Coag-Sense Prothrombin Time (PT) / INR Monitoring Device
K183255 · Coagusense, Inc. · GJS · Feb 21, 2019 · Hematology
Device Facts
Record ID
K183255
Device Name
Coag-Sense Prothrombin Time (PT) / INR Monitoring Device
Applicant
Coagusense, Inc.
Product Code
GJS · Hematology
Decision Date
Feb 21, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7750
Device Class
Class 2
Indications for Use
The Coag-Sense® Prothrombin Time (PT) / INR Monitoring System (Self-Test) is an in vitro diagnostic device that provides quantitative prothrombin time (PT) results, expressed in seconds and INR units. It uses fresh capillary whole blood. It is intended for use by properly selected and suitably trained patients or their caregivers on the order of the treating physician to monitor patients who are on anticoagulation therapy. Patients should be stabilized on warfarin-type (coumarin) anticoagulation therapy prior to self-testing. The device is not intended to be used for screening purposes.
Device Story
Portable, handheld PT/INR monitoring system; uses fresh capillary whole blood from finger stick. Device utilizes disposable test strips containing a rotating, spoked wheel; wheel draws blood into reaction well containing dried thromboplastin. Clot formation causes spokes to pick up clot, interrupting an infrared light beam. Meter detects beam interruption to determine PT endpoint. Results displayed as INR and seconds; stored in memory with date/time stamp. Used by patients/caregivers at home or clinicians at point of care. Provides quantitative monitoring to assist in managing warfarin-type anticoagulation therapy; helps patients maintain therapeutic INR levels.
Clinical Evidence
Method comparison study (n=203 patient samples) compared modified device to laboratory reference method (Diagnostica Stago analyzer) across 6 clinical sites. Results demonstrated acceptable correlation and bias. Usability study (n=15 professional users, n=30 home-use patients) confirmed interface and labeling clarity. Bench testing included electrical safety (IEC 61010-1, IEC 61010-2-101), vibration, drop, and transit testing.
Indicated for patients on warfarin-type (coumarin) anticoagulation therapy requiring PT/INR monitoring. Intended for use by trained patients or caregivers at home or by healthcare professionals at the point of care. Patients must be stabilized on therapy prior to self-testing. Not for screening.
Regulatory Classification
Identification
A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).
Coag-Sense Prothrombin Time (PT)/INR Monitoring System (K093243)
Submission Summary (Full Text)
{0}
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
I 510(k) Number:
K183255
II Applicant:
CoaguSense, Inc.
III Proprietary and Established Names:
Coag-Sense Prothrombin Time (PT) / INR Monitoring Device
IV Regulatory Information:
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| GJS | Class II | 21 CFR 864.7750 - Prothrombin Time Test | HE – Hematology |
V Submission/Device Overview:
A Purpose for Submission:
Modification of device(s) previously cleared for professional use under K050243 and patient self-testing under K093243. The hardware and operating system have been updated in the modified device.
B Measurand:
Prothrombin Time and International Normalized Ratio (INR)
C Type of Test:
Quantitative clotting assay
K183255 - Page 1 of 10
{1}
VI Intended Use/Indications for Use:
A Intended Use(s):
Same as Indication(s) for Use
B Indication(s) for Use:
Coag-Sense Prothrombin Time (PT) / INR Monitoring System (Professional Use)
The Coag-Sense Prothrombin Time (PT)/INR Monitoring System is an in vitro diagnostic device that provides quantitative prothrombin time (PT) results, expressed in seconds and INR units. It uses fresh capillary whole blood. It is intended for use by health care professionals at the point of care to monitor patients who are on warfarin-type (coumarin) anticoagulation therapy.
The device is not intended to be used for screening purposes.
Coag-Sense Prothrombin Time (PT) / INR Monitoring System (Self-Test)
The Coag-Sense Prothrombin Time (PT) / INR Monitoring System (Self-Test) is an in vitro diagnostic device that provides quantitative prothrombin time (PT) results, expressed in seconds and INR units. It uses fresh capillary whole blood. It is intended for use by properly selected and suitably trained patients or their caregivers on the order of the treating physician to monitor patients who are on anticoagulation therapy. Patients should be stabilized on warfarin-type (coumarin) anticoagulation therapy prior to self-testing.
