# Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set (K183051)

_Cook Incorporated · LJE · Jul 16, 2019 · GU · SESK_

**Canonical URL:** https://fda-staging.innolitics.com/device/K183051

## Device Facts

- **Applicant:** Cook Incorporated
- **Product Code:** [LJE](/productcode/LJE.md)
- **Decision Date:** Jul 16, 2019
- **Decision:** SESK
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** GU
- **Attributes:** Therapeutic

## Indications for Use

The Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set is used for nephrostomy drainage, ureteral drainage, and dynamic tamponade to prevent hemorrhage in nephrolithotomy and percutaneous stone removal procedures. The set sold without the stent component has the following intended use: The Kaye Nephrostomy Tamponade Balloon Catheter Set is used for nephrostomy drainage, and dynamic tamponade to prevent hemorrhage in nephrolithotomy and percutaneous stone removal procedures.

## Device Story

Device set includes balloon catheter, ureteral stent, flexible stylet, Y-connecting tube, and syringe. Used for nephrostomy/ureteral drainage and dynamic tamponade to prevent hemorrhage during percutaneous stone removal. Operated by clinicians in clinical settings. Balloon catheter (polyethylene) inflated via syringe to 12 mm diameter to provide tamponade. Stent (polyurethane) provides drainage. Device components inserted percutaneously. Output is mechanical tamponade and fluid drainage. Benefits include hemorrhage control and maintenance of urinary tract patency post-procedure. Single-use; maximum indwell 96 hours for stent, 48 hours for balloon catheter.

## Clinical Evidence

Bench testing only. Testing included biocompatibility, sterility, packaging/stability, shelf-life, and performance testing (dimensional verification, tensile strength, balloon burst pressure, flow rate, leakage, and radiopacity).

## Technological Characteristics

Balloon catheter: polyethylene balloon, ethylene-vinyl acetate copolymer tubing. Stent: polyurethane. Stylet: polyurethane. Y-connecting tube: polyvinylchloride. Radiopaque. Single lumen. Balloon inflated via Luerlock syringe. Sterile, single-use. Dimensions: 14.0 Fr shaft, 17.0 Fr tip, 25 cm length (catheter); 5.0 Fr diameter, 75 cm length (stent).

## Predicate Devices

- Malecot-Nephrostomy Tamponade Catheter, Cook Incorporated ([K915209](/device/K915209.md))
- Expel™ Nephroureteral Stent System with Twist-Loc Hub, Boston Scientific ([K141344](/device/K141344.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, with the letters "FDA" in a blue square. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

July 16, 2019

Cook Incorporated Ian Herrman Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47404

Re: K183051

> Trade/Device Name: Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set Kaye Nephrostomy Tamponade Balloon Catheter Regulation Number: N/A Regulation Name: N/A Regulatory Class: Unclassified Product Code: LJE Dated: June 10, 2019 Received: June 11, 2019

Dear Ian Herrman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Glenn B. Bell, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

#### 510(k) Number (if known) K183051

Device Name

Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set

Kaye Nephrostomy Tamponade Balloon Catheter Set

Indications for Use (Describe)

The Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set is used for nephrostomy drainage, and dynamic tamponade to prevent hemorrhage in nephrolithotomy and percutaneous stone removal procedures.

The Kaye Nephrostomy Tamponade Balloon Catheter Set is used for nephrostomy drainage, and dynamic tamponade to prevent hemorrhage in nephrolithotomy and percutaneous stone removal procedures.

| Type of Use (Select one or both, as applicable)                                                                           |
|---------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif; font-size: 10pt;">     Prescription Use (Part 21 CFR 801 Subpart D)  </span> |
| <span style="font-family: Arial, sans-serif; font-size: 10pt;">     Over-The-Counter Use (21 CFR 801 Subpart C)  </span>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in white, bold, sans-serif font. The word "MEDICAL" is written in white, smaller, sans-serif font and is located below the word "COOK".

COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 OMINGTON, IN 47402-04 W.COOKMEDIC

510(k) Summary

## K183051 Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set 21 CFR §807.92 Date Prepared: July 16, 2019

### Submitted By:

| Submission:        | Traditional 510(k) Premarket Notification                     |
|--------------------|---------------------------------------------------------------|
| Applicant:         | Cook Incorporated                                             |
| Primary Contact:   | Ian Herrman                                                   |
| Secondary Contact: | Karthik Pillai, Ph.D.                                         |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Phone Number:      | (812) 335-3575 x104034                                        |
| Fax Number         | (812) 332-0281                                                |

### Device Information:

| Submission:                  | Traditional 510(k) Premarket Notification                                                                    |
|------------------------------|--------------------------------------------------------------------------------------------------------------|
| Trade/Device Name:           | Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set<br>Kaye Nephrostomy Tamponade Balloon Catheter Set |
| Device Common Name:          | Catheter, Nephrostomy                                                                                        |
| Regulation Number:           | N/A                                                                                                          |
| Regulation Name:             | N/A                                                                                                          |
| Classification Product Code: | LJE                                                                                                          |
| Device Class:                | Unclassified                                                                                                 |
| Review Panel:                | Gastroenterology/Urology                                                                                     |

## Predicate Devices:

- Malecot-Nephrostomy Tamponade Catheter, Cook Incorporated, K915209
- Expel™ Nephroureteral Stent System with Twist-Loc Hub, Boston Scientific, K141344 .

## Device Description:

The Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set is comprised of a balloon catheter, ureteral stent, flexible stylet, Y-connecting tube and syringe. The

{4}------------------------------------------------

balloon catheter can be sold separately with the stylet and syringe. The radiopaque balloon catheter has an outer diameter of 14.0 French and a tip diameter of 17.0 French, with a working length of 25 centimeters. The clear polyethylene balloon measures 15 centimeters in length and has an inflated diameter of 12 millimeters. The flexible stylet is 8.5 French and is manufactured from polyurethane material. The flexible stylet, designed to be inserted through the proximal end of the balloon catheter, extends 2 millimeters from the balloon catheter tip when assembled. The stent is manufactured from radiopaque polyurethane with an outer diameter of 5.0 French and a length of 75 centimeters. The stent is designed with two adjustable and removable adapters located at the proximal end of the stent. The Y-connecting tube is made of non-radiopaque polyvinylchloride material with an outer diameter of 14.0 French, and a length of 30 centimeters. A 10 mL Luerlock syringe is also provided in the set and is used to inflate the balloon through the inflation check valve.

The Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set is supplied sterile in a Tyvek peel open pouch and labeled with a three-year shelf life. The subject device has a maximum indwell time of 96 hours (4 days) for the stent and 48 hours (2 days) for the balloon catheter and is labeled as a single-use device.

# Intended Use:

MEDICA

The Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set is used for nephrostomy drainage, ureteral drainage, and dynamic tamponade to prevent hemorrhage in nephrolithotomy and percutaneous stone removal procedures.

The set sold without the stent component has the following intended use:

The Kaye Nephrostomy Tamponade Balloon Catheter Set is used for nephrostomy drainage, and dynamic tamponade to prevent hemorrhage in nephrolithotomy and percutaneous stone removal procedures.

# Comparison to Predicates:

The Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set and the predicate devices, the Malecot-Nephrostomy Tamponade Catheter (K915209) and the Expel™ Nephroureteral Stent System with Twist-Loc Hub (K141344), are substantially equivalent in that these devices share similar indications for use, methods of operations and fundamental technological characteristics. The Malecot-Nephrostomy Tamponade Catheter (K915209) is the predicate specifically for the balloon catheter component of the subject device set. The Expel™ Nephroureteral Stent System with Twist-Loc Hub

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, bold, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font. The registered trademark symbol is to the right of the word "COOK".

COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM

(K141344) is the predicate specifically for stent component of the subject device set. Comparisons between each predicate and the corresponding subject device component are presented in the following tables.

