The QNext is a fully-automated, random-access instrument, intended for in vitro diagnostic use in clinical laboratories to perform hemostasis testing by detecting the changes in optical density. DG-PT is a thromboplastin reagent for the quantitative determination of Prothrombin Time on human plasma samples collected in 3.2% sodium citrate. The product is used for the evaluation of the extrinsic and common coagulation pathways in seconds and for the monitoring Oral Anticoagulant Therapy with warfarin in International Normalized Ratio (INR). For use with QNext. For clinical professional laboratory and prescription use only. For in vitro diagnostic use. The performance of this device has not been established in neonate and pediatric patient populations.
Device Story
QNext is a fully-automated, random-access hemostasis analyzer used in clinical laboratories by professional staff. It processes human citrated plasma samples using DG-PT reagent (rabbit brain thromboplastin). The device automates sample/reagent loading, dilution, incubation, and optical density measurement of coagulation reactions. It calculates analytical parameters from reaction curves to determine PT in seconds and INR. Results are transmitted to a Laboratory Information System (LIS). By automating these steps, the device reduces operator error, minimizes contamination risk, and improves throughput. It supports urgent sample analysis and quality control. The system provides clinicians with standardized coagulation data to evaluate extrinsic/common pathways and monitor oral anticoagulant therapy, aiding in patient management.
Clinical Evidence
Bench testing only. Precision/reproducibility evaluated per CLSI EP05-A3 across three sites (N=180-240). Method comparison study (N=360 US, N=271 OUS) against predicate showed Passing-Bablok regression slopes near 1.0. Total allowable difference (±15-18%) met. Analytical sensitivity for coagulation factors (FII, FV, FVII, FX) established. Interference testing performed for common substances and direct oral anticoagulants.
Technological Characteristics
Fully-automated, random-access photometric analyzer. Reagent: Lyophilized rabbit brain thromboplastin. Detection: Optical density change at 405 nm and/or 620 nm. Connectivity: Bi-directional LIS transmission. Software: Automated algorithm for reaction curve analysis and parameter calculation. Sterilization: Not specified.
Indications for Use
Indicated for quantitative determination of Prothrombin Time (PT) in human plasma (3.2% sodium citrate) for evaluation of extrinsic/common coagulation pathways and monitoring warfarin therapy. For clinical professional laboratory and prescription use. Not established for neonate or pediatric populations.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
A. 510(k) Number:
K183390
B. Purpose for Submission:
Clearance of a new device
C. Measurand:
Prothrombin Time (PT) and International Normalized Ratio (INR)
D. Type of Test:
Quantitative
E. Applicant:
Diagnostic Grifols S.A.
F. Proprietary and Established Names:
QNext and DG-PT
G. Regulatory Information:
1. Regulation section:
21 CFR 864.5425, Multipurpose system for in vitro coagulation studies
21 CFR 864.7750, Prothrombin time test
2. Classification:
Class II
3. Product code:
JPA, System, Multipurpose For In Vitro Coagulation
GJS, Test, Time, Prothrombin
{1}
4. Panel:
Hematology 81
H. Intended Use:
1. Intended use(s):
The QNext is a fully-automated, random-access instrument, intended for in vitro diagnostic use in clinical laboratories to perform hemostasis testing by detecting the changes in optical density.
DG-PT is a thromboplastin reagent for the quantitative determination of Prothrombin Time on human plasma samples collected in 3.2% sodium citrate.
The product is used for the evaluation of the extrinsic and common coagulation pathways in seconds and for the monitoring Oral Anticoagulant Therapy with warfarin in International Normalized Ratio (INR).
For use with QNext.
For clinical professional laboratory and prescription use only.
For in vitro diagnostic use.
The performance of this device has not been established in neonate and pediatric patient populations.
2. Indication(s) for use:
Same as intended use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
QNext
I. Device Description:
The QNext analyzer automatically performs the following procedures:
- Sample management: loading, positive identification, aspiration, dilution (if required) and dispensation into cuvettes.
- Reagent management: loading, positive identification, cooling, stirring, aspiration and dispensation into cuvettes.
