K183085 · Boston Scientific Corporation · EOQ · Mar 8, 2019 · Ear, Nose, Throat
Device Facts
Record ID
K183085
Device Name
CoreDx™ Pulmonary Mini-Forceps
Applicant
Boston Scientific Corporation
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Mar 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
The CoreDx Pulmonary Mini-Forceps are specifically designed to collect tissue endoscopically for histologic examination. These forceps can be used with endobronchial ultrasound endoscopes for ultrasound guided mini-forceps biopsy (MFB) of submucosal and extramural lesions of the tracheobronchial tree. These forceps should not be used for any purpose other than their intended use.
Device Story
Sterile, single-use biopsy forceps; distal radial jaws attached to flexible coil; proximal spool and thumb ring handle. Operated by physician during bronchoscopy; compatible with endobronchial ultrasound (EBUS) scopes (min 1.2 mm working channel). Device passes through scope to target lesion; jaws opened/closed via handle actuation to tear and retain tissue samples. Can be used post-EBUS-TBNA to access lymph nodes through existing needle puncture track. Enables histologic examination of submucosal/extramural lesions; facilitates diagnosis of lung periphery/lymph node pathologies.
Clinical Evidence
Bench testing only. Evaluated for biocompatibility (ISO 10993), sterilization (ISO 11135-1, ISO 10993-7), and mechanical performance (passability, pushability, working length, reliability, forceps operation/integrity, smooth edges). Ultrasound visibility confirmed via swine lung model user feedback assessment.
Technological Characteristics
Catheter-based biopsy forceps; stainless steel/polymer construction; echogenic distal section for ultrasound visibility. Actuation via proximal thumb ring handle. Working length 130cm; compatible with 1.2mm ID working channels. Sterilized via Ethylene Oxide. No software or electronic components.
Indications for Use
Indicated for endoscopic tissue collection for histologic examination in patients requiring biopsy of submucosal and extramural lesions of the tracheobronchial tree via endobronchial ultrasound endoscopes.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below.
March 8, 2019
Boston Scientific Corporation Elena Nieves Fellow, Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, Massachusetts 01752
Re: K183085
Trade/Device Name: CoreDx™ Pulmonary Mini-Forceps Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: February 5, 2019 Received: February 6, 2018
Dear Elena Nieves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183085
Device Name CoreDx Pulmonary Mini-Forceps
Indications for Use (Describe)
The CoreDx Pulmonary Mini-Forceps are specifically designed to collect tissue endoscopically for histologic examination. These forceps can be used with endoscopes for ultrasound guided mini-forcens biopsy (MFB) of submucosal and extramural lesions of the trachcobronchial tree. These for any purpose other than their intended use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.4
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
### 1. Submitter:
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752
Contact: Elena Nieves Fellow Regulatory Affairs Specialist Tel: 508-683-4347 Fax: 508-683-5939 Date Prepared: November 5, 2018
## 2. Proposed Device:
| Trade Name: | CoreDx™ Pulmonary Mini-Forceps |
|----------------------|--------------------------------------------------|
| Classification Name: | Bronchoscope (flexible or rigid) and accessories |
| Regulation Number: | 874.4680 |
| Product Code: | EOQ |
| Classification: | Class II |
### 3. Predicate Device:
| Trade Name: | Expect TM Pulmonary Endobronchial Ultrasound Transbronchial<br>Aspiration Needle |
|--------------------------|----------------------------------------------------------------------------------------------|
| Classification Name: | Bronchoscope (flexible or rigid) and accessories |
| Regulation Number: | 874.4680 |
| Product Code: | EOQ |
| Classification: | Class II |
| 510(k) Clearance Number: | K163248 |
#### 4. Reference Device:
| Trade Name: | Radial Jaw™ 4 Pulmonary Biopsy Forceps |
|--------------------------|--------------------------------------------------|
| Classification Name: | Bronchoscope (flexible or rigid) and accessories |
| Regulation Number: | 874.4680 |
| Product Code: | BWH |
| Classification: | Class II |
| 510(k) Clearance Number: | K121186 |
#### 5. Device Description:
The CoreDx Pulmonary Mini-Forceps is a sterile, single use device comprised of jaws at the distal tip, attached to a flexible coil with a spool and a thumb ring handle attached at the proximal end of the device. The two radial jaws are attached to an actuation mechanism and can be opened and closed by sliding the spool handle. The jaws are designed to tear and retain tissues within the jaws. Once the CoreDx Pulmonary Mini-Forceps is positioned at the target area, the radial jaws are opened and a sample of tissue is collected for histological examination. The CoreDx Pulmonary Mini-Forceps are designed to be compatible with scopes that have a working channel with a minimum inner diameter (ID) of 1.2 mm. The CoreDx™ Pulmonary Mini-Forceps is designed for
{4}------------------------------------------------
use in the pulmonary system. The proposed device can also be used following an EBUS Transbronchial Needle Aspiration (EBUS-TBNA) procedure to pass through the airway wall and access a lymph node via an EBUS scope.
