← Product Code [NKB](/productcode/NKB) · K183117

# PathLoc-L MIS Spinal System (K183117)

_L&K BIOMED Co., Ltd. · NKB · Nov 15, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K183117

## Device Facts

- **Applicant:** L&K BIOMED Co., Ltd.
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Nov 15, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The PathLoc-L MIS Spinal System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine. The PathLoc-L MIS Spinal System can be used in an open approach and a percutaneous approach. The PathLoc-L MIS Spinal System is intended for the following indications: Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) Spondylolisthesis Trauma (i.e., fracture or dislocation) Spinal stenosis Curvatures (i.e., scoliosis, kyphosis, lordosis) Tumor Pseudarthrosis Failed previous fusion

## Device Story

PathLoc-L MIS Spinal System is a pedicle screw system for thoracic, lumbar, and sacral spine stabilization. Components include poly screws, straight/curved rods, and set screws; used in open or percutaneous surgical approaches. Device provides mechanical immobilization of spinal segments to facilitate fusion. Operated by surgeons in clinical/OR settings. System accommodates patient anatomy via various diameters and lengths. Output is physical stabilization of the spine. Benefits include surgical flexibility and spinal segment immobilization for patients with instability or deformity.

## Clinical Evidence

No clinical data. Performance is supported by mechanical bench testing per ASTM F1717 (static compression, static tension, static torsion, dynamic fatigue) and ASTM F543 (axial pullout).

## Technological Characteristics

Materials: Titanium alloy (ASTM F136). Components: Poly screws, straight/curved rods, set screws. Sterilization: Gamma sterilization. Mechanical testing: ASTM F1717 and ASTM F543.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- PathLoc-L MIS Spinal System ([K161766](/device/K161766.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

November 15, 2018

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym with the full agency name on the right. The FDA part of the logo is in blue, with the acronym in a square and the words "U.S. FOOD & DRUG ADMINISTRATION" written in a sans-serif font.

L&K Biomed Co., Ltd. Ms. Jihyeon Seo Regulatory Affairs Associate #201. #202 16-25. Dongbaekjungang-ro 16 beon-gil. Giheung-gu, Yongin-si, Gyeonggi-do, 17015 KOREA

Re: K183117

Trade/Device Name: PathLoc-L MIS Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: November 8, 2018 Received: November 9, 2018

Dear Ms. Seo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Colin O'neill -S

for MNM

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K183117

Device Name PathLoc-L MIS Spinal System

The PathLoc-L MIS Spinal System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities of the thoracic, lumbar and sacral spine. The PathLoc-L MIS Spinal System can be used in an open approach and a percutaneous approach.

The PathLoc-L MIS Spinal System is intended for the following indications:

• Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies)

- · Spondylolisthesis
- · Trauma (i.e., fracture or dislocation)
- · Spinal stenosis
- · Curvatures (i.e., scoliosis, kyphosis, lordosis)
- Tumor
- Pseudarthrosis
- · Failed previous fusion

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):

- Submitter: L&K BIOMED Co., Ltd. 1. #201, #202 16-25, Dongbaekjungang-ro 16 beon-qil Giheung-gu, Yongin-si, Gyeonggi-do, 17015, Korea Phone. 82-2-6717-1983 e-mail: kate.seo@lnkbiomed.com Contact Person: Jihyeon Seo Date prepared November 8th, 2018
#### Device Identification 2.

| Trade Name                  | PathLoc-L MIS Spinal System                                  |
|-----------------------------|--------------------------------------------------------------|
| Regulatory Class            | Class II                                                     |
| Regulation Name/Common Name | Pedicle Screw Spinal System                                  |
| Classification Name         | Thoracolumbosacral pedicle screw<br>system (21 CFR 888.3070) |
| Panel                       | Orthopedic                                                   |
| Product Code                | NKB                                                          |

#### 3. Purpose of 510(k)

L&K BIOMED Co. Ltd., is submitting this Special 510(k) to add new components and to change part number of two set screws of the existing 510(k)-cleared device (K161766)

#### The name of the legally marketed (unmodified) device 4.

■ K161766 PathLoc-L MIS Spinal System (primary predicate)

#### Description of the Device 5.

PathLoc-L MIS Spinal System consists of poly screws, straight rods, curved rods and set screw components that can be used via percutaneous surgical approach. The components are available in a variety of diameters and lengths in order to accommodate patient anatomy and are made of titanium alloy (ASTM F136).

{4}------------------------------------------------

#### Indication for Use 6.

The PathLoc-L MIS Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The PathLoc-L MIS Spinal System can be used in an open approach and a percutaneous approach.

The PathLoc-L MIS Spinal System is intended for the following indications:

- Degenerative disc disease (defined as back pain of discogenic origin with • degeneration of the disc confirmed by history and radiographic studies)
- Spondylolisthesis
- . Trauma (i.e., fracture or dislocation)
- Spinal stenosis
- Curvatures (i.e., scoliosis, kyphosis, lordosis) .
- . Tumor
- . Pseudarthrosis
- . Failed previous fusion

## 7. Comparison of the technological characteristics of the subject and predicate devices

The new components of the PathLoc-L MIS Spinal System are considered substantially equivalent to the referenced device (K161766). The systems have the same design, material, scientific technology, and indications for use. The only difference is that there are additional components which use different sterilization method (gamma sterilization) and changed catalog number of two set screws of the existing device.

#### Performance Testing 8.

Mechanical performance of additional components are the same with the legally marketed device (unmodified) PathLoc-L MIS Spinal System (K161766). Therefore, we substitute mechanical test data of additional components of PathLoc-L MIS Spinal System with the one of unmodified device (K161766).

{5}------------------------------------------------

A rationale was provided as to why the gamma sterilization does not affect the performance of the device. They are the same product in all aspect, except sterilization method.

The PathLoc-L MIS Spinal System was tested according to ASTM F1717, specifically, static compression test, static tension test, static torsion test and dynamic fatigue test, and under ASTM F543, axial pullout test.

#### Conclusion 9.

The additional components of PathLoc-L MIS Spinal System are as safe, as effective and perform as well as the predicate device. Therefore, the additional components of PathLoc-L MIS Spinal System are substantially equivalent to the predicate device (K161766).

---

**Source:** [https://fda-staging.innolitics.com/device/K183117](https://fda-staging.innolitics.com/device/K183117)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
