← Product Code [FGB](/productcode/FGB) · K183188

# Flexible Sensor-Ureterorenoscopes BOA Vision EF, Flexible Sensor-Ureterorenoscopes COBRA Vision EF (K183188)

_Richard Wolf Medical Instruments Corporation · FGB · May 3, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K183188

## Device Facts

- **Applicant:** Richard Wolf Medical Instruments Corporation
- **Product Code:** [FGB](/productcode/FGB.md)
- **Decision Date:** May 3, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Flexible Sensor-Ureterorenoscopes BOA Vision EF and Flexible Sensor-Ureterorenoscopes COBRA Vision EF are used for visualizing body cavities and organs via natural and surgically created passages. These products are exclusively intended for use by specialized medical personnel and must only be used by medically qualified and adequately instructed persons. For examination, diagnostics and/or therapy in conjunction with endoscopic accessories. The product is used in the following disciplines: Urology: - Urogenital tract - Nephroscopy

## Device Story

Flexible video ureterorenoscopes (BOA Vision EF, COBRA Vision EF) used for visualization of urogenital tract and nephroscopy. Device consists of handle, flexible shaft, and active deflection tip. CMOS sensor at tip captures images; signal processed by external ENDOCAM Flex HD controller (not part of submission) for display on monitor. Integrated LED illumination. COBRA model includes secondary 0.8mm channel with laser adjustment unit for fiber alignment/stone contact. Used in clinical settings by urologists/specialized medical personnel. Provides visual access for diagnostic/therapeutic endoscopic procedures. Benefits include improved kink resistance and flexibility compared to predicate; laser adjustment unit facilitates precise stone contact.

## Clinical Evidence

No clinical data or animal studies were performed. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and mechanical/optical performance testing (ISO 8600).

## Technological Characteristics

Flexible video endoscope; CMOS imaging sensor; LED illumination. Materials: stainless steel, optical glass, polyurethane, PFA, epoxy. Working channel 1.2mm; COBRA model adds 0.8mm laser channel. Deflectable distal tip. Non-sterile; requires reprocessing. Complies with IEC 60601-1, IEC 60601-2-18, IEC 60601-1-6, IEC 60601-1-2, and ISO 8600.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Flexible Video Uretero Renoscope Flex XC ([K131369](/device/K131369.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 3, 2019

Richard Wolf Medical Instruments Corporation Mike Loiterman US Head of Regulatory - QA/QC 353 Corporate Woods Parkway Vernon Hills, IL 60061

Re: K183188

Trade/Device Name: Flexible Sensor-Ureterorenoscopes BOA Vision EF Flexible Sensor-Ureterorenoscopes COBRA Vision EF Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FGB, FGA, ODC Dated: April 9, 2019 Received: April 11, 2019

Dear Mike Loiterman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Glenn Bell, Ph.D. Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K183188

#### Device Name

Flexible Sensor-Ureterorenoscopes BOA Vision EF Flexible Sensor-Ureterorenoscopes COBRA Vision EF

Indications for Use (Describe)

The Flexible Sensor-Ureterorenoscopes BOA Vision EF and Flexible Sensor-Ureterorenoscopes COBRA Vision EF are used for visualizing body cavities and organs via natural and surgically created passages.

These products are exclusively intended for use by specialized medical personnel and must only be used by medically qualified and adequately instructed persons.

For examination, diagnostics and/or therapy in conjunction with endoscopic accessories. The product is used in the following disciplines: Urology:

- Urogenital tract

- Nephroscopy

Type of Use (Select one or both, as applicable):

| <span style="font-size: 14px;">☑ Registration Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                                        |

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 5. 510(k) Summary

#### I. SUBMITTER

RICHARD WOLF MEDICAL INSTRUMENTS CORP. 353 Corporate Woods Parkway Vernon Hills, Illinois, USA, 60061

Phone: (847) 913 1113 Fax: (847) 913 0924

Contact Person: Mr. Michael G. Loiterman Date Prepared: May 02, 2019

#### II. DEVICES

Trade Name: Flexible Sensor-Ureterorenoscope BOA VISION EF Flexible Sensor-Ureterorenoscope COBRA VISION EF

| Model numbers: BOA: | 73551071, 73551076                   |
|---------------------|--------------------------------------|
| COBRA:              | 73561071, 73561076                   |
| Accessories:        | 15369108, 15394144, 163914, 15023205 |

