MicroScan Dried Gram Negative MIC/Combo Panels with Meropenem/Vaborbactam (0.03/8-64/8ug/mL)
K183127 · Beckman Coulter, Inc. · LTT · Feb 12, 2019 · Microbiology
Device Facts
Record ID
K183127
Device Name
MicroScan Dried Gram Negative MIC/Combo Panels with Meropenem/Vaborbactam (0.03/8-64/8ug/mL)
Applicant
Beckman Coulter, Inc.
Product Code
LTT · Microbiology
Decision Date
Feb 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Indications for Use
To determine antimicrobial agent susceptibility
Device Story
MicroScan Dried Gram-Negative MIC/Combo Panels are miniaturized broth dilution susceptibility tests; dehydrated antimicrobial agents are rehydrated with water after inoculation with standardized organism suspensions. Used in clinical microbiology laboratories; panels incubated in non-CO2 incubators for 16-20 hours. Minimum inhibitory concentration (MIC) determined by identifying the lowest antimicrobial concentration inhibiting growth. Results used by clinicians to guide antimicrobial therapy for gram-negative bacterial infections. Compatible with autoSCAN-4 instrument or WalkAway system for reading.
Clinical Evidence
Performance evaluated using fresh, recent, and stock efficacy isolates and stock challenge strains. Compared against CLSI frozen reference panel. Overall Essential Agreement for meropenem/vaborbactam was 98.3%. Reproducibility and precision confirmed across different inoculum methods (Turbidity or Prompt) and instruments (autoSCAN-4 or WalkAway). Quality control testing yielded acceptable results.
Technological Characteristics
Miniaturized broth dilution susceptibility test; dehydrated antimicrobial agents in panels. Rehydrated with water. Compatible with autoSCAN-4 and WalkAway automated systems. Non-CO2 incubation required. Class II device.
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K183127
B. Purpose for Submission:
To obtain a substantial equivalence determination for the addition of Meropenem/Vaborbactam to the MicroScan Dried Gram-Negative MIC/Combo Panels with Meropenem/Vaborbactam (0.03/8 – 64/8 µg/mL).
C. Measurand:
Meropenem/Vaborbactam in the dilution range of 0.03/8 – 64/8 µg/mL.
D. Type of Test:
Quantitative Antimicrobial Susceptibility Test (AST)
E. Applicant:
Beckman Coulter, Inc.
F. Proprietary and Established Names:
MicroScan Dried Gram-Negative MIC/Combo Panels with Meropenem/Vaborbactam (0.03/8 – 64/8 µg/mL).
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1640 Antimicrobial Susceptibility Test Powder
2. Classification:
Class II
3. Product codes:
LTT – Panels, Test, Susceptibility, Antimicrobial
JWY – Manual Antimicrobial Susceptibility Test Systems
LRG – Instrument for Auto Reader and Interpretation of Overnight Susceptibility
{1}
Systems
LTW – Susceptibility Test Cards, Antimicrobial
4. Panel:
83 - Microbiology
H. Intended Use:
1. Intended use(s):
For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
2. Indication(s) for use:
The MicroScan Dried Gram-Negative MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. After inoculation, panels are incubated for 16 – 20 hours at 35°C ± 1°C in a non-CO₂ incubator, and read either visually or with MicroScan instrumentation, according to the package insert.
This particular submission is for the addition of the antimicrobial meropenem/vaborbactam at concentrations of 0.03/8 to 64/8 µg/mL to the test panel.
The gram-negative organisms which may be used for meropenem/vaborbactam susceptibility testing in this panel are as follows:
Meropenem/vaborbactam has been shown to be active both clinically and in vitro against the gram negative bacteria listed below according to the FDA drug approved label:
- Enterobacter cloacae complex
- Escherichia coli
- Klebsiella pneumoniae
Meropenem/vaborbactam has been shown to be active in vitro only against the gram-negative bacteria listed below according to the FDA drug approved label:
- Citrobacter freundii
- Citrobacter koseri
- Enterobacter aerogenes
- Klebsiella oxytoca
- Morganella morganii,
{2}
- Proteus mirabilis
- Providencia spp
- Serratia marcescens
3. Special conditions for use statement(s):
For Prescription use only.
The following limitations are included in the labeling:
- "Performance of meropenem/vaborbactam when testing Morganella morganii using the Prompt Inoculation system with the autoSCAN-4 or manual read methods were outside of essential agreement compared to the reference method and should be tested using the turbidity inoculation method."
