← Product Code [GEI](/productcode/GEI) · K183240

# Empower RF Catheter (K183240)

_Broncus Medical, Inc. · GEI · Feb 22, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K183240

## Device Facts

- **Applicant:** Broncus Medical, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Feb 22, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Empower RF Catheter is intended to be inserted through the working channel of a flexible bronchoscope and utilized in electrosurgical procedures involving removal/cutting of soft tissues (incision, vaporization, ablation, coagulation and hemostasis).

## Device Story

The Empower RF Catheter is a sterile, single-use monopolar electrosurgical device designed for use with flexible bronchoscopes (minimum 2.0 mm working channel). It features a stainless-steel distal electrode, a nitinol wire shaft with insulation, and a proximal banana plug. The device connects to a compatible electrosurgical unit and requires a separate patient return electrode. Operation is controlled via a footswitch. The device delivers RF energy to soft tissue, causing thermal necrosis to facilitate incision, vaporization, ablation, coagulation, and hemostasis. It is used by clinicians in bronchoscopic procedures to remove airway obstructions. The output is thermal energy; the clinician observes the tissue effect through the bronchoscope to guide treatment, enabling the removal of obstructive tissue to improve airway patency.

## Clinical Evidence

Bench testing only. No clinical data. Verification and validation included biocompatibility (ISO 10993-1, -5, -10, -11), sterilization validation (ISO 11137, AAMI TIR 33), electrical safety (IEC 60601-1, -2-2, -2-18), and performance bench testing (tensile, dimensional, electrical, simulated use, corrosion resistance, radiopacity, scope visualization, and ex-vivo ablation).

## Technological Characteristics

Monopolar RF electrosurgical catheter. Materials: stainless-steel electrode, nitinol wire shaft, polymeric insulation. Dimensions: 1.8 mm OD, 1445 mm minimum working length. Energy: RF via compatible electrosurgical generator. Connectivity: banana plug to electrosurgical unit. Sterilization: E-Beam (SAL 10^-6). Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-2-18, ISO 10993, ISO 11137.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- EC 2.7 Endoscopic Cutter ([K120909](/device/K120909.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left, and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

February 22, 2019

Broncus Medical, Inc Aradhana Dhanabalan Regulatory Affairs Manager 125 Nicholson Lane San Jose, California 95134

Re: K183240

Trade/Device Name: Empower RF Catheter Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: February 19, 2019 Received: February 21, 2019

Dear Aradhana Dhanabalan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by David David Krause -S Krause -S Krause -S
Date: 2019.02.22 14:41:28 -05'00' for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director

Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K183240

Device Name Empower RF Catheter

Indications for Use (Describe)

The Empower RF Catheter is a single-use, electrosurgical device designed to be used with flexible bronchoscopes. It is indicated for electrosurgical procedures involving soft tissue obstructions in the upper airways and tracheobronchial tree.

| Type of Use (Select one or both, as applicable)                                                                                       |                                                                                                                           |
|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

| APPLICANT             | Broncus Medical, Inc.<br>125 Nicholson Lane<br>San Jose, CA 95134                                                                           |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| PRIMARY CONTACT       | Aradhana Dhanabalan<br>Regulatory Affairs Manager<br>Broncus Medical, Inc.<br>Phone: 650-428-1600 Ext.311<br>Email: adhanabalan@broncus.com |
| SECONDARY CONTACT     | Liz Raphael<br>Senior Regulatory Affairs Manager<br>Broncus Medical, Inc.<br>Phone: 408-728-1037<br>Email: Iraphael@broncus.com             |
| DATE PREPARED         | February 15, 2019                                                                                                                           |
| TRADE NAME            | Empower RF Catheter                                                                                                                         |
| COMMON NAME           | Radiofrequency (RF) Catheter                                                                                                                |
| CLASSIFICATION NAME   | Electrosurgical cutting and coagulation devices and accessories                                                                             |
| DEVICE CLASSIFICATION | Class II, 21 CFR §878.4400                                                                                                                  |
| PRODUCT CODE          | GEI                                                                                                                                         |

### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION

The Empower RF Catheter is a sterile, single-use monopolar endoscopic device intended to be inserted through a flexible bronchoscope or a working channel of a flexible bronchoscope with an inner diameter of 2.0 mm.

{4}------------------------------------------------

### [Broncus Medical, Inc.- EMPOWER RF Catheter- Traditional 510(k)]

It has a stainless-steel electrode at the distal end, a flexible catheter shaft comprised of a nitinol wire covered by insulation tubing to prevent kinking and facilitate pushability during bronchoscopic procedures and a standard male banana plug at the proximal end.

The banana plug connects to a monopolar connecting cable of a compatible electrosurgical unit. Empower is activated by a footswitch accessory supplied with the electrosurgical unit. Empower is provided in a monopolar configuration and thus, must be used in conjunction with a commercially available compatible patient return electrode to complete the return path for the RF electrical current.

### INTENDED USE

The Empower RF Catheter is intended to be inserted through the working channel of a flexible bronchoscope and utilized in electrosurgical procedures involving removal/cutting of soft tissues (incision, vaporization, ablation, coagulation and hemostasis).

