K183139 · Intuitive Surgical, Inc. · KCT · Feb 1, 2019 · General Hospital
Device Facts
Record ID
K183139
Device Name
Endoscope Sterilization Tray
Applicant
Intuitive Surgical, Inc.
Product Code
KCT · General Hospital
Decision Date
Feb 1, 2019
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Intuitive Surgical Endoscope Sterilization Tray is intended for use to encase and protect compatible da Vinci Xi endoscopes for sterilization in any of the following sterilization machines/cycles: - STERRAD 100NX sterilization system using the Express and DUO cycles - STERRAD 100S sterilization system using the Standard cycle - Steris V-PRO maX using the Non Lumen, Flexible, or Lumen cycles - Steris V-PRO 1 Plus using the Non Lumen or Lumen cycles - Steris V-PRO 1 using the V-PRO/Lumen cycle The sterilization cycle parameters of the sterilizers are preset by the manufacturers and are not adjustable. The maximum product load per tray is one da Vinci Xi Endoscope. The length of the da Vinci Xi Endoscope is approximately 600 mm and the diameter of the shaft is 8.8 mm. The maximum weight of the tray and endoscope is 12.2 lbs. The Endoscope Sterilization Tray is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared STERRAD and Steris compatible sterilization wrap in order to maintain sterility of the enclosed medical instrument.
Device Story
Stainless steel and silicone tray with lid; designed to encase and protect da Vinci Xi endoscopes during transport and sterilization. Used in hospital settings by sterile processing staff. Device features grid pattern to allow penetration of sterilization gases (H2O2). Not intended to maintain sterility independently; must be used with FDA-cleared sterilization wrap. Protects endoscope from physical damage during handling and sterilization cycles. Benefits include ensuring instrument integrity and sterility maintenance for surgical use.
Clinical Evidence
No clinical data. Bench testing only. Testing included dimensional/functional verification, biocompatibility (ISO 10993-1), sterilization validation (STERRAD/V-PRO cycles), sterility maintenance after handling (spore aerosol challenge), limits of reuse (repeated cycles), and simulated shipping (ASTM D4169-09). All tests met pre-defined acceptance criteria.
Technological Characteristics
Materials: stainless steel, silicone. Form factor: tray with lid, grid pattern for gas penetration. Sterilization: H2O2 chemical sterilization. Compatibility: STERRAD 100NX, 100S; Steris V-PRO maX, V-PRO 1 Plus, V-PRO 1. Standards: ANSI/AAMI ST77 (bench/durability/corrosion/cleaning/sterilization), ISO 10993-1 (biocompatibility), ASTM D4169-09 (shipping).
Indications for Use
Indicated for encasing and protecting da Vinci Xi endoscopes during sterilization in specified STERRAD and Steris V-PRO systems. Patient population: N/A (device is a sterilization accessory). Contraindications: None stated.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Intuitive Surgical, Inc. Endoscope Sterilization Tray (K180964)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 1, 2019
Intuitive Surgical, Inc. Brandon Hansen Manager, Regulatory Submissions 1266 Kifer Road Sunnyvale, California 94086
Re: K183139
Trade/Device Name: Endoscope Sterilization Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: November 9, 2018 Received: November 13, 2018
#### Dear Brandon Hansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Clarence W. Murray lii III -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183139
Device Name Endoscope Sterilization Tray
Indications for Use (Describe)
The Intuitive Surgical Endoscope Sterilization Tray is intended for use to encase and protect compatible da Vinci Xi endoscopes for sterilization in any of the following sterilization machines/cycles:
