K183156 · Wontech Co., Ltd. · GEX · Apr 24, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183156
Device Name
V-Laser
Applicant
Wontech Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 24, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The V-Laser laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.
Device Story
V-Laser is a Nd:YAG laser system (1,064 nm and 532 nm) for surgical and aesthetic applications. System comprises main unit, optical fiber cable, handpiece, footswitch, and cradle. Operator controls fluence, frequency, and pulse via LCD touch pad. Device incorporates a Dynamic Cooling Device (DCD) using HFC 134a cryogen to protect skin surface during treatment. Used in clinical settings by medical professionals. Output delivered via handpiece to target tissue; cooling burst provides thermal protection. Benefits include precise laser delivery for dermatological/surgical procedures with integrated skin cooling.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including electrical safety (AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), software validation, and biocompatibility (ISO 10993-5, ISO 10993-10).
Technological Characteristics
Nd:YAG laser; 1,064 nm and 532 nm wavelengths; flashlamp-pumped solid-state rod. Includes Dynamic Cooling Device (DCD) using HFC 134a cryogen. Standards: AAMI ES60601-1 (Ed 3.1), IEC 60601-1-2 (Ed 3), IEC 60601-2-22 (Ed 3.1), IEC 60825-1 (Ed 3). Biocompatibility per ISO 10993-1. User interface via LCD touch pad.
Indications for Use
Indicated for surgical and aesthetic applications in dermatology and general and plastic surgery.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The full name is "U.S. Food & Drug Administration".
WONTECH Co., Ltd. Erin Park Staff of Regulatory Affair 64 Techno 8-Ro. Yuseong-Gu Daejeon, 305-500 KR
April 24, 2019
Re: K183156
Trade/Device Name: V-Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 9, 2018 Received: March 6, 2019
Dear Erin Park:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Digitally signed by Neil Neil R R Ogden -S Date: 2019.04.24 Oaden -S 14:51:18 -04'00' For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K183156
Image /page/2/Picture/1 description: The image shows the logo for WONTECH. The logo features a stylized letter "W" in orange and yellow gradient, followed by the word "WON" in red and "TECH" in gray. The text is in a sans-serif font.
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 305-500 Tel: +82-70-8730-8833 / Fax: +82-70-934-9491
# 005_Indications for Use Statement (FDA Form 3881)
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K183156
Device Name V-Lasaer
Indications for Use (Describe)
The V-Laser laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.
| Type of Use (Select one or both, as applicable) | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <table><tr><td><div> <span> <span style="border: 1px solid black; padding: 0px 2px;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div></td><td><div> <span> <span style="border: 1px solid black; padding: 0px 2px;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div></td></tr></table> | <div> <span> <span style="border: 1px solid black; padding: 0px 2px;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="border: 1px solid black; padding: 0px 2px;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
| <div> <span> <span style="border: 1px solid black; padding: 0px 2px;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="border: 1px solid black; padding: 0px 2px;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### K183156
Image /page/4/Picture/1 description: The image shows the logo for WONTECH. The logo consists of an orange and yellow circular graphic on the left, followed by the text "WONTECH" in red and gray. The "WON" part of the text is in red, while the "TECH" part is in gray.
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 34028 Tel: +82-70-8730-8833 / Fax: +82-70-934-9491
# 510(k) Summary
[As Required by 21 CFR 807.92]
#### 1. Date Prepared [21 CFR 807.92(a)(a)]
October 30, 2018
#### 2. Submitter's Information [21 CFR 807.92(a)(1)]
- Name of Sponsor: ● WON TECH Co., Ltd. Address: 64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, -305-500
- Contact Name: Erin Park / Staff of Regulatory Affair
- Telephone No.: -+82-70-8730-8833
- Fax No.: +82-70-7836-3348 -
- -Email Address: loveyougyuri@wtlaser.com
- Name of Manufacturer: Same as sponsor -Address: Same as sponsor
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Trade Name | V-Laser |
|-----------------------------|-----------------------------------|
| Device Classification Name | Powered Laser Surgical Instrument |
| Regulation Number | 21CFR878.4810 |
| Classification Product Code | GEX |
| Device Class | Class II |
| 510k Review Panel | General & Plastic Surgery |
{5}------------------------------------------------
## K183156
Image /page/5/Picture/1 description: The image shows the logo for WONTECH. The logo consists of a stylized letter "W" in orange and yellow, followed by the text "WONTECH" in red and gray. The "WON" part of the text is in red, while the "TECH" part is in gray.
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 34028 Tel: +82-70-8730-8833 / Fax: +82-70-934-9491
#### 4. Identification of Predicate Device [21 CFR 807.92(a)(3)[
The identified predicate device within this submission are shown as follow:
Primary Predicate
- K153671 ● 510(k) Number:
- Applicant: Cutera, Inc.
- Classification Name: ● Powered Laser Surgical Instrument
- Trade Name: Family of CoolGlide Aesthetic Lasers ●
#### 5. Description of the Device [21 CFR 807.92(a)(4)]
The V-Laser is a Nd:YAG laser operating at wavelengths of 1,064 nm and 532 nm. The V-Laser consists of the main body, optical fiber cable, user-undetachable laser handpiece tip, footswitch, and handpiece cable cradle. The laser output is delivered to the optical fiber terminated by the handpiece. The fluence (energy density), frequency and pulse are controlled from the LCD display/Touch Pad located on the front of the main unit. The LCD display is used to obtain feedback from the system, such as the number of pulses delivered or spot size selected. The Dynamic Cooling Device (DCD) protects the upper layers of the skin with a cooling burst of cryogen.
#### 6. Indications for Use [21 CFR 807.92(a)(5)]
The V-Laser laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for WONTECH. The logo features a stylized letter "W" in orange and yellow gradient, followed by the text "WON" in red and "TECH" in gray. The text is in a bold, sans-serif font.