### INDICATION FOR USE

The Empower RF Catheter is a single-use, electrosurgical device designed to be used with flexible bronchoscopes. It is indicated for electrosurgical procedures involving soft tissue obstructions in the upper airways and tracheobronchial tree.

### SUBSTANTIAL EQUIVALENCE TO

Empower is substantially equivalent to the legally marketed predicate device, the EC 2.7 Endoscopic Cutter (K120909). The subject device has the same intended use, indications for use and substantially the same technological characteristics.

The comparison table below, Table 1, presents side by side comparisons between both the devices. The table illustrates equivalence of the subject Empower RF Catheter to the predicate EC 2.7 Endoscopic Cutter.

| Device Name→<br>Device<br>Characteristics ↓ | Empower RF Catheter<br>(Subject Device) | EC 2.7 Endoscopic Cutter<br>(K120909)<br>(Predicate Device) | Differences |
|---------------------------------------------|-----------------------------------------|-------------------------------------------------------------|-------------|
| Device Classification                       | Class II                                | Class II                                                    | Same        |
| Code of Federal<br>Regulations              | 878.4400                                | 878.4400                                                    | Same        |
| Prescription or OTC                         | Prescription                            | Prescription                                                | Same        |
| Device Type                                 | Sterile, Single-Use                     | Sterile, Single-Use                                         | Same        |

|  |  |  | Table 1: Comparison of Kev Characteristics of Subject Device to the Predicate Device |  |  |
|--|--|--|--------------------------------------------------------------------------------------|--|--|
|  |  |  |                                                                                      |  |  |

{5}------------------------------------------------

| Device Name→<br>Device<br>Characteristics ↓ | Empower RF Catheter<br>(Subject Device)                                                                                                                                                                                                                                                            | EC 2.7 Endoscopic Cutter<br>(K120909)<br>(Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Differences                                                |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Intended Use                                | The Empower RF<br>Catheter is intended to<br>be inserted through the<br>working channel of a<br>flexible bronchoscope<br>and utilized in<br>electrosurgical<br>procedures involving<br>removal/cutting of soft<br>tissues (incision,<br>vaporization, ablation,<br>coagulation and<br>hemostasis). | The EC 2.7 Endoscopic<br>Cutter is intended to be<br>inserted through the<br>working channel of a<br>flexible bronchoscope<br>having an instrument<br>channel diameter of 2.8 mm<br>minimum, a working length<br>of 600 mm, activated by a<br>foot-switch connected to a<br>qualified electrosurgical<br>generator, and utilized in<br>electrosurgical procedures<br>involving removal/cutting<br>of soft tissues (excision,<br>incision, vaporization,<br>ablation) while also<br>providing electrosurgical<br>coagulation and hemostasis. | Same                                                       |
| Indications for use                         | The Empower RF<br>Catheter is a single-use,<br>electrosurgical device<br>designed to be used<br>with flexible<br>bronchoscopes. It is<br>indicated for<br>electrosurgical<br>procedures involving<br>soft tissue obstructions<br>in the upper airways<br>and tracheobronchial<br>tree.             | The EC 2.7 Endoscopic<br>Cutter is a single use<br>electrosurgical instrument<br>designed to be used with<br>flexible bronchoscopes and<br>qualified electrosurgical<br>units. It is indicated for<br>cutting of soft tissue<br>obstructions in upper<br>airways and<br>tracheobronchial tree, and<br>provision of electrosurgical<br>hemostasis during such<br>procedures.                                                                                                                                                                 | Same                                                       |
| Target Tissue                               | Soft Tissue obstructions<br>in the upper airways<br>and tracheobronchial<br>tree.                                                                                                                                                                                                                  | Soft Tissue obstructions in<br>the upper airways and<br>tracheobronchial tree.                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                       |
| Access Method                               | Flexible Bronchoscope<br>with minimum working<br>channel of 2.0 mm                                                                                                                                                                                                                                 | Flexible Bronchoscope with<br>minimum working channel<br>of 2.8 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Different,<br>discussion below                             |
| Energy Used                                 | Monopolar RF                                                                                                                                                                                                                                                                                       | Monopolar RF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                       |
| Device Name→<br>Device<br>Characteristics ↓ | Empower RF Catheter<br>(Subject Device)                                                                                                                                                                                                                                                            | EC 2.7 Endoscopic Cutter<br>(K120909)<br>(Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Differences                                                |
| Energy Source                               | Compatible<br>Electrosurgical<br>Generators                                                                                                                                                                                                                                                        | Compatible Electrosurgical<br>Generators                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                       |
| Operating Principle                         | Uses heat delivered via<br>a distal tip electrode to<br>facilitate treatment<br>through thermal<br>necrosis                                                                                                                                                                                        | Uses heat delivered via a<br>distal tip electrode to<br>facilitate treatment through<br>thermal necrosis                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                       |
| Operating Mode                              | Monopolar<br>configuration requires<br>patient return electrode                                                                                                                                                                                                                                    | Monopolar configuration<br>requires patient return<br>electrode                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                       |
| Device Components                           | Active electrode,<br>catheter shaft, electrical<br>connector                                                                                                                                                                                                                                       | Active electrode, catheter<br>shaft, electrical connector.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                       |
| Active Electrode                            | Stainless steel tip                                                                                                                                                                                                                                                                                | Similar biocompatible<br>electrode materials                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Similar<br>biocompatible<br>materials; hence<br>equivalent |
| Catheter Shaft                              | Flexible shaft,<br>polymeric materials                                                                                                                                                                                                                                                             | Flexible shaft, polymeric<br>materials                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar<br>biocompatible<br>materials; hence<br>equivalent |
| Electrical Connector                        | Banana Plug                                                                                                                                                                                                                                                                                        | Electrical adaptor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Similar electrical<br>port; hence<br>equivalent            |
| Catheter Outer<br>Diameter (OD)             | 1.8 mm                                                                                                                                                                                                                                                                                             | 2.68 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Different,<br>discussion below                             |
| Working Length                              | 1445 mm (minimum)                                                                                                                                                                                                                                                                                  | 750 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Different,<br>discussion below                             |
| Sterilization                               | E-Beam (SAL=10-6)                                                                                                                                                                                                                                                                                  | Ethylene Oxide (SAL=10-6)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Similar Sterility<br>Assurance Level;<br>hence equivalent  |

