This intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder.
Device Story
Hydrophilic-coated, phthalate-free PVC intermittent catheter; includes protective sleeve and introducer tip to shield against distal urethral bacteria; features integrated urine collection bag. Hydration achieved via internal water compartment. Designed for single-use management of urinary incontinence. Used by patients or caregivers in various settings; facilitates drainage when external receptacles are unavailable. Provides sterile, no-touch insertion; aids patients with limited dexterity via specialized packaging. Pocket version offers discreet storage.
Clinical Evidence
Bench testing only. Performance properties verified against EN 1618, EN 1616, EN 13868, ISO 8669-2, and ASTM-D1003-13. Biocompatibility confirmed per ISO 10993-1, 10993-5, and 10993-12. Sterilization validated per ISO 11137-1, ISO 11137-2, and AAMI/ANSI/ISO 11737-1/2.
Technological Characteristics
Phthalate-free PVC catheter; PVP-based hydrophilic coating; gamma irradiation sterilization. Features protective sleeve, introducer tip, and integrated urine collection bag. Hydration via internal water compartment. Standards: EN 1618, EN 1616, EN 13868, ISO 8669-2, ASTM-D1003-13, ISO 10993, ISO 11137, AAMI/ANSI/ISO 11737.
Indications for Use
Indicated for male, female, and pediatric patients requiring bladder urine drainage via urethral catheterization.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 18, 2018
Hollister Incorporated Michelle Schiltz-Taing Regulatory Affairs Manager 2000 Hollister Drive Libertyville, IL 60048
Re: K183253
> Trade/Device Name: ValPro 2 Plus and VaPro 2 Plus Pocket Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: GBM Dated: November 19, 2018 Received: November 27, 2018
Dear Michelle Schiltz-Taing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Glenn B. Bell -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| 510(k) Number (if known) | K183253 |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | VaPro 2 Plus and VaPro 2 Plus Pocket |
| Indications for Use (Describe) | This intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder. |
| Type of Use (Select one or both, as applicable) | <div> <span><label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label></span> </div> <div> <span><label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label></span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
ริสวเขตสิริ คลุกเบิน ตินิค HT JAヨH สิต สิโน ที่ 143MT ค.ศ. 1977 ludications for Use
See PRA Statement below.
Expiration Date: 06613012020
For
{3}------------------------------------------------
K183253
Page 1 of 3
Image /page/3/Picture/2 description: The image shows the Hollister logo. The logo consists of a red, stylized cross-like symbol on the left and the word "Hollister." in gray on the right. The logo is simple and modern, and the colors are eye-catching.
# 510(k) Summary
| Applicant: | Hollister Incorporated<br>2000 Hollister Drive<br>Libertyville, IL 60048 |
|----------------------|------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Michelle Schiltz-Taing<br>Hollister Incorporated<br>2000 Hollister Drive<br>Libertyville, IL 60018<br>(t) 847-680-2122 |
| Date Prepared: | 17 December 2018 |
| Trade Name: | VaPro 2 PlusTM and VaPro 2 Plus PocketTM |
| Device Common Name: | Catheter, Urethral |
| Classification Name: | Urological catheter and accessories |
| Regulation Number: | 21 CFR 876.5130 |
| Product Code: | GBM |
| Device Class: | II |
| Review Panel: | Gastroenterology/Urology |
## Predicate Device:
VaPro 2 Intermittent Catheter, K180824 by Hollister Incorporated.
## Reference Devices:
VaPro K141642 by Hollister Incorporated for VaPro 2 Plus. VaPro Plus Pocket K152268 by Hollister Incorporated for VaPro 2 Plus Pocket.
#### Indications for Use:
This intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder.
# Description of Applicant Device:
The VaPro 2 Plus and VaPro 2 Plus Pocket is a hydrophilic coated, single use catheter to be used as a means of managing urinary incontinence by draining urine from the bladder. VaPro 2 Plus and VaPro 2 Plus Pocket has the following features:
{4}------------------------------------------------
- Hydrophilic-coated PVC catheter (phthalate free) ●
- Two smooth catheter eyelets
- Protective sleeve ●
- . Protective introducer tip as a way to shield the sterile catheter from bacteria in the distal urethra during insertion
- Color-coded funnel
- Available with a Firmer Style Catheter (Standard packaging only with packaging ● differentiated as F-Style)
- . Urine Collection bag
The packaging contains a sealed water compartment chamber from which the water migrates to the catheter compartment and hydrates (lubricates) the catheter. The outer packaging is designed to facilitate access for those with limited dexterity. The Pocket version is designed to be discreet and easy to store.
The catheter is connected to a collection bag for use when drainage into a suitable receptacle is not feasible or practical.
## Technological Characteristics:
| | Subject Device<br>VaPro 2 Plus<br>VaPro 2 Plus Pocket | Predicate Device<br>VaPro 2 (K180824) |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| Indication for Use | This intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder | |
| Condition of Use | Single Use | |
| Pre-lubricated | Yes-by water vapor hydration | |
| Ready to use | Yes | |
| End Design | Funnel | |
| Sterile | Yes - Gamma Irradiation | |
| No touch design | Yes - contains sleeve | |
| Hydrophilic Coating | PVP Based (polyvinylpyrollidone) Coating | |
| Protective Introducer Tip | Yes | |
| Protective Cap | Yes | |
| Catheter Material | PVC (phthalate free) | |
| Urine Collection Bag | Yes | No |
The table below summarizes the technological characteristics of VaPro 2 Plus and VaPro 2 Plus Pocket as compared to the predicate device VaPro 2 Intermittent Catheter.
#### Brief Description of Non-Clinical Testing:
The physical performance properties of VaPro 2 Plus and VaPro 2 Plus Pocket met all applicable requirements of EN 1618, EN 1616, EN 13868, ISO 8669-2 and ASTM-D1003-13.
Biocompatibility testing met the requirements of ISO 10993-1, 10993-5 and 10993-12.
{5}------------------------------------------------
Sterilization met all requirements of ISO 11137-1, ISO 11137-2, AAMVANSI/ISO 11737-1 and AAMI/ANSI/ISO 11737-2.
# Conclusion:
It is concluded that the information supplied in this submission has demonstrated that VaPro 2 Plus and VaPro 2 Plus Pocket is substantially equivalent to the legally marketed device VaPro 2 Intermittent Catheter.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.