Bone Screw is used to stabilize and fixate bone grafts, bone filling material, and/or barrier membranes used for regeneration of bone in the oral cavity. Bone Tack is indicated for use to stabilize and support bone graft and/or fractured bone segments with or without bone plates or titanium mesh in oral and maxillofacial site defects.
Device Story
Bone Screw and Bone Tack are surgical fixation devices used in oral and maxillofacial procedures. Bone Screws act as anchors to secure bone plates or membranes for bone regeneration/remodeling; Bone Tacks stabilize membranes in Guided Bone Regeneration (GBR). Both devices are manufactured from Ti-6Al-4V titanium alloy. They are used by clinicians in dental/oral surgery settings to maintain the relative position of bone grafts or fractured segments. The devices are provided sterile and are intended for single use. By providing mechanical fixation, they facilitate bone healing and reconstruction in previously unsuitable sites.
Clinical Evidence
No clinical studies were submitted. Substantial equivalence is supported by non-clinical performance testing, including biocompatibility evaluation (leveraged from K172354), sterilization validation (ISO 11137, ISO 11607), and mechanical testing (ASTM F543).
Technological Characteristics
Materials: Ti-6Al-4V (ASTM F136). Principle: Mechanical fixation/anchoring. Dimensions: Bone Screw (1.4mm/1.95mm diameter; 4.0-16.0mm length); Bone Tack (2.5mm head diameter; 3.0mm length). Sterilization: Gamma radiation. Single-use, non-active, standalone devices.
Indications for Use
Indicated for patients requiring bone graft stabilization, bone filling material fixation, or barrier membrane support during bone regeneration in the oral cavity, or stabilization of fractured bone segments in oral and maxillofacial site defects.
Regulatory Classification
Identification
An intraosseous fixation screw or wire is a metal device intended to be inserted into fractured jaw bone segments to prevent their movement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is in blue text.
August 9, 2019
Osstem Implant Co., Ltd. Jungmin Yoo RA Manager 66-16, Bansong-ro 513beon-gil, Haeundae-gu Busan, 612-070 Kr
Re: K182881
Trade/Device Name: Bone Screw. Bone Tack Regulation Number: 21 CFR 872.4880 Regulation Name: Intraosseous Fixation Screw Or Wire Regulatory Class: Class II Product Code: DZL Dated: July 18, 2019 Received: July 18, 2019
Dear Jungmin Yoo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The address is 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea. The telephone number is +82 51 850 2500, the fax number is +82 51 861 4693, and the website address is www.osstem.com.
510(k) Number: K182881
Device Name: Bone Screw, Bone Tack
Indication for Use:
Bone Screw is used to stabilize and fixate bone grafts, bone filling material, and/or barrier membranes used for regeneration of bone in the oral cavity. Bone Tack is indicated for use to stabilize and support bone graft and/or fractured bone segments with or without bone plates or titanium mesh in oral and maxillofacial site defects.
X Prescription Use (21 CFR 801 Subpart D) Over-The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
QS-QI-505-7(Rev.0)
{3}------------------------------------------------
DSSTEM Implant Co., Ltd 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com
## 510(K) Summary
## K182881
### Date: August 9, 2019
### 1. Company and Correspondent making the submission
| - Submitter's Name<br>- Address<br>- Contact<br>- Phone | : OSSTEM IMPLANT Co., Ltd.<br>: 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, 612-070, Republic of Korea<br>: Ms. Jungmin Yoo<br>: +82-51-850-2500 |
|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| - Correspondent's Name<br>- Address<br>- Contact<br>- Phone | : HIOSSEN Inc.<br>: 85 Ben Fairless Dr.<br>Fairless Hills, PA, 19030<br>: PETER LEE<br>: 267-759-7031 |
- 2. Device
| - Trade or (Proprietary) Name | : Bone Screw, Bone Tack |
|-------------------------------|---------------------------------|
| - Classification Name | : Screw, Fixation, Intraosseous |
| - Regulation Number | : 21CFR872.4880 |
| - Devce Classification | : Class II |
| - Classification Product Code | : DZL |
## 3. Predicate Device
- -Primary Predicate Fixation Screw of Straumann GBR System, INSTITUT STRAUMANN AG (K011698)
- Reference Devices -
Bone Screw of Jeil Bone Fixation System, Jeil Medical Corporation (K050669) truTACK of ACE Tru-FIX Implant System, ACE Surgical Supply Co., Inc. (K080074) Osstem Implant System, OSSTEM Implant Co., LTd. (K161604) OssBuilder System, OSSTEM Implant Co., Ltd. (K172354)
{4}------------------------------------------------
Image /page/4/Picture/0 description: This image shows the logo and contact information for OSSTEM Implant Co., Ltd. The address is 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea. The telephone number is +82 51 850 2500, and the fax number is +82 51 861 4693. The website address is www.osstem.com.
