K182965 · Ast Products, Inc. · MSS · Mar 29, 2019 · Ophthalmic
Device Facts
Record ID
K182965
Device Name
BL-Cart IOL Delivery Cartridge
Applicant
Ast Products, Inc.
Product Code
MSS · Ophthalmic
Decision Date
Mar 29, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4300
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The BL-Cart™ IOL Delivery Cartridge is a single use, sterile device intended to insert an approved single-piece foldable acrylic intraocular lens (IOL) into the human eye through a surgical procedure. The cartridge is intended to be used in conjunction with Alcon Monarch® III IOL Delivery System injector. The BL-Cart™ IOL Delivery Cartridge is only for the insertion of Lenstec Softec I IOL and IOL models validated for use with this device as indicated in the IOL approved labeling.
Device Story
Single-use, sterile cartridge for intraocular lens (IOL) delivery; used during cataract surgery. Device loads into Alcon Monarch® III IOL Delivery System injector; surgeon pushes IOL through cartridge into eye. Cartridge features LubriMATRIX™ hydrophilic/lubricious polymer coating to facilitate lens folding and passage. Used in hospital/surgery center settings by ophthalmologists. Output is the successful implantation of the IOL into the eye. Benefits include safe, consistent delivery of foldable IOLs without damaging lens optical or physical properties.
Clinical Evidence
No clinical data. Bench testing only. Simulated surgical manipulation evaluated IOL optical/dimensional properties, surface homogeneity, and lens opening time post-delivery. Biocompatibility (ISO 10993-5, -10, -11), sterilization, and shelf-life testing performed. Results confirmed successful delivery of Lenstec Softec I IOLs without lens damage.
Technological Characteristics
Single-use polypropylene cartridge with LubriMATRIX™ hydrophilic/lubricious polymer coating. Dimensions/form factor compatible with Alcon Monarch® III injector. Sterilized via ethylene oxide (EtO) to SAL 10^-6. Complies with ISO 11979-3:2012, ISO 10993-5:2009, and ISO 10993-10:2010.
Indications for Use
Indicated for patients undergoing cataract surgery requiring the implantation of an approved single-piece foldable acrylic intraocular lens (IOL), specifically Lenstec Softec I IOL or other models validated for use with the device.
Regulatory Classification
Identification
An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.
Predicate Devices
Monarch® III IOL Delivery System (D Cartridge) (K063155)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
March 29, 2019
AST Products, Inc. % Dr. David Lim VP. R&D and Regulatory Affairs 9 Linnell Circle Billerica. MA 01821
Re: K182965
Trade/Device Name: BL-Cart™ IOL Delivery Cartridge Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular Lens Guide Regulatory Class: Class I Product Code: MSS Dated: February 14, 2019 Received: February 19, 2019
Dear Dr. Lim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours.
Jennifer N. Brown -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182965
Device Name BL-Cart™ IOL Delivery Cartridge
#### Indications for Use (Describe)
The BL-Cart™ IOL Delivery Cartridge is a single-use, sterile device intended to insert an approved single-piece foldable acrylic intraocular lens (IOL) into the human eye through a surgical procedure. The cartridge is intended to be used in conjunction with Alcon Monarch® III IOL Delivery System injector.
The BL-Cart™ IOL Delivery Cartidge is only for the insertion of Lenstec Softec I IOL and IOL models validated for use with this device as indicated in the approved IOL labeling.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <span style="display:inline-block;"> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |
| <span style="display:inline-block;"> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
As required by the Safe Medical Devices Act (SMDA) of 1990 and in accordance with 21 CFR § 807.92, 510(k) summary is provided.
## 1. Submitter
AST Products, Inc. 9 Linnell Circle Billerica, MA 01821 USA Tel: (978) 667-4500 Fax: (978) 667-9778
## 2. Official Correspondence/Contact Person
William Lee. Ph.D. VP, R&D and Regulatory Affairs AST Products, Inc. 9 Linnell Circle Billerica, MA 01821 USA Tel: (978) 667-4500 Fax: (978) 667-9778 Email: wlee@astp.com
## 3. 510(k) Preparer
Regulatory Doctor™ RCTC, LLC 3955 Riner Road, Box 148 Riner, VA 24149 Tel: (800) 321-8567 Email: info@regulatorydoctor.com
## 4. Product Name
- Trade Name: BL-Cart™ IOL Delivery Cartridge (Model: Type D)
- Common Name: Folders and Injectors, Intraocular Lens (IOL) |
## 5. Device Classification
- Classification Name: Class I (21 CFR § 886.4300 Intraocular lens guide)
- Classification Panel: Ophthalmic (86)
- Product Code: MSS
## 6. Legally Marketed Predicates
- K063155: Monarch® III IOL Delivery System (D Cartridge)
## 7. Device Description
BL-Cart™ IOL Delivery Cartridge (Model: Type D) is an intraocular lens (IOL) delivery
{4}------------------------------------------------
cartridge solution capable of supplying the ophthalmologist surgeon with a cartridge precoated with lubricious coating for use during a typical cataract surgery. The BL-Cart™ cartridge is used to deliver an approved IOL lens.
