Full Digital Colour Doppler Ultrasonic Diagnostic System
Applicant
Wuhan Zoncare Bio-medical Electronics Co., Ltd.
Product Code
IYN · Radiology
Decision Date
Jul 10, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Indications for Use
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Device Story
The ZONCARE-M5 is a portable, battery-powered diagnostic ultrasound system. It uses a piezoelectric transducer array to convert electronic voltage pulses into ultrasonic pressure waves, which are transmitted into body tissues. Reflected echoes are detected by the transducer, converted back into electrical signals, and processed by the system to generate diagnostic images or perform Doppler flow analysis. The device includes analysis packages for measuring anatomical structures (e.g., distance, volume, stenosis, velocity, heart rate). It is operated by qualified clinicians in clinical settings to assist in diagnosis. The system features a 12.1-inch display and physical user interface buttons. It provides real-time visualization and quantitative data to support clinical decision-making.
Clinical Evidence
Bench testing only. No clinical data was required or provided. Substantial equivalence was demonstrated through non-clinical verification and validation, including electrical safety, electromagnetic compatibility, biocompatibility (ISO 10993), and acoustic output testing per FDA guidance.
Technological Characteristics
Portable laptop-style ultrasound system. Piezoelectric transducer array. Modes: B-Mode, M-Mode, Color Doppler, PDI, PW, CW. Connectivity: 100-240V power supply, rechargeable battery. Biocompatibility: ISO 10993-1, -5, -10. Safety standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD2, NEMA UD3.
Indications for Use
Indicated for use by qualified physicians or allied health professionals for ultrasound evaluations in abdominal, gynecological (including endovaginal), obstetric, cardiac, small parts (breast, testes, thyroid), urological, musculoskeletal, and peripheral vascular applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Wuhan Zoncare Bio-medical Electronics Co., Ltd. % Long Yang, COO Shenzhen Hlongmed Biotech Co., Ltd. 1002, 10th Floor, Zhongxing Administrative Building Zhongxing Industrial Zone, Chuangye Road, Nanshan Shenzhen, Guangdong 518054 CHINA
July 10, 2019
### Re: K183041
Trade/Device Name: Full Digital Colour Doppler Ultrasonic Diagnostic System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: May 31, 2019 Received: June 5, 2019
### Dear Long Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Section 10
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K183041
Device Name
#### Full Digital Colour Doppler Ultrasound Diagnostic System
#### Indications for Use (Describe)
The ZONCARE-M5 Ultrasound system is intended for use by a qualified physician or allied health professional for ultrasound evaluations. Specific clinical applications include:
Abdominal Gynecology(including endovaginal) Obstetric Cardiac Small parts(Breast,Testes,Thyroid,etc.) Urology Musculoskeletal Peripheral vascular
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Full Digital Color Doppler Ultrasonic Diagnostic System
| Clinical Application | | Mode of Operation | | | | | | | |
|-----------------------------|------------------------------------|----------------------|---|-----|-----|------------------|-----------------------|--------------------|--|
| General<br>(Track1 Only) | Specific<br>(Track 1&3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal/Obstetrics | N | N | | | N | N | | |
| | Abdominal | N | N | N | N | N | N | | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro<br>logical) | | | | | | | | |
| Fetal<br>Imaging<br>& other | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ(Specify)* | N | N | | | N | N | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | N | N | | | N | N | | |
| | Trans-vaginal | N | N | | | N | N | | |
| | Trans-urethral | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | N | N | | | N | N | | |
| | Musculo-skeletal<br>(Superficial) | N | N | | | N | N | | |
| | Intravascular | | | | | | | | |
| | Other(Specify)** | N | N | | | N | N | | |
| | Adult Cardiac | N | N | N | N | N | N | | |
| | Pediatric Cardiac | | | | | | | | |
| Cardiac | Intravascular(cardiac) | | | | | | | | |
| | Trans-esoph.(cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| Peripheral<br>vascular | Peripheral vascular | N | N | | | N | N | | |
| | Other(Specify) | | | | | | | | |
| Clinical Application | | Mode of<br>Operation | | | | | | | |
| General | Specific | B | M | PWD | CWD | Color | Combined | Other | |
| (Track1 Only) | (Track 1&3) | | | | | Doppler | (Specify) | (Specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal/Obstetrics | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro | | | | | | | | |
| | logical) | | | | | | | | |
| Fetal | Laparoscopic | | | | | | | | |
| Imaging | Pediatric | | | | | | | | |
| & other | Small Organ(Specify)* | N | N | N | N | N | N | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Musculo-skeletal | N | N | N | N | N | N | | |
| | (Conventional) | | | | | | | | |
| | Musculo-skeletal | N | N | N | N | N | N | | |
| | (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other(Specify)** | | | | | | | | |
| | Adult Cardiac | | | | | | | | |
| | Pediatric Cardiac | | | | | | | | |
| Cardiac | Intravascular(cardiac) | | | | | | | | |