- "Elevated MICs with beta-lactam antimicrobials (e.g. aztreonam, ceftazidime/avibactam, Ceftolozane/tazobactam, meropenem/vaborbactam) may be observed if panels are over-inoculated with microorganisms such as Pseudomonas aeruginosa, Serratia spp., Proteus spp., Morganella spp., and Providencia spp. Inoculum concentration is critical with these antimicrobials as their mechanism of action involves disruption of bacterial cell wall synthesis. The user should pay careful attention to inoculum preparation, especially with manual methods that are technique dependent such as the Prompt system or inoculum prepared without the aid of a photometric device".
- "The ability of the MicroScan Dried Negative Panels to detect resistance to meropenem/vaborbactam is unknown with C. koseri, E. aerogenes, K. oxytoca, M. morganii, P. mirabilis, Providencia species and S. marcescens because resistant strains were not available at the time of comparative testing. If such isolates are observed, they should be tested on an alternate method and/or submitted to a reference lab".
- "Meropenem/vaborbactam is not active against bacteria that produce metallo-beta lactamases or oxacillinases with carbapenemase activity".
4. Special instrument requirements:
MicroScan panels can be read either manually or automatically on the WalkAway or autoSCAN-4 instrument systems.
I. Device Description:
The MicroScan Dried Gram-Negative MIC/Combo panel with meropenem/vaborbactam is used to determine the quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram negative bacilli. After inoculation, panels are incubated for 16-20 hours at $35^{\circ}\mathrm{C} \pm 1^{\circ}\mathrm{C}$ in a
{3}
non- $\mathrm{CO}_{2}$ incubator and read either visually or with MicroScan instrumentation according to the package insert.
Inoculation methods: Prompt Inoculation System (primary inoculation method) and Turbidity (alternate inoculation method).
Read methods: MicroScan WalkAway System (primary read method) and Manual, MicroScan autoSCAN-4 (alternate read methods).
# J. Substantial Equivalence Information:
1. Predicate device name(s):
MicroScan Dried Gram Negative MIC/Combo Panels - ceftazidime/avibactam
2. Predicate $510(\mathbf{k})$ number(s):
K172337
3. Comparison with predicate:
Table 1. Comparison with the Predicate Device
| Similarities | | |
| --- | --- | --- |
| Item | Device (K183127) | Predicate (K172337) |
| Device | MicroScan Dried Gram Negative MIC/Combo Panels - meropenem/vaborbactam | MicroScan Dried Gram Negative MIC/Combo Panels - ceftazidime/avibactam (K172337) |
| Technology | Overnight microdilution MIC susceptibility | Same |
| Result Reported | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | Same |
| Read Methods | Manual and automated | Same |
| Inoculation Methods | Turbidity and Prompt | Same |
| Instrumentation | WalkAway or autoSCAN-4 Instrument systems | Same |
| Specimen | Isolated colonies from cultures | Same |
| Incubation Temperature | 35°C ± 1°C | Same |
| Incubation Atmosphere | Aerobic | Same |
| Incubation Time | 16-20 hours | Same |
| Differences | | |
| Item | Device | Predicate |
| Intended Use | Determination of susceptibility to meropenem/vaborbactam with gram-negative bacteria | Determination of susceptibility to ceftazidime/avibactam with gram-negative bacteria |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device (K183127) | Predicate (K172337) |
| Antimicrobial Agent | Dried Meropenem/vaborbactam 0.03/8–64/8 μg/mL | Dried Ceftazidime/avibactam 0.25/4 – 64/4 μg/mL |
## K. Standard/Guidance Document Referenced (if applicable):
1. Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA
2. CLSI M07-11th Edition. Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically, January, 2018.
3. CLSI M100-S28. Performance Standards for Antimicrobial Susceptibility Testing; Twenty-Eight Informational Supplement, January 2018.
## L. Test Principle:
The antimicrobial susceptibility tests are dehydrated miniaturizations of the broth dilution susceptibility test. Various antimicrobial agents are diluted in Mueller Hinton broth supplemented with calcium and magnesium to concentrations spanning the range of clinical interest. Breakpoint Combo panels use concentrations equivalent to the categorical breakpoints of FDA and/or CLSI. After inoculation and rehydration with a standardized suspension of organism and incubation at 35°C for a minimum of 16 hours, the minimum inhibitory concentration (MIC) for the test organism is determined by observing the lowest antimicrobial concentration showing inhibition of growth.
