Endosee System

K183020 · CooperSurgical, Inc. · HIH · Jan 31, 2019 · Obstetrics/Gynecology

Device Facts

Record IDK183020
Device NameEndosee System
ApplicantCooperSurgical, Inc.
Product CodeHIH · Obstetrics/Gynecology
Decision DateJan 31, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1690
Device ClassClass 2

Indications for Use

The Endosee System is used to permit viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include: Abnormal bleeding Infertility and pregnancy wastage Evaluation of abnormal hysterosalpingogram Intrauterine foreign body Amenorrhea Pelvic pain The Endosee System can also be used to permit viewing of the adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of performing diagnostic procedures.

Device Story

Handheld, battery-operated portable endoscope; consists of reusable LCD touchscreen display module and sterile, single-use flexible cannula. Cannula contains distal LED light source and camera; captures images/video of diagnostic area. Signals transferred via electrical connector/cable to display module for real-time viewing. Used in OR, outpatient, or office settings by clinicians. Includes docking station for battery recharging and image/video download to external computer via USB. Wi-Fi capability included. Provides direct visualization to assist clinicians in diagnostic decision-making; minimally invasive approach via natural orifices benefits patients by enabling office-based diagnostics.

Clinical Evidence

Bench testing only. Testing included sterility (ISO 11135-1), biocompatibility (ISO 10993), shelf-life (ASTM F1980-16), ship testing (ISTA 3A), and functional testing (image quality, optical distortion, flow rate, leak, bend, LED temperature, drop test, IPX2 ingress). Electrical safety and EMC testing performed per ANSI/AAMI ES 60601-1 and IEC 60601-1-2. Software verification conducted per IEC 62304.

Technological Characteristics

Handheld, battery-operated endoscope. Cannula materials: PEEK, ink, glass, adhesive, lens coating, surface black coating. Light source: LED. Field of view: 100° ± 5°; Direction of view: 20° ± 3°. Connectivity: Wi-Fi, USB. Sterilization: Ethylene oxide (SAL 10^-6). Software: IEC 62304 compliant. Dimensions: 276mm working length, 4.5mm outer diameter.

Indications for Use

Indicated for adult patients requiring diagnostic visualization of the cervical canal, uterine cavity, or urinary bladder. Used for conditions including abnormal bleeding, infertility, pregnancy wastage, abnormal hysterosalpingogram, intrauterine foreign bodies, amenorrhea, and pelvic pain.