Common or Usual Name: Flexible Video Ureterorenoscope, Nephroscope, Endoscope Channel Accessories

Classification Name: Endoscope and Accessories (21 CFR 876.1500)

II Regulatory Class:

Product Code: FGB (Ureteroscope and Accessories, flexible/rigid) ODC (Endoscope Channel Accessory) FGA (Kit, Nephroscope)

#### III. PREDICATE DEVICE

Flexible Video Uretero Renoscope Flex XC (K131369)

This predicate has not been subject to a design-related recall.

{4}------------------------------------------------

# 5.1 Description

The Flexible Sensor-Ureterorenoscope BOA VISION EF (short name: BOA) and Flexible Sensor-Ureterorenoscopes COBRA VISION EF (short name: COBRA) are flexible video endoscopes used for visualizing body cavities and organs via natural and surgically created passages.

The Flexible Sensor-Ureterorenoscopes compromise a handle and a flexible shaft connected to the handle with an active controlled deflection section at the endoscope tip.

The video image of the Flexible Sensor-Ureterorenoscopes is produced by a CMOS imaging sensor located at the tip of the insertion shaft together with the objective lens. The imaging sensor with the electronics in the handle generates a signal that is transferred to the designated camera controller ENDOCAM Flex HD (cleared with K161204), which processes the signals to display them on a monitor (not part of this submission).

The Flexible Sensor-Ureterorenoscopes use integrated LED lights for illumination.

The Flexible Sensor-Ureterorenoscopes provide a working channel with 1.2 mm diameter and a working length of 680 mm length. A Luer sealing cap may be used to seal the provided Luer connectors if not used.

The Flexible Sensor-Ureterorenoscope COBRA VISION EF has a second separate channel with 0.8 mm diameter for laser fibers or continuous irrigation. The laser fiber channel is equipped with a laser adjustment unit for housing, clamping and aligning the laser fiber, which support to move the clamped laser fiber sensitively by turning an adjustment ring for contacting the stone.

Parts of the Flexible Sensor-Ureterorenoscopes have direct patient contact. When used as intended, the duration of tissue contact is limited (up to 24 hours). The used raw material with direct or indirect patient consist of Stainless steel, optical glass, Polyurethane, Perfluoroalkoxy (PFA) and adhesives (epoxy).

# 5.2 Indications for Use

The Flexible Sensor-Ureterorenoscope BOA VISION EF and Flexible Sensor-Ureterorenoscope COBRA VISION EF are used for visualizing body cavities and organs via natural and surgically created passages.

These products are exclusively intended for use by specialized medical personnel and must only be used by medically qualified and adequately instructed persons.

For examination, diagnostics and/or therapy in conjunction with endoscopic accessories.

The product is used in the following disciplines:

Urology:

- Urogenital tract -
- Nephroscopy -

{5}------------------------------------------------

# 5.3 Comparison with Predicate Device

The new Flexible Sensor-Ureterorenoscope BOA VISION EF and Flexible Sensor-Ureterorenoscope COBRA VISION EF and the predicate device are optoelectronic devices used for visualizing body cavities and organs via natural and surgically created passages, for diagnostic and therapeutic use on human patients in the urogenital tract and kidney. Thus, the intended use can be declared as same.

The new Flexible Sensor-Ureterorenoscopes BOA VISION EF and Flexible Sensor-Ureterorenoscopes COBRA VISION EF have equivalent technical characteristics and fundamental design as the predicate device. The operating principle, mechanical design, dimensions and device materials are same or equivalent. The submitted devices and the predicate device are delivered non-sterile and must be reprocessed before and after each use.

The small differences between the new Flexible Sensor-Ureterorenoscopes and the predicate device are mainly related to the following aspects:

#### Design/Material:

The Flexible Sensor-Ureterorenoscope BOA VISION EF and Flexible Sensor-Ureterorenoscope COBRA VISION EF with their working channel have similar diameter, length and deflection angle compared to the predicate device. Different material was used for the new Flexible Sensor-Ureterorenoscopes to improve kink characteristics and flexibility of the insertion part and deflectable distal tip tubing.