## M. Performance Characteristics:
### 1. Analytical performance:
#### a. Precision/Reproducibility:
A reproducibility study was conducted at three external sites using 14 gram negative strains that were consistent with the intended use. The range of Meropenem/vaborbactam dilutions tested was 0.03/8 to 64/8 μg/mL. Isolates were tested in triplicate over three days (total of 27 replicates per strain). The isolates tested in the reproducibility study included C. freundii (one isolate), E. aerogenes (one isolate), E. cloacae (one isolate), E. coli (three isolates), K. oxytoca (one isolate), K. pneumoniae (three isolates), P. aeruginosa (two isolates), P. mirabilis (one isolate), S. marcescens (one isolate). Inocula were prepared using both the turbidity and Prompt inoculum methods and results were read manually and with the WalkAway (WA) and autoSCAN-4 (AS4) instruments.
The majority of data points were on-scale and within ± one doubling dilution
{5}
agreement as compared to the mode MIC (Table 2). One data point was off-scale using the Prompt inoculation method and the WA read method, three data points were off-scale using the Prompt inoculation method and the AS4 read method and four data points were off-scale using the Turbidity inoculation method and the AS4 reading method. As shown in Table 2, the data was analyzed taking into consideration best case and worst case scenarios as described in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems:
Table 2. Reproducibility with all Inoculation and Read Methods (All Sites).
| Read Method | | Prompt Inoculation | Turbidity Inoculation |
| --- | --- | --- | --- |
| WalkAway | Best Case | 363/378 (96.0%) | 371/378 (98.1%) |
| | Worst Case | 363/378 (96.0%) | 371/378 (98.1%) |
| autoSCAN-4 | Best Case | 362/378 (95.8%) | 371/378 (98.1%) |
| | Worst Case | 359/378 (95.0%) | 367/378 (97.1%) |
| Manual | Best Case | 366/378 (96.8%) | 367/378 (97.1%) |
| | Worst Case | 366/378 (96.8%) | 367/378 (97.1%) |
The reproducibility results were acceptable at $\geq 95\%$ .
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Inoculum Density Check. A spectrophotometric device, the MicroScan Turbidity Meter was used to ensure quality control of the turbidity inoculum method. The turbidity inocula were prepared using the MicroScan Turbidity Meter with a reading of $0.08 \pm 0.02$ (equivalent to a 0.5 McFarland barium sulfate turbidity standard). The digital reading was recorded each day of use at each clinical trial site.
Organism density data was also collected during the Reproducibility phase of the study for suspensions prepared using the Prompt inoculum preparation method. Colony counts were performed weekly using the quality control strain $E$ coli ATCC 25922. Colony counts were also done once at each site for each reproducibility isolate. All results were acceptable.
Purity Check. Purity check plates were performed on all isolates tested during the clinical study for meropenem/vaborbactam. If a purity plate indicates mixed culture, the isolate was retested.
Growth Failure Rate. During the clinical trial, there were no growth failures with both the MicroScan panel and the CLSI frozen reference method with Meropenem/vaborbactam.
Quality Control Testing. The CLSI and FDA recommended QC organisms (E. coli
{6}
ATCC 25922, P. aeruginosa ATCC 27853, E. coli ATCC35218 and K. pneumoniae ATCC 700603) were tested using the meropenem/vaborbactam dilution range of $0.03/8 - 64/8 \mu \mathrm{g} / \mathrm{mL}$ for both the device and reference panel. Testing was performed with both Turbidity and Prompt inoculation methods and all read methods (manual, WalkAway instrument and autoSCAN-4 instrument). The reference panel was inoculated using the turbidity inoculation method only. Results are outlined in Table 3.
