K183357 · Xablecath, Inc. · PDU · Jan 21, 2019 · Cardiovascular
Device Facts
Record ID
K183357
Device Name
XableCath Crossing Catheter
Applicant
Xablecath, Inc.
Product Code
PDU · Cardiovascular
Decision Date
Jan 21, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K183357 · Jan 21, 2019
XableCath Crossing Catheter
Xablecath, Inc.
Clinical patient cohort (49 patients)
The sponsor performed a clinical assessment of 49 patients undergoing peripheral arterial disease re-canalization to verify device performance in real-world conditions, citing limitations of animal modeling.
Peripheral Arterial Disease; Chronic Total Occlusions; Real-world assessment; Clinical cohort
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Real-world assessment of XableCath Catheter; Retrospective/Observational clinical assessment; Follow-up/Duration: 30-45 day post-treatment follow-up (5.6 week average)
49 patients with peripheral arterial disease (PAD); 57% male, 43% female; mean age 71 years; 100% with comorbidities.; Sample Size: 49 patients
Not applicable for this study
Primary: Successful passage within the target lesion; Secondary: Absence of major adverse event attributable to XableCath Catheter use.
Indications for Use
The XableCath Catheter is intended for use with a guidewire to access discrete regions of the peripheral vasculature. The XableCath™ Crossing Catheter is intended for use to facilitate the intraluminal placement of conventional guidewires beyond stenotic peripheral lesions (including chronic total occlusions) prior to placement of other interventional devices.
Device Story
Over-the-wire (OTW) single-lumen catheter; blunt or abrasion distal tip designs. Used in peripheral vasculature to facilitate guidewire placement across stenotic lesions/chronic total occlusions. Operated by physicians in hospital/surgical suites under fluoroscopic imaging. Manual proximal manipulation transmits axial/torsional forces to distal tip for lesion crossing. Enables subsequent interventional device delivery (angioplasty/stenting). Benefits include improved access to challenging, calcified, or occluded peripheral arterial segments where standard wire/balloon passage fails.
Clinical Evidence
Prospective clinical assessment of 49 patients (mean age 71; 57% male) with PAD (Rutherford Class 3-6). Treated 52 lesions (54% CTO, 81% moderate/severe calcification). Primary endpoint: successful lesion passage for subsequent therapy achieved in 98% of cases. Secondary endpoint: absence of major adverse events (perforation, thrombosis, embolization) achieved in 100% of cases. No adverse events were attributable to the device.
Indicated for patients with peripheral artery disease (PAD) requiring access to discrete regions of the peripheral vasculature to facilitate intraluminal guidewire placement across stenotic lesions, including chronic total occlusions, prior to interventional procedures.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services USA logo, which features a stylized caduceus. To the right is the FDA U.S. Food & Drug Administration logo, with the FDA acronym in a blue square and the full name in blue text.
January 21, 2018
XableCath, Inc. Mr. Rick Gaykowski Chief Regulatory Officer 417 S. Wakara Way, Ste 3510 Salt Lake Citv. Utah 84108-1457
Re: K183357
Trade/Device Name: XableCath Crossing Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PDU Dated: November 19, 2018 Received: December 4, 2018
Dear Mr. Gaykowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
| | Gregory |
|-----|------------------|
| For | O'Connell |
| | 2019.01.21 |
| | 10:37:48 -05'00' |
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183357
Device Name XableCath™ Crossing Catheter
#### Indications for Use (Describe)
The XableCath Catheter is intended for use with a guidewire to access discrete regions of the peripheral vasculature. The XableCath™ Crossing Catheter is intended for use to facilitate the intraluminal placement of conventional guidewires beyond stenotic peripheral lesions (including chronic total occlusions) prior to placement of other interventional devices.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Section 6 510(k) SUMMARY
## XableCath Crossing Catheter
| Date prepared: | November 03, 2018 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| Submitter information [21 CFR 807.929(a)(1)] | |
| Name: | XableCath, Inc. |
| Address: | 417 S. Wakara Way, Ste 3510, Salt Lake City, Utah 84108-1457 |
| Phone Number: | (617) 447-4000 Mobile |