The device is not intended to be used for screening purposes.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
Not applicable
VII Device Description:
A Device Description:
The Coag-Sense Prothrombin Time (PT)/INR Monitoring System is a modification to the Immedia Prothrombin Time System cleared for professional use (K050243) and the Coag-Sense Prothrombin Time (PT)/INR Monitoring System cleared for patient self-testing (K093243). The hardware and operating system have been upgraded in the modified device to address parts and technology obsolescence, and to address current communications technologies. The modified Coag-Sense Prothrombin Time (PT)/INR Monitoring System uses the identical test strip, method of detection, thromboplastin reagent and calibration approach as found in the predicate system. The intended use and technological characteristics remain unchanged from that of the predicate
K183255 - Page 2 of 10
{2}
device. The new Coag-Sense Prothrombin Time (PT)/INR Monitoring System is available in two formats; one for Professional Use and one for Patient Self-Testing.
## B Principle of Operation:
The Coag-Sense Prothrombin Time (PT)/INR Monitoring System measures the PT of fresh capillary whole blood using micro-mechanical end-point detection. The test is performed by inserting a test strip into the meter and applying a drop of blood to the sample receptacle of the test strip. The disposable test strip contains a rotating, spoked wheel that draws the sample into the reaction well after it is applied to the sample receptacle. The reaction well contains a pre-measured and dried quantity of thromboplastin. The wheel spokes rotate across the optical path of an infrared light beam. The spokes mix the patient sample with the reagent. As the patient sample transforms into a solid clot, it is picked up by the continuously-rotating spokes and is carried across the path of the infrared light beam interrupting the beam pattern. The time at which the clot interrupts the light beam is reported as the prothrombin time (PT). The PT seconds result is a true Prothrombin Time (patient's actual clotting time) and not a calculated value. The INR is calculated from the PT in seconds result, the Mean Normal Prothrombin Time (MNPT) and International Sensitivity Index (ISI) values for the reagent are determined by comparison to a WHO reference preparation.
## C Instrument Description Information:
| Modes of Operation | Yes | No |
| --- | --- | --- |
| Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐ |
| Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☑ | ☐ |
| Software | | |
| FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types. | ☑ | ☐ |
1. **Instrument Name:**
Coag-Sense Prothrombin Time (PT)/INR Monitoring System
2. **Specimen Identification:**
Patient ID information must be scanned using the optional barcode scanner or manually entered in the meter when prompted.
3. **Specimen Sampling and Handling:**
The Coag-Sense Prothrombin Time (PT)/INR test strip is intended for single-use only. Once the test strip is inserted into the meter, a drop of fresh capillary whole blood sample collected by fingerstick is manually applied to the test strip target area and analyzed by the Coag-Sense Prothrombin Time (PT)/INR meter.
K183255 - Page 3 of 10
{3}
K183255 - Page 4 of 10
4. Calibration:
The calibration parameters for each master lot of reagent is obtained during production by calibrating a subset of the test strips against a working International Reference Preparation (IRP) standard traceable to the tilt tube reference method adopted by the World Health Organization.
For each test strip lot, the calibration (MNPT and ISI) information used to convert PT in seconds to INR is conveyed to the meter on a lot specific Near Field Communication (NFC) tag, which is read using a built-in wireless NFC reader on the meter. This tag also contains strip lot expiration date and allowable low and high QC strip plasma clotting time ranges. The NFC Tag is located on the test strip box and is a micro data chip with antenna that contains the required test strip kit information. It allows transmission of the test strip kit information to the meter.
5. Quality Control:
High Control and Low Control Strips are provided for user quality control purposes and contain the same thromboplastin reagent as the patient test strips as well as plasma of known INR. The CoaguSense control strips form a clot using actual plasma of known INR and the identical lot of thromboplastin used in the manufacturing of the patient test strips. The process of running this liquid quality control tests the reagents' ability to generate a clot as well as the analyzer's ability to detect a clot. Control testing should be completed for each new lot of test strip received.