Table 3. Quality Control Results with the Extended Dilution Range for all Sites
| Organism | Conc. (μg/mL) | Ref | Prompt Inoculation Method | | | Turbidity Inoculation Method | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Manual | WalkAway | AS4 | Manual | WalkAway | AS4 |
| E. coliATCC25922 | ≤0.03* | 121 | 102 | 121 | 119 | 120 | 119 | 120 |
| | 0.06 | | 19 | | 2 | | | |
| | 0.12 | | | | | | | |
| | 0.25 | | | | | | 1 | 1 |
| | 0.5 | | | | | 1 | | |
| | 1 | | | | | | | |
| | 2 | | | | | | | |
| | 4 | | | | | | | |
| | 8 | | | | | | | |
| | 16 | | | | | | | |
| | 32 | | | | | | | |
| | 64 | | | | | | | |
| >64 | | | | | | | | |
| | | | | | | | | |
| P. aeruginosaATCC27853 | ≤0.03 | | | | | | | |
| | 0.06 | | | | | | | |
| | 0.12 | | 2 | | | 3 | | 3 |
| | 0.25 | 53 | 79 | 75 | 80 | 79 | 69 | 73 |
| | 0.5 | 55 | 31 | 32 | 32 | 31 | 42 | 36 |
| | 1 | 13 | 6 | 10 | 6 | 8 | 10 | 9 |
| | 2 | | 2 | 2 | 2 | | | |
| | 4 | | | | | | | |
| | 8 | | | | | | | |
| | 16 | | | | | | | |
| | 32 | | | | | | | |
| | 64 | | | 1 | 1 | | | |
| >64 | | 1 | | | | | | |
| | | | | | | | | |
| E. coliATCC35218 | ≤0.03* | 118 | 117 | 116 | 117 | 118 | 117 | 118 |
| | 0.06 | | | | | | | |
| | 0.12 | | | | | | | |
| | 0.25 | | | | | | | |
| | 0.5 | | | | | | | |
| | 1 | | | | | | | |
| | 2 | | | | | | | |
| | 4 | | | | | | | |
| | 8 | | | | | | | |
| | 16 | | | | | | | |
{7}
| Organism | Conc. (μg/mL) | Ref | Prompt Inoculation Method | | | Turbidity Inoculation Method | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Manual | WalkAway | AS4 | Manual | WalkAway | AS4 |
| | 32 | | | | | | | |
| | 64 | | | | | | | |
| | >64 | | 1 | 1 | 1 | | | |
| | | | | | | | | |
| K. pneumoniae1ATCC 700603Expected range 0.016/8-0.06/8 | ≤0.03* | 112 | 11 | 10 | 14 | 2 | 1 | 1 |
| | 0.06 | 5 | 101 | 102 | 98 | 110 | 111 | 111 |
| | 0.12 | | 1 | 1 | 1 | 3 | 3 | 3 |
| | 0.25 | | 1 | 1 | 1 | 2 | 2 | 2 |
| | 0.5 | 1 | 3 | 3 | 3 | | | |
| | 1 | | | | | | | |
| | 2 | | | | | | | |
| | 4 | | | | | | | |
| | 8 | | | | | | | |
| | 16 | | 1 | 1 | 1 | | | |
| | 32 | | | | | | | |
| | 64 | | | | | | | |
| | >64 | | | | | | | |
| K. pneumoniae2ATCC BAA-17050.008/8-0.064/8 | ≤0.03* | 114 | 91 | 105 | 104 | 97 | 94 | 96 |
| | 0.06 | 3 | 25 | 12 | 12 | 14 | 17 | 15 |
| | 0.12 | | | | 1 | 4 | 4 | 4 |
| | 0.25 | | | | | 1 | 2 | 1 |
| | 0.5 | | | | | | | 1 |
| | 1 | | | | | 1 | | |
| | 2 | | | | | | | |
| | 4 | | | | | | | |
| | 8 | | | | | | | |
| | 16 | | | | | | | |
| | 32 | | | | | | | |
| | 64 | | | | | | | |
| | >64 | | | | | | | |
*The lowest end point of the MicroScan Meropenem/Vaborbactam MIC range is $\leq 0.03\mu \mathrm{g / mL}$ . Obtaining this value was considered as an indicator that the quality control test results were acceptable.
$^{1}$ E. coli ATCC 25922, E. coli ATCC 35218, and K. pneumoniae ATCC 700603 are not required for routine testing.
$^{2}$ K. pneumoniae ATCC BAA-1705 is recommended for routine QC with Meropenem/Vaborbactam. The QC range for K. pneumoniae ATCC BAA-1705 is $\leq 0.03 / 8 - 0.06 / 8 \mu \mathrm{g} / \mathrm{mL}$ (Table 3a).
Because of the meropenem/vaborbactam dilution range of $0.03/8 - 64/8 \mu \mathrm{g} / \mathrm{mL}$ , all MIC values for the QC strains tested were off-scale, except for $P_{\cdot}$ aeruginosa ATCC 27853. Therefore, only $P_{\cdot}$ aeruginosa ATCC 27853 is recommended for QC testing of meropenem/vaborbactam. In addition, $K.$ pneumoniae BAA-1705 is recommended for routine QC with meropenem/vaborbactam and with meropenem alone. The strains included in the Quality Control section of the MicroScan Dried Gram Negative Procedural Manual are shown in Table 3a.
Table 3a. Recommended QC strains for Meropenem/Vaborbactam
| Antimicrobial Agent | Organism | ATCC # | Range (μg/mL) |
| --- | --- | --- | --- |
| Meropenem/Vaborbactam | P. aeruginosa | 27853 | 0.12/8-1/8 |
| | K. pneumoniae* | BAA-1705 | ≤0.03/8-0.06/8 |
{8}
*to ensure that the plasmid encoding β-lactamase has not been lost, this strain was tested using a single β-lactam agent (meropenem) as instructed in CLSI M100, 28th ed. QC performance of meropenem on the frozen reference panel confirmed the integrity of this QC strain to reliably QC meropenem/vaborbactam at all sites.
