← Product Code [QEW](/productcode/QEW) · K183179

# CAPERE Thrombectomy System (K183179)

_Vascular Medcure, Inc. · QEW · Feb 25, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K183179

## Device Facts

- **Applicant:** Vascular Medcure, Inc.
- **Product Code:** [QEW](/productcode/QEW.md)
- **Decision Date:** Feb 25, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5150
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The CAPERE™ Thrombectomy System is indicated for: - Non-surgical removal of soft emboli and thrombi from blood vessels. - Injection, infusion and/or aspiration of contrast media and other fluids into blood vessel. The CAPERE™ Thrombectomy System is intended only for use in the peripheral vasculature and is not intended for use in the pulmonary arteries.

## Device Story

CAPERE Thrombectomy System is a percutaneous mechanical thrombectomy device for peripheral vasculature. System comprises 8Fr delivery catheter and funnel catheter. Physician accesses vasculature transfemorally or via jugular vein. Device utilizes nitinol wire basket to capture and mechanically remove thrombi/emboli; does not rely on aspiration for clot removal, though side port allows fluid injection/aspiration. Procedure involves advancing catheters to therapy site, unsheathing funnel, and deploying nitinol basket past obstruction. Basket is retracted and extended to surround clot, then withdrawn into funnel catheter for removal. Device provides non-surgical alternative for clot management, potentially reducing need for invasive interventions.

## Clinical Evidence

No human clinical data. Evidence includes biocompatibility, package integrity, accelerated aging, and functional bench testing (ISO 10555-1, ISO 594-1). Simulated use testing confirmed clot retrieval without basket rupture or catheter damage. Confirmatory non-GLP in vivo animal study validated system usability, delivery, and deployment in 8Fr configuration.

## Technological Characteristics

Materials: Nitinol basket. Components: 8Fr funnel catheter, delivery catheter. Energy: Mechanical (manual operation). Connectivity: None. Sterilization: 2X Ethylene Oxide (EO). Standards: ISO 10555-1, ISO 594-1. Form factor: Percutaneous catheter system.

## Regulatory Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

## Predicate Devices

- CAPERE™ Thrombectomy System ([K180722](/device/K180722.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 25, 2019

Vascular Medcure, Inc. % Ms. Janice Hogan Regulatory Counsel Hogan Lovells US LLP 1735 Market Street, Floor 23 Philadelphia, Pennsylvania 19103

Re: K183179

Trade/Device Name: CAPERE Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: January 28, 2019 Received: January 28, 2019

Dear Ms. Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801 and Part 809: medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone

(1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory O'Connell 2019.02.25 13:26:14
Gregory O'Connell 205'00' For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K183179

Device Name CAPERE Thrombectomy System

Indications for Use (Describe) The CAPERE Thrombectomy System is indicated for:

· Non-surgical removal of soft emboli and thrombi from blood vessels.

- · Injection, infusion and/or aspiration of contrast media and other fluids into blood vessel.
The CAPERE™ Thrombectomy System is intended only for use in the peripheral vasculature and is not intended for use in the pulmonary arteries.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Vascular Medcure, Inc.'s CAPERE™ Thrombectomy System

## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Jeffrey P. DuMontelle

Vascular Medcure, Inc.

1500 S. Sunkist Street, Suite H

Anaheim, CA 92806

Office: 657-549-5175

714-915-0886 Cell:

Date Prepared: November 16, 2018

## Name of Device

CAPERE™ Thrombectomy System

## Common or Usual Name

Embolectomy Catheter

## Classification

21 CFR 870.5150, Class II, product code DXE

### Predicate Devices

CAPERE™ Thrombectomy System (K180722)

## Intended Use / Indications for Use

The CAPERE™ Thrombectomy System is indicated for:

- Non-surgical removal of soft emboli and thrombi from blood vessels. o
- Injection, infusion and/or aspiration of contrast media and other fluids into blood vessel. O

The CAPERE™ Thrombectomy System is intended only for use in the peripheral vasculature and is not intended for use in the pulmonary arteries.

#### Device Description

The CAPERE™ Thrombectomy System primarily consists of an 8Fr Delivery Catheter and Funnel Catheter. The CAPERE™ Thrombectomy System is delivered percutaneously via transfemoral or jugular venous access. Once delivered, the System's fine mesh nitinol wire basket is used to capture and mechanically remove emboli and thrombi. The CAPERE™ System does not use aspiration to pull out the thrombus but does have a side port in the funnel catheter that allows aspiration or injection of saline or fluids if needed.