The BL-Cart™ IOL Delivery Cartridge and referenced predicate device. Monarch® III D cartridge, utilize the same lens folding system design in which IOLs are folded and/or delivered from an injector system and the resulting optical and physical properties of the IOL remain unchanged as a result of folding/delivery.
The BL-Cart™ IOL Delivery Cartridge comes with one pre-coated cartridge, packaged in a Tyvek pouch, placed in a 10 unit box as intended for marketing, and sterilized by ethylene oxide (EtO).
The BL-Cart™ IOL Delivery Cartridge and the referenced legally marketed predicate device share the same principal method to implant an intraocular lens into the eve.
Note: LubriMATRIX™ is a proprietary treatment composed of a hydrophilic polymer and a lubricious polymer engrafted onto the surface of the cartridge via a chemical polymerization process. Refer to MAF-1963 for further information. LubriMATRIX™ is applied to AST Products' current line of IOL Delivery Systems, lioli™ IOL Delivery System (K142056) and pioli™ IOL Delivery System (K172228).
#### 8. Intended Use and Indications for Use
The BL-Cart™ IOL Delivery Cartridge is a single use, sterile device intended to insert an approved single-piece foldable acrylic intraocular lens (IOL) into the human eye through a surgical procedure. The cartridge is intended to be used in conjunction with Alcon Monarch® III IOL Delivery System injector.
The BL-Cart™ IOL Delivery Cartridge is only for the insertion of Lenstec Softec I IOL and IOL models validated for use with this device as indicated in the IOL approved labeling.
The BL-Cart™ IOL Delivery Cartridge and the referenced legally marketed predicate device share the same principal folding method to implant an intraocular lens into the eye and similar intended use and indications for use for lens insertion into the eye, and thus does not raise any new questions or concerns regarding safety and effectiveness. Refer to Table 1.
{5}------------------------------------------------
| Trade<br>Name | Manufacturer | Regulation<br>No. | Product<br>Code | Regulation<br>Description | 510(k) No. | Intended Use &<br>Indications for Use |
|-------------------------------------------------------------|-------------------------|-------------------|-----------------|---------------------------|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| BL-Cart™<br>IOL Delivery<br>Cartridge<br>(Model:<br>Type D) | AST Products,<br>Inc. | 886.4300 | MSS | Intraocular<br>Lens Guide | New device | The BL-Cart™ IOL Delivery Cartridge<br>is a single use, sterile device intended<br>to insert an approved single-piece<br>foldable acrylic intraocular lens (IOL)<br>into the human eye through a surgical<br>procedure. The cartridge is intended<br>to be used in conjunction with Alcon<br>Monarch® III IOL Delivery System<br>injector.<br>The BL-Cart™ IOL Delivery Cartridge<br>is only for the insertion of Lenstec<br>Softec I IOL and IOL models validated<br>for use with this device as indicated in<br>the IOL approved labeling. |
| Monarch® III<br>IOL Delivery<br>System (D<br>Cartridge) | Alcon Research,<br>Ltd. | 886.4300 | MSS | Intraocular<br>Lens Guide | K063155 | The MONARCH® III IOL Delivery<br>System is used for implantation of<br>ALCON® qualified ACRYSOF®<br>intraocular lenses into the eye<br>following cataract removal. |
Table 1. Device Comparisons – Intended Use and Indications for Use
{6}------------------------------------------------
#### 9. Technological Characteristics
The BL-Cart™ IOL Delivery Cartridge is an intraocular lens delivery cartridge used for progressive folding in combination with a reusable handpiece to deliver an intraocular lens into the eye for replacement of the human crystalline lens during cataract surgery. The device consists of a lubricious coated cartridge in preparation for intraocular lens loading and lens folding.
The BL-Cart™ IOL Delivery Cartridge shares similar technological characteristics for lens insertion into the eye with the predicate device referenced in this 510(k) premarket notification application. Refer to Tables 2 and 3.
