Retrospective clinical case series; Published clinical literature (Miller et al., 2003)
A retrospective case series of 19 patients was used to evaluate post-operative adhesion rates and bleeding. These results were compared to historical adhesion rates reported in literature for the predicate device (MeroGel) to support substantial equivalence.
Retrospective case series; Adhesion rate; FESS; Historical comparison
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective case series of PuraSinus; Retrospective case series; Follow-up/Duration: Post-operative period
PuraSinus® is a space occupying gel stent intended to separate and prevent adhesions between mucosal surfaces, help to control minimal bleeding following surgery or nasal trauma, and act as an adjunct to aid in the natural healing process. PuraSinus® is indicated following nasal/sinus surgery or trauma to prevent lateralization of the middle turbinate and formation of nasal adhesions during the post-operative period.
Device Story
PuraSinus® is a sterile, synthetic peptide-based hydrogel provided in a prefilled syringe; used as a space-occupying wound dressing in ENT procedures. Upon contact with physiological fluids (blood), the peptide solution forms a transparent gel without volume expansion. Applied by clinicians post-nasal/sinus surgery or trauma to separate mucosal surfaces, control minimal bleeding, and support healing. The gel is non-animal/non-plant derived and contains no preservatives. It is removed by gentle saline flushing, which allows for non-traumatic removal of secondary dressings. The device acts as a physical barrier to prevent adhesions and lateralization of the middle turbinate. Benefits include a moist wound environment and reduced post-operative adhesion formation.
Clinical Evidence
Retrospective case series of 19 patients undergoing functional endoscopic sinus surgery (FESS). Primary endpoint: post-operative adhesion formation and re-bleeding. Results: 0% re-bleeding; 16% adhesion rate (3/19 patients), with adhesions described as mild, soft, and easy to separate. Compared favorably to literature-reported 27% adhesion rate for the predicate MeroGel device in similar procedures.
Technological Characteristics
Synthetic peptide-based hydrogel; 2.5% peptide content. Provided in 1, 3, or 5 ml cyclo-olefin polymer syringes with HDPE plunger and butyl rubber components. Delivered via polypropylene applicator nozzle. Terminally sterilized. Resorbable. Standalone device.
Indications for Use
Indicated for patients following nasal/sinus surgery or trauma to prevent middle turbinate lateralization and nasal adhesion formation.
Regulatory Classification
Identification
Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.
Predicate Devices
Medtronic Xomed, Inc. MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel (K122434)
{0}------------------------------------------------
In accordance with 21 CFR 807.87(h) and 21 CFR 807.92, the 510(k) Summary for PuraSinus® is provided below.
#### SUBMITTER 1.
Dr. Lisa Spirio, Chief Scientific Officer, Co-Founder 3-D Matrix, Inc. 250 First Ave., Suite 205 Needham, MA 02494
Contact Person: Stephen Rhodes, Affiliate, Biologics Consulting Group, Inc. Phone: 301-579-3700 Email: stephen.rhodes@streamlineregulatory.com
Date Prepared: April 11, 2019
#### DEVICE 2.
Name of Device: PuraSinus® Common Name: polymer, ear, nose and throat, synthetic, absorbable Classification Regulation: 21 CFR 874.3620 Regulatory Class: II Product Code: NHB Panel: Ear Nose & Throat
## PREDICATE DEVICE 3.
Predicate Device: Medtronic Xomed, Inc. MeroGel Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel (K122434)
Reference Device: 3-D Matrix, Inc. PuraDerm Gel (K143058)
## DEVICE DESCRIPTION 4.
PuraSinus® is a sterile gel composed of a synthetic peptide and sterile water for injection. It is provided as a prefilled syringe (2.5% peptide content) ready for use as a wound dressing with or without the optional sterile application nozzle. PuraSinus® forms a moist wound environment that is supportive of the healing process and allows non-traumatic removal of the secondary dressing without damaging newly formed tissue.
PuraSinus® is completely non-animal and non-plant derived and contains no preservatives that might present a risk of allergic reaction or skin irritation.
Exposure to physiological fluids such as blood causes the peptide solution to quickly form a transparent gel without expansion in volume. PuraSinus® can be easily rinsed away by gently flushing the wound with sterile saline, without causing trauma to the underlying wound.
{1}------------------------------------------------
## 5. INDICATIONS FOR USE
PuraSinus® is a space occupying gel stent intended to separate and prevent adhesions between mucosal surfaces, help to control minimal bleeding following surgery or nasal trauma, and act as an adjunct to aid in the natural healing process.
PuraSinus® is indicated following nasal/sinus surgery or trauma to prevent lateralization of the middle turbinate and formation of nasal adhesions during the post-operative period.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS 6.
The similarities and differences in technological characteristics between the subject device (PuraSinus®), predicate device (MeroGel® - K122434), and the reference device (PuraDerm Gel - K143058) are summarized below.