### Balloon Catheter and Stylet:

| Device<br>Characteristics  | Predicate Device<br>K915209 - Malecot-<br>Nephrostomy Tamponade<br>Catheter                                                                                                                  | Subject Device<br>K183051 - Kaye<br>Nephrostomy<br>Tamponade Balloon<br>Catheter                                                                           | Comparison                                                                                                                                                                                                                                                                                                                                                      |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use        | The Malecot-Nephrostomy<br>Tamponade Catheter is used<br>for nephrostomy drainage<br>and low pressure dynamic<br>tamponade to prevent<br>hemorrhage following<br>percutaneous stone removal. | Used for nephrostomy<br>drainage, and dynamic<br>tamponade to prevent<br>hemorrhage in<br>nephrolithotomy and<br>percutaneous stone<br>removal procedures. | The subject and the predicate<br>device are both used for<br>nephrostomy drainage, and<br>dynamic tamponade to prevent<br>hemorrhage following percutaneous<br>stone removal. The slight difference<br>in wording does not raise new<br>questions of safety and<br>effectiveness (S&E) as the balloons<br>are labeled with the same max<br>volume and pressure. |
| Placement Method           | Percutaneous                                                                                                                                                                                 | Identical                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                       |
| Maximum Indwell<br>Time    | 5 days                                                                                                                                                                                       | 2 days (48 hours)                                                                                                                                          | The subject device has a shorter<br>maximum indwell time than the<br>predicate. This shorter indwell time<br>does not raise any new questions of<br>S&E.                                                                                                                                                                                                        |
| General Design             | Balloon Catheter with stylet<br>and syringe accessories                                                                                                                                      | Identical                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                       |
| Materials                  | Balloon: Polyethylene<br>Catheter Tubing:<br>Polyethylene, non-<br>radiopaque tubing                                                                                                         | Balloon: Polyethylene<br>Catheter Tubing:<br>Ethylene-vinyl acetate<br>copolymer, radiopaque                                                               | Material differences in these<br>catheters do not raise new questions<br>of S&E as demonstrated by<br>biocompatibility and performance<br>testing.                                                                                                                                                                                                              |
| Balloon Length             | 15 cm                                                                                                                                                                                        | Identical                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                       |
| Balloon Diameter           | 12 mm                                                                                                                                                                                        | Identical                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                       |
| Max inflation<br>pressure  | 40 psi                                                                                                                                                                                       | Identical                                                                                                                                                  | Identical                                                                                                                                                                                                                                                                                                                                                       |
| Catheter Length            | 25.7 cm                                                                                                                                                                                      | 25 cm                                                                                                                                                      | The slight difference in length does<br>not raise questions of S&E.                                                                                                                                                                                                                                                                                             |
| Catheter Tip               | 16 Fr malecot tip                                                                                                                                                                            | 17 Fr dual sideport tip                                                                                                                                    | The difference in tip diameter and<br>drainage opening type does not<br>raise different questions of S&E                                                                                                                                                                                                                                                        |
| Catheter Shaft<br>Diameter | 16 Fr                                                                                                                                                                                        | 14 Fr                                                                                                                                                      | The subject device catheter shaft<br>has a smaller outer diameter than<br>the predicate device. Performance<br>testing indicates the difference in<br>outer diameter raises no new<br>questions of S&E.                                                                                                                                                         |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the contact information for Cook Incorporated. The address is 750 Daniels Way, P.O. Box 489, Bloomington, IN 47402-0489 U.S.A. The phone number is 812.339.2235, and the toll-free number is 800.457.4500. The website is www.cookmedical.com.

The total length of the predicate stent is 65 cm with a working length

subject device is 75cm with a

of 22-28 cm. The total length of the

Regarding technological characteristics, the subject and predicate balloon catheter (K141344) devices have similarities in their general designs. However, differences do exist as described in the table above (e.g., tip drainage design, dimensions, materials, etc.) The differences identified do not raise different questions of safety and/or effectiveness as compared to the predicate device, as stated in the table.

| Device<br>Characteristics | Predicate Device<br>K141344 - ExpelTM<br>Nephroureteral Stent<br>System with Twist-Loc<br>Hub                                                                                                                                                                                     | Subject Device<br>K183051 - Kaye<br>Nephrostomy<br>Tamponade Balloon<br>Catheter Stent                                                                                           | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use       | The Expel Nephroureteral<br>Stents are delivered<br>percutaneously and are<br>intended to establish<br>internal drainage from the<br>ureteropelvic junction to<br>the bladder while<br>maintaining external<br>access to the stent, as well<br>as providing external<br>drainage. | Used for nephrostomy<br>drainage, ureteral<br>drainage, and dynamic<br>tamponade to prevent<br>hemorrhage in<br>nephrolithotomy and<br>percutaneous stone<br>removal procedures. | The subject device has a different<br>indication for use statement<br>compared to predicate K141344.<br>However, the stent included in the<br>subject device has the same<br>intended use as the predicate<br>K141344, which is internal drainage<br>to the bladder and external drainage<br>from the ureteropelvic junction<br>while maintaining external access to<br>the stent.                                                                                           |
| Maximum Indwell<br>Time   | 30 days                                                                                                                                                                                                                                                                           | 4 days (96 hours)                                                                                                                                                                | The subject device stent component<br>has a 4 day maximum indwell time<br>which is shorter than that of the<br>predicate. Because the subject<br>device maximum indwell time falls<br>within that of the predicate device<br>no new questions of S&E are raised.                                                                                                                                                                                                             |
| General Design            | Single lumen<br>nephroureteral stent with<br>distal and proximal<br>retention pigtails with<br>sideports                                                                                                                                                                          | Single lumen straight<br>nephroureteral stent with<br>sideports.                                                                                                                 | The predicate and subject device are<br>similar in general design. The main<br>difference in design is that the<br>predicate device has pigtails in the<br>bladder and ureteropelvic junction<br>for retention. In the subject, device<br>retention is assured either through<br>connection to the balloon catheter<br>through the proximal adapter and/or<br>anchoring the device to the patient.<br>Thus, the difference in design does<br>not raise new questions of S&E. |
| Materials                 | Radiopaque Flexithane                                                                                                                                                                                                                                                             | Radiopaque Polyurethane                                                                                                                                                          | Material differences between the<br>predicate and subject devices do not<br>raise new questions of S&E as<br>demonstrated by biocompatibility<br>and performance testing.                                                                                                                                                                                                                                                                                                    |