Regulatory Classification

Identification

A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. January 31, 2019 CooperSurgical, Inc. Roaida Johnson Director. RA New Product Development 95 Corporate Drive Trumbull, Connecticut 06611 Re: K183020 Trade/Device Name: Endosee® System Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope And Accessories Regulatory Class: Class II Product Code: HIH, FAJ Dated: October 31, 2018 Received: November 1, 2018 Dear Roaida Johnson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Sharon M. Andrews -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K183020 Device Name Endosee® System Indications for Use (Describe) The Endosee System is used to permit viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include: - Abnormal bleeding - Infertility and pregnancy wastage - · Evaluation of abnormal hysterosalpingogram - · Intrauterine foreign body - · Amenorrhea - · Pelvic pain The Endosee System can also be used to permit viewing of the adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of performing diagnostic procedures. Type of Use (Select one or both, as applicable) | <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| |----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K183020 Page 1/5 # oper Surgical # 510(k) Summary K183020 #### Submitter Information | Company Name: | CooperSurgical Inc. | |------------------|------------------------------------------------------------------------------------------| | Company Address: | 95 Corporate Drive<br>Trumbull, CT 06611<br>Telephone: 203-601-5200<br>Fax: 203-601-9870 | | Contact Person: | Roaida Johnson | |-----------------|------------------| | Date Prepared: | January 28, 2019 | #### Device Information | Trade Name: | Endosee® System | |--------------------|------------------------------| | Common Name: | Hysteroscope, Cystoscope | | Regulation Number: | 21 CFR 884.1690 | | Regulation Name: | Hysteroscope and Accessories | | Product Code: | HIH; FAJ | | Regulatory Class: | II | #### Predicate Device Information The subject Endosee System is substantially equivalent to the predicate Endosee Hysteroscope with Disposable Diagnostic (Dx) Cannula (K170660). The predicate device has not been subject to a design-related recall. ## Device Description The Endosee System is a handheld, battery-operated, portable endoscope that consists of a reusable display module with an LCD touchscreen monitor and a sterile, single-use cannula. It is intended for use in viewing the adult cervical canal, uterine cavity, or female urinary tract, including the bladder, to perform diagnostic hysteroscopy or cystoscopy procedures in an operating room, outpatient, or office setting. The cannula has a light source and camera at the distal end, which are used for visualization and to capture image and video of the diagnostic area. The image and video signals are {4}------------------------------------------------ electronically transferred from the cannula to the display module via an electrical connector and cable, and the LCD monitor is used for viewing. The Endosee is provided with a docking station that recharges the battery in the display module and allows the user to download images/video from the internal device memory to an external computer. #### Indications for Use The Endosee System is used to permit viewing of the adult cervical canal and uterine cavity for the purpose of performing diagnostic procedures. Generally recognized indications for diagnostic hysteroscopy include: - Abnormal bleeding - Infertility and pregnancy wastage - Evaluation of abnormal hysterosalpingogram - Intrauterine foreign body - Amenorrhea - Pelvic pain The Endosee System can also be used to permit viewing of the adult urinary bladder through a minimally invasive approach by utilizing natural orifices for the purpose of performing diagnostic procedures. ## Substantial Equivalence Discussion Table 1 provides a comparison of the subject and predicate devices. | Attribute | Subject Endosee System | Predicate Endosee Hysteroscope | Significant Differences | |--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------| | 510(k)<br>Number | Not yet assigned | K170660 | N/A | | Manufacturer | CooperSurgical, Inc. | CooperSurgical, Inc. | N/A | | Indications for<br>Use | The Endosee System is used<br>to permit viewing of the<br>cervical canal and uterine<br>cavity for the purpose of<br>performing diagnostic<br>procedures. Generally<br>recognized indications for<br>diagnostic hysteroscopy<br>include: Abnormal bleeding,<br>Infertility and pregnancy<br>wastage, Evaluation of | The Endosee Hysteroscope<br>and Disposable Diagnostic (Dx)<br>Cannula are used to permit<br>viewing of the cervical canal<br>and uterine cavity for the<br>purpose of performing<br>diagnostic procedures.