K182976 · Eneura®, Inc. · OKP · Feb 25, 2019 · Neurology
Device Facts
Record ID
K182976
Device Name
SpringTMS
Applicant
Eneura®, Inc.
Product Code
OKP · Neurology
Decision Date
Feb 25, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5808
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The eNeura® Spring TMS® is designed and intended to deliver brief duration, pulsed, magnetic fields that are externally directed at spatially discrete regions of the brain to induce electric currents in the brain (Product Code OKP). The eNeura Inc. SpringTMS® is indicated for the acute and prophylactic treatment of migraine headache in adolescents (age 12 and older) and adults.
Device Story
Portable, hand-held device; delivers brief single pulse of magnetic energy (0.9 Tesla) to back of head; induces electrical current in occipital cortex to treat migraine. Used in home or office settings; self-administered by patient; prescription only. No changes in design, manufacturing, or functionality from predicate. Output is magnetic pulse; intended to stop or lessen migraine effects. Benefits include non-invasive acute and prophylactic migraine management.
Clinical Evidence
No new clinical testing conducted for this submission. Relies on data from predicate K162797, which included the prospective, single-arm, non-randomized ESPOUSE study.
Technological Characteristics
Portable, hand-held transcranial magnetic stimulator. Delivers 0.9 Tesla magnetic pulse. Induces electrical current in occipital cortex. No changes in materials or design from predicate.
Indications for Use
Indicated for acute and prophylactic treatment of migraine headache in adolescents (age 12+) and adults.
Regulatory Classification
Identification
A transcranial magnetic stimulator device for headache is a device that delivers brief duration, rapidly alternating, or pulsed, magnetic fields that are externally directed at spatially discrete regions of the brain to induce electrical currents for the treatment of headache.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing must demonstrate electromagnetic compatibility, electrical safety, and thermal safety.
(2) Appropriate verification, validation, and hazard analysis must be performed on the device software and firmware.
(3) The elements of the device that contact the patient must be assessed to be biocompatible.
(4) Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. This includes full characterization of the magnetic pulse output and resulting magnetic field map. This also includes characterization of the sound level of the device during use.
(5) Clinical testing must demonstrate that the device is safe and effective for treating headache in the indicated patient population.
(6) The physician and patient labeling must include the following:
(i) A summary of the clinical performance testing, including any adverse events and complications.
(ii) The intended use population in terms of the types of headaches appropriate for use with the device.
(iii) Information on how to report adverse events and device malfunctions.
(iv) A diagram or picture depicting the proper placement of the device on the user.
1. Appropriate analysis/testing must demonstrate electromagnetic compatibility (EMC), electrical safety, and thermal safety.
2. Appropriate verification, validation, and hazard analysis must be performed on the device software and firmware.
3. The elements of the device that contact the patient must be assessed to be biocompatible.
4. Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. This includes full characterization of the magnetic pulse output and resulting magnetic field map. This also includes characterization of the sound level of the device during use.
5. Clinical testing must demonstrate that the device is safe and effective for treating headache in the indicated patient population.
6. The physician and patient labeling must include the following:
a. A summary of the clinical performance testing, including any adverse events and complications.
b. The intended use population in terms of the types of headaches appropriate for use with the device.
c. Information on how to report adverse events and device malfunctions.
d. A diagram or picture depicting the proper placement of the device on the user.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 25, 2019
eNeura® Inc. % Larry Getlin Regulatory Consultant for eNeura Inc. Larry W. Getlin 2690 Pheasant Road Orono, Minnesota 55331
Re: K182976
Trade/Device Name: SpringTMS Regulation Number: 21 CFR 882.5808 Regulation Name: Transcranial magnetic stimulator for headache Regulatory Class: Class II Product Code: OKP Dated: January 17, 2019 Received: January 18, 2019
Dear Larry Getlin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# John Marler -- C Digitally signed by John Marler -S
Date: 2019.02.25 16:58:23 -05'00'
For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182976
Device Name SpringTMS®
#### Indications for Use (Describe)
The eNeura Inc. Spring TMS® is indicated for the acute and of migraine headache in adolescents (age 12 and older) and adults.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K182976
#### GENERAL INFORMATION [807.92(a)(1)]
#### Applicant:
eNeura® Inc. 715 North Pastoria Avenue Sunnyvale, CA 94085 U.S.A. Phone: 408-245-6400 FAX: 408-245-6424
#### Contact Person:
Larry W. Getlin Regulatory Consultant for eNeura® Inc. 2690 Pheasant Road Orono, MN 55331 USA Phone: 612-850-8144
### Date Prepared:
October 24, 2018
#### DEVICE INFORMATION [807.92(a)(2)]
Trade Name: SpringTMS®
Generic/Common Name: Transcranial magnetic stimulator for headache
Classification: 21 CFR§882.5808
Product Code: OKP
#### PREDICATE DEVICE(S) [807.92(a)(3)]
The eNeura SpringTMS is substantially equivalent to the eNeura SpringTMS predicate device K162797. These products are technically identical. This traditional 510(k) is for a labeling expansion only.