Total: 75

(adjustable)

Working length: 33 cm

### Stent:

Stent Length

Total: 65

Working length: 22-28 cm

Image /page/6/Picture/4 description: The image shows the logo for Cook Medical. The word "COOK" is in white, bold, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, bold, sans-serif font on a red background.

{7}------------------------------------------------

COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 OMINGTON. IN 47402-0489 U.S.A. 2235 TOLL EREE: 800-457.4500 VWW.COOKMEDICAL.COM

Image /page/7/Picture/1 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, block letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, block letters on a red background. The red background is shaped like a rectangle with a diagonal cut on the bottom left corner.

| Device<br>Characteristics | Predicate Device<br>K141344 - Expel™<br>Nephroureteral Stent<br>System with Twist-Loc<br>Hub | Subject Device<br>K183051 - Kaye<br>Nephrostomy<br>Tamponade Balloon<br>Catheter Stent | Comparison                                                                                                                                                                                                                                                                                                                                           |
|---------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                           |                                                                                              |                                                                                        | working length adjustable up to 33<br>cm. The total length and working<br>length of the subject device stent<br>encompasses the predicate stent and<br>does not generate any new questions<br>of S&E.                                                                                                                                                |
| Stent Diameter            | 8.3 and 10.3 Fr                                                                              | 5 Fr                                                                                   | The subject device stent has a<br>smaller outer diameter compared to<br>the predicate device stent. The<br>smaller stent outer diameter ensures<br>that the predicate can be used in the<br>same patient anatomy as the<br>predicate. Performance testing has<br>demonstrated that this difference<br>does not generate any new questions<br>of S&E. |

As shown above, the indications for use statement of the subject device set including the stent component is not identical to the predicate device; however, the differences do not represent a new intended use as both the predicate device stent and subject device stent have the same intended use.

Regarding the technological characteristics, the subject and predicate devices have similarities in their general designs. However, differences do exist as described in the table above (e.g., dimensions, materials, indwell time, etc.) The differences identified do not raise different questions of safety and effectiveness as compared to the predicate device as stated in the table.

## Technological Characteristics:

The following tests were performed to demonstrate that The Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set met applicable design and performance requirements and support a determination of substantial equivalence:

- 1. Biocompatibility
- 2. Sterility
- 3. Packaging: Distribution and Stability
- 4. Shelf-life
- 5. Balloon Catheter:
	- a. Compatibility

{8}------------------------------------------------

COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. 9.2235 TOLLEREE: 800-457.4500 WWW.COOKMEDICAL.COM

Image /page/8/Picture/1 description: The image shows the Cook Medical logo. The word "COOK" is in white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font.

- b. Dimensional Verification
- Balloon Inflation Volume C.
- d. Balloon Catheter Rated Burst Pressure
- e. Tensile Test
- Radiopacity f.
- 6. Stent:
	- Compatibility a.
	- b. Dimensional Verification
	- c. Tensile Test
	- d. Flow Rate Test
	- Kink and Lumen Patency e.
	- Radiopacity f.
- 7. Stylet:
	- a. Compatibility
	- b. Dimensional Verification
	- c. Tensile Test
- 8. Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set:
	- a. Liquid Leakage and Gravity Flow

# Conclusion:

The results of these tests provide reasonable assurance that the Kaye Nephrostomy Tamponade Balloon Catheter and Stent Set will perform as intended. The subject device does not raise different questions of safety and/or effectiveness as compared to the predicate devices. In conclusion, the results of these tests support a determination of substantial equivalence to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K183051](https://fda-staging.innolitics.com/device/K183051)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).