← Product Code [ISD](/productcode/ISD) · K183296

# REAL Immersive System (K183296)

_Penumbra, Inc. · ISD · Mar 18, 2019 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K183296

## Device Facts

- **Applicant:** Penumbra, Inc.
- **Product Code:** [ISD](/productcode/ISD.md)
- **Decision Date:** Mar 18, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5360
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The REAL Immersive System is an immersive virtual reality and display system that interactively displays and tracks upper-extremity rehabilitation exercises for adult patients using a combination of virtual environments and full presence tracked avatars for visual feedback. These rehabilitation exercises are intended to be conducted in a seated position in a clinical environment and prescribed and supervised by a medical professional trained in rehabilitation therapy.

## Device Story

REAL Immersive System is a digital VR platform for physical and neuro rehabilitation. Input: patient upper-extremity movement tracked via electromagnetic (EM) sensors (Wireless Transmitter Module, Wireless Sensor Module, sensor bands). Transformation: system translates physical movement into real-time visual feedback within virtual environments using full-presence tracked avatars. Output: visual feedback for patient; kinematic parameters (e.g., joint angular changes) and performance metrics for clinician. Usage: clinical environment; operated by medical professional trained in rehabilitation therapy. Benefit: provides interactive, immersive rehabilitation exercises to assist in patient recovery; allows clinicians to assign activities and monitor progress via tablet and HMD interfaces.

## Clinical Evidence

No clinical study conducted. Bench testing (packaging, battery, HMD/sensor accuracy, simulated use, usability) deemed sufficient. Literature review of similar devices supported the proposed indications for use.

## Technological Characteristics

Wireless VR system using electromagnetic (EM) tracking. Components: All-in-One HMD, controller, wireless transmitter/sensor modules, tablet, router, sensor bands. Compliance: ANSI/AAMI 60601-1:2005, IEC 60601-1-2:2014, 60601-1-6:2010, IEC 62366:2007. Biocompatibility per ISO 10993. FCC certified.

## Regulatory Identification

Measuring exercise equipment consist of manual devices intended for medical purposes, such as to redevelop muscles or restore motion to joints or for use as an adjunct treatment for obesity. These devices also include instrumentation, such as the pulse rate monitor, that provide information used for physical evaluation and physical planning purposes., Examples include a therapeutic exercise bicycle with measuring instrumentation, a manually propelled treadmill with measuring instrumentation, and a rowing machine with measuring instrumentation.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a measuring exerciser or an interactive rehabilitation exercise device for prescription use only, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

## Predicate Devices

- MindMotion Pro ([K162748](/device/K162748.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 19, 2019

Penumbra, Inc. Mary Rose Director of Regulatory Affairs One Penumbra Place Alameda, California 94502

Re: K183296

Trade/Device Name: REAL Immersive System Regulation Number: 21 CFR 890.5360 Regulation Name: Measuring exercise equipment Regulatory Class: Class II Product Code: ISD Dated: February 6, 2019 Received: February 7, 2019

Dear Mary Rose:

This letter corrects our substantially equivalent letter of March 18, 2019.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Vivek J. Pinto -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K183296

Device Name REAL(TM) Immersive System

### Indications for Use (Describe)

The REAL Immersive System is an immersive virtual reality and display system that interactively displays and tracks upper-extremity rehabilitation exercises for adult patients using a combination of virtual environments and full presence tracked avatars for visual feedback. These rehabilitation exercises are intended to be conducted in a seated position in a clinical environment and prescribed and supervised by a medical professional trained in rehabilitation therapy.

Type of Use (Select one or both, as applicable)

| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☑</span> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------------------------------------------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                                                                                   |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The word "Penumbra" is written in a bold, sans-serif font in a dark red color. To the right of the name is a circular logo, also in dark red, with a stylized "P" inside.

#### SUBMITTER I.

Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA

Phone: (510) 748-3346 FAX: (510) 217-6414

Contact Person: Mary Rose, Director of Regulatory Affairs Date of Prepared: March 18, 2019

#### II. DEVICE

| Name of Device:       | REAL Immersive System         |
|-----------------------|-------------------------------|
| Common or Usual Name: | Virtual Reality based Therapy |
| Classification Name:  | Measuring Exercise Equipment  |
| Regulatory Class:     | II                            |
| Product Code:         | ISD                           |

#### III. PREDICATE DEVICE

MindMotion Pro, K162748

This predicate has not been subject to a design-related recall

No reference devices were used in this submission.

#### DEVICE DESCRIPTION IV.

REAL Immersive System is a digital hardware and software medical device platform using a combination of virtual environments and full presence tracked avatars for visual feedback. It is designed for use in healthcare and focusing on physical and neuro rehabilitation. The use of the device is intended to be in a clinical environment supervised by a medical professional trained in rehabilitation therapy.