VIII Substantial Equivalence Information:
A Predicate Device Name(s):
Immedia Prothrombin Time System
Coag-Sense Prothrombin Time (PT)/INR Monitoring System
B Predicate 510(k) Number(s):
K050243
K093243
C Comparison with Predicate:
| Device & Predicate Device(s): | Subject Device
K183255 | Predicate Device
K050243 (Professional Use)
K093243 (Self-Test Use) |
| --- | --- | --- |
| Inventor | Inventor | Inventor |
| Intended Use/Indications for Use | Professional Use | Professional Use (K050243) |
{4}
K183255 - Page 5 of 10
| Device & Predicate Device(s): | Subject Device
K183255 | Predicate Device
K050243 (Professional Use)
K093243 (Self-Test Use) |
| --- | --- | --- |
| | The Coag-Sense Prothrombin Time (PT)/INR Monitoring System is an in vitro diagnostic device that provides quantitative prothrombin time (PT) results, expressed in seconds and INR units. It uses fresh capillary whole blood. It is intended for use by health care professional at the point of care to monitor patients who are on warfarin-type (coumarin) anticoagulation therapy.
The Coag-Sense PT/INR System may be used for multiple-patient testing in a professional healthcare setting. The device is not intended to be used for screening purposes. | The Coag-Sense Prothrombin Time (PT)/INR Monitoring System is an in vitro diagnostic device that provides quantitative prothrombin time (PT) results, expressed in seconds and INR units. It uses fresh capillary whole blood. It is intended for use by health care professional at the point of care to monitor patients who are on warfarin-type (coumarin) anticoagulation therapy.
The Coag-Sense PT/INR System may be used for multiple-patient testing in a professional healthcare setting. The device is not intended to be used for screening purposes. |
| | **Self-Testing**
The Coag-Sense Prothrombin Time (PT)/INR Monitoring System (Self-Test) is an in vitro diagnostic device that provides quantitative prothrombin time (PT) results, expressed in seconds and INR units. It uses fresh capillary whole blood.
It is intended for use by properly selected and suitably trained patients or their caregivers on the order of the treating physician to monitor patients who are on anticoagulation therapy.
Patients should be stabilized on warfarin-type (coumarin) anticoagulation therapy prior to self-testing. The device is not intended to be used for screening purposes. | **Self-Testing (K093243)**
The Coag-Sense Prothrombin Time (PT)/INR Monitoring System (Self-Test) is an in vitro diagnostic device that provides quantitative prothrombin time (PT) results, expressed in seconds and INR units. It uses fresh capillary whole blood.
It is intended for use by properly selected and suitably trained patients or their caregivers on the order of the treating physician to monitor patients who are on anticoagulation therapy.
