P. aeruginosa  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Challenge | 55 | 53 | 96.4% | 22 | 20 | 90.9% | 52 | 94.5% | 39 | 15 | 3 | 0 | 0  |

To address testing of non-indicated species the sponsor included the following statement in the Precautions section of the device labeling:

Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.

Device failure: During the clinical study two episodes of device failure occurred; both were resolved with an instrument restart. Isolates being processed at the time of device failure were discarded and retested.

{10}

Resistance Mechanisms. Challenge isolates of *P. aeruginosa* and *Enterobacteriaceae* harboring the following resistance mechanisms were evaluated: KPC, OXA, IMP, VIM, SPM, modified OprD, VEB, CARB-3, TEM, PDC, GES, PER, NDM, OmpK, OmpC, ACT/MIR, IMI, VEB-1B, and SHV.

## MIC Trending

An analysis of trending was conducted using the combined clinical and challenge data for each organism group. This trending calculation takes into account MIC values that are determined to be one or more doubling dilutions lower or higher compared to the reference method irrespective of whether the device MIC values are on-scale or not. Results that are not clearly at least one dilution lower, at least one dilution higher or in exact agreement with the CLSI reference method are not considered in the trending analysis.

Trending results are shown in Table 6; results were stratified by species to determine if species-related trends were observed. Species for which the difference between the percentage of isolates with higher vs. lower readings was ≥30 and for which the confidence interval was determined to be statistically significant were considered to show evidence of trending. Trending that provides higher or lower MIC values compared to the reference is addressed in labeling.

A trend toward higher MIC readings was observed for E. coli with the autodilution inoculation method on VITEK 2, and for *E. cloacae/E. cloacae* complex with all dilution methods on VITEK 2 and VITEK 2 Compact. A trend toward lower MIC readings was observed for *Klebsiella* spp. for manual dilution with VITEK 2 and VITEK 2 Compact (Table 6). The sponsor included the following footnote to the performance table to address the trending observed for imipenem.

Overall, MIC values for *Enterobacteriaceae* were within one dilution of the CLSI reference broth microdilution method. However, imipenem MIC values tended to be in exact agreement or at least one dilution higher when testing E. coli with autodilution, *E. cloacae/E. cloacae* complex with manual dilution and autodilution and at least one dilution lower when testing *Klebsiella* spp. with manual dilution as compared to the CLSI reference broth microdilution

11

{11}

Table 6. MIC Trending for all Dilution and Read Methods

|  Inoculation/ Read Method | Organism | Total Evaluable for Trending | ≥1 Dilution lower No. (%) | Exact No. (%) | ≥1 Dilution Higher No. (%) | Percent Difference (CI) | Trending Noted  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Autodilution/ VITEK 2 | Acinetobacter spp. | 6 | 0 | 4 (66.7) | 1 (33.3) | 33.3 (-12.3-70.0) | Yesa  |
|   |  Citrobacter spp. | 8 | 2 (25.0) | 4 (50.0) | 2 (25.0) | 0 | No  |
|   |  E. cloacae/E. cloacae complex | 40 | 1 (2.5) | 21 | 18 (45.0) | 42.5 (24.8-57.8) | Yes  |
|   |  E. coli | 10 | 2 (20.0) | 1 (10.0) | 7 (70.0) | 50.0 (6.7-74.0) | Yes  |
|   |  Klebsiella spp. | 50 | 16 (32.0) | 22 (44.0) | 12 (24.0) | -8.0 | No  |
|   |  Enterobacteriaceae | 108 | 21 (19.44) | 48 (44.4) | 39 36.1) | 16.7 | No  |
|   |  P. aeruginosa | 90 | 26 (28.9) | 36 (40.0) | 28 (31.1) | 2.2 | No  |
|  Manual Dilution VITEK 2 | Acinetobacter spp. | 5 | 1 (20.0) | 2 (40.0) | 2 (40.0) | 20.0 | No  |
|   |  Citrobacter spp. | 0 | - | - | - | - | -  |
|   |  E. cloacae/E. cloacae complex | 13 | 0 | 6 (46.2) | 7 (53.9) | 53.9 (20.2-76.8) | Yes  |
|   |  E. coli | 3 | 0 | 1 (33.3)0 | 2 (66.7) | 66.7 (-5.9-93.9) | Yesa  |
|   |  Klebsiella spp. | 36 | 16 (44.4) | 16 (44.4) | 4 (11.1) | -33.3 (-50.7 - -12.8) | Yes  |
|   |  Enterobacteriaceae | 52 | 16 (30.8) | 23 (44.2) | 13 (25.0) | -5.8 | No  |
|   |  P. aeruginosa | 23 | 6 (26.1) | 10 (43.5) | 7 (30.4) | 4.5 | No  |
|  Manual Dilution Compact | Acinetobacter spp. | 5 | 1 (20.0) | 3 (60.0) | 1 (20.0) | 0 | No  |
|   |  Citrobacter spp. | 0 | - | - | - | - | -  |
|   |  E. cloacae/E. cloacae complex | 15 | 0 | 7 (46.7) | 8 (53.3) | 53.3 (22.4-75.2) | Yes  |
|   |  E. coli | 3 | 0 | 1 (33.3) | 2 (66.7) | 66.7 (-5.9-93.9) | Yesa  |
|   |  Klebsiella spp. | 46 | 20 (43.5) | 22 (47.8) | 4 (8.7) | -34.8 (-5.0- -17.1) | Yes  |
|   |  Enterobacteriaceae | 64 | 20 (31.2) | 30 (46.9) | 14 (21.9) | -9.4 | No  |
|   |  P. aeruginosa | 22 | 6 (27.3) | 9 (40.9) | 7 (31.8) | 4.6 | No  |

a Not statistically significant
b. Matrix comparison:

N/A

3. Clinical studies:

a. Clinical Sensitivity:

N/A

{12}

b. Clinical specificity:

N/A

c. Other clinical supportive data (when a. and b. are not applicable):

N/A

4. Clinical cut-off:

N/A

5. Expected values/Reference range:

Table 7. Interpretive Categories for Imipenem (FDA STIC Webpage and CLSI M100)

|  Organism | Interpretive Categories for Imipenem MIC (μg/mL)a  |   |   |
| --- | --- | --- | --- |
|   |  S | I | R  |
|  Enterobacteriaceae | ≤1 | 2 | ≥4  |
|  P. aeruginosa | ≤2 | 4 | ≥8  |
|  Acinetobacter spp. | ≤2 | 4 | ≥8  |

a FDA STIC Webpage https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm

# N. Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device

# O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K183415](https://fda-staging.innolitics.com/device/K183415)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).