K182963 · Zimmer Medizinsysteme GmbH · NGX · Jun 21, 2019 · Physical Medicine
Device Facts
Record ID
K182963
Device Name
emFieldPro
Applicant
Zimmer Medizinsysteme GmbH
Product Code
NGX · Physical Medicine
Decision Date
Jun 21, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
emFieldPro is indicated to be used for: - Improvement of abdominal tone, strengthening of the abdominal muscles, development for firmer abdomen. - Strengthening, toning and firming of buttocks and thighs.
Device Story
emFieldPro is a mobile, non-invasive powered muscle stimulator; generates electromagnetic fields via applicators to induce nerve action potentials, resulting in muscle contraction. Device comprises a system console with touch-screen interface, computer, and electromagnetic field generators; two applicator sizes (large/small) available. Operated by clinician in clinical setting; user controls frequency, time, and intensity via rotary knob and LCD. Device intended to strengthen, tone, and firm abdomen, buttocks, and thighs. Benefits include non-invasive muscle conditioning. Safety features include housing to protect against electrical/mechanical injury; tissue heating studies confirm no appreciable temperature rise at maximum intensity.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included magnetic field intensity verification, tissue heating studies, and adherence to electrical safety and electromagnetic compatibility standards (IEC 60601 series).
Technological Characteristics
Powered muscle stimulator; generates magnetic fields via applicators. Energy source: 100-240 VAC, 50/60 Hz. Output: Symmetrical Biphasic Sine Wave. Pulse repetition rate: 1-150 Hz. Magnetic field intensity: 0.5-1.5T (large applicator), 0.5-2.0T (small applicator). Connectivity: Standalone mobile unit with touch-screen interface. Software: Firmware-controlled. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-10, IEC 62366-1, IEC 62304, ISO 14971.
Indications for Use
Indicated for individuals seeking improvement of abdominal tone, strengthening of abdominal muscles, development of a firmer abdomen, and strengthening, toning, and firming of buttocks and thighs.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
June 21, 2019
Zimmer MedizinSysteme GmbH % Scott Blood Principal Regulatory Consultant Quality and Regulatory Services 22 Nichols Street #2 Salem, Massachusetts 01970
Re: K182963
Trade/Device Name: emFieldPro Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: May 15, 2019 Received: May 16, 2019
Dear Scott Blood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Vivek Pinto, PhD Assistant Director, Acute Injury Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182963
Device Name emFieldPro
Indications for Use (Describe)
emFieldPro is indicated to be used for:
- · Improvement of abdominal tone, strengthening of the abdominal muscles, development for firmer abdomen.
· Strengthening, toning and firming of buttocks and thighs.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The word "zimmer" is in bold, gray font. Underneath the word "zimmer" is the word "MedizinSysteme" in a smaller, gray font.
# 510(k) Summarv K182963
- 1. Basic Information-Submitter:
510(k) Owner:
Zimmer MedizinSysteme GmbH Junkersstrasse 9 89231 Neu-Ulm Germany Establishment Registration: 8010720
- Official Contact:
Mrs. Ute Hauss Manager Regulatory Affairs Phone: +49-731-9761-216 Fax: +49-731-9761-118 E-mail: u.hauss@zimmer.de
Date Summary Prepared:
14 May 2019
- 2. Device Name:
Trade Name: Common Name: Classification Name:
Regulation Number: Product Code: Classification:
emFieldPro Powered muscle stimulator Stimulator, Muscle, Powered, For Muscle Conditioning 21 CFR 890.5850 NGX Class II
- 3. Predicate Devices: BTL 799-2 – K180813
- 4. Device Description:
emFieldPro is a non-invasive therapeutic device. The device produces electromagnetic field that interacts with the tissues of the human body. By muscle stimulation, the emFieldPro helps to strengthen, tone and firm the abdomen, buttocks and thighs.
The device housing protects the patient from electrical shock and mechanical injuries. The device is mobile standalone equipment with four wheels. Two applicators are available for therapy: a large and a small one. The main body of emFieldPro is used to control function of magnetic stimulation. It is operated with parameters such as Frequency, Time and Intensity. These parameters can be controlled by the user on LCD and with the help of a rotary knob at the user control panel.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "z". Below the word "zimmer" is the text "MedizinSysteme" in a smaller, sans-serif font. The logo is in grayscale.
- 5. Indications for Use Statement:
emFieldPro is indicated to be used for:
- Improvement of abdominal tone, strengthening of the abdominal muscles, development for ● firmer abdomen.
- . Strengthening, toning and firming of buttocks and thighs.
- 6. Technological Characteristics:
The emFieldPro device has the same indications for use and similar technological characteristics and principles of operation as its predicate device. The emFieldPro device and its predicate are comprised of a system console and applicator. The system console consists of the electromagnetic field generators, computer, and the touch-screen control panel.