- Cuvette management: loading, transport, incubation during the reactions and management of used cuvettes.
- Management of test requests.
2
{2}
- Execution of test procedures.
- Result management: optical measurement of the reactions, algorithm calculation of analytical parameters from reaction curves, validation of results, traceability, bidirectional transmission of requests and results to the LIS.
- Management of disposable components.
The data analyzed can be stored, displayed and printed. Additionally, the analyzer conducts integrated functions such as the analysis of urgent samples or the Quality Control module.
DG-PT consists of a glass vial containing lyophilized thromboplastin (tissue factor and phospholipids) from rabbit brain tissue, buffer, calcium ions and preservative. The closure system includes a stopper and a screw cap. The DG-PT reagent is used to perform PT tests to screen the extrinsic and common coagulation pathways (Factors II, V, VII and X), and monitor oral anticoagulant therapy (OAT) with vitamin K antagonists.
The assay is based on the activation of the extrinsic coagulation pathway by the addition of the reagent to the plasma sample. The thromboplastin interacts with FVII and calcium ions activating a series of specific enzymes that comprise the extrinsic and common pathways of the coagulation cascade ultimately leading to the formation of a fibrin clot. The reader in the QNext analyzer measures the light change produced during the reaction.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
ACL TOP 700
HemosIL PT-Fibrinogen HS PLUS
2. Predicate 510(k) number(s):
K160276
K060931
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device QNext K183390 | Predicate ACL TOP 700 K160276 |
| Intended Use | The QNext is a fully-automated, random-access instrument, intended for in vitro diagnostic use in clinical laboratories to perform hemostasis testing by detecting the changes in | The ACL TOP is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device QNext K183390 | Predicate ACL TOP 700 K160276 |
| | optical density. | fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The system provides results for both direct hemostasis measurements and calculated parameters. |
| Sample type | 3.2% Citrated Plasma | Same |
| Application | Clotting Tests | Same |
| Modes of operation | Automated and continuous | Same |
| Sample predilution | Yes | Same |
| Sample liquid level detection | Yes | Same |
| Temperature Control for Sample Incubation | 37 ± 2°C | 37 ±1°C |
| Differences | | |
| --- | --- | --- |
| Item | Device QNext K183390 | Predicate ACL TOP 700 K160276 |
| Sample load capacity | 60 | 120 |
| Reagent load capacity | 30 | 60 |
| Wavelength(s) Used in Analysis | Bichromatic: 405 nm and/or 620 nm | Bichromatic: 405 nm and/or 671 nm |
| Volume needed for reaction | 150 – 500 μL | 150 – 600 μL |
| Reaction container | Qcell cuvettes | Cuvettes |
| Reaction container capacity | 400 | 264 |
| Sample throughput | ~200 PT/hour | ~360 PT/hour |
| Similarities | | |
| --- | --- | --- |
| Item | Device DG-PT K183390 | Predicate HemosIL PT-Fibrinogen HS PLUS K060931 |
| Intended Use | DG-PT is a thromboplastin reagent for the quantitative determination of Prothrombin Time on human plasma samples collected in 3.2% sodium | The HemosIL PT-Fibrinogen HS PLUS is a very high sensitivity calcium thromboplastin for simultaneous determinations of |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device DG-PT K183390 | Predicate HemosIL PT-Fibrinogen HS PLUS K060931 |
