K182910 · Micro-Tech (Nanjing) Co., Ltd. · ESW · May 29, 2019 · General, Plastic Surgery
Device Facts
Record ID
K182910
Device Name
Segmented Esophageal Stent System
Applicant
Micro-Tech (Nanjing) Co., Ltd.
Product Code
ESW · General, Plastic Surgery
Decision Date
May 29, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3610
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Segmented Esophageal Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic, or extrinsic malignant tumors only, and occlusion of concurrent esophageal fistula.
Device Story
Segmented Esophageal Stent System maintains esophageal luminal patency in patients with malignant strictures and concurrent fistulas. Device consists of a self-expanding Nitinol wire frame with PTFE connecting loops, fully covered by a silicone membrane to prevent tissue ingrowth and occlude fistulas. Flanges at both ends provide migration resistance. Radiopaque bands (4 at ends, 2 at middle) facilitate fluoroscopic visualization. Stent is pre-loaded in a delivery system allowing desheathing and recapture during placement. Used by trained healthcare professionals in clinical settings. Stent conforms to esophageal curvature due to segmented structure, reducing lumen stimulation. Retrieval loop allows repositioning during initial procedure. Benefits include restored patency and fistula occlusion, improving patient quality of life.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (cytotoxicity, sensitization, irritation, toxicity, pyrogen, mutagenicity, muscle implant), and performance testing (visual inspection, dimension, deployment/expansion/compression force, corrosion, tensile strength, sterility, shelf life, and MR compatibility).
Technological Characteristics
Nitinol wire frame, PTFE connecting loops, silicone membrane covering. Segmented design for flexibility. Radiopaque markers. EO sterilized (SAL 10^-6). Biocompatibility per ISO 10993-1. MR compatibility per ASTM F2503, F2052, F2119, F2182, F2213. Single-use, prescription only.
Indications for Use
Indicated for patients with esophageal strictures caused by intrinsic or extrinsic malignant tumors and concurrent esophageal fistula. Prescription use only.
Regulatory Classification
Identification
An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 29, 2019
Micro-Tech (Nanjing) Co., Ltd. Becky Li Quality and Regulatory Affairs Director No. 10 Gaoke Third Road Nanjing, Jiangsu 210032 China
Re: K182910
> Trade/Device Name: Segmented Esophageal Stent System Regulation Number: 21 CFR§ 878.3610 Regulation Name: Esophageal Prosthesis Regulatory Class: II Product Code: ESW Dated: April 29, 2019 Received: May 2, 2019
Dear Becky Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Shani P. Haugen, Ph.D. Acting, Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K182910
Device Name Segmented Esophageal Stent System
### Indications for Use (Describe)
The Segmented Esophageal Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic, or extrinsic malignant turnors only, and occlusion of concurrent esophageal fistula.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="padding-right: 20px;"> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows a logo with the letters "MT" in a stylized, blocky font, colored in blue. A horizontal line is placed beneath the logo. Below the line, the text "Tab 2" is written in a large, bold font. The logo appears to be a trademarked symbol, indicated by the circled 'R' symbol in the upper right corner.
# 510K Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: K182910
#### Date of Preparation: 1.
2019-05-22
#### 2. Sponsor Identification
Micro-Tech (Nanjing) Co., Ltd.
No.10 Gaoke Third Road, Nanjing National Hi-Tech Industrial Development
Zone, Nanjing, Jiangsu Province, PRC
## Establishment Registration Number: 3004837686
Contact Person: Becky Li
Position: Quality & Regulatory Affairs Director
Tel: +86-25-58646378
Fax: +86-25-58350006
Email: Beckyli_mt@outlook.com
#### 3. ldentification of Proposed Device
### Trade Name: Segmented Esophageal Stent System
Common Name: Prosthesis, Esophageal
### Regulatory Information
Classification Name: Esophageal Prosthesis
Classification: 2
Product Code: ESW
Regulation Number: 878.3610
Review Panel: Gastroenterology/Urology
#### ldentification of Predicate Device 4.
510(k) Number: K072094
Product Name: CHOOSTENT™ Covered Esophageal Stent
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional design. The letters are bold and appear to be interconnected, with the "M" slightly overlapping the "T". A small "R" in a circle, indicating a registered trademark, is positioned in the upper right corner of the logo.