| Fax number: | N/A |
| Establishment Registration<br>Number: | Yet to be secured |
| Name of contact person: | Rick Gaykowski, Chief Regulatory Officer |
| Name of the device [21 CFR 807.92(a)(2)] | |
| Trade or proprietary name: | XableCath Crossing Catheter |
| Common or usual name: | Peripheral Vascular Catheter |
| Classification name: | Percutaneous Catheter |
| Classification number: | 21 CFR 870.1250, Procode PDU |
| Legally marketed device(s) to which equivalence is claimed<br>[21 CFR 807.92(a)(3)] | |
| eV3/Medtronic Viance Crossing® Catheter (K120533) | |
| Device description [21 CFR 807.92(a)(4)] | |
| The XableCath Crossing Catheter device is an over-the-wire (OTW) single-lumen catheter, with blunt and abrasion tip<br>design. The catheter is offered in a product family configuration, with size models as outlined below: | |
| Catalog Number | | Dimensions | |
|-----------------|------------------|---------------------------|-----------------------------------------|
| AbrasionTip (A) | Blunt<br>Tip (B) | Working<br>Length<br>(cm) | Recommended<br>Guidewire<br>Inches (mm) |
| XC-014-065A | XC-014-065B | 65 | 0.014" (0.36 mm) |
| XC-014-090A | XC-014-090B | 90 | 0.014" (0.36 mm) |
| XC-014-145A | XC-014-145B | 145 | 0.014" (0.36 mm) |
| XC-018-065A | XC-018-065B | 65 | 0.018" (0.46 mm) |
| XC-018-090A | XC-018-090B | 90 | 0.018" (0.46 mm) |
| XC-018-145A | XC-018-145B | 145 | 0.018" (0.46 mm) |
| XC-035-065A | XC-035-065B | 65 | 0.035" (0.89 mm) |
| XC-035-090A | XC-035-090B | 90 | 0.035" (0.89 mm) |
{4}------------------------------------------------
### XableCath Crossing Catheter Traditional 510(k), K183357
| XC-035-145A | XC-035-145B | 145 | 0.035" (0.89 mm) |
|-------------|-------------|-----|------------------|
|-------------|-------------|-----|------------------|
The catheter shaft consists of a cortically reinforced co-polymeric material blend, with radiopague cobalt chromiumtunqsten-nickel distal tip. The proximal portion is comprised of an olefin strain relief and polycarbonate hub, which includes a female luer connection with the catheter lumen. The catheter pass-through lumen is used to advance the catheter over an appropriately sized guidewire for intravascular advancement to the target site. Component connections are ensured through inter-mechanical locks and utilization of medical grade adhesives. Catheter quidewire compatibility size/length are declared on the XableCath Crossing Catheter strain relief. All provided family member models are compatible with ≥4Fr introducer sheaths. All product configurations are provided sterile, single-use only.
#### Indications for Use [21 CFR 807.92(a)(5)]
The XableCath Catheter is intended for use with a quidewire to access discrete regions of the peripheral vasculature.
The XableCath Crossing Catheter is intended for use to facilitate the intraluminal placement of conventional quidewires beyond stenotic peripheral lesions (including chronic total occlusions) prior to placement of other interventional devices.
#### Comparison of the technological characteristics with the predicate device [21 CFR 807.92(a)(6)]
#### The subject devices and predicate devices are based on the following similar technological elements:
The XableCath Crossing Catheter indication for use is comparable to the predicate device, with direct overlap for peripheral application. Both device systems provide support access to the vasculature, utilize similar insertion and deployment principles, and allow for quidewire and other interventional device exchanges. Both products are for use in endovascular catheter utilization and deployment. Both products are configured with uniquely design distal tip regions for intravascular passage across stenotic regions. Furthermore, both catheters have an 0.014" product offering possessing similarly offered working lengths, and identical guidewire compatibility. Additionally, the subject XableCath Crossing Catheter offers additional dimensional working versions as well (i.e., 0.018", and 0.035") within the product family.
Both products have similar intended therapeutic applications, used in similar patient populations, and employ equivalent technological characteristics in achieving desired technical methods-of-action through manual advancement under fluoroscopic imaging. Similar categories of product materials comprise the subject and predicate products, with both containing hub/luer, shaft body, and distal tip design reqions.