{4}------------------------------------------------

#### Technological Characteristics

The CAPERE™ Thrombectomy System has the same technological characteristics as its predicate device CAPERE Thrombectomy Catheter (K180722). Both CAPERE "1" Thrombectomy Systems have either the same components or substantially equivalent components. Each device system includes a funnel or quide catheter, aelivery catheter, and a mechanism for capturing/removing the soft emboli or thrombi. The quide catheter of the predicate uses a 20F guide catheter while the proposed CAPERE™ Thrombectomy System uses an 8Fr funnel catheter. In addition, both systems use guidewires (0.035" for the predicate; 0.014" for the CAPERE™ Thrombectomy System) during the procedure.

The CAPERE "" System includes a Funnel Catheter and a Delivery Catheter. In the same manner as the previously cleared device, the Funnel Catheter and Delivery Catheter are advanced to the therapy site. The funnel catheter is unsheathed, and the Delivery Catheter is then advanced passed the obstruction. A nitinol basket connected to the Delivery Catheter is deployed by retracting the outer sheath of the catheter.

To capture the soft emboli or thrombi ("clot"), the deployed basket is retracted while simultaneously being extended to surround and capture the length of the clot. The Delivery Catheter and clot-load is then retracted into the Funnel Catheter where the clot is captured and can be removed.

The basket design in the CAPERE™ Thrombectomy Systems is exactly the same (nitinol, 10mm diameter, 11.4cm length) as the basket design in the predicate.

Therefore, the subject CAPERE™ Thrombectomy System has the same technological characteristics as its predicate.

#### Performance Data

The following nonclinical performance testing has been conducted, using the established test methods used in the cleared predicate device, to support the substantial equivalence of the CAPERE™ Thrombectomy System to its predicate device. In all instances, the CAPERE Thrombectomy System functioned as intended under the previously established test methods.

- Biocompatibility of the patient-contacting components of the device was established.
- . Package integrity and accelerated aging studies were completed and demonstrated that the packaging was capable of maintaining a sterile barrier and protecting the device after 2X EO sterilization, simulated handling and shipping, environmental conditioning, and 6-months of accelerated aging in support of a 6-month shelf-life.
- Functional bench testing was conducted (including demonstrated compliance with ● relevant standards such as ISO 10555-1 and ISO 594-1) after 2X EO sterilization, simulated handling and shipping, environmental conditioning and 6-months of aging.
- Simulated Use Testing after 2X EO sterilization, simulated handling and shipping, . environmental conditioning and 6-months of aging was completed to demonstrate that simulated clot can be retrieved under simulated conditions without capture basket rupture, catheter damage, or other adverse device effect.

{5}------------------------------------------------

- Confirmatory In vivo testing in an animal model was performed. Although not strictly required to establish substantial equivalence to the predicate given the minor nature of changes to the predicate (e.q. pre-assembled configuration, smaller diameters), the testing was used to confirm the usability of the system in the pre-assembled 8Fr configuration and gather user feedback on ease of use, etc. in the previously established acute animal model conditions. While, non-GLP, due to the confirmatory nature of the testing, the testing was performed in accordance with all established procedures and methods of the test facility; just without QA audit. The in-vivo testing was also used to confirm that there were no issues with introduction of the delivery catheter to, and deployment of, the funnel and basket, or with the interaction of the system within the vascular blood flow at the treatment site. While all issues were considered low risk, testing was performed out of an abundance of caution and all aspects of the evaluation met the acceptance criteria established by the protocol.
#### Substantial Equivalence

The CAPERE"" Thrombectomy System has the same intended use and indications for use, technological characteristics, and principles of operation as its predicate devices. Nonclinical testing, including functional testing, simulated use testing, and confirmatory in vivo testing in an animal model, demonstrated that the minor differences between the device and predicate (e.g. 8Fr, device length) do not raise new types of safety or effectiveness questions. Thus, the CAPERE™ Thrombectomy System is substantially equivalent to the predicate devices.

### Conclusion

Vascular Medcure's CAPERE" is an Embolectomy Catheter, Class II device that has been evaluated in nonclinical testing in accordance with FDA's recognized standards and preestablished acceptance criteria. Testing demonstrated that the device performs as intended. The CAPERE™ is substantially equivalent to its predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K183179](https://fda-staging.innolitics.com/device/K183179)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