PuraSinus® and the predicate MeroGel® (K122434) are both space-occupying gel stents intended to be used in nasal surgery to separate and prevent adhesions between mucosal surfaces, help control minimal bleeding, and act as an adjunct to aid in the natural healing process.
The PuraSinus® solution is sterile-filtered and filled into 5-ml syringes made of cyclo-olefin polymers with a high-density polyethylene plunger and a butyl rubber head cap and gasket. Each syringe is filled with either 1, 3, or 5 ml of gel. The device is terminally sterilized, and the resorbable gel is delivered to the intended application site(s) via a polypropylene applicator nozzle tip.
The predicate MeroGel® is also a space-occupying gel provided in a prefilled syringe. The gel is cross-linked polymers of hyaluronic acid and is resorbable. Like the predicate device, it is terminally sterilized.
Both devices need to deliver the gel via a syringe and both need to function as a temporary space-occupying gel stent.
In terms of technological differences, PuraSinus® consists of a synthetic peptide-based hydrogel material provided in a prefilled syringe. The predicate MeroGel® is made from cross-linked polymers of hyaluronic acid. Both devices are hydrogels designed to be used as a barrier to prevent adhesion formation and bleeding after nasal and/or sinus surgical procedures. The differences in the material compositions of the two devices raises no new questions of safety and effectiveness. Additionally, the difference in material between the devices was addressed by a side-by-side comparison of the devices in a rabbit model. The difference in the material did not impact the safety and effectiveness of PuraSinus® when compared to the MeroGel® predicate.
Overall, the differences in technological characteristics of the subject and predicate devices do not raise any new questions of safety and effectiveness.
Lastly, the subject PuraSinus® is the identical product cleared as the PuraDerm Gel reference device (K140358) although for different indications.
{2}------------------------------------------------
#### PERFORMANCE DATA 7.
The following bench tests were conducted on PuraSinus® and the MeroGel® predicate:
- Complex Modulus ●
- Complex Viscosity
- Complex Modulus, Simulated Body Fluid
- Complex Viscosity, Simulated Body Fluid
- . Iniection Force.
The difference in material between the devices was addressed by a side-by-side comparison of the devices in a rabbit model. The difference in the material did not impact the safety and effectiveness of PuraSinus® when compared to the MeroGel® predicate.
PuraSinus® was also evaluated in a retrospective case series of 19 sequential patients requiring functional endoscopic sinus surgery (FESS). There were no adverse events related to the operation/usage of PuraSinus®. In all 19 patients, no post-operative re-bleeding was observed. Three of 19 patients developed adhesions for a rate of 16 %. The adhesions that formed in 3 patients were determined to be mild, soft and easy to separate. The adhesion results were compared to a literature article that reported adhesion rates with the predicate MeroGel® for similar procedures. [Miller RS. Steward DL. Tami TA et al. The clinical effects of hyaluronic acid ester nasal dressing (MeroGel®) on intranasal wound healing after functional endoscopic sinus surgery. Otolaryngology Head and Neck Surgery 2003 Jun; 128(6):862-9.] In this study, an adhesion rate of 27% was found after MeroGel® use following ethmoid surgery. The adhesion rate of 16% in this case series using PuraSinus® compares favorably to the 27% rate of adhesion formation with the predicate MeroGel® device.
# CONCLUSIONS 8.
The subject PuraSinus® has the same intended use and similar indications for use at the predicate MeroGel® Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel (K122434). Both devices are space occupying gel stents that are intended to separate and prevent adhesions between mucosal surfaces, help control minimal bleeding following surgery or nasal trauma, and act as an adjunct to aid in the natural healing process.
The indications for use for the subject are a subset of the predicate device, with the difference being that the predicate device includes indications for specific surgical procedures not included in the subject device indications for use. Thus, PuraSinus® does not have a new intended use.
The subject and predicate device are both bioresorbable gels that are intended to fill ENT cavities after trauma or surgery. Both gel devices are provided sterile in syringes.
The subject device is the identical product as the PuraDerm reference device (K143058). The biocompatibility testing and product characterization studies performed on the PuraDerm reference device apply to PuraSinus®.
In conclusion. PuraSinus® is substantially equivalent to the predicate MeroGel® Injectable Bioresorbable Nasal Dressing and Sinus Stent and Otologic Gel (K122434).
{3}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 12, 2019
3-D Matrix, Inc. % Stephen Rhodes Affiliate Biologics Consulting Group, Inc. 1555 King Street, Suite 300 Alexandria, Virginia 22314
Re: K183015
Trade/Device Name: PuraSinus Regulation Number: 21 CFR 874.3620 Regulation Name: Ear, Nose, And Throat Synthetic Polymer Material Regulatory Class: Class II Product Code: NHB Dated: February 28, 2019 Received: March 4, 2019
Dear Stephen Rhodes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{5}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.