<br>Generally recognized<br>indications for diagnostic<br>hysteroscopy include:<br>Abnormal bleeding, Infertility | Same | | | | | | | | abnormal<br>hysterosalpingogram,<br>Intrauterine foreign body,<br>Amenorrhea, Pelvic Pain<br><br>The Endosee System can<br>also be used to permit<br>viewing of the urinary<br>bladder through a minimally<br>invasive approach by<br>utilizing natural orifices for<br>the purpose of performing<br>diagnostic procedures. | and pregnancy wastage,<br>Evaluation of abnormal<br>hysterosalpingogram,<br>Intrauterine foreign body,<br>Amenorrhea, Pelvic Pain<br><br>The Endosee and Disposable<br>Diagnostic (Dx) Cannula can<br>also be used to permit viewing<br>of the urinary bladder through<br>a minimally invasive approach<br>by utilizing natural orifices for<br>the purpose of performing<br>diagnostic procedures. | | | Technology | Handheld, battery-operated<br>endoscope that consists of a<br>reusable, detachable<br>Display Module and sterile,<br>disposable Cannula with a<br>handle; Wi-Fi capability | Handheld battery-operated<br>endoscope that consists of a<br>reusable Handheld Monitor<br>and a sterile, disposable<br>Cannula | Different. The subject<br>device provides<br>updated technology<br>features and<br>component<br>configurations. | | Cannula Type | Flexible | Flexible | Same | | Patient<br>Contacting<br>Materials | PEEK, ink, glass, adhesive,<br>lens coating, surface black<br>coating | Glass, stainless steel, nylon,<br>acrylic and adhesive | Different. The subject<br>device contains PEEK,<br>ink, and coatings. | | Cannula<br>Working<br>Length | 276mm | 276mm | Same | | Cannula<br>Largest Outer<br>Diameter | 4.5mm | 4.8mm | Different. The subject<br>cannula largest outer<br>diameter is smaller<br>than that of the<br>predicate. | | Light Source | LED | LED | Same | | Field of View | $100° ± 5°$ | $100° ± 5°$ | Same | | Direction of<br>View | $20° ± 3°$ | $20° ± 3°$ | Same | | Image Display | Handheld LCD display<br>module | Handheld LCD display module | Same | | Cannula<br>Sterilization | Ethylene oxide, SAL⁻⁶ | Ethylene oxide, SAL⁻⁶ | Same | | Number of<br>Uses | Cannula: single-use,<br>disposable<br>Display Module: reusable | Cannula: single-use,<br>disposable<br>Display Module: reusable | Same | | Cannula Shelf-<br>Life | 6 months | 3 years | Different. The subject | | | | | shelf-life than the<br>predicate. | | Packaging | Cannula: Individually<br>pouched or individually<br>pouched with a pipette<br>Display Module: Boxed with<br>Docking Station, power<br>cord, and USB cable | Cannula: Individually pouched<br>or individually pouched with a<br>pipette<br>Display Module: Boxed with<br>Docking Station, power cord,<br>and USB cable | Same | Table 1: Subject and Predicate Device Comparison {5}------------------------------------------------ {6}------------------------------------------------ The subject Endosee System has the same indications for use as the predicate device. Both devices have the same method of sterilization, number of uses, and patient contact duration. As described above, there are differences in materials and technology between the subject and predicate device. However, these differences do not raise different questions of safety or efficacy. #### Non-Clinical Performance Testing As part of demonstrating substantial equivalence to the predicate the following non- clinical tests were performed: #### ● Sterility Testing per ISO 11135-1:2014 - EO Residuals per ISO 10993-7:2008 O - Bioburden per ISO 11737-1:2006 & 11737-2:2009 O - Biocompatibility Testing per ISO 10993-1:2009 - Cytotoxicity per ISO 10993-5:2009 o - Sensitization per ISO 10993-10:2010 O - Irritation per ISO 10993-10:2010 o - Shelf Life of Cannula ● ● - Aged to 6-months per ASTM F1980-16 o - O The following tests were performed on 6-month accelerated aged samples: - . Seal Tensile Strength - . Seal Peel - Functional (Image Quality, Flow Rate, Leak, Bend) - Ship Testing per ISTA 3A:2008 - Bubble Leak per ASTM 2096-11 O #### ● Functional Testing - Image Quality Test (Including Resolution) o - Optical Distortion Test o - Saline Flow Rate Test o - O Leak Test - 90° Bend Test O - 3-Point Bend Test O - LED Temperature Test O - O Drop Test per IEC 60601-1-15 - IPX2 Ingress Test per IEC 60529:2004 O - System Level Verification O {7}------------------------------------------------ - EMC and Electrical Safety . - o UL Electrical Safety per ANSI/AAMI ES 60601-1:2005/(R)2012 and IEC 60601-2-18:2009 - o UL EMC per IEC 60601-1-2:2014 and IEC 60601-2-18:2009 - Software Verification per IEC 62304:2006 and as recommended in the FDA guidance ● "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" - Cybersecurity/Intraoperability Information provided as recommended in FDA . guidance, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" - . Wireless information provided as recommended in FDA guidance, "Radio Frequency Wireless Technology in Medical Devices Guidance" - Design Validation #### Conclusion The results of the testing described above demonstrate that the subject Endosee System is substantially equivalent to the predicate.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...