{7}------------------------------------------------
| Trade<br>Name | Design | Operating Principle | Biocomp. | Sterilization<br>Method | Setting<br>Used | Patient<br>Contact | Recommended<br>for Use With | Human<br>Factors | Performance<br>Criteria Met? |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|---------------------------------------|--------------------------------|-------------------------------------------------------|------------------------------|-------------------------------|------------------------------|
| BL-Cart™<br>IOL<br>Delivery<br>Cartridge<br>(Model:<br>Type D) | A sterile,<br>single- use<br>disposable<br>coated<br>cartridge | The one-piece<br>disposable cartridge<br>is loaded into a<br>reusable handpiece<br>injection system and<br>the IOL is pushed<br>through the cartridge<br>into the eye | Yes | ETO sterilization<br>SAL level 1x10-6 | Hospital/<br>Surgery<br>center | Distal end<br>of cartridge<br>to place<br>lens in eye | FDA-approved<br>viscoelastic | For use by<br>doctors<br>only | Yes |
| Monarch®<br>III IOL<br>Delivery<br>System (D<br>Cartridge) | Monarch® III<br>reusable<br>handpiece and<br>single-use,<br>sterile coated<br>cartridge | The one-piece<br>disposable cartridge<br>is loaded into a<br>reusable handpiece<br>injection system and<br>the IOL is pushed<br>through the cartridge<br>into the eye | Yes | ETO sterilization | Hospital/<br>Surgery<br>center | Distal end<br>of cartridge<br>to place<br>lens in eye | FDA-approved<br>viscoelastic | For use by<br>doctors<br>only | Yes |
Table 2. Device Comparisons – Technological Characteristics
## Table 3. Device Comparisons – Materials
| Trade Name | Device Materials | Standards Met |
|-------------------------------------------------|---------------------------------------------------------------------------------------|------------------------------------------------------------|
| BL-Cart™ IOL Delivery Cartridge (Model: Type D) | Cartridge - Polypropylene (RJ880MO-12) | ISO 11979-3:2012<br>ISO10993-5: 2009<br>ISO 10993-10: 2010 |
| Monarch® III IOL Delivery System (D Cartridge) | Cartridge- Polypropylene with a polyvinylpyrrolidone (PVP) coating on the inner lumen | ISO 11979-3:2012<br>ISO10993-5: 2009<br>ISO 10993-10: 2010 |
{8}------------------------------------------------
#### 10. Performance Data (Nonclinical and/or Clinical)
- Non-clinical Tests:
The BL-Cart™ IOL Delivery Cartridge contacts the patient's eye directly for a limited (≤24hr) time. The device also has limited contact with the IOL. Based on the contact duration, cytotoxicity, sensitization and ocular irritation tests were performed. In addition, acute systemic toxicity and material-mediated pyrogenicity were also performed. The biocompatibility testing was performed in accordance with International Standard Organization (ISO) Biocompatibility evaluation of medical devices-parts 5. 10 and 11. The biocompatibility testing was found acceptable.
All lenses were delivered through the BL-Cart™ IOL Delivery Cartridge according to the loading and delivery instructions.
The IOLs were evaluated for optical properties, dimensional properties and overall surface and bulk homogeneity before and after being surgically manipulated using the BL-Cart™ IOL Delivery Cartridge, as well as lens opening time after folding. The BL-Cart™ was also evaluated for its cartridge performance, such as overall cartridge surface and bulk homogeneity.
IOL optical properties and overall surface and bulk homogeneity inspection were conducted in accordance with ISO 11979-3:2012, Ophthalmic implants – Intraocular lenses - Part 3: Mechanical properties and test methods.
After delivery, lenses were observed for possible damages or scratches using a microscope and showed passing results, and were within dimensional specifications. BL-Cart™ also showed no damage after lens delivery.
The resulting data from simulated surgical manipulation of BL-Cart™ IOL Delivery Cartridge (Model: Type D) to deliver intraocular lenses showed that the device can successfully deliver Lenstec's hydrophilic Softec 1 IOLs of low to high diopters and IOL models validated for use with this device as indicated in the IOL approved labeling without affecting the functionality of the lens.
- Clinical Tests: Not required for this device.
#### 11. Conclusion
The resulting data from simulated surgical manipulation of BL-Cart™ IOL Delivery Cartridge to deliver intraocular lenses showed that BL-Cart™ (Model: Type D) can successfully deliver Lenstec Softec I IOL and IOL models validated for use with this device as indicated in the IOL approved labeling without affecting the functionality of the lens.
Based on the assessment of these findings, along with the results of Biocompatibility, Sterilization and Shelf-Life testing, we conclude that the BL-Cart™ IOL Delivery Cartridge is substantially equivalent to the referenced legally marketed predicate device provided in this 510(k) premarket notification submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.