| | citrate. The product is used for the evaluation of the extrinsic and common coagulation pathways in seconds and for the monitoring Oral Anticoagulant Therapy with warfarin in International Normalized Ratio (INR). For use with QNext. For clinical professional laboratory and prescription use only. For in vitro diagnostic use. The performance of this device has not been established in neonate and pediatric patient populations. | Prothrombin Time (PT) and Fibrinogen (Fib), for the evaluation of the extrinsic coagulation pathway and monitoring Oral Anticoagulant Therapy in human citrated plasma on the IL Coagulation Systems. |
| Sample type | 3.2% Citrated Plasma | Same |
| Reagent Contents | Lyophilized rabbit brain thromboplastin | Same |
| Measurement | PT in seconds and INR | Same |
| Detection principle | Photometric | Same |
| Differences | | |
| --- | --- | --- |
| | Device DG-PT K183390 | Predicate HemosIL PT-Fibrinogen HS PLUS K060931 |
| Fibrinogen determination | Not for use in fibrinogen determination | Intended for use in fibrinogen determination |
| Quality Control | Two levels | Three levels |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A3, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline
CLSI EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline
{5}
CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline
CLSI C28-A3c, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline
CLSI H21-A5, Collection, Transport, and Processing of Blood Specimens for Testing Plasma-Based Coagulation Assays and Molecular Hemostasis Assays; Approved Guideline – Fifth Edition
CLSI H47-A2, One-Stage Prothrombin Time (PT) Test and Activated Partial Thromboplastin Time (APTT) Test; Approved Guideline-Second Edition
CLSI H54-A, Procedures for Validation of INR and Local Calibration of PT/INR Systems; Approved Guideline
## L. Test Principle:
The DG-PT test contains thromboplastin (tissue factor and phospholipids) from rabbit brain tissue. The test is based on the activation of the extrinsic coagulation pathway by the addition of thromboplastin reagent to the plasma sample. The thromboplastin interacts with FVII and calcium ions which activate a series of specific enzymes that comprise the extrinsic and common pathways of the coagulation cascade ultimately leading to the formation of a fibrin clot.
The reader on the QNext analyzer measures the light change produced in the cuvette from the mixing of the plasma with the thromboplastin reagent until a fibrin clot is formed. The light change produced in the cuvette due to the clot formation is simultaneously read at two wavelengths (405 nm and 620 nm) obtaining as a result a sigmoidal curve. Once the sigmoidal curve (absorbance-time) is obtained, the clotting time in seconds is determined with an algorithm.
The INR is automatically calculated using the MNPT (Mean Normal Prothrombin Time) and the ISI (International Sensitivity Index) from the current lot of PT reagent. The value is obtained according to the following equation: INR = (PT Patient (s) / MNPT (s))¹ˢᵗ.
## M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision studies were designed in accordance with CLSI EP05-A3.
Repeatability and within-laboratory precision were evaluated in a single site study using seven native plasma samples with three lots of reagent over 20 days with two runs per day and two replicates per run.
6
{6}
The reproducibility study was performed in three sites using seven native plasma samples on two QNext analyzers over five days with two runs per day and three replicates per run.
## Summary of Repeatability Results for PT in Seconds
| Sample | Mean Value | N | Repeatability | | Between-run | | Between-day | | Between-lot | | Total variability | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 10.65 | 240 | 0.15 | 1.4 | 0.12 | 1.1 | 0.08 | 0.8 | 0.14 | 1.4 | 0.25 | 2.4 |