- · STERRAD 100NX sterilization system using the Express and DUO cycles
- · STERRAD 100S sterilization system using the Standard cycle
- · Steris V-PRO maX using the Non Lumen, Flexible, or Lumen cycles
- · Steris V-PRO 1 Plus using the Non Lumen or Lumen cycles
- · Steris V-PRO 1 using the V-PRO/Lumen cycle
The sterilization cycle parameters of the sterilizers are preset by the manufacturers and are not adjustable. The maximum product load per tray is one da Vinci Xi Endoscope. The length of the da Vinci Xi Endoscope is approximately 600 mm and the diameter of the shaft is 8.8 mm. The maximum weight of the tray and endoscope is 12.2 lbs. The Endoscope Sterilization Tray is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared STERRAD and Steris compatible sterilization wrap in order to maintain sterility of the enclosed medical instrument.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------------------------------------------|
| | <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
| | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K183139
| 510(k) Owner: | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086 |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Brandon Hansen<br>Manager, Regulatory Submissions<br>Phone Number: 408-523-7485<br>Fax Number: 408-523-8907<br>Email: Brandon.Hansen@intusurg.com |
| Date Summary Prepared: | January 17, 2019 |
| Trade Name: | Endoscope Sterilization Tray |
| Common Name: | Sterilization Tray |
| Classification: | Class II<br>21 CFR 880.6850, Sterilization Wrap |
| Product Codes: | KCT |
| Classification Advisory Committee: | General Hospital |
| Predicate Device: | Intuitive Surgical, Inc. Endoscope Sterilization Tray<br>(K180964) |
Image /page/3/Picture/4 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and in a light gray color. There is a yellow dot above the "I" in "INTUITIVE". Below "INTUITIVE" is the word "SURGICAL" in a smaller font and also in light gray.
{4}------------------------------------------------
### Device Description
The Endoscope Sterilization Tray is a stainless steel tray with silicone inserts and a stainless steel lid. The tray and lid have a grid pattern to allow sterilization gases to penetrate the tray and sterilize the endoscope. The Endoscope Sterilization Tray is designed to encase and protect da Vinci endoscopes during transport and sterilization. The tray is compatible with the STERRAD 100NX and 100S and Steris V-PRO maX V-PRO 1 Plus and V-PRO 1 sterilization systems.
#### Intended Use:
To encase and protect da Vinci endoscopes for sterilization.
### Indications for Use:
The Intuitive Surgical Endoscope Sterilization Tray is intended for use to encase and protect compatible da Vinci Xi endoscopes for sterilization in any of the following sterilization machines/cycles:
- . STERRAD 100NX sterilization system using the Express and DUO cycles
- STERRAD 100S sterilization system using the Standard cycle
- Steris V-PRO maX using the Non Lumen, Flexible, or Lumen cycles
- Steris V-PRO 1 Plus using the Non Lumen or Lumen cycles
- . Steris V-PRO 1 using the V-PRO/Lumen cycle
The sterilization cycle parameters of the sterilizers are preset by the manufacturers and are not adjustable. The maximum product load per tray is one da Vinci Xi Endoscope. The length of the da Vinci Xi Endoscope is approximately 600 mm and the diameter of the shaft is 8.8 mm. The maximum weight of the tray and endoscope is 12.2 lbs. The Endoscope Sterilization Tray is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared STERRAD and Steris compatible sterilization wrap in order to maintain sterility of the enclosed medical instrument.