Quality control results of all QC organisms tested met the acceptance criteria of > 95% agreement with the expected value for all inoculation and read methods and demonstrated that the system could produce the expected quality control results.
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The results obtained with the MicroScan Dried Gram-Negative MIC/Combo Panel with meropenem/vaborbactam were compared to results obtained using a frozen broth microdilution reference panel at three testing sites in the U.S.
The reference panel was prepared according to CLSI M07-A10 guidelines except for the use of Pluronic-F in the inoculum water for the reference panel. A validation study was performed to demonstrate equivalence between reference panels inoculated with organisms suspended in water supplemented with Pluronic-F and reference panels inoculated with autoclaved distilled water without Pluronic-F. The essential agreement (EA) of MIC values obtained using Pluronic-F as the diluent as compared to MIC values obtained using autoclaved distilled water as the diluent was 100% and the categorical agreement (CA) was 92.9%.
The Enterobacteriaceae clinical data was collected on a dried test panel which included twelve dilutions of meropenem/vaborbactam (0.03/8-64/8 μg/mL). The data analysis was performed by comparing test panel results to the reference results.
For each organism tested, MicroScan panels and reference panels were inoculated using the same standardized suspension further diluted into 25 mL of water with either Pluronic-D (for the MicroScan dried panels) or Pluronic-F (for the frozen reference panels). Panels were inoculated using both the Prompt System and by the Turbidity method; MicroScan panels were read using the WalkAway, the autoSCAN4 and Manual Read. The reference panels were read manually. Performance was
9
{9}
evaluated using FDA breakpoints for Meropenem/vaborbactam, and results were analyzed based on the guidelines provided in the AST Class II Special Controls Guidance Document. The primary read has been identified as the WalkAway system with the prompt inoculation system. The clinical data are also presented on the two secondary read methods collected at the three sites, which includes MicroScan and autoSCAN-4 instruments and manual reads with both Prompt and turbidity inoculation methods.
## Clinical Study:
Clinical testing was performed at three US sites. A total of 560 Enterobacteriaceae clinical isolates were tested including C. freundii (63 isolates), C. koseri (57 isolates), E. aerogenes (55 isolates), E. cloacae complex (67 isolates) E. coli (64 isolates), K. oxytoca (49 isolates), K. pneumoniae (59 isolates), M. morganii (29 isolates), P. mirabilis (32 isolates), Providencia species (51 isolates) and S. marcescens (34 isolates). Of the 560 clinical isolates, 358 (69.10%) were fresh isolates, 136 (24.3%) were recent isolates and 66 (11.8%) were stock isolates.
## Challenge Study:
Challenge testing was performed at one internal site. A total of 95 challenge isolates of Enterobacteriaceae were evaluated. The Enterobacteriaceae isolates included C. freundii (4 isolates), C. koseri (3 isolates), E. aerogenes (5 isolates), E. cloacae complex (13 isolates), E. coli (10 isolates), K. oxytoca (4), K. pneumoniae (30 isolates), M. morganii (4 isolates), P. mirabilis (5 isolates), Providencia species (11 isolates), and S. marcescens (6 isolates).
The performance of the 655 clinical and challenge isolates is summarized in Tables 3 and 4.