| 2 | 17.02 | 238 | 0.27 | 1.6 | 0.38 | 2.3 | 0.00 | 0.0 | 0.36 | 2.1 | 0.59 | 3.5 |
| 3 | 27.68 | 240 | 0.47 | 1.7 | 0.50 | 1.8 | 0.10 | 0.4 | 0.70 | 2.5 | 0.99 | 3.6 |
| 4 | 29.40 | 240 | 0.56 | 1.9 | 0.31 | 1.1 | 0.26 | 0.9 | 1.02 | 2.4 | 0.99 | 3.4 |
| 5 | 37.48 | 240 | 0.66 | 1.8 | 0.56 | 1.5 | 0.13 | 0.3 | 1.19 | 3.2 | 1.48 | 3.9 |
| 6 | 53.82 | 240 | 0.93 | 1.7 | 0.63 | 1.2 | 0.71 | 1.3 | 1.99 | 3.7 | 2.39 | 4.4 |
| 7 | 85.08 | 240 | 1.76 | 2.1 | 0.93 | 1.1 | 1.10 | 1.3 | 4.25 | 5.0 | 4.82 | 5.7 |
## Summary of Repeatability Results for INR
| Sample | Mean Value | N | Repeatability | | Between-run | | Between-day | | Between-lot | | Total variability | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.729 | 240 | 0.012 | 1.6 | 0.009 | 1.2 | 0.007 | 0.9 | 0.002 | 0.2 | 0.016 | 2.2 |
| 2 | 1.250 | 238 | 0.022 | 1.7 | 0.032 | 2.6 | 0.007 | 0.5 | 0.028 | 2.3 | 0.049 | 3.9 |
| 3 | 2.186 | 240 | 0.440 | 2.0 | 0.045 | 2.1 | 0.012 | 0.5 | 0.072 | 3.3 | 0.096 | 4.4 |
| 4 | 2.343 | 240 | 0.052 | 2.2 | 0.027 | 1.1 | 0.025 | 1.1 | 0.085 | 3.6 | 0.106 | 4.5 |
| 5 | 3.098 | 240 | 0.063 | 2.0 | 0.053 | 1.7 | 0.130 | 0.4 | 0.141 | 4.5 | 0.164 | 5.3 |
| 6 | 4.698 | 240 | 0.095 | 2.0 | 0.061 | 1.3 | 0.070 | 1.5 | 0.258 | 5.5 | 0.291 | 6.2 |
| 7 | 7.959 | 240 | 0.194 | 2.4 | 0.090 | 1.1 | 0.117 | 1.5 | 0.602 | 7.6 | 0.650 | 8.2 |
## Summary of Reproducibility Results for PT in Seconds
| | | | Repeatability | | Between run | | Between day | | Between instrument | | Between site | | Total variability | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean Value | N | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 10.70 | 180 | 0.24 | 2.2 | 0.12 | 1.2 | 0.09 | 0.8 | 0.08 | 0.7 | 0.15 | 1.4 | 0.33 | 3.1 |
| 2 | 16.78 | 180 | 0.26 | 1.6 | 0.27 | 1.6 | 0.26 | 1.6 | 0.10 | 0.6 | 0.24 | 1.4 | 0.53 | 3.1 |
| 3 | 27.08 | 180 | 0.51 | 1.9 | 0.58 | 2.1 | 0.39 | 1.5 | 0.50 | 1.8 | 0.55 | 2.0 | 1.14 | 4.2 |
| 4 | 29.06 | 180 | 0.50 | 1.7 | 0.40 | 1.4 | 0.56 | 1.9 | 0.54 | 1.9 | 0.58 | 2.0 | 1.17 | 4.0 |
| 5 | 36.87 | 180 | 0.61 | 1.7 | 0.35 | 0.9 | 0.49 | 1.3 | 0.42 | 1.1 | 1.11 | 3.0 | 1.47 | 4.0 |
| 6 | 52.27 | 180 | 0.90 | 1.7 | 1.83 | 3.5 | 0.46 | 0.9 | 1.32 | 2.5 | 1.66 | 3.2 | 2.97 | 5.7 |
| 7 | 80.65 | 180 | 2.46 | 3.1 | 1.15 | 1.4 | 0.78 | 1.0 | 0.00 | 0.0 | 3.89 | 4.8 | 4.81 | 6.0 |
{7}
Summary of Reproducibility Results for INR
| | | | Repeatability | | Between run | | Between day | | Between instrument | | Between site | | Total variability | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean Value | N | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 0.712 | 180 | 0.019 | 2.6 | 0.010 | 1.4 | 0.008 | 1.1 | 0.015 | 2.2 | 0.029 | 4 | 0.039 | 5.5 |
| 2 | 1.212 | 180 | 0.023 | 1.9 | 0.023 | 1.9 | 0.022 | 1.8 | 0.029 | 2.4 | 0.046 | 3.8 | 0.067 | 5.5 |
| 3 | 2.133 | 180 | 0.047 | 2.2 | 0.054 | 2.5 | 0.038 | 1.8 | 0.074 | 3.5 | 0.100 | 4.7 | 0.148 | 6.9 |
| 4 | 2.318 | 180 | 0.048 | 2.1 | 0.039 | 1.7 | 0.053 | 2.3 | 0.082 | 3.5 | 0.110 | 4.7 | 0.159 | 6.9 |
| 5 | 3.070 | 180 | 0.061 | 2.0 | 0.035 | 1.1 | 0.049 | 1.6 | 0.087 | 2.8 | 0.166 | 5.4 | 0.206 | 6.7 |
| 6 | 4.637 | 180 | 0.094 | 2.0 | 0.191 | 4.1 | 0.055 | 1.2 | 0.177 | 3.8 | 0.291 | 6.3 | 0.405 | 8.7 |
| 7 | 7.735 | 180 | 0.285 | 3.7 | 0.129 | 1.7 | 0.095 | 1.2 | 0.133 | 1.7 | 0.615 | 7.9 | 0.709 | 9.2 |