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 34028 Tel: +82-70-8730-8833 / Fax: +82-70-934-9491
#### 7. Determination of Substantial Equivalence [21 CFR 807.92(a)(6) and 21 CFR 807.92(b)]
There are no significant differences in the technological characteristics of this device compared to the predicate device which adversely affect safety or effectiveness. Provided below is a table summarizing and comparing the technological characteristics of the V-Laser and the predicate device:
| | Proposed Device | Predicate Device |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K183156 | K153671 |
| Manufacturer | WON TECH Co., Ltd. | Cutera, Inc. |
| Model | V-Laser | Family of CoolGlide Aesthetic Lasers |
| Product Code | GEX | GEX |
| Intended Use/<br>Indications for Use | The V-Laser laser system is intended for use<br>in surgical and aesthetic applications in the<br>medical specialties of dermatology and<br>general and plastic surgery. | Excel V is Intended use is for surgical and<br>aesthetic applications in the medical<br>specialties of dermatology and general and<br>plastic surgery. |
| Laser Type | Nd:YAG Laser | Nd:YAG Laser |
| Wavelength | 1064 nm<br>532 nm | 1064 nm<br>532 nm |
| Spot Size | 1064 nm: 2 to 20 mm<br>532 nm: 2 to 20 mm<br>Genesis Mode: 8 mm | 1064 nm: 3 to 18 mm<br>532 nm: 2 to 12 mm<br>Genesis Mode: 8 mm |
| Output | 1064 nm: Max. 50 J<br>532 nm: Max. 10 J | 1064 nm: Max. 100 J<br>532 nm: Max. 12 J |
| Fluence | 1064 nm: 2 to 300 J/cm2<br>532 nm: 1.8 to 42 J/cm2<br>Genesis Mode: 4 to 7 J/cm2 | 1064 nm: 2 to 300 J/cm2<br>532 nm: 1.8 to 42 J/cm2<br>Genesis Mode: 4 to 7 J/cm2 |
| Pulse Duration | 1064 nm: Max. 60 ms<br>532 nm: Max. 40 ms<br>Genesis Mode: Max. 0.3 ms | 1064 nm: Max. 60 ms<br>532 nm: Max. 40 ms<br>Genesis Mode: Max. 0.3 ms |
| Repetition Rate | Max. 10 Hz | Max. 10 Hz |
| Laser Media | Flashlamp-pumped solid state rod | Flashlamp-pumped solid state rod |
| Aiming Beam | 635 nm | 630 to 680 nm |
| Dynamic Cooling Device (DCD) | | |
| Cryogen | HFC 134a | Not applicable |
| DCD Spray Duration | 10 to 100 ms | |
| DCD Delay Duration | Max. 100 ms | |
| DCD Post Spray<br>Duration | 0 to 20 ms | |
The key differences between V-Laser and the predicate are (1) Spot Size and (2) Output, and which do not raise any new safety and effectiveness issues. The V-Laser and predicate device, K153671 have similar applicable fluence ranges, which are given in joules/cm2.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for WONTECH. The logo consists of a stylized "W" in orange and yellow, followed by the text "WON" in red and "TECH" in gray. The "W" is made up of curved lines that create a sense of movement.
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 34028 Tel: +82-70-8730-8833 / Fax: +82-70-934-9491
Verification and validation activities were conducted to establish the performance and safety characteristics of the V-Laser. The results of these activities show that there are no any new safety and effectiveness issues. Therefore, the V-Laser is considered substantially equivalent to the predicate.
#### Non-Clinical Test Summary [21 CFR 807.92(b)(1)]
1) Electrical Safety, Electromagnetic Compatibility and Performance
Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
| Standards No. | Standards<br>Organization | Standard Title | Version | Publication<br>Year |
|---------------|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|---------------------|
| ES60601-1 | AAMI | Medical Electrical Equipment - Part 1: General Requirements for Basic<br>Safety and Essential Performance | Edition 3.1 | 2012 |
| 60601-1-2 | IEC | Medical Electrical Equipment - Part 1-2: General Requirements for<br>Safety - Collateral Standard: Electromagnetic Compatibility -<br>Requirements and Tests | Edition 3 | 2007 |
| 60601-2-22 | IEC | Medical Electrical Equipment - Part 2-22: Particular Requirements for<br>Basic Safety and Essential Performance of Surgical, Cosmetic,<br>Therapeutic and Diagnostic Laser Equipment | Edition 3.1 | 2012 |
| 60825-1 | IEC | Safety of Laser Products - Part 1: Equipment Classification, and<br>Requirements | Edition 3 | 2014 |
- 2) Software Validation
The V-Laser contains MODERATE level of concern software was designed and developed according to a software development process and was verified and validated.
The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.
- 3) Biocompatibility
According to the recommendations of ISO 10993-1 and FDA Blue Book Memo #G95-1, the following tests were performed:
- -Cytotoxicity test according to ISO 10993-5:2009
- -Intracutaneous (intradermal) reactivity test according to ISO 10993-10:2010
- -Skin sensitization test according to ISO 10993-10:2010
#### Clinical Test Summary [21 CFR 807.92(b)(2)]
No clinical studies were considered necessary and performed.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for WONTECH. The logo consists of a stylized letter "W" in orange and yellow gradient, followed by the text "WONTECH" in red and gray. The "WON" part of the text is in red, while the "TECH" part is in gray.
64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea, 34028 Tel: +82-70-8730-8833 / Fax: +82-70-934-9491
### 8. Conclusion [21 CFR 807.92(b)(3)]
Based on the information provided in this premarket notification WON TECH Co., Ltd., concludes that the V-Laser is substantially equivalent in safety and effectiveness to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.