K183190 · Beeken Biomedical, LLC · POD · Sep 25, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183190
Device Name
NuStat
Applicant
Beeken Biomedical, LLC
Product Code
POD · General, Plastic Surgery
Decision Date
Sep 25, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4454
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NuStat is indicated for temporary control of internal organ space bleeding for patients displaying class III or class IV bleeding. It may also be used for control of severely bleeding wounds and traumatic injuries.
Device Story
NuStat is a hemostatic wound dressing composed of 65% continuous filament silica and 35% bamboo cellulose rayon fibers, incorporating a radiopaque thread for X-ray detection. Used in clinical settings by healthcare providers, the device is applied to bleeding sites (internal organ spaces, surgical wounds, traumatic injuries). Its principle of operation is physical: the cellulose component absorbs excess fluid, facilitating the physical aggregation of blood cells and clotting factors at the application site. The silica fibers further influence the coagulation cascade via contact activation. The dressing provides temporary hemostasis, aiding in the management of severe bleeding. It is provided sterile in various sizes and configurations (foil, Tyvek, or LDPE pouches). The device is intended for prescription use.
Clinical Evidence
No human clinical data. Evidence includes bench testing and GLP animal studies (pig liver resection model). Hemostasis achieved within 3 minutes. Radiographic visualization confirmed at placement and 48 hours. No evidence of vascular obstruction, embolization, or infection. Biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, genotoxicity) showed no adverse findings. In-vitro clot assessment (PT/aPTT) showed accelerated clotting. Particulate release was substantially equivalent to the predicate.
Technological Characteristics
Hemostatic gauze composed of 65% silica fiber and 35% bamboo cellulose rayon fiber. Includes radiopaque thread (PP-BaSO4). Dimensions range from 2"x2" to 12"x12". Sterilized via gamma radiation. Mechanism: physical fluid absorption and contact activation of the coagulation cascade. Biocompatibility per ISO 10993. Packaging: Tyvek or LDPE pouches.
Indications for Use
Indicated for patients with class III or class IV bleeding requiring temporary control of internal organ space bleeding, and for control of severely bleeding wounds and traumatic injuries.
Regulatory Classification
Identification
A non-absorbable, hemostatic gauze for temporary internal use is a prescription device intended to be placed temporarily for control of severely bleeding wounds such as surgical wounds and traumatic injuries. The gauze is coated or impregnated with a hemostatic material which may enhance hemostasis by physical means. The device is intended to be removed once the patient is stabilized.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Animal performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Specifically testing must:
(i) Demonstrate that the device is able to achieve hemostasis;
(ii) Demonstrate that the device can be radiographically detected; and
(iii) Assess pertinent safety endpoints including vascular obstruction and adhesion formation.
(2) The device must be demonstrated to be biocompatible.
(3) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following tests must be performed:
(i) In vitro clot assessment;
(ii) Particulate release testing;
(iii) Physical characterization, including swelling percent and particulate size;
(iv) Chemical characterization;
(v) Radiopacity testing; and
(vi) Mechanical integrity testing, including tensile strength and tear strength.
(4) Performance data must demonstrate the sterility of the device.
(5) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(6) Labeling must include the following:
(i) Instructions for use, including an instruction to remove all visible device components by irrigation;
(ii) The maximum amount of time the device may be left within the body;
(iii) A shelf life;
(iv) A contraindication for intravascular use of the device; and
(v) A warning regarding the potential for adhesion formation.
In combination with the general controls of the FD&C Act, the non-absorbable, hemostatic gauze for temporary internal use is subject to the following special controls:
1. Animal performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Specifically testing must:
a) Demonstrate that the device is able to achieve hemostasis;
b) Demonstrate that the device can be radiographically detected; and
c) Assess pertinent safety endpoints including vascular obstruction and adhesion formation.
2. The device must be demonstrated to be biocompatible.
3. Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following tests must be performed:
a) In vitro clot assessment;
b) Particulate release testing;
c) Physical characterization, including swelling percent and particulate size;
d) Chemical characterization;
e) Radiopacity testing; and
f) Mechanical integrity testing, including tensile strength and tear strength.