The emFieldPro is equipment that generates a magnetic field by applying a strong current to an applicator. The essential technical characteristics of the emFieldPro and its predicate device, particularly the type of energy, type of operation, therapy time and the Magnetic Field Intensity are equivalent.
Compared to the predicate, the emFieldPro device is equipped with an arm which allows adjustment of the applicator right in front of the patient in an optimized position for treatment.
The technological differences between the emFieldPro device and the predicate device do not raise any new types of safety or effectiveness questions.
| Technological<br>Characteristics | SUBJECT DEVICE<br>Zimmer MedizinSyteme GmbH<br>emFieldPro<br>K182963 | PREDICATE DEVICE<br>BTL Industries, Inc.<br>BTL 799-2<br>K180813 |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code and<br>Regulation | Physical Medicine<br>21 CFR 890.5850<br>NGX - Stimulator Muscle,<br>Powered, Muscle Conditioning | Physical Medicine<br>21 CFR 890.5850<br>NGX - Stimulator Muscle,<br>Powered, Muscle Conditioning |
| Indications for Use | The emFieldPro is indicated to be<br>used for:<br>• Improvement of abdominal<br>tone, strengthening of the<br>abdominal muscles,<br>development of firmer<br>abdomen.<br>• Strengthening, Toning and<br>Firming of buttocks and thighs. | BTL 799-2 is indicated to be used<br>for:<br>• Improvement of abdominal<br>tone, strengthening of the<br>abdominal muscles,<br>development of firmer<br>abdomen.<br>• Strengthening, Toning and<br>Firming of buttocks and thighs. |
| Primary Function | Muscle stimulation | Muscle stimulation |
| Principle of Action | Initiating action potential of nerves | Initiating action potential of nerves |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "z". Below the word "zimmer" is the text "MedizinSysteme" in a smaller, sans-serif font.
| Technological<br>Characteristics | SUBJECT DEVICE<br>Zimmer MedizinSyteme GmbH<br>emFieldPro | PREDICATE DEVICE<br>BTL Industries, Inc.<br>BTL 799-2 | |
|--------------------------------------------------------|-----------------------------------------------------------------|-------------------------------------------------------|--|
| | K182963 | K180813 | |
| | results in muscle contraction | results in muscle contraction | |
| Electrical Protection | Class I, BF | Class II, BF | |
| User Interface | Touch screen | Touch screen | |
| Touch screen size | 7" | 15.6" (39.6 cm) | |
| Firmware Controlled | Yes | Yes | |
| Type of Energy | Magnetic field | Magnetic field | |
| Number of outputs | 2 | 2 | |
| Number of Magnetic | 1 | 1 | |
| Coils in the Applicator<br>Magnetic Field<br>Intensity | Large applicator 0.5 - 1.5T +/-20%<br>Small applicator 0.5 - 2T | 0.5-1.8 T | |
| Total Induced Current<br>in Tissue (mA) | 251 | 285 | |
| Type of Operation | Continuous | Continuous | |
| Pulse Repetition Rate | 1 – 150 Hz | 1 – 150 Hz | |
| Pulse Duration | 400 ps +/- 20% | 280 +/- 20% µs | |
| Pulse Amplitude | 0 - 100 % | 0 - 100 % | |
| Selection of<br>parameters (Intensity,<br>Time) | Yes | Yes | |
| Therapy Time | Up to 60 min | Up to 60 min | |
| Shape of Stimulation<br>Pulse | Symmetrical Biphasic Sine Wave | Symmetrical Biphasic Sine Wave | |
| Energy Source | 100 - 240 VAC, 50/60 Hz | 100 - 240 V AC, 50-60 Hz | |
| System Dimensions<br>(WxHxD) | 542x501x993mm | 500×1380×580 mm<br>(20×55×23 in) | |
| Operating Ambient<br>Temperature | 10°C to 30 °C | -10°C to +55°C | |
| Environmental<br>Specifications | For indoor use only | For indoor use only | |
The Zimmer MedizinSysteme GmbH emFieldPro has the same technological characteristics as the predicate device except for the following:
| Technological<br>Characteristic | Characteristic difference<br>between emFieldPro and<br>Predicate Device | Discussion on why this difference does not affect<br>the overall safety and effectiveness of the subject<br>device when compared to the predicate device |
|---------------------------------|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Protection | Class I versus Class II | Electrical safety is covered as class 1 device equal to a<br>class 2 device. |
| Touch Screen Size | 7" versus 15.6" (39.6 cm) | Even the size of the touch screen of emFieldPro is<br>smaller than the predicate device all icons and data are<br>displayed clearly and usability is passed successfully. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a line above the "z". Below the word "zimmer" is the text "MedizinSysteme" in a smaller, lighter font. The logo is in gray color.