Patients should be stabilized on warfarin-type (coumarin) anticoagulation therapy prior to self-testing. The device is not intended to be used for screening purposes. |
{5}
| Device & Predicate Device(s): | Subject Device
K183255 | Predicate Device
K050243 (Professional Use)
K093243 (Self-Test Use) |
| --- | --- | --- |
| Test Principle | Rotating wheel physically lifts the clot at the time coagulation occurs and is detected by interruption of infrared light beam | Same |
| Clotting Reagent | Dried recombinant rabbit thromboplastin, stabilizers and preservatives | Same |
| Patient Test Strips | Disposable plastic strip with rotating wheel | Same |
| Blood Sample Size | 10–12 μL | Same |
| Control Strips | Hi and Low-test control strips are provided with each carton of strips; each control strips contains plasma of known INR | Same |
| Quality Control | Internal self-check protocol to assure room temperature, timing functions, battery level and optical and mechanical functions are within specification | Same |
| Measurement Range | 0.8 – 8.0 INR | Same |
| Hematocrit Range | 15–60% | Same |
| Replacement | Replaced with 1000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 | Same |
| Memory Capacity | • 2000 patient test results with date and time
• 500 control test results with date and time
• 1000 Operator accounts | 100 tests with time, date, calibration code, controls and errors |
| PC and Printer Connectivity Method | Wired - Ethernet, USB Wireless - WiFi, Bluetooth | Wired - RS-232 |
| Communications Port | USB | RS-232 |
| User Interface Screen | Color LCD Touch Screen | Monochrome LCD display |
| Battery Type/Replaceable | Lithium Polymer battery. Not user replaceable | 4 x 1.5V (AA) alkaline batteries. Replaceable by the user |
IX Standards/Guidance Documents Referenced:
IEC 60601-1-2: 2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
CLSI POCT14-A: 2014 Point-of-Care Monitoring of Anticoagulation Therapy; Approved Guideline
K183255 - Page 6 of 10
{6}
X Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
The precision of the Coag-Sense Prothrombin Time (PT)/INR Monitoring System was evaluated during the clearance of the device for professional use using control material, therapeutic samples (plasma and whole blood), and capillary whole blood. Refer to submission K050243 for the details of this study.
2. Linearity/Assay Reportable Range:
The linear range of the Coag-Sense Prothrombin Time (PT)/INR Monitoring System was established during the clearance of the device for professional use. The system was found to be linear up to an INR of 8.0. Refer to submission K050243 for the details of the linearity study.
3. Analytical Specificity/Interference:
An interference study was conducted to determine the level of interference with the INR results generated by the device for the following interfering substances: bilirubin, hemolysis, heparin, low molecular weight heparin, and lipemia. Refer to submission K050243 for the details of this study.
4. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Stability
The stability of the Coag-Sense PT/INR test strips and the control test strips (high and low) were tested at designated time intervals, under varying storage conditions during the clearance of the device for professional use. Refer to submission K050243 for the details of this study.
Traceability
Each lot of Coag-Sense PT/INR test strips is factory calibrated to a reference lot of human recombinant thromboplastin traceable to the WHO International Reference Preparation, rTF/16.
5. Detection Limit:
Detection limits for the Coag-Sense Prothrombin Time (PT)/INR Monitoring System were established during the clearance of the device for professional use. Refer to submission K050243 for the details of this study.
6. Assay Cut-Off:
K183255 - Page 7 of 10
{7}
Not applicable
## B Comparison Studies:
### 1. Method Comparison with Predicate Device:
A study was conducted comparing the modified device to the predicate device at a single-site using whole blood from normal subjects not on oral anticoagulation therapy as well as from subjects on oral anticoagulation therapy. To validate the performance of the system across the measuring range, plasma samples ranging from 0.9 to 7.4 INR were also tested.
Test meters representing the predicate device (PT1) and the modified device (PT2) were paired in three sets (PT1 vs. PT2) analyzing identical test samples. The test samples included contrived plasma control samples covering medical decision levels of PT/INR measurements as well as whole blood samples from therapeutic patients and normal subjects (not on warfarin therapy). Plasma samples spanning the INR measuring range of the device (INR 0.9–7.4) were included in the study. Each test sample in this study was tested five times on each meter, for a total of 15 data points for PT1 and 15 data points for PT2.
In addition, whole blood samples were obtained from five different therapeutic patients and five different normal subjects. Three separate fingerstick samples were collected, each fingerstick providing two samples for each set of meters. Bland-Altman bias plots were created to evaluate agreement between the two quantitative measurements. The resulting graph shows the difference between the two paired measurements (PT1 -PT 2) plotted against the average of these measures ((PT1+PT2)/2). The analysis demonstrates that the acceptance criteria were met, and the modified device has acceptable bias across samples and INR ranges when compared to the predicate device.