#### Technological Characteristics Comparison:
| Characteristic | Subject Device<br>Intuitive Surgical, Inc. Endoscope<br>Sterilization Tray<br>(K183139) | Predicate Device<br>Intuitive Surgical, Inc.<br>Endoscope Sterilization Tray<br>(K180964) |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Intuitive Surgical, Inc. | Intuitive Surgical, Inc. |
| Characteristic | Subject Device<br>Intuitive Surgical, Inc. Endoscope<br>Sterilization Tray<br>(K183139) | Predicate Device<br>Intuitive Surgical, Inc.<br>Endoscope Sterilization Tray<br>(K180964) |
| Trade Name | Endoscope Sterilization Tray | Endoscope Sterilization Tray |
| 510(k) No. | K183139 | K180964 |
| Common Name | Sterilization Tray | Identical |
| Regulation No. | 21 CFR 880.6850 | Identical |
| Product Code | KCT | Identical |
| Device Class/<br>Regulation Name | Class II/ Sterilization Wrap | Identical |
| Classification<br>Advisory<br>Committee | General Hospital | Identical |
| Sterilization<br>Method | H2O2 chemical sterilization | Identical |
| Sterility /<br>Disposable or<br>Multiple use | Multiple use | Identical |
| Intended Use | Intended to encase and protect da<br>Vinci endoscopes for sterilization<br>in the STERRAD 100NX<br>sterilization system using the<br>Express and DUO cycles settings<br>and the 100S sterilization system<br>using the Standard cycle setting<br>and in the STERIS V-PRO max,<br>V-PRO 1 Plus and V-PRO 1<br>systems. | Identical |
| Characteristic | Subject Device<br>Intuitive Surgical, Inc. Endoscope<br>Sterilization Tray<br>(K183139) | Predicate Device<br>Intuitive Surgical, Inc.<br>Endoscope Sterilization Tray<br>(K180964) |
| Indications for<br>Use | The Intuitive Surgical Endoscope<br>Sterilization Tray is intended for use<br>to encase and protect compatible da<br>Vinci Xi endoscopes for sterilization<br>in any of the following sterilization<br>machines/cycles:<br>• STERRAD 100NX sterilization<br>system using the Express and DUO<br>cycles<br>• STERRAD 100S sterilization<br>system using the Standard cycle<br>• Steris V-PRO maX using the Non<br>Lumen, Flexible, or Lumen cycles<br>• Steris V-PRO 1 Plus using the Non<br>Lumen or Lumen cycles<br>• Steris V-PRO 1 using the V-<br>PRO/Lumen cycle<br>The sterilization cycle parameters of<br>the sterilizers are preset by the<br>manufacturers and are not adjustable.<br>The maximum product load per tray is<br>one da Vinci Xi Endoscope. The<br>length of the da Vinci Xi Endoscope<br>is approximately 600 mm and the<br>diameter of the shaft is 8.8 mm. The<br>maximum weight of the tray and<br>endoscope is 12.2 lbs. The Endoscope<br>Sterilization Tray is not intended to<br>maintain sterility; it is intended to be<br>used in conjunction with a legally<br>marketed, validated, FDA-cleared<br>STERRAD and Steris compatible<br>sterilization wrap in order to maintain<br>sterility of the enclosed medical<br>instrument. | The Intuitive Surgical Endoscope<br>Sterilization Tray is intended for use to<br>encase and protect compatible da Vinci<br>Xi endoscopes for sterilization in any of<br>the following sterilization<br>machines/cycles:<br>• STERRAD 100NX sterilization<br>system using the Express and DUO<br>cycles<br>• STERRAD 100S sterilization system<br>using the Standard cycle<br>• Steris V-PRO maX using the Non<br>Lumen, Flexible, or Lumen cycles<br>• Steris V-PRO 1 Plus using the Non<br>Lumen or Lumen cycles<br>• Steris V-PRO 1 using the V-<br>PRO/Lumen cycle<br>The sterilization cycle parameters of the<br>sterilizers are preset by the<br>manufacturers and are not adjustable.<br>The maximum product load per tray is<br>one da Vinci Xi Endoscope. The length<br>of the da Vinci Xi Endoscope is<br>approximately 600 mm and the diameter<br>of the shaft is 8.8 mm. The maximum<br>weight of the tray and endoscope is 8.9<br>lbs. The Endoscope Sterilization Tray is<br>not intended to maintain sterility; it is<br>intended to be used in conjunction with<br>a legally marketed, validated, FDA-<br>cleared STERRAD and Steris<br>compatible sterilization wrap in order to<br>maintain sterility of the enclosed<br>medical instrument. |
| Where used<br>(hospital, home,<br>ambulance, etc.) | Hospital | Identical |
| Materials; | stainless steel, silicone | ULTEMTM, silicone, stainless steel |
| Characteristic | Subject Device<br>Intuitive Surgical, Inc. Endoscope<br>Sterilization Tray<br>(K183139) | Predicate Device<br>Intuitive Surgical, Inc.<br>Endoscope Sterilization Tray<br>(K180964) |
| Compatibility<br>with the<br>environment and<br>other devices | Compatible with da Vinci<br>endoscopes | Identical |
| Sterilization<br>system<br>compatibility | STERRAD 100NX and 100S<br>Steris V-PRO maX, V-PRO 1 Plus<br>and V-PRO 1 | Identical |
Image /page/4/Picture/16 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in a larger, sans-serif font above the word "SURGICAL," which is in a smaller, sans-serif font. There is a yellow dot above the word "INTUITIVE".