| Technological<br>Characteristic | Characteristic difference<br>between emFieldPro and<br>Predicate Device | Discussion on why this difference does not affect<br>the overall safety and effectiveness of the subject<br>device when compared to the predicate device |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Magnetic Field<br>Intnesity | 0.5 - 1.5T +/- 20% for the large<br>applicator ad 0.5 – 2.0T for the<br>small applicator versus 0.5 to<br>1.8T | The maximum field intensity for the emfieldPro's small<br>applicator is 2.0T as compared to the 1.8T of the<br>predicate device. This difference is within 20% of the<br>acceptable variability in field intensity. |
| Total Induced Current<br>in Tissue (mA) | 251 mA versus 285 mA | The total induced current on the tissue for the<br>emFieldPro is less than the predicate device and<br>therefore safer with the same effect. The difference<br>between the two induced current measurements are<br>within just over 10%, which is not great enough of a<br>variation to produce a different clinical effect. |
| Pulse Duration | 400 $\mu$ s +/- 20% for the large<br>applicator and 250 $\mu$ s +/- 20%<br>for the small applicator versus<br>280 +/- 20% $\mu$ s at the predicate<br>device | The pulse width of the emFieldPro device and the<br>predicate device are in the typical clinical range of 50 to<br>500 $\mu$ s<br>Reference: "the effect of stimulus current pulse width on<br>nerve fiber size recruitment patterns" by Robert B.<br>Szlavik and Hubert de Bruin |
| System Dimensions<br>(WxHxD) | 501x993x542 mm versus<br>580×1380×580 mm | Different dimensions have no influence on the safety or<br>effectiveness of the device. The usability is passed<br>successfully. |
| Operating Ambient<br>Temperature Range | 10°C to 30 °C versus<br>-10°C to +55°C | The emFieldPro labeling allows for device usage in an<br>environment that is more typical than the environment<br>stated in the predicate labeling |
Any differences in their technological characteristics are explained to demonstrate in this submission that these differences do not raise any new questions of safety and effectiveness.
## 7. Performance data
The emFieldPro device has been investigated and tested against and complies with the following voluntary standards:
| Standards | Standards<br>Organization | Standards Title |
|------------|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 60601-1 | IEC | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance |
| 60601-1-2 | IEC | Medical electrical equipment – Part 1-2: General requirements<br>for basic safety and essential performance – Collateral standard:<br>Electromagnetic disturbances – Requirements and tests |
| 60601-1-6 | IEC | Medical electrical equipment - Part 1-6: General requirements<br>for basic safety and essential performance - Collateral<br>standard: Usability |
| 60601-2-10 | IEC | Medical electrical equipment – Part 2–10: Particular<br>requirements for the Basic Safety and Essential Performance of<br>Nerve and Muscle Stimulators |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "i" and "e". Below the word "zimmer" is the phrase "MedizinSysteme" in a smaller, sans-serif font. The logo is simple and modern, and the use of a bold font makes it easily recognizable.
| Standards | Standards<br>Organization | Standards Title |
|-----------|---------------------------|---------------------------------------------------------------------------------------------------------------------|
| 62366-1 | IEC | Medical devices – Application of usability engineering to medical<br>devices |
| 62304 | IEC | Medical devices software -software life cycle processes |
| 14971 | ISO | Medical devices – Application of risk management to medical<br>devices |
| 10993-1 | ISO | No Biocompatibility testing performed as the applied doesn't get<br>in contact with the patient<br>(see section 16) |
The following table shows a comparison of the performance testing in comparison to the predicate devices:
| Standards | SUBJECT DEVICE<br>Zimmer MedizinSyteme GmbH<br>emFieldPro<br>K182963 | PREDICATE DEVICE<br>BTL Industries, Inc.<br>BTL 799-2<br>K180813 |
|---------------|----------------------------------------------------------------------|------------------------------------------------------------------|
| IEC 60601-1 | X | X |
| IEC60601-2-10 | X | X |
| IEC 60601-1-2 | X | X |
| ISO 10993-1 | Not applicable<br>(no patient-contacting materials) | X |
According to this comparison table all required performance tests were conducted and show substantial equivalence with the predicate devices.
## Preclinical Testing Results
The following tests were performed on the subject device in addition to the testing listed above:
- Magnetic Field testing ●
- Tissue Heating study
The testing above confirmed that the large and small applicator each operate within the magnetic field intensity specifications for each applicator and that the tissue being treated by the device does not present an appreciable rise in temperature at maximum intensity to cause a risk to the patient.
Testing has been performed and all components, subassemblies and/or full devices and systems have met the required specifications for the completed tests.
### 8. 510(k) Summary:
Zimmer MedizinSysteme GmbH has demonstrated that the emFieldPro device is substantially equivalent to the predicate device listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.