The total allowable difference was analyzed for DG-PT in seconds and INR using the data from method comparison studies to address the concerns regarding variability in the performance of the DG-PT and to support use of the device as a reference method for PT/INR meters (both professional-use and prescription home-use PT/INR meters). According to CLIA and CMS(42 CFR 493.941), the percentage of samples should be $\pm 15\%$ within an allowable error. For DG-PT/INR, $95.2\%$ of the samples that are within a total allowable difference of $\pm 18\%$ relatively to HemosIL thromboplastin PT-Fibrinogen HS Plus reagent on ACL TOP 700 instrument and for DG-PT in seconds $94.3\%$ of the samples that are within a total allowable difference of $\pm 15\%$ relatively to HemosIL PT-Fibrinogen HS Plus on ACL TOP 700 instrument.
b. Linearity/assay reportable range:
Linearity is not applicable for prothrombin time and INR. The INR for the DG-PT is calculated mathematically using a standardized thromboplastin based on log of clotting time in seconds.
The reportable range for the DG-PT, 0.9–8.0 INR, was established by assay performance in the method comparison study against the predicate device (HemosIL PT-Fibrinogen HS Plus on ACL TOP 700).
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Shelf-life Stability:
The shelf-life stability claim for DG-PT is supported by the stability studies performed in accordance with CLSI EP25-A “Evaluation of Stability of In Vitro Diagnostic Reagents”. Three lots of reagents were evaluated at different time points (0, 6, 12, 18, 24 and 26 months) at $2 - 8^{\circ}\mathrm{C}$. The study was conducted by testing a total of four samples: one native normal plasma pool, one frozen plasma pool collected from patients on AVK (anti-Vitamin K treatment) and commercial quality control material (normal and pathological levels). The shelf-life stability study has been
{8}
completed up to 26 months for DG-PT, and found to be acceptable to support a 24-month shelf-life stability claim at 2–8°C.
## In-Use Stability:
The in-use stability claim for DG-PT is supported by the stability studies performed in accordance with CLSI EP25-A “Evaluation of Stability of In Vitro Diagnostic Reagents”. The study was conducted by testing a total of three samples: one native normal plasma pool, and commercial quality control material (normal and pathological levels). One lot of reagent was evaluated at different time points and at different temperature ranges with different reagent conditions (e.g., capped, uncapped).
The in use stability study has been completed up to 16 days for DG-PT, and found to be acceptable to support a 10-day in-use stability claim at 2–8°C.
To support the in-use stability claim at 25°C, different time points (12, 24, 36, 48, 60 and 72 hours) were evaluated. The in-use stability study has been completed up to 72 hours for DG-PT, and found to be acceptable to support a 2-day (48-hour) in use stability claim at 25°C.
To support the on-board stability claim, different time points (1, 4, 6, 8, and 11 days) were evaluated. The on-board stability study has been completed up to 11 days for DG-PT, and found to be acceptable to support a 5-day on-board stability claim.
## Transport stability:
Transport stability was established by evaluating one lot of reagent for one week at ≤ -20°C and 40°C, at the beginning of the shelf-life study and following reagent storage at the recommended storage conditions (2–8°C) to undergo shelf-life and in-use testing, as indicated below.
- Shelf-life stability study at 5 ± 3°C
- In-use stability study at 5 ± 3°C (capped)
- In-use stability study at 25°C ± 2°C / 40% RH (relatively humidity) ± 5% (uncapped)
- In-use stability study at 25°C ± 2°C (capped)
- In-use stability onboard the QNext instrument (uncapped)
All acceptance criteria for the transportation stability were met.