# 4. Description
Bone Screw is using as anchor to fix the bone plate, membrane that cover bone material or block bone for bone regeneration or remodeling. Bone Tack is used to fix the membrane in GBR.
| Item | Content | | |
|------------|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Bone Screw | Description | Bone Screw is using as anchor to fix the bone plate, membrane that cover bone<br>material or block bone for bone regeneration or remodeling. Ø1.4 diameter<br>screw is for fixing membrane or bone plate, and Ø2.0 diameter screw is for<br>fixing block bone. | |
| | Material | Ti-6Al-4V (ASTM F 136) | |
| | Dimension<br>(mm) | D (Ø, implanted) | 1.4, 1.95 |
| | | Length | 4.0, 6.0, 8.0, 10, 12, 14, 16 |
| Bone Tack | Description | Bone Tack is a stabilizer that can stably fix the membrane in GBR. | |
| | Material | Ti-6Al-4V (ASTM F 136) | |
| | Dimension<br>(mm) | D (Ø, Head) | 2.5 |
| | | Length | 3.0 |
# 5. Substantial Equivalence Matrixes
| | Bone Screw | Straumann<br>GBR System | Jeil Bone<br>Fixation System | Remark |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | Proposed | Primary Predicate<br>(K011698) | Reference Device<br>(K050669) | - |
| Manufacturer | Osstem Implant<br>Co., Ltd. | Straumann USA | Jeil Medical<br>Corporation | - |
| Design | Image: Bone Screw | Image: Straumann GBR System | Image: Jeil Bone Fixation System | Similarity |
| Indications<br>for Use<br>Statement | Bone Screw is used<br>to stabilize and fixate<br>bone grafts, bone<br>filling material,<br>and/or barrier<br>membranes used for<br>regeneration of bone<br>in the oral cavity. | The Straumann<br>Guided Bone<br>Regeneration (GBR)<br>System is used to<br>stabilize and fixate<br>bone grafts, bone<br>filling materials,<br>and/or barrier | The device is<br>intended for use in<br>stabilizing and<br>fixating bone grafts,<br>bone filling material<br>and/or barrier<br>membranes used for<br>guided bone/tissue | Similarity.<br>Small changes<br>are due to<br>differences in<br>component-<br>specific<br>language. |
| | membranes used for<br>regeneration of bone<br>in the oral cavity.<br>The GBR technique<br>can make it possible<br>for the placement of<br>dental implants in<br>previously unsuitable<br>sites. | regeneration in the<br>oral cavity. Single<br>patient use only. | | |
| Material | Titanium Alloy<br>(Ti-6Al-4V,<br>ASTM F136) | Titanium<br>(Ti Grade 4,<br>ASTM F67) | Titanium Alloy<br>(Ti-6Al-4V,<br>ASTM F136) | Identical as that<br>of the reference<br>device |
| Screw<br>Diameter | 1.4 mm<br>1.95 mm | 1.2 mm<br>1.5 mm | 1.4 mm<br>1.6 mm<br>2.0 mm | Different but the<br>range of screw<br>diameter and the<br>screw length is<br>within the<br>permissible |
| Screw Length | Ø1.4:<br>4.0 mm<br>6.0 mm<br>8.0 mm<br>Ø1.95:<br>8.0 mm<br>10.0 mm<br>12.0 mm<br>14.0 mm<br>16.0 mm | 3.0 mm<br>4.0 mm<br>5.0 mm<br>7.0 mm<br>8.0 mm<br>9.0 mm<br>10.0 mm<br>12.0 mm<br>14.0 mm | 4.0 mm<br>5.0 mm<br>6.0 mm<br>8.0 mm<br>10.0 mm<br>12.0 mm<br>14.0 mm<br>16.0 mm | range of the<br>predicates<br>devices;<br>therefore, the<br>dimensions for<br>the proposed<br>devices are<br>substantially<br>equivalent to the<br>predicates<br>devices |
| Single Use | Yes | Yes | Yes | Similarity |
| Sterility | Sterile<br>(Shelf life: 8 years) | Non-sterile<br>(Sterilize before use) | Non-sterile<br>(Sterilize before use) | Different |
| Principle of<br>Operation | The function of<br>proposed devices is<br>used in application<br>for maintaining the<br>relative position of<br>and/or bone grafts in<br>reconstruction of<br>maxillary and/or<br>mandibular areas | The function of<br>predicate devices is<br>used in application<br>for maintaining the<br>relative position of<br>and/or bone grafts in<br>reconstruction of<br>maxillary and/or<br>mandibular areas | The function of<br>predicate devices is<br>used in application<br>for maintaining the<br>relative position of<br>and/or bone grafts in<br>reconstruction of<br>maxillary and/or<br>mandibular areas | Similarity |
| S.E. | Similarities | The proposed devices have similar indications for use with that of the primary predicate,<br>Straumann GBR System of Straumann USA (K011698). The proposed devices have<br>similar principle of operation which is to function for maintaining the relative position of<br>and/or bone grafts in reconstruction of maxillary and/or mandibular areas. The proposed<br>devices are made with same material and have similar designs compared to that of the<br>predicates. The proposed devices are single use only which is the same way as the | | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is in orange, with a small orange circle above the "M". Below "OSSTEM" is the word "IMPLANT" in gray. The logo is simple and modern.
OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea
Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is written in orange, with a small superscript "6" next to the "M". Below "OSSTEM", the word "IMPLANT" is written in gray. The logo is simple and modern.
OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea
Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com
| predicates. |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Differences</b> |
| The proposed devices have different dimensions compared to that of the predicates, but<br>the range of screw diameter and screw length is within the permissible range of the<br>predicates devices, Straumann GBR System of Straumann USA (K011698) and Jeil<br>Bone Fixation System of Jeil Medical Corporation (K050669). The proposed devices are<br>gamma sterilized while the predicates are delivered in non-sterile which required<br>sterilization prior to use. |
| .: While given dimensions are different with that of the predicates (K011698 and<br>K050669), the proposed devices have similar indications for use, similar shape and<br>principle of operation; and are made with same materials compared to that of the<br>described predicates. Therefore, the proposed devices are substantially equivalent to the<br>predicates above. |
| | Bone Tack | ACE Tru-FIX Implant System | Remark |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510(K) | Proposed | Reference Device (K080074) | - |
| Manufacturer | Osstem Implant Co., Ltd. | ACE Surgical Supply Co., Inc. | Different |
| Design | Image: Bone Tack | Image: ACE Tru-FIX Implant System | Similarity |
| Indications<br>for Use<br>Statement | It is indicated for use to stabilize<br>and support bone graft and/or<br>fractured bone segments with or<br>without bone plates or titanium<br>mesh in oral and maxillofacial<br>site defects. | It is indicated for use to stabilize<br>and support bone graft and/or<br>fractured bone segments with or<br>without bone plates or titanium<br>mesh in oral and maxillofacial<br>site defects. | Similarity |
| Material | Titanium Alloy<br>(Ti-6Al-4V, ASTM F136) | Titanium Alloy<br>(Ti-6Al-4V, ASTM F136) | Similarity |
| Head Diameter | 2.5 mm | 2.5 mm | Similarity |
| Length | 3.0 mm | 3.0 mm<br>5.0 mm | Similarity |
| Single Use | Yes | Yes | Similarity |
| Sterility | Sterile<br>(Shelf life: 8 years) | Sterile<br>(Shelf life: unknown) | Similarity |
| Principle of<br>Operation | The function of proposed device<br>is used in application for fixating<br>or stabilizing barrier membranes<br>in GBR | The function of proposed device<br>is used in application for fixating<br>or stabilizing barrier membranes<br>in GBR | Similarity |
| S.E. | Similarities | | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is in orange, with the letters connected in a stylized way. Below "OSSTEM" is the word "IMPLANT" in gray, in a smaller, sans-serif font. The logo is simple and modern, with a focus on the company name.
SSTEM Implant Co.. Ltd.
16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea 'el: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com
| | The proposed device has same indication and function, and is made of same material<br>compared to that of the predicate device, ACE Tru-FIX Implant System of ACE Surgical<br>Supply Co., Inc. (K080074). Proposed device has same head diameter and length, and<br>supplied in sterile compared to that of the predicate device. |
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Differences |
| | The proposed device has one length option while the predicates have two options in<br>length with having same diameter. |
| | : The proposed device has same dimension, made of same material, and has similar<br>indications and priniciple of operation compared to that of the predicate device;<br>therefore, the proposed device is substantially equivalent to the predicate above. |
- 6. Indications for Use
Bone Screw is used to stabilize and fixate bone grafts, bone filling material, and/or barrier membranes used for regeneration of bone in the oral cavity. Bone Tack is indicated for use to stabilize and support bone graft and/or fractured bone segments with or without bone plates or titanium mesh in oral and maxillofacial site defects.
- 7. Summary of Non-Clinical Performance Testing
Non-clinical testing data are submitted, referenced, or relied upon to demonstrate substantial equivalence.
## Biocompatibility evaluation
Biocompatibility testing was considered following the FDA Guidance Document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"and the ISO 10993 suite of standards and leveraged from reference device K172354.
### Sterilization Validation and Shelf-life
All subject devices are delivered sterile. Sterilization validation and shelf-life was considered following the ISO 11137 and ISO 11607 suites of standards. Endotoxin batch testing is in place per the FDA Guidance Document Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile.
Mechanical Properties Mechanical testing was evaluated following ASTM F543 with the worst case scenario.
- 8. Summary of Clinical Testing
No clinical studies are submitted.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The logo includes the company name in bold, black font, with the word "OSSTEM" also appearing in orange next to the word "IMPLANT" in gray. Below the logo is the company's address: 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea, along with their telephone number (+82 51 850 2500), fax number (+82 51 861 4693), and website address (www.osstem.com).
## 9. Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, OSSTEM IMPLANT Co., Ltd. concludes that the Bone Screw and Bone Tack are substantially equivalent to the predicated devices as herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.