Table 4. Performance of MicroScan Dried Gram-Negative Panels with Meropenem/vaborbactam with Enterobacteriaceae, Prompt Inoculation Method, All Read Methods
| | Tot | No. EA | EA % | Eval EA Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R | No. S | Min (%) | Maj (%) | Vmj (%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| WalkAway | | | | | | | | | | | | | |
| Clinical | 560 | 553 | 98.8 | 145 | 138 | 95.2 | 560 | 100 | 1 | 559 | 0 (0%) | 0 (0%) | 0 (0%) |
| Challenge | 95 | 91 | 95.8 | 72 | 70 | 97.2 | 89 | 93.7 | 30 | 60 | 6 (6.3%) | 0 (0%) | 0 (0%) |
| Combined | 655 | 644 | 98.3 | 217 | 208 | 95.9 | 649 | 99.1 | 31 | 619 | 6 (0.9%) | 0 (0%) | 0 (0%) |
| | | | | | | | | | | | | | |
| autoSCAN-4 | | | | | | | | | | | | | |
| Clinical | 560 | 549 | 98 | 139 | 129 | 92.8 | 560 | 100 | 1 | 559 | 0 (0%) | 0 (0%) | 0 (0%) |
| Challenge | 95 | 88 | 92.6 | 72 | 68 | 94.4 | 89 | 93.7 | 30 | 60 | 6 (6.3%) | 0 (0%) | 0 (0%) |
| Combined | 655 | 637 | 97.3 | 211 | 197 | 93.4 | 649 | 99.1 | 31 | 619 | 6 (0.9%) | 0 (0%) | 0 (0%) |
| | | | | | | | | | | | | | |
| Manual | | | | | | | | | | | | | |
{10}
Table 5. Performance of MicroScan Dried Gram-Negative Panels with Meropenem/vaborbactam with Enterobacteriaceae, Turbidity Inoculation Method, All Read Methods
| | Tot | No. EA | EA % | Eval EA Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R | No. S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| WalkAway | | | | | | | | | | | | | |
| Clinical | 560 | 555 | 99.1 | 145 | 140 | 96.6 | 560 | 100 | 1 | 559 | 0 (0%) | 0 (0%) | 0 (0%) |
| Challenge | 95 | 88 | 92.6 | 71 | 65 | 91.5 | 85 | 89.5 | 30 | 60 | 9 (6.5%) | 0 (0%) | 1 (3.3%) |
| Combined | 655 | 643 | 98.2 | 216 | 205 | 94.9 | 645 | 98.5 | 31 | 619 | 9 (1.4%) | 0 (0%) | 1 (3.2%) |
| | | | | | | | | | | | | | |
| autoSCAN-4 | | | | | | | | | | | | | |
| Clinical | 560 | 554 | 98.9 | 138 | 132 | 95.7 | 560 | 100 | 1 | 559 | 0 (0%) | 0 (0%) | 0 (0%) |
| Challenge | 95 | 87 | 91.6 | 72 | 65 | 90.3 | 85 | 89.5 | 30 | 60 | 9 (9.5%) | 0 (0%) | 1 (3.3%) |
| Combined | 655 | 641 | 97.9 | 210 | 197 | 93.8 | 645 | 98.5 | 31 | 619 | 9 (1.4%) | 0 (0%) | 1 (3.2%) |
| | | | | | | | | | | | | | |
| Manual Read | | | | | | | | | | | | | |
| Clinical | 560 | 553 | 98.8 | 146 | 140 | 95.9 | 560 | 100 | 1 | 559 | 0 (0%) | 0 (0%) | 0 (0%) |
| Challenge | 95 | 88 | 92.6 | 72 | 66 | 91.7 | 85 | 89.5 | 30 | 60 | 9 (9.5%) | 0 (0%) | 1(3.3%) |
| Combined | 655 | 641 | 97.9 | 218 | 206 | 94.5 | 645 | 98.5 | 31 | 619 | 9 (1.4%) | 0 (0%) | 1 (3.2%) |
EA - Essential Agreement (+/- 2 dilutions)
CA - Category Agreement
EVAL - Evaluable isolates
R or NS - Resistant or non-susceptible isolates
min - minor discrepancies
maj - major discrepancies
vmj - very major discrepancies
Essential Agreement (EA) occurs when there is agreement between the result of the reference method and that of the MicroScan panel within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the MicroScan panel and the reference method. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the MicroScan panel.
## Overall Performance:
The overall essential agreement and category agreement of the MicroSCAN Meropenem/Vaborbactam for all Enterobacteriaceae is acceptable at $\geq 90.0\%$ for both inoculation methods across all read methods (Tables 4 and 5). The essential agreement of evaluable results was also acceptable with both the prompt and the turbidity inoculation methods across all read methods.
{11}
With the Prompt inoculation method across all reads, there was no very major or major discrepancies; (Table 4). The minor discrepancy rate was 0.9% (6/655).
The high very major error rate of 3.3% observed with the turbidity inoculation method across all read methods (Table 5) was attributed to Citrobacter freundii strains which had one of two challenge resistant isolates false susceptible result with an MIC of 4μg/mL with test panel and an MIC of 16μg/mL by the reference method. The following footnote was added to the Procedural Manual:
“One of the two resistant Citrobacter freundii strains had discrepant results compared to the reference method when using the turbidity inoculation method across all read methods”.
Since the number the number of resistant isolates tested during the comparative studies was low (31), 25 additional resistant isolates were evaluated during verification on all read and inoculation methods. The combined data of 56 resistant isolates had no additional very major errors. The expansion of the resistant isolates pool resulted in a very major error rate less than 2% which is acceptable.