| | Trans-esoph.(cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| Peripheral | Peripheral vascular | N | N | N | N | N | N | | |
| vascular | Other(Specify) | | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | | |
| General<br>(Track1 Only) | Specific<br>(Track 1&3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal/Obstetrics | N | N | | | N | N | | |
| | Abdominal | | | | | | | | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neurological) | | | | | | | | |
| Fetal<br>Imaging<br>& other | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ(Specify)* | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | N | N | | | N | N | | |
| | Trans-vaginal | N | N | | | N | N | | |
| | Trans-urethral | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other(Specify)** | | | | | | | | |
| | Adult Cardiac | | | | | | | | |
| | Pediatric Cardiac | | | | | | | | |
| Cardiac | Intravascular(cardiac) | | | | | | | | |
| | Trans-esoph.(cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| Peripheral<br>vascular | Peripheral vascular | | | | | | | | |
| | Other(Specify) | | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | | |
| General<br>(Track1 Only) | Specific<br>(Track 1&3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal/Obstetrics | N | N | | N | N | N | | |
| | Abdominal | N | N | | N | N | N | | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| Fetal<br>Imaging<br>& other | Pediatric | | | | | | | | |
| | Small Organ(Specify)* | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other(Specify)** | N | N | | N | N | N | | |
| | Adult Cardiac | | | | | | | | |
| | Pediatric Cardiac | | | | | | | | |
| Cardiac | Intravascular(cardiac) | | | | | | | | |
| | Trans-esoph.(cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| Peripheral<br>vascular | Peripheral vascular | | | | | | | | |
| | Other(Specify) | | | | | | | | |
Intended use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Additional comments: Combined mode: B+M
Note: * Small organ includes Thyroid, Testes, Breast
{4}------------------------------------------------
#### ZONCARE-M5 with TL40 Transducer
Intended use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M
Note:* Small organ includes Thyroid, Testes, Breast
{5}------------------------------------------------
#### ZONCARE-M5 with TC10 Transducer
Intended use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Additional comments: Combined mode: B+M
Note:* Small organ includes Thyroid, Testes, Breast
** Other use includes Urology, Gynecology of the county of the comment of the
{6}------------------------------------------------
#### ZONCARE-M5 with TC50 Transducer
Intended use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Additional comments: Combined mode: B+M
Note:* Small organ includes Thyroid, Testes, Breast
{7}------------------------------------------------
### ZONCARE-M5 with TP16 Transducer
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track1 Only) | Specific<br>(Track 1&3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& other | Fetal/Obstetrics | | | | | | | |
| | Abdominal | N | N | N | N | N | N | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ(Specify)* | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other(Specify)** | | | | | | | |
| Cardiac | Adult Cardiac | N | N | N | N | N | N | |
| | Pediatric Cardiac | | | | | | | |
| | Intravascular(cardiac) | | | | | | | |
| | Trans-esoph.(cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| Peripheral<br>vascular | Peripheral vascular | | | | | | | |
| | Other(Specify) | | | | | | | |
Intended use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Additional comments: Combined mode: B+M
Note:* Small organ includes Thyroid, Testes, Breast
{8}------------------------------------------------
# ZONCarE中旗
Section 9
# 510(k) Summary
(as required by 807.92(c))
The assigned 510(K) number is:_ K183041
Date of Summary:
### 1. Submitter information
Manufacturer Name: Wuhan Zoncare Bio-medical Electronics Co., Ltd
Address:Zoncare Building,#380,High-tech 2ND Road,Eastlake high-tech district ,Wuhan,
Hubei , China
Contact Person and Title: Chenglin Tian/Manager of Medical Regulation
Tel: (86)-27-86637765
Fax: (86)-27-87174399
Email: fda ce@zoncae.cn
### 2. Contact person
### 2.1 Primary Contact Person
Long Yang (COO)
Shenzhen Hlongmed Biotech Co.,Ltd
1002, 10th Floor, Zhongxing Administrative Building, Zhongxing Industrial Zone,
Chuangye Road, Nanshan, Shenzhen, P.R.China
Tel: 0086-755-86664986
Fax: 0086-755-86664933
E-mail: yanglong@hlongmed.com
### 2.2 Secondary Contact Person
{9}------------------------------------------------
# ZONCarE中旗
Chenglin Tian/Manager of Medical Regulation
Wuhan Zoncare Bio-medical Electronics Co., Ltd
Zoncare Building, #380, High-tech 2ND Road, Eastlake high-tech district ,Wuhan,
Hubei , China
Tel: (86)-27-86637765
#### 3. Device information
Device name: Full Digital Colour Doppler Ultrasonic Diagnostic System
Model:ZONCARE-M5
Common Name:Diagnostic Ultrasound System with Accessories
#### Classification Name and Product Code:
21 CFR 892.1550 System, Imaging, Pulsed Doppler, Ultrasonic
Product code: IYN
21 CFR 892.1560 Ultrasonic, Pulsed echo, Imaging
Product code: IYO
21 CFR 892.1570 Transducer, Ultrasonic, Diagnostic
Product code: ITX
Regulatory Class: class II
#### 4. Predicate device information
Manufacturer:Edan Instruments, Inc.