The following table illustrates comparable features of the subject and predicate products:
| Substantial Equivalence Comparison | | |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Information | Subject | Predicate |
| | XableCath Crossing Catheter | Medtronic Viance Crossing®<br>Catheter |
| 510(k) # | K183357<br>This Submission | K120533 |
| FDA Classification | Class II | Class II |
| Regulation Number | 870.1250 | 870.1250 |
| Product Code | PDU (projected) | PDU |
| Indications for Use | The XableCath Catheter is<br>intended for use with a guidewire<br>to access discrete regions of the<br>peripheral vasculature.<br><br>The XableCath Crossing Catheter<br>is intended for use to facilitate the<br>intraluminal placement of | The Viance Catheter is intended for<br>use with a guidewire to access<br>discrete regions of the peripheral<br>vasculature.<br><br>When used as part of the Covidien<br>Peripheral System, the Viance<br>Catheter is indicated for use to |
| | | |
| Proximal Luer/Hub | HDPE | Polycarbonate |
| Strain Relief | Olefin | None |
| Catheter Shaft | PEBAX/SS braiding/Nylon, Single<br>lumen | PEBAX / SS braiding, Single lumen |
| Markers | Cobalt-chromium/SS | SS |
| Coating | None | Hydrophilic |
| Distal Tip | Atraumatic (Blunt, Abrasion<br>versions) | Atraumatic |
| Infusion<br>Pressure/Burst<br>Strength | ≥300 psi | ≥300 psi |
| Outer Diameter | Varies by Model<br>Proximal: 0.043", 0.053", 0.062"<br>Distal: 0.060", 0.070", 0.080"<br>Tip: 0.052", 0.063", 0.071<br>0.014", 0.018", 0.035" | 0.039" |
| Guidewire<br>Compatibility | 0.014", 0.018", 0.035" | 0.014" |
| Guide Catheter (max) | ≥4Fr | ≥6Fr |
| Access Sheath (max) | ≥4Fr | ≥6Fr |
| Effective Working<br>Length | 65, 90, 145cm | 135cm |
| Deployment | OTW - Manual | OTW - Manual |
| Mechanism of Action | Manual proximal manipulation via<br>axial and torsional applied forces<br>transmitted through catheter length | Manual proximal manipulation via<br>axial and torsional applied forces<br>transmitted through catheter length |
| System Design | Proximal Hub/Handle/Luer<br>Catheter Shaft<br>Distal Tip | Proximal Hub/Handle/Luer<br>Catheter Shaft<br>Distal Tip |
| Duration of Therapy | Short-term use | Short-term use |
| Sterilization Method | Gamma<br>(SAL – 10-6) | Gamma<br>(SAL - 10-6) |
| Single Use, Sterile | Yes | Yes |
| Labeling | Individual IFU: Warning, Cautions,<br>Contraindications, tables, images,<br>organized outline. | Individual IFU: Warning, Cautions,<br>Contraindications, tables, images,<br>organized outline, clinical outcomes. |
| Packaging | Sterile thermal sealed Tyvek/PET<br>Pouch, SBS Carton | Sterile thermal sealed pouch, SBS<br>Carton |
| | physician. Hospital, Lab/Surgical<br>Suite | physician. Hospital, Lab/Surgical<br>Suite |
| Performance Testing | ISO 10555-1 Second Edition 2013-<br>07-01. Intravascular catheters --<br>Sterile, single-use intravascular<br>catheters -- Part 1: General<br>Requirements | ISO 10555-1 Second Edition 2013-<br>07-01. Intravascular catheters --<br>Sterile, single-use intravascular<br>catheters -- Part 1: General<br>Requirements (as applicable) |
| | Tensile, Dimensional,<br>Insertion/Withdrawal, Distal Tip,<br>Kink, Torque, Surface Defects,<br>Corrosion, Luer and Hub testing,<br>Radiopacity, Packaging,<br>Biocompatibility ISO 10993 | Tensile, Dimensional,<br>Insertion/Withdrawal, Distal Tip,<br>Kink, Coating, Torque, Surface<br>Defects, Corrosion, Luer and Hub<br>testing, Radiopacity, Packaging,<br>Biocompatibility ISO 10993 |
#### Substantial Equivalence Comparison
{5}------------------------------------------------
{6}------------------------------------------------
#### XableCath Crossing Catheter Traditional 510(k), K183357
#### The following technological differences exist between the subject and predicate devices:
- The subject and predicate devices have slightly differing distal tip configurations, unique to their particular design, . and overall product composition. Nevertheless, both product distal tip designs are intended to achieve the same outcomes and have been shown to be commensurate in performance through comparative assessment.
- . The predicate product contains an exterior hydrophilic coating over the distal length, whilst the subject product does not. Nevertheless, ease of insertion and tracking within applied models yields equivalent product delivery performance through comparative assessment.
- The subject and predicate products are offered in 0.014" workhorse models, whilst the subject product is also . offered in 0.018", and 0.035" versions as well.
Despite these noted differences between the subject and predicate products, these attributes are determined to be cumulatively insignificant and do not raise new questions of safety and effectiveness of these products as demonstrated through direct product comparative assessment.