Morganella morganii was observed to have low essential agreement with Prompt inoculation method when read on the autoSCAN-4 instrument (84.8%) and manually (72.7%), whereas categorical agreement was 100%. However, the performance of this organism was acceptable when read by the WalkAway with an essential agreement of 93.9%. The following limitation are included in the procedural manual:
“Performance of meropenem/vaborbactam when testing Morganella morganii using the Prompt Inoculation system with the autoSCAN-4 or manual read methods were outside of essential agreement compared to the reference method and should be tested using the turbidity inoculation method”.
“Elevated MICs with beta-lactam antimicrobials (e.g. aztreonam, ceftazidime/avibactam, Ceftolozane/tazobactam, meropenem/vaborbactam) may be observed if panels are over-inoculated with microorganisms such as Pseudomonas aeruginosa, Serratia spp., Proteus spp., Morganella spp., and Providencia spp. Inoculum concentration is critical with these antimicrobials as their mechanism of action involves disruption of bacterial cell wall synthesis. The user should pay careful attention to inoculum preparation, especially with manual methods that are technique dependent as the Prompt system or inoculum prepared without the aid of a photometric device”.
To address the testing of non-indicated species the following statement is included in the device labeling:
“The safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the
12
{12}
indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved".
## Resistant Organisms:
A total of 31 resistant isolates were identified out of 655 organisms tested (4.7%) in the combined challenge and clinical study of with Meropenem/vaborbactam. However, the following indicated organisms had no resistant isolates available during comparative testing: Citrobacter koseri, Enterobacter aerogenes, Klebsiella oxytoca, Morganella morganii, Proteus mirabilis, Providencia species and Serratia marcescens. This was addressed by adding the following limitation in the labeling:
"The ability of the MicroScan Dried Negative Panels to detect resistance to meropenem/vaborbactam is unknown with C. koseri, E. aerogenes, K. oxytoca, M. morganii, P. mirabilis, Providencia species and S. marcescens because resistant strains were not available at the time of comparative testing. If such isolates are observed, they should be tested on an alternate method and/or submitted to a reference lab".
## Trending:
An analysis of trending was conducted using the combined clinical and challenge data for Enterobacteriaceae and for each claimed species within the Enterobacteriaceae group with meropenem/vaborbactam 0.03/8-64/8 μg/mL for both inoculation methods (Prompt and Turbidity).
This trending calculation takes into account MIC values that are determined to be one or more doubling dilutions lower or higher compared to the reference method irrespective of whether the device MIC values for both systems are on-scale or not. A difference of ≥30% between the percentage of isolates with higher versus lower readings compared to the reference method was considered evidence of significant trending. Trending that provides higher or lower MIC values compared to the reference is addressed in labeling.
The analysis for trending takes into account only results that are evaluable for trending. Results that cannot be evaluated because it is unclear if they are at least one dilution lower, at least one dilution higher or in exact agreement with the CLSI reference method are not taken into account in the trending analysis.
The analysis for which significant high trending was observed is outlined in Tables 6 and 7 below. Table 6 demonstrates the results for Prompt inoculation method by reading method for each species and by manual reading for all Enterobacteriaceae. Table 7 demonstrates the results for Turbidity inoculation method by reading method for Providencia species.