REAL Immersive System contains the following components:

- . All-in-One HMD with Software Application
- . HMD Controller
- Wireless Transmitter Module (WTM or Large Sensor) .
- Wireless Sensor Module (WSM or Small Sensor) .

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Penumbra, Inc., a healthcare company. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern, and the red color gives it a sense of energy and urgency.

- . Tablet with Software Application
- . Sensor Charger (charging station)
- Router ●
- Router Battery ●
- Sensor Bands .

#### V. INDICATIONS FOR USE

The REAL Immersive System is an immersive virtual reality and display system that interactively displays and tracks upper-extremity rehabilitation exercises for adult patients using a combination of virtual environments and full presence tracked avatars for visual feedback. These rehabilitation exercises are intended to be conducted in a seated position in a clinical environment and prescribed and supervised by a medical professional trained in rehabilitation therapy.

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH VI. THE PREDICATE DEVICE

The REAL Immersive System and the predicate device MindMotion PRO, have similar technical characteristics. Both systems are software-based devices intended for a computing platform that utilizes sensors for tracking the patient's upper extremity movements. Both systems are interactive, immersive, and translate the physical movements to moving images on the display, along with audio-visual feedback. Both systems provide the patient's performance metrics for the healthcare professional. The subject and predicate devices are non-invasive and intended for upper extremity rehabilitation.

## Principles of Operation

The MindMotion Pro is a wireless Virtual Reality (VR) rehab system utilizing optically based tracking. MindMotion is an interactive medical device that combines optically based motion capture and virtual reality technologies together with advanced cognitive principles related to body perception.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the left side of the "P" to the edge of the circle.

The REAL Immersive System is a wireless VR rehab system utilizing Electromagnetic (EM) tracking. The REAL Immersive system, utilizing EM tracking, is an interactive system that allows a medical professional to assign movement activities to patients. It tracks patient movement providing visual feedback and reports on kinematic parameters like joint angular changes during movement.

#### PERFORMANCE DATA VII.

The following performance data were provided in support of the substantial equivalence determination.

## Biocompatibility Testing

The biocompatibility evaluation for the REAL Immersive System was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, " May 1, 1995, and International Standard ISO 10993-1 'Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process,' as recognized by FDA.

The following tests were successfully performed on the device:

- Cytotoxicity .
- Irritation .
- . Sensitization

In summary, non-clinical testing has not found the REAL Immersive System to be biologically incompatible according to according to ISO 10993 requirements.

## Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the REAL Immersive System. The system complies with the requirements of ANSI/AAMI 60601-1:2005, IEC 60601-1-2:2014, 60601-1-6:2010, and IEC 62366:2007. REAL Immersive System also underwent FCC testing and is FCC certified.

## Software Verification and Validation Testing

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The logo features the word "Penumbra" in a bold, red font, followed by a circular emblem. The emblem contains a stylized "P" in white against a red background, creating a visually distinct and recognizable brand identity.

Software verification and validation testing and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern. Prior to mitigation of hazards, there is a small but non-zero risk of Minor injury to the user.

## Bench-Top Performance (Design Verification)

Design Verification testing was conducted to evaluate the physical and mechanical properties and to demonstrate substantial equivalence to the predicate device. The tests were performed included; Packaging, Battery performance, HMD and Sensor Accuracy, Simulated Use, and Usability Interface Summative.

All tests passed successfully.

## Animal Studies

No Animal study was conducted as bench testing was determined sufficient for verification and validation purposes.

## Clinical Studies

No clinical study was conducted as bench testing was determined sufficient for verification and validation purposes. A review was conducted considering published clinical study articles that featured devices with similar intended use. The literature review was used to support the proposed indications for use by leveraging clinical outcomes from devices that are considered technologically similar.

## VIII. CONCLUSIONS

The non-clinical data support the safety of the device and the hardware and software verification and validation demonstrate that the REAL Immersive System should perform as intended in the specified use conditions.

The subject REAL Immersive System is substantially equivalent to the predicate device MindMotion Pro. The subject device has the same intended use as the predicate device. The subject and the predicate devices differ slightly in regards to technological variations, while maintaining the same fundamental scientific technology. However, these differences do not raise different questions of safety and effectiveness.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a circular logo, also in red, with a white "P" inside. The logo is a stylized representation of the letter "P" within a circle, creating a visual element that complements the text.

The device testing described in the 510(k) Summary demonstrate the subject devices are substantially equivalent to the predicate device in regards to operating principle, fundamental technology, and device performance.

---

**Source:** [https://fda-staging.innolitics.com/device/K183296](https://fda-staging.innolitics.com/device/K183296)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