## 6. Indications for Use:
The CoreDx Pulmonary Mini-Forceps are specifically designed to collect tissue endoscopically for histologic examination. These forceps can be used with endobronchial ultrasound endoscopes for ultrasound guided mini-forceps biopsy (MFB) of submucosal and extramural lesions of the tracheobronchial tree. These forceps should not be used for any purpose other than their intended use.
# 7. Technological Characteristics:
The proposed CoreDx Pulmonary Mini-Forceps has different technological characteristics compared to the primary predicate Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle (K163248). However, both the proposed and primary predicate device can pass through the scope to the target position. In addition, the materials of the distal section of the proposed and primary predicate are echogenic for visibility under ultrasound.
The proposed CoreDx Pulmonary Mini-Forceps has the same indications for use as the primary predicate Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle (K163248) and reference device Radial Jaw™ 4 Pulmonary Biopsy Forceps (K121186). However, the proposed device's working length is 130cm which is longer than the predicate devices in order to access the lung periphery. The proposed CoreDx will also be compatible with currently available digital and ultrasound bronchoscopes. In addition, the proposed device is designed to be compatible with scopes that have a working channel with a minimum inner diameter (ID) of 1.2 mm. Therefore, the dimensions of the proposed device better suit the airway anatomy to ensure market needs are adequality addressed, as the proposed CoreDx Pulmonary Mini-Forceps can be used in the lung periphery.
The proposed CoreDx Pulmonary Mini-Forceps and the primary predicate Expect Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle are intended to be used to obtain tissue samples from the lymph node (one sample acquired and removed in each pass) through the EBUS-TBNA needle puncture hole within and adjacent to the major lumens of the airway that can be identified and targeted using the ultrasound bronchoscope. The proposed CoreDx will follow the same track through the scope and puncture hole as that of the EBUS-TBNA needle. Both the CoreDx and the Expect Pulmonary can reach lesions that require extension beyond the bronchial wall. Extension bevond the bronchial wall may be required in order to access lymph nodes or anatomies such as the lung that a physician is interested in obtaining a biopsy sample. Accessing the lung may be intentional and is considered a standard EBUS-TBNA procedure.
When the proposed CoreDx™ Pulmonary Mini-Forceps is compared to the primary predicate Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle, both devices have technological characteristics which include catheter based
{5}------------------------------------------------
technology with actuating handles that slightly differ as the proposed includes a handle with a thumb ring while the primary predicate includes a handle with a base. Both the proposed and primary device serve the same function and are intended to be used to obtain tissue samples from the lymph node (one sample acquired and removed in each pass) through the EBUS-TBNA needle puncture hole within and adjacent to the major lumens of the airway that can be identified and targeted using the ultrasound bronchoscope. The difference is that the proposed CoreDx has jaws to obtain tissue samples while the primary predict Expect has a needle to obtain tissue samples, therefore, both devices serve the same function of taking tissue samples.
When the proposed CoreDx is compared to the reference device Radial Jaw™ 4 Pulmonary Biopsy Forceps, the technological characteristics are similar as both devices have a catheter base and the same technology of actuating the handle, opening and closing the jaw. The only technological difference is the length of the proposed device which is longer and was evaluated through bench testing (i.e. Working Length).
## 8. Performance Data:
The proposed device meets the requirements of ISO 10993 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing". Testing that was performed include Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, and USP Rabbit Pyrogen Testing; ISO 11135-1 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices", and ISO 10993-7 "Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals".
The following bench tests were performed on the CoreDx Pulmonary Mini-Forceps: Passability; Pushability; Working Length: Device Reliability; Forceps Operation; Forceps Integrity; and Smooth Edges.
In addition, the proposed device was evaluated for visibility under ultrasound after a standard EBUS-TBNA procedure in a swine lung model which was confirmed through user feedback assessment.
The non-clinical testing that was performed and submitted demonstrated that the proposed CoreDx Pulmonary Mini-Forceps is substantially equivalent to the primary predicate and reference devices as the performance of the proposed device meets the requirements of its pre-defined acceptance criteria and intended use.
## 9. Conclusion:
Boston Scientific Corporation has demonstrated that the proposed CoreDx Pulmonary Mini-Forceps is substantially equivalent to the currently cleared Expect™ Pulmonary Endobronchial Ultrasound Transbronchial Aspiration Needle (K163248) as the performance of the proposed device meets the requirements of its pre-defined acceptance criteria and intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.