13
{13}
Table 6. Trending Analysis Stratified by Species (Clinical and Challenge Combined) using the Prompt Inoculation Method
| Total Evaluable For Trending | ≥2 dil lower | 1 dil lower | Exact | 1 dil higher | ≥2 dil higher |
| --- | --- | --- | --- | --- | --- |
| Morganella morganii | | | | | |
| WAa | | | | | |
| 32 | 0 | 1 | 18 | 11 | 2 |
| | 1 (3.13%) | | | 13 (40.63%) | |
| Manuallyb | | | | | |
| 32 | 0 | 0 | 10 | 13 | 9 |
| | 0 (0%) | | | 22 (68.75%) | |
| Providencia spp | | | | | |
| WAc | | | | | |
| 60 | 0 | 2 | 26 | 26 | 6 |
| | 2 (3.33%) | | | 32 (53.33%) | |
| AS4d | | | | | |
| 58 | 0 | 3 | 25 | 24 | 6 |
| | 3 (5.17%) | | | 30 (51.72%) | |
| Manuale | | | | | |
| 58 | 0 | 2 | 18 | 34 | 4 |
| | 2 (3.45%) | | | 38 (65.52%) | |
| Proteus mirabilis | | | | | |
| Manualf | | | | | |
| 37 | 0 | 5 | 15 | 13 | 4 |
| | 5 (13.51%) | | | 17 (45.95%) | |
| Enterobacter cloacae complex | | | | | |
| WAg | | | | | |
| 23 | 0 | 0 | 10 | 2 | 11 |
| | 0 (0%) | | | 13 (56.52%) | |
| AS4h | | | | | |
| 22 | 0 | 0 | 10 | 10 | 2 |
| | 0 (0%) | | | 12 (54.55%) | |
| Manuali | | | | | |
| 25 | 0 | 0 | 10 | 13 | 2 |
| | 0 (0%) | | | 15 (60%) | |
| Enterobacteriaceae | | | | | |
| Manual | | | | | |
| 272 | 0 | 21 | 101 | 127 | 23 |
| | 21 (7.72%) | | | 150 (55.15%) | |
a Percent difference between the higher and lower dilution trends is: -37.50%
b Percent difference between the higher and lower dilution trends is: -68.75%
Percent difference between the higher and lower dilution trends is: -50%
d Percent difference between the higher and lower dilution trends is: -46.55%
e Percent difference between the higher and lower dilution trends is: -62.07%
f Percent difference between the higher and lower dilution trends is: -32.43%
g Percent difference between the higher and lower dilution trends is: -56.52%
h Percent difference between the higher and lower dilution trends is: -54.55%
i Percent difference between the higher and lower dilution trends is: -60%
j Percent difference between the higher and lower dilutions trends is: -47.43%
{14}
Table 7. Trending Analysis for Providencia spp (Clinical and Challenge Combined) using the Turbidity Inoculation Method
| Total Evaluable For Trending | ≥2 dil lower | 1 dil lower | Exact | 1 dil higher | ≥2 dil higher |
| --- | --- | --- | --- | --- | --- |
| Providencia spp | | | | | |
| AS4a | | | | | |
| 56 | 0 | 2 | 34 | 20 | 0 |
| | 2 (3.57%) | | | 20 (35.71%) | |
| Manualb | | | | | |
| 57 | 0 | 0 | 35 | 21 | 1 |
| | 0 (0%) | | | 22 (38.60%) | |
aPercent difference between the higher and lower dilution trends for AS4 is: -32.14%
bPercent difference between the higher and lower dilution trends for Manual is: -38.60%
Based on the above results, the following footnotes were added to the meropenem/vaborbactam Procedural Manual to address the observed trending:
"Meropenem/vaborbactam MIC values tended to be one or more doubling dilution higher when compared to the reference broth microdilution method for $P$ mirabilis., Providencia spp., Enterobacter cloacae complex and Morganella morganii with the WalkAway, autoSCAN-4 and manual reads when using the Prompt Inoculation system. Additionally, Meropenem/vaborbactam MIC values tended to be in exact agreement or at least one doubling dilution higher when testing Enterobacteriaceae when using the Prompt Inoculation system and the manual read compared to the CLSI reference broth method"
"Meropenem/vaborbactam MIC values tended to be in exact agreement or at least one doubling dilution higher when compared to the reference broth microdilution for Providencia spp. with autoSCAN-4 and manual reads when using the turbidity Inoculation method".
# Resistance Mechanisms:
The meropenem/vaborbactam approved drug label indicates activity against specific resistance mechanisms. Therefore, resistance mechanisms for indicated Enterobacteriaceae isolates were tested in the challenge phase and provided in the submission. They consisted mostly of $\beta$ etalactamases and Extended-spectrum Beta-lactamases (Table 7).
Table 8. Summary of Enterobacteriaceae Beta-Lactamases and Extended-spectrum Beta-Lactamases Resistance Mechanisms Tested
| Resistance Mechanism | Number Tested* |
| --- | --- |
| KPC | 18 |
| SME | 5 |
| TEM | 28 |
| SHV | 29 |
| CTX-M | 26 |
| ACT | 6 |
| OXA | 23 |
*Isolates tested may harbor one or more resistance mechanism
{15}
Based on the resistance mechanism information included in the approved drug label for meropenem/vaborbactam, the following limitation is added to the procedural manual:
"Meropenem/Vaborbactam is not active against bacteria that produce metallo-beta-lactamases, oxacillinases with carbapenemase activity".
b. Matrix comparison:
N/A
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data:
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Table 9. FDA Recognized Interpretive Criteria for Meropenem/Vaborbactam (μg/mL)
| Organism | Susceptible (S) | Intermediate (I) | Resistant (R) |
| --- | --- | --- | --- |
| Enterobacteriaceae | ≤ 4/8 | 8/8 | ≥ 16/8 |
N. Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.