## Traceability/Calibrator:
DG-PT is calibrated using WHO International Reference Standards, per recommendations of the WHO Technical Report Series N° 979, Annex 6, 2013, Guidelines for thromboplastins and plasma used to control oral anticoagulant therapy with vitamin K antagonists. Two types of calibration are performed: 1) Calibration of secondary standards using an appropriate International Reference Preparation and 2) Calibration of DG-PT against the corresponding secondary standard (“lot-to-lot”
9
{9}
calibration). All the lots used for the performance of DG-PT have been calibrated against the corresponding secondary standard.
## Quality Controls:
There are two Quality Control materials for use with the DG-PT tests on the QNext analyzer – DG-C1, which is a normal lyophilized control (PT at ~12 seconds, within reference interval for the tests), and DG-C2, which is an abnormal lyophilized control (prolonged PT at ~20 seconds, which is the upper end of the reference interval for the prothrombin time tests).
## d. Detection limit:
A factor sensitivity study was designed in accordance with CLSI H47-A2. For the factor sensitivity study, PT determinations were made for factors II, V, VII and X in a set of samples spanning from <1% to 100% of a specific coagulation factor, in order to establish the minimum concentration at which the screening test rises above the upper limit of the established reference interval.
A set of 13 samples obtained from dilutions of normal pooled plasma with the appropriate factor deficient plasma was tested. Two replicates of each sample were tested using two lots of DG-PT. No clotting time result was obtained for the <1 IU/dL sample level for factor II, factor V, factor VII and factor X.
According to the interpolation of the upper limit of the reference interval in the quadratic regression model using the log-log transformation, the sensitivity estimate for each factor was determined to be the values reported in the table below.
| Factor | Sensitivity (IU/dL) |
| --- | --- |
| FII | 29 |
| FV | 45 |
| FVII | 44 |
| FX | 43 |
## e. Analytical specificity:
The study was designed in accordance with the recommendations in CLSI EP7-A2 and the statistical analyses were conducted in accordance with the recommendations in CLSI EP07-A3. The following interferents were evaluated: hemoglobin, triglycerides, bilirubin, LMWH (low molecular weight heparin), UFH (unfractioned heparin), argatroban, dabigatran and rivaroxaban. Two sample levels were tested (pooled normal samples and pooled abnormal samples from patients on oral anticoagulant). Each potential interfering substance was added to each sample pool.
The following interferents: hemoglobin, triglycerides, bilirubin, LMWH, UFH showed no significant interference (≤10% observed effect) up to the concentrations indicated in the table below:
10
{10}
| Interferent | Concentration |
| --- | --- |
| Hemoglobin | 1000 mg/dL |
| Triglycerides | 3000 mg/dL |
| Bilirubin (conjugated) | 43 mg/dL |
| LMWH (low molecular weight heparin) | 1.8 IU/mL |
| UFH (unfractioned heparin) | 0.7 IU/mL |
The following interferents: argatroban, dabigatran and rivaroxaban were evaluated and showed interference at the specified concentration in citrated plasma samples.
| Interferent | Normal Sample | Abnormal Sample |
| --- | --- | --- |
| Rivaroxaban (ng/mL) | 313 | 156 |
| | 625 | 313 |
| | 938 | 469 |
| | 1250 | 625 |
| Argatroban (ng/mL) | 375 | 150 |
| | 750 | 300 |
| | 1125 | 450 |
| | 1500 | 600 |
| Dabigatran (ng/mL) | 150 | 75 |
| | 300 | 150 |
| | 450 | 225 |
| | 600 | 300 |
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The method comparison study was conducted at three clinical sites. Samples were collected from adult patients (19-95 years of age) and covered the claimed analytical measurement range (AMR). Samples were tested in parallel on the subject device, QNext with DG-PT and the predicate device, ACL TOP 700 with HemosIL PT Fibrinogen HS Plus. The results of the study were evaluated using a Passing-Bablok regression analysis for each site and all sites combined. A summary of results are included in the tables below.