#### Performance Data [21 CFR 807.92(b)]
| Summary of non-clinical tests conducted for determination of substantial equivalence<br>[21 CFR 807.92(b)(1)] | | | |
|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|--------------------------------------|
| ● | Direct product bench in-vitro comparison testing has shown the subject & predicate products to be substantially<br>equivalent, via assessment within the following areas: | | |
| | ○ | Visual & technical inspections | Dimensional assessment & comparisons |
| | ○ | Luer syringe compatibility | Guidewire compatibility |
| | ○ | Sheath compatibility | Leak Testing |
| | ○ | Guidewire retraction/reinsertion | Catheter kink-resistance |
| | ○ | Relative radiopacity comparison | Tensile testing (proximal & distal) |
| | ○ | Torque transmission & capability | Corrosion testing |
| | ○ | Simulated use (iliac model) | General packaging, shelf-life/expiry |
| ● | Full panel biocompatibility was successfully performed in accord with product classification, under GLP rigors,<br>demonstrating that all utilized materials and methods of construction/processing passed biocompatibility rigors.<br>Conducted test included: | | |
| | ○ | Cytotoxicity | Sensitization |
| | ○ | Irritation/Intracutaneous Reactivity | Systemic Toxicity (Acute) |
| | ○ | Hemolysis | Thromboresistance |
| | ○ | Complement Activation | Partial Thromboplastin Time |
| | ○ | Platelet and Leukocyte Count | Materials Mediated Pyrogenicity |
{7}------------------------------------------------
- Packaging integrity, transport challenge testing were applied and successfully completed in . accordance with established acceptance criteria demonstrating configurational adequacy.
Summary of clinical tests conducted for determination of substantial equivalence or of clinical information [21 CFR 807.92(b)(2)]
No clinical tests were mandated, yet were performed on the subject product to verify actual performance of the device in real-world conditions, secondary to limitations of effective animal modeling for peripheral vascular disease. From real-world assessment. 49 patients underwent lesion modifying therapeutic intravascular treatment with XableCath Catheters for peripheral arterial disease re-canalization. Primary endpoint was successful passage within the target lesion, enabling subsequent therapy (angioplasty and/or stenting), whilst the secondary endpoint was absence of major adverse event attributable to XableCath Catheter product use (perforation, thrombosis, or distal embolization).
Treatment cohort demographics were 21 females (43%) and 28 males (57%). with a mean age of 71 years (range of 48-94 vears). Presenting comorbidities (at least one) within the population was 100%, with Diabetes Mellitus (25 (51%) being the most common. Peripheral Artery Disease (PAD) classification using the Rutherford Scale (1986, revised 1997) was applied, with Rutherford Class 3-4 in 29 (59%) of the cohort, and Rutherford Class 5-6 in 20 (41%) of the treatment population. The number of lesions treated were 52, associating >100 arterial segments traversed, with treated mean lesion length (cm) of targeted occlusions being 5.6cm. Additionally, vascular lesions treated included liac 8 (15%), Femoral – AK Popliteal 36 (68%), BK Popliteal 2 (4%), and Tibial 7 (13%), being those common as PAD therapeutic targets. Lesion characteristics were of an occlusive magnitude that wire and balloon passage was not possible in 15% and 73% of cases respectively, with 54% being Chronic Total Occlusions (CTO) and 81% categorized as moderate/severe calcification.
From these clinical applications, a 98% primary endpoint result was achieved with successful PTA/Stenting & technical success with lesion/claudication crossing. Additionally, 100% secondary endpoint achievement was attained, with routine minor dissection in 10% of treatment subjects experiencing an access site complication both being successfully treated with standard remedial maneuvers, without any residual clinical sequelae at 30 - 45-day post-treatment follow-up (5.6 week average). Of significance, none (0%) of the recorded minor adverse events were either caused by or attributable to the XableCath Crossing Catheter deployment. In addition, human factors engineering and usability assessments were performed to ensure subject product features could be safely and effectively used.
#### Conclusions drawn [21 CFR 807.92(b)(3)]
Based upon the overlapping intended use, basal product designs, methods of deployment, target population and anatomical site overlap, and direct bench comparative assessment, the preponderance within this pre-market notification application demonstrates the XableCath Crossing Catheter is shown to be substantially equivalent to the predicate device. Furthermore, the XableCath Crossing Catheter raises no new questions of safety or effectiveness when compared directly to the predicate device, and is therefore justifiably concluded to be substantially equivalent for declared intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.