K182926 · Philips Healthcare Informatics, Inc. · LLZ · Dec 6, 2018 · Radiology
Device Facts
Record ID
K182926
Device Name
IntelliSpace Radiology
Applicant
Philips Healthcare Informatics, Inc.
Product Code
LLZ · Radiology
Decision Date
Dec 6, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
IntelliSpace Radiology Software (client) is a software package intended to be used by trained professionals, including but not limited to Radiologists, physicians, administrators and medical technicians. The software is used with general purpose computing hardware for the presentation, processing, measurement and distribution of images and associated data throughout a clinical environment. IntelliSpace Radiology contains an Advanced Mammography module for functionality specific to Mammography. Lossy compressed mammographic images and digitized film/screen images must not be reviewed for primary image interpretations. Mammography images may only be interpreted using display monitors approved for digital mammography.
Device Story
IntelliSpace Radiology is a client-server software system for radiology image management. It receives, processes, displays, and distributes medical images (CT, MR, NM, US, XA, PET, CR, DX, DR, RF, RT, MG, SC, VL, OP) via DICOM. Operated by clinicians on general-purpose hardware, it provides tools for image navigation, measurement (linear, angle, Cobb, ROI), and annotation. It supports advanced mammography, tomosynthesis slice indicators, and slab views. The system integrates with hospital/radiology information systems and third-party applications like PowerScribe. Clinicians use the output for diagnostic review and clinical decision-making. The device facilitates efficient workflow, remote reading, and teleradiology, potentially improving diagnostic accuracy and patient care delivery.
Clinical Evidence
No clinical data required. Substantial equivalence supported by non-clinical verification and validation testing, including software life cycle processes (IEC 62304), risk management (ISO 14971), and usability engineering (IEC 62366-1).
Technological Characteristics
Software-based PACS; client-server architecture; Windows OS; TCP/IP networking; DICOM compliant. Supports high-resolution diagnostic monitors. Features include image processing, measurement tools, and Advanced Mammography module. Software level of concern: moderate. Sterilization: N/A (software).
Indications for Use
Indicated for use by trained professionals (radiologists, physicians, administrators, technicians) for the presentation, processing, measurement, and distribution of medical images and data in clinical environments. Includes an Advanced Mammography module. Contraindicated for primary interpretation of lossy compressed mammographic images or digitized film/screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Philips Healthcare Informatics, Inc. % Mark Job Official Correspondent Regulatory Technology Services, LLC 1394 25th Street. Nw BUFFALO, MN 55313
December 6, 2018
Re: K182926
Trade/Device Name: IntelliSpace Radiology Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: November 27, 2018 Received: November 28, 2018
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K182926
Device Name IntelliSpace Radiology
### Indications for Use (Describe)
IntelliSpace Radiology Software (client) is a software package intended to be used by trained professionals, including but not limited to Radiologists, physicians, administrators and medical technicians.
The software is used with general purpose computing hardware for the presentation, processing, measurement and distribution of images and associated data throughout a clinical environment.
IntelliSpace Radiology contains an Advanced Mammography module for functionality specific to Mammography. Lossy compressed mammographic images and digitized film/screen images must not be reviewed for primary image interpretations.
Mammography images may only be interpreted using display monitors approved for digital mammography.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Philips Healthcare Informatics, Inc.
Image /page/3/Picture/2 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. A thin black line is visible at the bottom of the image.
5 510(K) SUMMARY
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and aligned horizontally. The font is sans-serif and appears to be a standard typeface. The background is white, providing a strong contrast to the blue letters.
# 510(k) Summary
# IntelliSpace Radiology
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92.
Date Prepared: September 5, 2018
# I. Submitter's name and address
| Manufacturer: | Philips Healthcare Informatics, Inc.<br>4430 Rosewood Drive, Suite 200<br>Pleasanton, CA 94588<br>USA<br>Establishment Registration Number: 2954704 |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Elfriede Pagan<br>Sr. Manager, Regulatory Affairs<br>Phone: (321) 339-9144<br>E-mail: phiiregulatory@philips.com |
### II. Device information
### Subject Device:
IntelliSpace Radiology Device Name: Common/Usual Name: Radiology Image and Information Management System Classification: Classification name: Picture Archiving and Communications System Device class: Class II Classification regulation: 21 CFR 892.2050 Classification panel: Radiology Product Code: LLZ
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and appear to be a sans-serif font. A thin black line is present underneath the word, spanning its entire length.