4. Performance data must demonstrate the sterility of the device.
5. Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
6. Labeling must include the following:
a) Instructions for use, including an instruction to remove all visible device components by irrigation;
b) The maximum amount of time the device may be left within the body;
c) A shelf life;
d) A contraindication for intravascular use of the device; and
e) A warning regarding the potential for adhesion formation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 25, 2019
Beeken Biomedical, LLC % Mary McNamara Vice President of Regulatory Affairs Alira Health 1 Grant Street, Suite 400 Framingham, MA 01702
Re: K183190
Trade/Device Name: NuStat Regulation Number: 21 CFR 878.4454 Regulation Name: Non-Absorbable, Hemostatic Gauze for Temporary Internal Use Regulatory Class: Class II Product Code: POD Dated: August 29, 2019 Received: August 30, 2019
Dear Mary McNamara:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183190
Device Name NuStat
Indications for Use (Describe)
NuStat is indicated for temporary control of internal organ space bleeding for patients displaying class III or class IV bleeding. It may also be used for control of severely bleeding wounds and traumatic injuries.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span style="white-space: nowrap;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> <span style="white-space: nowrap;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| 510(k) Number: | K183190 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Beeken Biomedical, LLC<br>378 Page Street, Suite 201<br>Stoughton, MA 02072 |
| Contact Person: | Richard A. Kendall |
| Preparation Date: | September 23, 2019 |
| Device Name: | NuStat® |
| Device Classification: | Common Name: Temporary, Internal Use Hemostatic Wound Dressing<br>Generic Name: Non-absorbable, hemostatic gauze for temporary<br>internal use<br>Device Classification: Class II<br>Regulation Number: 21 CFR 878.4454<br>Product Code: POD |
| Special Controls: | Complies with the Non-absorbable, hemostatic gauze for temporary<br>internal use as identified in 21 CFR 878.4454 |
| Intended Use / | NuStat is indicated for temporary control of internal organ space bleeding | |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Indications for Use: | for patients displaying class III or class IV bleeding. It may also be used<br>for control of severely bleeding wounds such as surgical wounds and<br>traumatic injuries. | |
| Device Description: | The NuStat is a hemostatic wound dressing that is composed of<br>continuous filament silica and bamboo cellulose rayon fiber and is<br>provided with radiopaque thread. The distribution of cellulose and silica<br>fibers in each dressing is 65% silica fiber and 35% bamboo cellulose<br>rayon fiber. The dressings are available in various sizes and provided<br>sterile in either foil, Tyvek, or LDPE pouched configurations. The NuStat<br>hemostatic wound dressings have a number of hemostatic properties that<br>enhance the ability of the dressing to temporarily control bleeding.<br>NuStat Hemostatic Dressing's mode of action is via absorption of fluid,<br>which results in a physical aggregation of blood cells and clotting factors<br>at the site of application. The radiopaque element allows for detection via<br>x-ray. | |
| Predicate Device: | Z-Medica, LLC's D2 Dressing (DEN160012), decision date June 30,<br>2017. | |
| Beeken Biomedical 510k) Summary | NuStat K183190 | 1 Page |
| Reference Device: | Beeken Biomedical, LLC's NuStat XR (K160578), decision date June<br>29, 2016. | |
| Special Controls: | The performance testing demonstrates the device performs as intended<br>under anticipated conditions of use, per the special controls for the non-<br>absorbable, hemostatic gauze for temporary internal use as identified in<br>21 CFR 878.4454 | |
| Mechanism of Action: | The principle of operation for NuStat is via absorption of fluid, resulting<br>in a physical aggregation of blood cells and clotting factors.<br>The cellulose component absorbs excess fluid which results in a<br>physical aggregation of blood cells and clotting factors at the site of<br>application. | |
| Comparison of<br>Technological<br>Characteristics: | Fundamental scientific technology, including design, are equivalent to<br>the predicate device D2 Dressing (DEN160012) and identical to<br>reference device NuStat XR (K160578). The key technological and<br>performance similarities examined among devices are as follows: | |
| | Mechanism of Action - Equivalent to the predicate device and identical<br>to the reference device. | |
| | Packaging Materials – Identical to the reference device and equivalent to<br>the predicate device. | |
| | Sterilization Method - Identical to the predicate and reference devices. | |
| Substantial<br>Equivalence: | Predicate device, Z-Medica, LLC's D2 Dressing (DEN160012), is a class<br>II device per decision date June 30, 2017. Reference device, Beeken<br>Biomedical, LLC's NuStat XR (K160578) is the exact same identical<br>device as the subject device. Substantial equivalence to the predicate<br>device is based on intended use, physical and technological<br>characteristics, and comparative device information. This submission<br>seeks to expand the Rx indication for use for the NuStat device. Table 1<br>below demonstrates the substantial equivalence between NuStat and the<br>predicate and reference device. | |
{4}------------------------------------------------
{5}------------------------------------------------
| Substantial Equivalence | | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Predicate Device | Reference Device | Equivalence |