{11}
DG-PT Prothrombin Time (seconds) and INR, U.S. Sites:
| Assay | Range | N | Slope (95 % CI) | Intercept (95 % CI) |
| --- | --- | --- | --- | --- |
| PT (Seconds) | 12.9–71.6 | 360 | 0.913 (0.893, 0.938) | 1.118 (0.706, 1.477) |
| PT (INR) | 0.91–6.62 | 360 | 1.034 (1.010, 1.060) | -0.024 (-0.054, 0.009) |
DG-PT Prothrombin Time (seconds) and INR from Outside U.S. Site:
| Assay | Range | N | Slope (95 % CI) | Intercept (95 % CI) |
| --- | --- | --- | --- | --- |
| PT (Seconds) | 13.2–81.7 | 271 | 0.985 (0.962, 1.003) | 0.272 (-0.078, 0.716) |
| PT (INR) | 0.90–7.76 | 271 | 1.000 (0.978, 1.022) | -0.070 (-0.099, -0.042) |
b. Matrix comparison:
A matrix comparison study was conducted to demonstrate equivalence between fresh and frozen citrated plasma samples. The study assessed the comparability of sample results when tested freshly and after 7 days (1 week) of storage at $\leq -70^{\circ}\mathrm{C}$ (fresh vs frozen plasma samples). Testing was performed at one external site using one reagent lot. Fifty four (54) fresh patient samples covering the clinical decision points and both normal and abnormal levels of the measuring interval were tested. Results were analyzed using Passing-Bablok regression analysis and Bland-Altman plots. The study results demonstrated comparability between fresh and frozen samples and met the pre-defined acceptance criteria.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
{12}
5. Expected values/Reference range:
The reference interval study was designed in accordance with CLSI C28-A3.
The study was performed at two sites using one lot of reagents. A total of 243 reference samples obtained from reference individuals (apparently healthy men and women adults, $\geq 21$ years of age) were tested. Results from two sites were pooled and all reference intervals were established by calculating the non-parametric $95\%$ confidence interval $(2.5^{\mathrm{th}}$ to $97.5^{\mathrm{th}}$ percentiles). The calculated normal reference range for PT is 12.1 to 16.1 seconds and for INR is 0.83 to 1.17.
N. Instrument Name:
QNext analyzer
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X ☐ or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ X ☐ or No ☐ X
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
3. Specimen Identification:
Specimens can be manually identified or identified using an internal radiofrequency identification system. The sample tubes are inserted in the analyzer using specially-designed holders, the Qsample holders, which have an internal radiofrequency identification system. When sample identification is manual, the analyzer manages the sample tube using the corresponding Qsample holder internal identification as it passes through the Internal Identification Area of Qsample holders.
4. Specimen Sampling and Handling:
The following sample management processes are conducted by the QNext analyzer: loading, positive identification, aspiration, dilution (if required) and dispensation into
13
{13}
cuvettes. The Samples Desk manages the sample tubes once they have been inserted in the analyzer using specially-designed holders, the Qsample holders. The Sample Arm is a Multi-task robotic arm which allows cap piercing of the samples tubes, aspiration, dispensation, dilution and homogenization of samples in the Qcells cuvettes.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Sample Stability Study:
A total of 42 samples (normal and abnormal) were tested for frozen stability at $< -70^{\circ}\mathrm{C}$ at three different timepoints (0, 2 and 13 months), 45 samples (normal and abnormal) were tested for on-board stability (uncapped) at room temperature at three different timepoints (0, 16 and 25 hours) and 20 samples (normal and abnormal) were tested capped onboard at three different timepoints (0, 13 and 25 hours). The result at each time point was compared to the respective baseline result and found to be acceptable to support the claim for 24 hours on board for both capped and uncapped sample collection tubes and for two months at $< -70^{\circ}\mathrm{C}$ for frozen platelet-poor plasma samples.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.
R. Conclusion:
The submitted information in this premarket notification is complete and support a substantial equivalence decision.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.