#### Laser fiber channel:

The Flexible Sensor-Ureterorenoscope COBRA VISION EF provides an additional channel for laser fibers or continuous irrigation. The use of laser fiber and irrigation are possible on the predicate device also through the working channel. The laser fiber channel is equipped with a laser adjustment unit for housing, clamping and aligning the laser fiber, which supports the clamped laser fiber to move sensitively by turning an adjustment ring for contacting the stone. Without the laser adjustment unit, the laser fiber still can be guided manually by the user the same way as it is done in the predicate device.

### 5.4 Performance Testing

The following performance data were provided in support of the substantial equivalence determination.

#### Biocompatibility

Biological Risk Assessment, considering FDA Guidance -Use of International Standard ISO 10993, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", Guidance for Industry and Food and Drug Administration Staff, issued June 16, 2016, was performed and necessary tests were conducted to ensure the biocompatibility of the product.

{6}------------------------------------------------

Considering the categorization of the Flexible Sensor-Ureterorenoscopes as externally communicating devices in limited (≤24h) contact with tissue, the following endpoints were tested according to ISO 10993 standards: Cytotoxicity, Sensitization, Intracutaneous reactivity/irritation, acute systemic toxicity, material mediated pyrogenicity.

This demonstrates the biocompatibility of the Flexible Sensor-Ureterorenoscopes when used as intended.

#### Cleaning and Sterilization

Validation of reprocessing was performed in accordance with the FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling", Guidance for Industry and Food and Drug Administration Staff, issued March 17, 2015.

#### Electrical safety and Electromagnetic Compatibility

Richard Wolf's Flexible Sensor-Ureterorenoscope BOA VISON EF and Flexible Sensor-Ureterorenoscope COBRA VISION EF comply with the standards regarding electrical safety and electromagnetic compatibility:

- IEC 60601-1: 2005/AMD1:2012: Medical electrical equipment Part 1: General requirements for basic safety and essential performance; with US national differences according to ANSI/AAMI ES60601-1: 2005 / A2:2010.
- IEC 60601-2-18:2009: Part 2-18: Particular requirements for basic safety and essential performance of endoscopic equipment
- IEC 60601-1-6:2010/AMD1:2013: Medical electrical equipment Part 1-6: General requirements for safety - Collateral Standard: Usability
- IEC 60601-1-2 :2014 (Limit class: B); / ANSI AAMI IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic Disturbances - Requirements and test

#### Photobiological safety

The used LEDs in submitted Flexible Sensor-Ureterorenoscope BOA VISION EF and Flexible Sensor-Ureterorenoscope COBRA VISION EF were tested according to the following standard:

- IEC 62471:2006, modified + EN 62471:2008: Medical electrical equipment, Photobiological safety of lamps and lamp systems
#### Mechanical and Optical Performance

The Flexible Sensor-Ureterorenoscope BOA VISION EF and Flexible Sensor-Ureterorenoscope COBRA VISION EF were designed to comply with applicable parts of ISO 8600. Optical measurements were performed according to applicable part of ISO 8600 standard.

Mechanical characteristics were tested and include leakage tightness, bending, deflection endurance, withstand of channel.

In addition, comparative testing related to image quality parameters was performed for submitted Flexible Sensor-Ureterorenoscopes and the predicate device to support substantial equivalence.

{7}------------------------------------------------

#### Software Verification and Validation Testing

The submitted Flexible Sensor-Ureterorenoscope BOA VISION EF and Flexible Sensor-Ureterorenoscope COBRA VISION EF do not contain software.

#### Animal and Clinical Study

Animal Study and Clinical Study were not performed for submitted products.

### 5.5 Conclusion

Richard Wolf's Flexible Sensor-Ureterorenoscope BOA VISION EF, Flexible Sensor-Ureterorenoscope COBRA VISION EF and their accessories have the same intended use as the predicate device. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as or better than the legally marketed device. For these reasons, Richard Wolf's Flexible Sensor-Ureterorenoscope BOA VISION EF and Flexible Sensor-Ureterorenoscope COBRA VISION EF and their accessories are substantially equivalent to the legally marketed device.

---

**Source:** [https://fda-staging.innolitics.com/device/K183188](https://fda-staging.innolitics.com/device/K183188)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).