#### III. Predicate device information
## Predicate Device:
| Trade name: | IntelliSpace PACS 4.x |
|----------------------------|---------------------------------------------|
| Manufacturer: | Philips Healthcare Informatics, Inc. |
| 510(k) clearance: | K111804 |
| Classification name: | Picture Archiving and Communications System |
| Device class: | Class II |
| Classification regulation: | 21 CFR 892.2050 |
| Classification panel: | Radiology |
| Product code: | LLZ |
#### IV. Device Description
The subject device IntelliSpace Radiology product described in this section is a modification of the previously cleared IntelliSpace PACS 4.x software device (K111804). Changes made or being implemented are described in Section 12 'Substantial Equivalence Discussion'.
IntelliSpace Radiology is a software package intended to be used by trained professionals, including but not limited to physicians, administrators and medical technicians.
The software is used with general purpose computing hardware for the presentation, processing. measurement and distribution of images and associated data throughout a clinical environment.
IntelliSpace Radiology software supports receiving, sending, printing, and displaying studies received from the following modality types via DICOM: CT, MR, NM, US, XA, PET, CR, DX, DR, RF, RT, MG, SC, VL, and OP as well as hospital/radiology information systems.
IntelliSpace Radiology contains an Advanced Mammography module for functionality specific to Mammography. Lossy compressed mammographic images and digitized film/screen images must not be reviewed for primary image interpretations.
# V. Indications for Use
IntelliSpace Radiology Software (client) is a software package intended to be used by trained professionals, including but not limited to Radiologists, physicians, administrators and medical technicians.
The software is used with general purpose computing hardware for the presentation, processing, measurement and distribution of images and associated data throughout a clinical environment.
IntelliSpace Radiology contains an Advanced Mammography module for functionality specific to Mammography. Lossy compressed mammographic images and digitized film/screen images must not be reviewed for primary image interpretations.
Mammography images may only be interpreted using display monitors approved for digital mammography.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and appear to be a sans-serif font. The background is white, providing a strong contrast that makes the word stand out.
#### Comparison of Technological Characteristics with the Predicate Device VI.
The subject device IntelliSpace Radiology (ISR) product described in this section is a modification of the previously cleared IntelliSpace PACS 4.x software device (K111804). The software is used with general purpose computing hardware for the presentation, processing, measurement and distribution of images and associated data throughout a clinical environment by healthcare professionals.
IntelliSpace Radiology product uses the standard principles of operation typically seen in PACS systems such as database and image management systems, image processing tools, and standard measurement tools. Both the subject device and the predicate device provide Diagnostic Review Solution for radiology and other areas where images are utilized, such as Visible Light. ISR is utilizing client-server technology, communication and interoperability with hospital systems, such as radiology workflow providers and image archives.
A comparison matrix below (please see Table 5-1 below) provides a comparison which outlines a high-level overview of the differences and similarities between IntelliSpace Radiology and the predicate device, IntelliSpace PACS 4.x (K111804).
| # | Specification / Feature | IntelliSpace<br>Radiology<br>(subject device) | Predicate<br>IntelliSpace<br>PACS 4.x<br>(K111804) |
|------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | Software Image and Information management<br>system | Yes | Yes |
| 2. | Hardware Platform requirements | Yes | Yes |
| System Configuration | | | |
| 3. | Windows Operating System | Yes | Yes |
| 4. | TCP-IP Network Protocol | Yes | Yes |
| 5. | Supports High Resolution Diagnostic Monitors | Yes | Yes |
| 6. | Storage capabilities | Not supported | Yes |
| 7. | Multiple monitor support | Yes | Yes |
| Communication and Interoperability with other image management systems | | | |