**Calibration Comparison: RTF/09 vs. RTF/16**
During the clearance of the predicate device (PT1), reagents used in the test and control strips were calibrated against International Reference Preparation (IRP) rTF/09. Due to the phasing out of IRP rTF/09, reagents used in the test and control strips for the modified device (PT2) were calibrated to the current IRP rTF/16. To ensure that the change in IRP did not impact the performance of the modified device, an internal study was conducted to compare strip lots calibrated against rTF/09 against strip lots calibrated against rTF/16. Nine different plasma samples spanning the AMR were used in the study. The samples were tested on both PT1 and PT2 meters using test strips containing reagent that were calibrated using IRP rTF/09 and a current test strip lot containing reagents calibrated using IRP rTF/16. Linear regression and bias analysis demonstrate that the two lots of test strips, rTF/09 and rTF/16, correlate well with acceptable performance across the AMR, and that the modified PT2 meter performs the same as the predicate PT1 meter.
**Comparison to Reference Method**
Performance of the Coag-Sense PT/INR Monitoring System (modified device) was evaluated by comparison to the laboratory reference method. Customer conducted correlation studies were performed at six clinical sites using 12 POC operators. A fingerstick sample was
K183255 - Page 8 of 10
{8}
collected from each patient and the capillary whole blood was applied to the Coag-Sense test strip. A venous sample was collected at the same time and sent for testing on the plasma-based Diagnostica Stago analyzer. All Coag-Sense test strips used in these studies were calibrated against IRP rTF/16. A total of 203 patient samples were analyzed across the six sites, including INR values ranging between 0.9–5.1.
The pooled customer data was analyzed using regression analysis; the correlation coefficient (r), slope and intercept were determined for the data set, and graphs of the regression and bias analysis results were generated. The data demonstrate acceptable performance.
2. Matrix Comparison:
Not applicable
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable
2. Clinical Specificity:
Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Usability Testing
A usability study was conducted to verify changes in the user interface. To gain feedback on the general user interface flow and labeling materials, a formative evaluation was performed by 10 users unfamiliar with the use of home-use devices. The formative evaluation was followed by a summative validation study which was performed by two groups: 1) professional users and 2) home use patients. The professional user group included 10 users with some experience on the original Coag-Sense meter and five users who had familiarity with fingerstick point-of-care devices (n= 15 professional users). The study was performed at three sites for the professional group. The home use patient group included 20 users experienced with fingerstick testing at home and 10 inexperienced users, not familiar with home test monitoring devices (n= 30 home use patients).
Once testing concluded, both groups of users were asked to complete a questionnaire to evaluate the ease of use for various features of the meter and the clarity of the instructions for use. All usability testing results were determined to be acceptable.
D Clinical Cut-Off:
Not applicable
E Expected Values/Reference Range:
K183255 - Page 9 of 10
{9}
A normal range study was conducted on 120 healthy subjects not on anticoagulation therapy. The PT reference interval was established as 11.6–14.5 seconds and the INR range was established as 0.8–1.2 during the clearance of the device for professional use. Refer to submission K050243 for the details of this study.
## F Other Supportive Instrument Performance Characteristics Data:
### Validation and Verification
Validation and verification studies including bench testing, transportation testing, and electrical safety testing were performed on the modified Coag-Sense PT/INR Monitoring System. Bench testing included the following studies: Viral Inactivation Study, Cleaning/Disinfection Study, Component Testing (including power consumption, battery, clock, temperature, humidity, memory, and communication), Drop Test, and Vibration Testing. In addition, Transit Testing was conducted to demonstrate that the meter can withstand shipping simulation, including drop and shock conditions during transportation.
### Electrical Safety Testing
Electrical safety testing was performed per the applicable sections of the following electrical safety standards:
- IEC 61010-1:2010; Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
- IEC 61010-2-101:2015; Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
### Software Verification and Validation
Software verification and validation was performed on functions of the software to ensure that the software performs as intended. Verification and validation testing of the Coag-Sense PT/INR Monitoring System demonstrated that the software performed as intended and demonstrated acceptable performance. Cybersecurity documentation was prepared according to FDA Guidance Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (October 2, 2014).
The verification and validation tests passed all predetermined acceptance criteria. The modified device met design input requirements, product specifications and relevant standards.
## XI Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## XII Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K183255 - Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.