| Applicant | Beeken Biomedical,<br>LLC | Z-Medica, LLC | Beeken Biomedical,<br>LLC | N/A |
| Device Name | NuStat | D2 Dressing | NuStat XR | N/A |
| 510(k)/De Novo<br>Number | TBD | DEN160012 | K160578 | N/A |
| Decision Date | TBD | June 30, 2017 | June 29, 2016 | N/A |
| Device Classification<br>Name | Non-absorbable,<br>hemostatic gauze for<br>temporary internal<br>use; | Non-absorbable,<br>hemostatic gauze for<br>temporary internal use | Dressing, Wound,<br>Drug | Equivalent |
| Regulation Number | 878.4454; | 878.4454 | Unclassified | Equivalent |
| Product Code | POD | POD | FRO | Equivalent |
| 510(k) Review Panel | General & Plastic<br>Surgery | General & Plastic<br>Surgery | General & Plastic<br>Surgery | Identical |
| Device Description | | | Equivalent. | |
| The NuStat is a<br>hemostatic wound<br>dressing that is<br>composed of<br>continuous filament<br>silica and bamboo<br>cellulose with<br>radiopaque thread.<br>The distribution of<br>cellulose and silica<br>fibers in each<br>dressing is 65% silica<br>fiber and 35%<br>cellulose. The<br>dressings are<br>available in various<br>sizes in either Tyvek<br>or LDPE pouched<br>configurations. The<br>NuStat range of<br>hemostatic wound<br>dressings have a<br>number of hemostatic<br>properties that<br>enhance the ability of<br>the dressing to<br>temporarily control<br>bleeding. The<br>radiopaque element<br>allows for detection<br>via x-ray. The<br>principle of operation<br>for NuStat is via<br>absorption of fluid,<br>resulting in a physical<br>aggregation of blood<br>cells and clotting<br>factors.<br><br>The cellulose<br>component absorbs<br>excess fluid which<br>results in a physical<br>aggregation of blood<br>cells and clotting<br>factors at the site of<br>application. | QuikClot Control+<br>consists of a white to<br>off-white to yellow<br>sterile, x-ray<br>detectable hemostatic<br>dressing and is<br>packaged for aseptic<br>removal.<br><br>QuikClot Control+ is<br>impregnated with<br>kaolin, a naturally<br>occurring, inorganic<br>mineral that<br>accelerates the body's<br>natural clotting<br>process. It's the same<br>active ingredient used<br>in QuikClot Combat<br>Gauze®. Because<br>kaolin contains no<br>animal or human<br>proteins, no thrombin,<br>fibrinogen, botanicals,<br>or shellfish products,<br>there is no risk of<br>allergic responses.<br>Biocompatibility<br>studies have shown<br>that it is safe and has<br>no negative effect on<br>tissues. Integrated<br>double X-ray<br>indicators facilitate<br>detection and removal,<br>reducing the risk of<br>lost or retained<br>product. | The NuStat is a<br>hemostatic wound<br>dressing that is<br>composed of<br>continuous filament<br>silica and bamboo<br>cellulose, and are<br>available with or<br>without radiopaque<br>thread. The<br>distribution of<br>cellulose and silica<br>fibers in each dressing<br>is 65% silica fiber and<br>35% cellulose. The<br>dressings are<br>available in various<br>sizes in either Tyvek<br>or LDPE pouched<br>configurations. The<br>NuStat range of<br>hemostatic wound<br>dressings have a<br>number of hemostatic<br>properties that<br>enhance the ability of<br>the dressing to<br>temporarily control<br>bleeding. The<br>cellulose and<br>continuous filament<br>silica influence the<br>contact activation<br>pathway of the<br>coagulation cascade<br>by absorbing blood<br>fluids, resulting in the<br>localized<br>concentration of<br>platelets and clotting<br>factors. The<br>negatively charged<br>fibers of the<br>continuous filament<br>silica simulate the<br>negative ions secreted<br>by activated platelets,<br>which further<br>influence the<br>coagulation cascade.<br>The radiopaque<br>element allows for | Both are hemostatic<br>wound dressings that<br>utilize a matrix of<br>fibers to absorb blood<br>and accelerate the<br>clotting process. Both<br>use hemostatic<br>material that may<br>enhance hemostasis<br>by physical means.<br><br>This difference does<br>not raise any new<br>issues of safety or<br>effectiveness. | |
| Intended Use / | Rx: | Rx: | Rx: | Equivalent. |
| Indications for Use | NuStat is indicated<br>for temporary control<br>of internal organ<br>space bleeding for<br>patients displaying<br>class III or class IV<br>bleeding. It may also<br>be used for control of<br>severely bleeding<br>wounds such as<br>surgical wounds and<br>traumatic injuries. | D2 Dressing is<br>indicated for<br>temporary control of<br>internal organ space<br>bleeding for patients<br>displaying class III or<br>class IV bleeding. It<br>may also be used for<br>control of severely<br>bleeding wounds such<br>as surgical wounds<br>and traumatic injuries. | NuStat is a single-use<br>hemostatic wound<br>dressing applied<br>externally with<br>mechanical<br>compression to<br>temporarily control<br>bleeding in<br>lacerations, punctures,<br>abrasions, surgical<br>wounds (operative,<br>postoperative,<br>dermatological, etc.)<br>and traumatic injuries.<br>OTC:<br>NuStat is indicated to<br>temporarily control<br>bleeding in minor<br>cuts, lacerations,<br>punctures, abrasions<br>and incisions. | Subject device uses<br>both intended use /<br>indications for use of<br>predicate and<br>reference device.<br>Additional testing<br>was performed in<br>support of expanding<br>the Rx indication for<br>use in this<br>submission. |
| Mechanism of<br>Action | The gauze is knit<br>from silica and<br>cellulose fibers,<br>hemostatic material<br>that may enhance<br>hemostasis by<br>physical means.<br>Knitted material<br>forms a physical<br>structure which acts<br>as a fluid absorbent,<br>aggregating platelets<br>and red blood cells. | The gauze is<br>impregnated with<br>kaolin, hemostatic<br>material that may<br>enhance hemostasis by<br>physical means.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.