Address:3/F-B, Nanshan Medical Equipments Park, Nanhai Rd 1019#, Shekou,
Nanshan Shenzhen, 518067 P.R. China
Device name:Acclarix Diagnostic Ultrasound System
510(k)number:k150999
#### 5. Indications for Use
The ZONCARE-M5 Ultrasound system is intended for use by a qualified physician or
{10}------------------------------------------------
# ZONCarE中旗
allied health professional for ultrasound evaluations. Specific clinical applications include:
- Abdominal Gynecology (including endovaginal) Obstetric Cardiac Small parts(Breast, Testes, Thyroid, etc.) Urology Musculoskeletal Peripheral vascular
#### 6. Device Description
The ZONCARE-M5 Ultrasound system consists of a main system along with associated transducers.
The system circuitry generates an electronic voltage pulse, which is transmitted to the transducer. In the transducer, a piezo electric array converts the electronic pulse into an ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body tissues. The waves are then reflected within the body and detected by the transducer, which then converts back to an electrical signal. The ZONCARE-M5 system then analyzes the returned signal to generate an image or conduct Doppler processing.
The ZONCARE-M5 system gives the operator the ability to measure anatomical structures, and offers analysis packages that provide information used by competent health care professionals to make a diagnosis.
The system provides hardware buttons for the User Interface.
#### 7. Comparison to Predicate Devices
Wuhan Zoncare Bio-medical Electronics Co., Ltd believes the ZONCARE-M5 Ultrasound System described in this submission is substantially equivalent to the
{11}------------------------------------------------
predicate device as follow:
Acclarix AX8 Diagnostic Ultrasound System(k150999)
The following table shows similarities and differences between our device and the
predicate devices.
| Item | Proposed Device | Predicate Device |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | Full Digital Colour Doppler<br>Ultrasound Diagnostic System | Acclarix Diagnostic Ultrasound<br>System |
| Model | ZONCARE-M5 | AX8 |
| 510k<br>submitter | Wuhan Zoncare Bio-medical<br>Electronics Co.,Ltd. | Edan Instruments,Inc |
| 510K Number | / | K150999 |
| Intended use | Diagnostic ultrasound imaging or<br>fluid flow analysis of the human<br>body | Diagnostic ultrasound imaging or<br>fluid flow analysis of the human<br>body |
| Indication for<br>use | The ZONCARE-M5 Ultrasound<br>system is intended for use by a<br>qualified physician or allied health<br>professional for ultrasound<br>evaluations.<br>Specific clinical applications<br>include:<br>Abdominal<br>Gynecology (including endovaginal)<br>Obstetric<br>Cardiac<br>Small parts(Breast, Testes,Thyroid,<br>etc.) | The Edan Acclarix AX8 Ultrasound<br>system is intended for use by a<br>qualified physician or allied health<br>professional for ultrasound<br>evaluations.<br>Specific clinical applications<br>include:<br>Abdominal<br>Gynecology(including endovaginal)<br>Obstetric<br>Cardiac<br>Small parts (Breast, Testes,Thyroid,<br>etc.) |
| | Urology | Urology |
| | Musculoskeletal | Musculoskeletal |
| | Peripheral vascular | Peripheral vascular<br>Intra-operative. |
| Installation<br>and use | Portable(laptop)Mobile Equipment | Portable(laptop)Mobile Equipment |
| | IEC 60601-1, | IEC 60601-1, |
| | IEC 60601-1-2, | IEC 60601-1-2, |
| Safety<br>standards | IEC 60601-2-37, | IEC 60601-2-37, |