| 8. | Supports DICOM studies received from different<br>modalities types | Yes -<br>CT, MR, NM, US,<br>XA, PET, CR, DX,<br>DR, RF, RT, MG,<br>SC, VL, and OP | Yes-<br>CT, MR, NM, US,<br>XA, PET, DX, DR,<br>RF, RT, MG, SC,<br>VL |
| 9. | Support for IHE Profiles for Key Image Notes (KIN)<br>and Digital Breast Tomosynthesis (DBT) | Yes | No |
| Operating Platform requirements | | | |
| 10. | Client-server technology | Yes | Yes |
| 11. | Thin client installer | Yes | Yes |
| 12. | Multiple concurrent user support | Yes | Yes |
| Interfaces with other Image management systems | | | |
| # | Specification / Feature | IntelliSpace<br>Radiology<br>(subject device) | Predicate<br>IntelliSpace<br>PACS 4.x<br>(K111804) |
| 13. | Supported Data and Multi Modalities | Supports<br>receiving,<br>sending, printing,<br>storing and<br>displaying studies<br>received from the<br>following modality<br>types via DICOM:<br>CT, MR, NM, US,<br>XA, PET, CR, DX,<br>DR, RF, RT, MG,<br>SC, VL, and OP<br>as well as<br>hospital/<br>Radiology<br>information<br>systems. | Supports<br>receiving,<br>sending, printing,<br>storing and<br>displaying studies<br>received from the<br>following modality<br>types via DICOM:<br>CT, MR, NM, US,<br>XA, PET, DX, DR,<br>RF, RT, MG, SC,<br>VL as well as<br>hospital/<br>Radiology<br>information<br>systems. |
| 14. | Federation | Yes | Yes |
| 15. | Vendor Neutral Archive for Client | Yes | Yes |
| | Patient-Exam Worklist Browser | | |
| 16. | Patient Exam Query Tool | Yes | Yes |
| 17. | Patient Exam Worklist Filters | Yes | Yes |
| | Image Display/Formatting/Navigation | | |
| 18. | Multiple Monitor layout option | Yes | Yes |
| 19. | Scales image to window | Yes | Yes |
| 20. | Next/Previous Image/ Series/ Page Options | Yes | Yes |
| 21. | Cine Toolset | Yes | Yes |
| 22. | Image/Series Linking | Yes | Yes |
| 23. | Image Rotate/Flip | Yes | Yes |
| 24. | Image Zoom/Pan | Yes | Yes |
| 25. | Scout lines Tool | Yes | Yes |
| 26. | Image Stacking | Yes | Yes |
| 27. | Hanging Protocols | Yes | Yes |
| 28. | Grayscale Softcopy Presentation State (GSPS) | Yes | Yes |
| 29. | Tomosynthesis Slice Indicator | Yes | No |
| 30. | Bidirectional Tomo Localizer | Yes | No |
| 31. | Tomo Slab View | Yes | No |
| 32. | Tomo Localizer Mode | Yes | No |
| | Measurement and Annotations | | |
| 33. | Measurement Tools | Yes | Yes |
| 34. | Linear Measurements (Ruler) | Yes | Yes |
| 35. | Angle | Yes | Yes |
| 36. | Cobb Measurement | Yes | Yes |
| 37. | ROI Circle | Yes | Yes |
| 38. | ROI Freehand | Yes | Yes |
| # | Specification / Feature | IntelliSpace<br>Radiology<br>(subject device) | Predicate<br>IntelliSpace<br>PACS 4.x<br>(K111804) |
| 40. | Point Value | Yes | Yes |
| 41. | Calibrating Images | Yes | Yes |
| 42. | Angle Line Ratio | Yes | No |
| 43. | Ultrasound Measurements | Yes | Yes |
| 44. | Annotation Tools | Yes | Yes |
| | Results | | |
| 45. | View/Print Diagnostic Reports from RIS | Yes | Yes |
| | Printing/Export/Save Images... | | |
| 46. | Printing images to Paper | Yes | Yes |
| 47. | Printing images to Film | Yes | Yes |
| 48. | Save images as BMP, JPEG, TIF files | Yes | Yes |
| 49. | Export of Patient-Exams | Yes | Yes |
| | General Usage | | |
| 50. | Designed for diagnostic review of images inside<br>and outside of Radiology | Yes | Yes |
| | Clinical Tools | | |
| 51. | Remote Reading Tools / Teleradiology | Yes | Yes |
| 52. | Image Filters | Yes | Yes |
| 53. | Variable Slice Thickness for CT (Slab Reading) | Yes | Yes |
| 54. | Exam Notes | Yes | Yes |
| 55. | Key Image Series | Yes | Yes |
| 56. | Patient Merging | Yes | Yes |
| 57. | Standard Mammography | Yes | Yes |
| 58. | Advanced Mammography Module | Yes | No |
| | Management Tools | | |
| 59. | Patient Management | Yes | Yes |
| 60. | Exam Management | Yes | Yes |
| 61. | Exception Study | Yes | Yes |
| 62. | User Security Access Management | Yes | Yes |
| 63. | Dictionary Management | Yes | Yes |
| 64. | Auditing Tools | Yes | Yes |
| 65. | Billing | Yes | Yes |
| 66. | Easy Client Installer | Yes | Yes |
| 67. | Data Migration | Yes | Yes |
| 68. | Workflow Layer | Yes | Yes |
| 69. | System Monitoring | Yes | Yes |
| | Third Party Application Support | | |
| 70. | PowerScribe | Yes | Yes |
| 71. | Cardiology Enterprise Viewer | Yes | Yes |
| 72. | Application Programming Interface | Yes | Yes |
| 73. | Universal Data Manager | Yes | No |
| 74. | IntelliSpace PACS Anywhere | Yes | No |
Table 5-1 Technological characteristics comparison
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "PHILI…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.