#### DEVICE DESCRIPTION [807.92(a)(4)]
The SpringTMS® is a portable, hand-held device that is designed and intended to deliver a brief single pulse of magnetic energy at 0.9 Tesla to the back of the head to induce an electrical current in a portion of the brain called the occipital cortex to stop or lessen the effects of migraine headaches. Since a single pulse of magnetic stimulation is emitted, this method of
{4}------------------------------------------------
stimulation is called single pulse transcranial magnetic stimulation or sTMS. The SpringTMS is indicated for the acute and prophylactic treatment of migraine headache. The device is designed for patient use where treatments are self-administered and can be delivered in a variety of settings including the home or office. The device is intended for prescription use only.
The SpringTMS delivers the same energy and maintains the same operational characteristics as the SpringTMS device cleared in K162797. No changes in design or manufacturing process have been made that could affect device functionality. All functional aspects of the device remain the same as K162797, including the strength and nature of the magnetic field generated and the pulse generation.
## INTENDED USE/INDICATIONS FOR USE [807.92(a)(5)]
The eNeura® Spring TMS® is designed and intended to deliver brief duration, pulsed, magnetic fields that are externally directed at spatially discrete regions of the brain to induce electric currents in the brain (Product Code OKP).
The eNeura Inc. SpringTMS® is indicated for the acute and prophylactic treatment of migraine headache in adolescents (age 12 and older) and adults.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a)(6)]
The eNeura SpringTMS is substantially equivalent to the previously cleared SpringTMS device (K162797). It has the same Intended Use (to deliver externally directed, pulsed, magnetic fields to induce electric currents in spatially discrete regions of the brains of patients with migraine headache). No changes in materials, design or manufacturing process have been made that could affect device functionality. All functional aspects of the device remain the same, as described in K162797. including the strength and nature of the magnetic field generated and the pulse generation. Thus, the technological characteristics remain the same as the predicate device.
| Feature | Subject Device | Predicate Device | Analysis of Differences |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | SpringTMS | SpringTMS | Same |
| Manufacturer | eNeura Inc. | eNeura Inc. | Same |
| 510(k) Number | Not Assigned | K162797 | N/A |
| Regulation Number | 21 CFR§882.5808 | 21 CFR§882.5808 | Same |
| Class | Class II | Class II | Same |
| Classification | Transcranial magnetic stimulator for<br>headache | Transcranial magnetic stimulator for<br>headache | Same |
| Product Code | OKP | OKP | Same |
| Indications for Use | The eNeura Inc. SpringTMS® is<br>indicated for the acute and<br>prophylactic treatment of migraine<br>headache in adolescents (age 12 and<br>older) and adults. | The eNeura Inc. SpringTMS® is<br>indicated for the acute and<br>prophylactic treatment of migraine<br>headache. | Expansion of<br>indications for use does<br>not raise new issues of<br>safety and effectiveness.<br>Support for expansion<br>of intended use is<br>provided in this<br>submission. |
| | | Table 1: Substantial Equivalence Table – Regulatory Information |
|--|--|------------------------------------------------------------------|
| | | |
{5}------------------------------------------------
| Feature | Subject Device | Predicate Device | Analysis of Differences |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Intended Use | The eNeura® SpringTMS® is designed and intended to deliver brief duration, pulsed, magnetic fields that are externally directed at spatially discrete regions of the brain to induce electric currents in the brain. | The eNeura® SpringTMS® is designed and intended to deliver brief duration, pulsed, magnetic fields that are externally directed at spatially discrete regions of the brain to induce electric currents in the brain. | Same |
| Fundamental<br>Scientific Technology | Portable, hand-held device that is designed and intended to deliver a brief single pulse of magnetic energy at 0.9 Tesla to the back of the head to induce an electrical current in a portion of the brain called the occipital cortex to stop or lessen the effects of migraine headaches. | Portable, hand-held device that is designed and intended to deliver a brief single pulse of magnetic energy at 0.9 Tesla to the back of the head to induce an electrical current in a portion of the brain called the occipital cortex to stop or lessen the effects of migraine headaches. | Same |
Table 1: Substantial Equivalence Table - Regulatory Information (Cont.)
## SUBSTANTIAL EQUIVALENCE
The Intended Use/Indications for Use for the predicate device (K162797) is substantially equivalent to the proposed Intended Use/Indications for Use for the SpringTMS device. There are no differences in the technological characteristics between the devices so no new issues of safety or effectiveness are raised. Thus, the subject SpringTMS is substantially equivalent to the predicate SpringTMS device.
## PERFORMANCE DATA [807.92(b)]
The SpringTMS device is the same as the predicate device and no technological changes have occurred. All previous performance testing continues to apply.
## Clinical Testing Summary [807.92(b)(2)]
The SpringTMS device is the same as the predicate device and no technological changes have occurred. Thus, clinical testing was not conducted for this premarket notification, nor was required to support the safety and performance of the SpringTMS for the expanded Intended Use/Indications for Use population. Please refer to K162797 for further information regarding the prospective, single-arm, non-randomized Non-Significant Risk clinical study (ESPOUSE) conducted by eNeura using the SpringTMS device.
## CONCLUSIONS [807.92(b)(3)]
The SpringTMS device is the same as the predicate device and no technological changes have occurred. There are no differences in the technological characteristics between the devices, therefore no new issues of safety or effectiveness are raised. Thus, the SpringTMS is substantially equivalent to the predicate device.
## SUMMARY
The SpringTMS is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.