510K Summary-2019-05-29 updated
Manufacturer: M. I. Tech Co., Ltd.
#### 5. Indications for Use
The Segmented Esophageal Stent System is intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic, or extrinsic malignant tumors only, and occlusion of concurrent esophageal fistula.
#### 6. Device Description
The stent is provided pre-loaded in the delivery system. The stent is made up of several segments. The main frame is woven by Nitinol wire, and the connecting part is PTFE connecting loop. The stent with this segmented structure can flexibly conform to the curvature of the human body lumen, thereby reducing the stimulation of the end to the lumen. Silicone membrane applied to the stent covers the complete body of the stent along both flanges, which means the segmented esophageal stent is a fully covered stent. The stent is formed with a flange at either end. The increased diameter of the stent at the stent ends helps provide resistance to migration. The covering is intended to reduce the risk of tissue ingrowth and provides a occlusion for esophageal fistulas. To aid in visibility under fluoroscopy there are 4 bands at either end of the stent and 2 bands at the middle of the stent. The stent has a retrieval loop which can be used to reposition the stent during the initial placement procedure if desired. The delivery system allows for desheathing to deploy the stent and recapturing the stent during stent deployment. The device is supplied sterile, intended for single use only and is available for prescription use only. Use of this device is restricted to a trained healthcare professional.
#### 7. Comparison of Technological Characteristics with a Predicate Device
The Segmented Esophageal Stent System incorporates substantially equivalent device materials, design, configuration, packaging fundamental technology, manufacturing processes, sterilization process and features the same intended use with the predicate device, namely the Esophageal Stent System manufactured by M. I. Tech Co., Ltd., under K072094.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional design. The letters are connected and appear to be leaning slightly to the right. A small "R" in a circle, indicating a registered trademark, is positioned in the upper right corner of the logo. The logo has a clean and modern look.
The proposal device is substantially equivalent to the currently marketed devices, CHOOSTENT™ covered esophageal stent, K072094.
The following table is the technological comparison between the proposed device and predicate device CHOOSTENT™ covered esophageal stent (K072094).
Table 2 Technological comparison between the proposed device and predicate device
| ITEM | Proposed Device<br>Segmented<br>Esophageal Stent<br>System | Predicate Device<br>(K072094)<br>CHOOSTENTTM covered<br>esophageal stent | Remark |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product Code | ESW | ESW | Same |
| Regulation No. | 878.3610 | 878.3610 | Same |
| Class | 2 | 2 | Same |
| Configuration | Stent and delivery<br>system | Stent and delivery system | Same |
| Indications for<br>Use | Segmented<br>Esophageal Stent<br>System is intended<br>for maintaining<br>esophageal luminal<br>patency in<br>esophageal strictures<br>caused by intrinsic,<br>or extrinsic malignant<br>tumors only, and<br>occlusion of<br>concurrent<br>esophageal fistula. | The CHOOSTENTTM<br>covered esophageal stent<br>is intended for<br>maintaining esophageal<br>luminal patency in<br>esophageal strictures<br>caused by intrinsic and or<br>extrinsic malignant<br>tumors only and<br>occlusion of concurrent<br>esophageal fistula. | Same |
| Single Use | Yes | Yes | Same |
| Diameter of<br>Stent (mm) | 18,20,22 | 18 | Different |
| Length of<br>Stent (mm) | 60, 80, 100, 120, 140 | 80, 110, 140, 170 | Different |
| Maximum OD<br>(D) of Delivery<br>System (mm) | 6.7 | 6.0 | Different |
| Working<br>Length (mm) | 650 | 700 | Different |