{5}------------------------------------------------
{6}------------------------------------------------
Image /page/6/Picture/3 description: The image shows the logo for Intuitive Surgical. The logo is in gray and consists of the words "INTUITIVE" on the top line and "SURGICAL" on the bottom line. There is a registered trademark symbol after the word "SURGICAL" and a yellow dot above the word "INTUITIVE".
{7}------------------------------------------------
The subject and predicate devices have the same intended use, indications for use, and materials.
## Summary of Non-Clinical Testing:
Performance test data in accordance with ANSI/AAMI ST77 (bench, durability, corrosion resistance, cleaning, and sterilization tests) demonstrate that the subject device is substantially equivalent to the predicate device and that the design output meets the design input requirements. The testing conducted consisted of dimensional measurements, functional verification, biological safety assessment, cleaning and sterilization and package testing.
The subject Endoscope Sterilization Tray underwent verification testing of dimensional and functional specifications. Trays were assessed for sharp edges, compatible materials, latch security, lid and base interchangeability, weight, and tray markings. The subject device successfully passed all requirements described in the protocol. These results demonstrate that the Endoscope Sterilization Tray meets the functional requirements.
Based upon the toxicological assessment of the material characterization, the established biocompatibility results of the predicate device for intracutaneous injection, sensitization and hemolysis testing for the same materials and same intended clinical use, and the provisions of ISO 10993-1, the biological risk evaluation has determined that the subject Endoscope Sterilization Tray is biocompatible for its intended use as a containment device.
Sterilization validation testing was conducted in the STERRAD 100NX sterilization system using the Express and DUO cycles, in the STERRAD 100S sterilization system using the Standard cycle and in the V-PRO maX using the Non Lumen cycle. The subject
{8}------------------------------------------------
Endoscope Sterilization Tray, in combination with an FDA cleared sterilization wrap, passed the sterilization validations.
The subject Endoscope Sterilization Tray (P/N 400490-04) was evaluated for the ability of the wrapped tray to maintain sterility during routine handling that would be encountered in the clinical setting. The sterilized wrapped tray was showered with an aerosol of spores and incubated. After incubation the tray was tested for positive growth. The tray successfully demonstrated negative growth. These results verify that, when wrapped and sterilized, the tray will maintain sterility after routine handling.
The subject Endoscope Sterilization Tray underwent Limits of Reuse testing. The Iimits of reuse study consisted of repeated sterilization cycles. Each tray was inspected for damage to the cover, base and latches at the beginning of the study and again after completion of all sterilization cycles. The subject tray passed all requirements as described in the protocol.
The subject Endoscope Sterilization Tray was packaged and subjected to simulated shipping testing per ASTM D4169-09 to ensure that both the product and packaging are validated for customer shipment. Packaged Endoscope Sterilization Trays were subjected to environmental conditioning, initial manual handling (drop testing), vehicle vibration, vehicle stacking (compression), etc. The packaging and the Endoscope Sterilization Tray passed all inspection criteria.
#### Conclusion:
The conclusions drawn from the comparison of the intended use, and technological characteristics and the results of the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device K180964.
Image /page/8/Picture/8 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and in a light gray color. Below that, the word "SURGICAL" is in a smaller font and also in light gray, with the registered trademark symbol next to it. There is a yellow dot above the "I" in Intuitive.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.