| | ISO 10993-1,-5,-10,-12 | ISO 10993-1,-5,-10,-12 |
| | NEMA UD2 | AIUM,NEMA<br>UD2,UD3 |
| | NEMA UD3 | |
| Patient contact<br>materials | Complies with ISO 10993 | Complies with ISO 10993 |
| General<br>Imaging mode | B-Mode,M-Mode,Color,PDI,PW,CW | B-Mode,M-Mode,Color,PDI/DPDI,PW,CW |
| | B-Mode:Distance,Circ/Area,Angle,<br>Volume,Stenosis | B-Mode:Distance,Circ/Area,Angle,<br>Volume,Stenosis |
| Measurements | M-Mode:Distance, Time,Slope and<br>Heart Rate | M-Mode:Distance, Time,Slope and<br>Heart Rate |
| | D-Mode:Velocity, RI,Time, PI, Heart<br>Rate,Auto Trace PG,S/D,<br>△V,Acceleration,PHT, VTI | D-Mode:Velocity,RI,Time, PI, Heart<br>Rate, Auto Trace PG,S/D,<br>△V,Acceleration,PHT, VTI |
| Principle of<br>Operation | Applying high voltage burst to the<br>Piezoelectric material in the | Applying high voltage burst to the<br>Piezoelectric material in the |
| | transducer and detect reflected echo | transducer and detect reflected echo |
| | to construct diagnostic image | to construct diagnostic image |
| | Track 3: MI, TIS, TIC,TIB (TI | Track 3: MI, TIS, TIC,TIB (TI |
| | Range 0-6.0) | Range 0-6.0) |
| Acoustic<br>output | Derated ISPTA: 720W/cm2<br>maximum,Mechanic Index ≤1.9 | Derated ISPTA: 720W/cm2<br>maximum,Mechanic Index≤1.9 |
| | maximum or Derated ISPPA 190<br>W/cm² max | maximum or Derated ISPPA 190<br>W/cm² max |
| | Convex Array | Convex Array |
| Transducer<br>Types | Linear Array | Linear Array |
| | Phased Array | Phased Array |
| | | Micro Convex Array |
| Transducer<br>Frequency | 2.0-12.0MHz | 2.5-15.0MHz |
| Display | Primary Screen:12.1inch(1024*768) | Primary Screen:15 inch(1920x1080) |
| | 370mm(W)*470mm(L)*380mm(H) | 407mm(W)*388mm(L)*77mm(H) |
| Dimensions/<br>Weight | Weight:nearly 9kg (with<br>Rechargeable battery, without<br>power adaptor or transducers) | Weight: ≤9.1kg(with<br>Rechargeable battery, without<br>power adaptor or transducers) |
| Power<br>Supply | 100-240V<br>50/60Hz | 100-240V<br>50/60Hz |
| Rechargeable<br>battery | Yes | Yes |
{12}------------------------------------------------
{13}------------------------------------------------
The subject device has same intended use, similar product design, same performance effectiveness, and performance safety as the predicate device.
The differences between the subject device and predicate device do not affect the basic design principle, usage, effectiveness and safety of the subject device. And no question is
{14}------------------------------------------------
raised regarding to effectiveness and safety.
### 8. Effectiveness and Safety Considerations
### Clinical test:
Clinical testing is not required
#### Non-clinical test:
The ZONCARE-M5 Ultrasound System complies with
(1)AAMI/ANSI ES60601-1 Electrical Safety
(2) IEC 60601-1-2 Electromagnetic Compatibility
(3) Acoustic output testing as per the guideline "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" dated September 9, 2008.
The following biocompatibility standards are conducted on the subject device:
(1) ISO 10993-1, ISO 10993-5 and ISO 10993-10
The tests were selected to show substantial equivalence between the subject device and the predicate.
### 9. Substantial Equivalence Conclusion
Verification and validation testing has been conducted on the ZONCARE-M5 Ultrasound System.This premarket notification submission demonstrates that ZONCARE-M5 Ultrasound System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.