| ITEM | Proposed Device<br>Segmented<br>Esophageal Stent<br>System | Predicate Device<br>(K072094)<br>CHOOSTENT™ covered<br>esophageal stent | Remark |
| Covering | Fully Covered | Fully Covered | Same |
| Shelf life | 2 years | 3 years | Different |
| Packaging | EO sterilized pouch<br>with one device per<br>pouch | EO sterilized pouch with<br>one device per pouch | Same |
| Biocompatibility | Conform to ISO<br>10993-1 | Conform to ISO 10993-1 | Same |
| Sterilization | EO Sterilized,<br>SAL:10-6 | EO Sterilized, SAL:10-6 | Same |
| MRI<br>information | Comply with ASTM F<br>2503, ASTM F 2052,<br>ASTM F2119, ASTM<br>F2182, ASTM F2213 | Comply with ASTM F<br>2503, ASTM F 2052,<br>ASTM F2119, ASTM<br>F2182, ASTM F2213 | Same |
| Labeling | Conform to 21 CFR<br>part 801 | Conform to 21 CFR part<br>801 | Same |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a blue and white logo with the letters "MT" in a stylized, three-dimensional design. The "M" is formed by two angled lines that meet at a point, and the "T" is a vertical line with a horizontal line across the top. The logo has a three-dimensional effect, with the letters appearing to be raised and angled slightly. A circled "R" symbol is located in the upper right corner of the logo, indicating that it is a registered trademark.
510K Summary-2019-05-29 updated
#### 8. Performance Data
The Segmented Esophageal Stent System was conducted in accordance with, ISO 10993-1: 2009 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process", ISO 11135 'Sterilization of Health Care Products-Ethylene Oxide-Part 1:Requirements for Development, Validation, and Routine Control of Sterilization Processes for Medical Devices', and FDA's biocompatibility guidance, Use of International Standard ISO-10993-7, "Biological Evaluation of Medical Devices7: Ethylene Oxide Sterilization Residuals', the following tests were conducted:
### Stent Biocompatibility Testing:
- Vitro Cytotoxicity
- Skin Sensitization
- Irritation
- Acute Systemic Toxicity
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a blue logo with the letters 'MT' stacked on top of each other in a stylized, three-dimensional font. The 'M' is on the bottom and the 'T' is on the top. The logo has a registered trademark symbol '®' in the upper right corner. The logo appears to be for a company or organization with the initials 'MT'.
- Pyrogen ●
- Salmonella Reverse Mutation Test
- In Vitro Mammalian Cell Gene Mutation Test
- Muscle Implant
### Delivery System Biocompatibility Testing:
- Vitro Cytotoxicity
- Skin Sensitization
- Irritation
- Acute Systemic Toxicity
The device specific guidance document was consulted in preparing this premarket submission, VIII Performance testing-Bench of Guidance for the Content of Premarket Notifications for Esophageal and Tracheal Prostheses issued April 28th, 1998, the following tests were conducted for the subject device:
- ゃ Visual Inspection
- ゃ Dimension Testing
- ゃ Deployment Force Testing
- ゃ Expansion Force Testing
- ゃ Compression Force Testing
- ゃ Corrosion Testing
- や Tensile Strength Testing
- や Sterility Testing
- ゃ Shelf Life Testing
- ゃ MR Compatibility Testing
The testing performed demonstrated that the proposed device and predicate device are equivalent.
#### 9. Clinical Test Conclusion
No clinical study is included in this submission.
#### 10. Substantially Equivalent (SE) Conclusion
Based on the indications for use, technological characteristics, and safety and
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows a blue logo with the letters 'MT' in a stylized, three-dimensional design. The letters are connected and appear to be angled slightly, giving the logo a dynamic look. A small 'R' in a circle, indicating a registered trademark, is positioned in the upper right corner of the logo.
510K Summary-2019-05-29 updated
performance testing, the Segmented Esophageal Stent System has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared the CHOOSTENT™ Covered Esophageal Stent (K072094).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.