### [Broncus Medical, Inc.- EMPOWER RF Catheter- Traditional 510(k)]

{6}------------------------------------------------

### [Broncus Medical, Inc.- EMPOWER RF Catheter-Traditional 510(k)]

A smaller OD (outer diameter) and a longer minimum working channel allows Empower to be used with a wide range of bronchoscopes and the longer working length allows greater flexibility and convenience during equipment setup and scope access. Since the difference between the subject and predicate device include minor differences in dimensional characteristics such as lengths and diameters, it does not alter the device's intended use or principle of operation. As such, these differences do not raise additional questions on safety and effectiveness, and therefore Empower RF Catheter is considered substantially equivalent to the predicate.

{7}------------------------------------------------

### PERFORMANCE DATA

Performance testing has been completed to demonstrate substantial equivalence of the subject device to the predicate device. Empower was subjected to the following verification and validation tests, as applicable:

### Biocompatibility

Biocompatibility verification was performed for patient contacting components of Empower in accordance with:

- -ISO 10993-1:2009 Biological evaluation of medical devices-Part 1: Evaluation and testing within risk management process
- -ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for In vitro cytotoxicity
- -ISO 10993-10:2010 Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization
- -ISO 10993-11:2006 Biological evaluation of medical devices-Part 11: Tests for systemic toxicity

### Sterilization and Shelf Life Testing

Empower was validated to achieve sterility assurance level of 10 and adopted into a validated Ebeam sterilization cycle. Packaging validation was performed to ensure that the device maintains package integrity, sterility and labeling requirements.

Sterilization validation was performed in accordance with:

- -AAMI TIR 33:2005 Sterilization of Health Care Products-Radiation- Substantiation of a selected sterilization dose - Method VDmax
- -ISO 11137-1:2006/Amd 2013 Sterilization of health care products -- Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- -ISO 11137-2:2013 Sterilization of health care products -- Radiation -- Part 2: Establishing the sterilization dose
- -ISO 11137-3:2006 Sterilization of health care products -- Radiation -- Part 3: Guidance on dosimetric aspects

### Electrical Safety Testing

Electrical safety testing was performed for the applicable components of Empower. The results demonstrated compliance to all applicable requirements of the below standards:

- -IEC 60601-1:2006 General requirements for basic safety and essential performance
- -IEC 60601-2-2:2009 Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
- -IEC 60601-2-18:2015 Particular requirements for the basic safety and essential performance of endoscopic equipment

{8}------------------------------------------------

### Performance Bench Testing

The following performance bench tests were performed:

- -Tensile Testing
- Packaging Inspection -
- -Dimensional Inspection
- -Electrical Inspection
- -Simulated Use testing
- -Corrosion Resistance
- Radiopacity verification -
- -Scope Visualization
- Ex-vivo ablation testing -

Biocompatibility testing, sterilization and shelf life testing, electrical safety testing, and performance bench testing were performed on Empower. The results from this testing demonstrates that Empower meets the defined design requirements and supports its safety and effectiveness for its intended use and its substantial equivalence to the predicate device.

### CONCLUSION

The Empower RF Catheter and the predicate device share intended use, indications for use, and performance characteristics. The minor differences in dimensional characteristics between the subject and the predicate device do not raise any new concerns of safety and effectiveness. Furthermore, the results of verification and validation demonstrate reasonable assurance of safety and effectiveness.

The data and information presented within this 510(k) premarket notification supports a determination of substantial equivalence to the predicate device, and market clearance of the Empower RF Catheter.

---

**Source:** [https://fda-staging.innolitics.com/device/K183240](https://fda-staging.innolitics.com/device/K183240)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
