K182845 · Alpinion Medical Systems Co., Ltd. · IYN · Dec 27, 2018 · Radiology
Device Facts
Record ID
K182845
Device Name
minisono
Applicant
Alpinion Medical Systems Co., Ltd.
Product Code
IYN · Radiology
Decision Date
Dec 27, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal; Pediatric, Small Organ (breast, testes, thyroid); Musculo-skeletal (Conventional); Musculo-skeletal (Superficial); Peripheral Vascular (PV); and Urology (including prostate).
Device Story
The minisono is a portable diagnostic ultrasound imaging system consisting of a transducer and a USB cable, designed to interface with a tablet PC (Microsoft Surface Pro). It captures ultrasound signals to produce B-mode, M-mode, Pulsed Wave Doppler (PWD), Color Doppler, Power Doppler, and Tissue Harmonic Imaging outputs. Operated by qualified physicians in clinical settings, the device provides real-time imaging for soft tissue and blood flow evaluation. The system utilizes image processing technologies (Xpeed, SRI, Spatial Compounding) to enhance image quality. The output is displayed on the tablet screen, allowing clinicians to perform diagnostic assessments, which aids in clinical decision-making and patient management. The device is intended for use in various clinical applications including fetal, abdominal, pediatric, small organ, musculoskeletal, peripheral vascular, and urological exams.
Clinical Evidence
No clinical data required. Substantial equivalence supported by bench testing, including biocompatibility, acoustic output, thermal, electrical, electromagnetic, and mechanical safety testing, and compliance with NEMA UD2/UD3 and IEC standards.
Indicated for qualified physicians to evaluate soft tissue and blood flow in fetal, abdominal, pediatric, small organ (breast, testes, thyroid), musculoskeletal (conventional/superficial), peripheral vascular, and urological (including prostate) clinical applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are displayed side-by-side.
Alpinion Medical Systems Co., Ltd. Boyeon CHO QARA Manager 5fl, i dong, 77, heungan-daero 81 beon-gil dongan-gu Anyang-si, Gyeonggi-do, 14117 REPUBLIC OF KOREA
Re: K182845
Trade/Device Name: minisono Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: November 30, 2018 Received: December 3, 2018
Dear Boyeon CHO:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
December 27, 2018
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael D.'Hara For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
#### 510(k) Number (if known)
#### K182845
Device Name minisono
Indications for Use (Describe)
The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal; Pediatric, Small Organ (breast, testes, thyroid); Musculo-skeletal (Conventional); Musculo-skeletal (Superficial); Peripheral Vascular (PV); and Urology (including prostate).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Diagnostic Ultrasound Indications for Use
## minisono Ultrasound Imaging System: minisono C1-6
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | |
|------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | N | N | N | | N | N | N | N | |
| Abdominal | N | N | N | | N | N | N | N | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | N | |
| Small Organ | | | | | | | | | |
| (breast, testes, thyroid) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal | N | N | N | | N | N | N | N | |
| (Conventional) | | | | | | | | | |
| Musculo-skeletal | N | N | N | | N | N | N | N | |
| (Superficial) | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Peripheral vessel | N | N | N | | N | N | N | N | |
| Urology (including prostate) | N | N | N | | N | N | N | N | |
N = new indication; P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH
Prescription User (Per 21 CFR 801.109)
{4}------------------------------------------------
## Diagnostic Ultrasound Indications for Use
## minisono Ultrasound Imaging System: minisono L3-12
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | |
|------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal | N | N | N | | N | N | N | N | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | | N | N | N | N | |
| Small Organ | N | N | N | | N | N | N | N | |
| (breast, testes, thyroid) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal | N | N | N | | N | N | N | N | |
| (Conventional) | | | | | | | | | |
| Musculo-skeletal | N | N | N | | N | N | N | N | |
| (Superficial) | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Intravascular (Cardiac) | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Intra-cardiac | | | | | | | | | |
| Peripheral vessel | N | N | N | | N | N | N | N | |
| Urology (including prostate) | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under appendix
* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH
Prescription User (Per 21 CFR 801.109)
{5}------------------------------------------------
# 510(k) Summary
In accordance with 21CFR807.92, the following summary of information is provided;
| Date | Oct. 5th, 2018 |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | ALPINION MEDICAL SYSTEMS Co., Ltd.<br>Address: 5fl, i dong ,77, heungan-daero 81beon-gil, dongan-gu, anyang-<br>si, gyeonggi-do , REPUBLIC OF KOREA, 14117 |
| Primary Contact<br>Person | Boyeon CHO<br>Quality Management Representative(QMR)<br>Address: 5fl, i dong ,77, heungan-daero 81beon-gil, dongan-gu, anyang-<br>si, gyeonggi-do , REPUBLIC OF KOREA, 14117<br>Phone: +82 70 7465 2104<br>Fax: +82 2 851 5593<br>Email: qa_ra@alpinion.com |
| Secondary Contact<br>Person | Kevin Chun<br>Address: 21312 30th Dr SE Ste 100 Bothell, WA 98021, United States<br>Phone: 425 949 4904<br>Fax: 425 949 4910<br>Email: kevin.chun@alpinionusa.com |
| Device Trade Name: | minisono |
| Common/Usual<br>Name: | Ultrasonic Pulsed Doppler Imaging System |
| Classification Names | System, Imaging, Pulsed Doppler Ultrasonic |
| Product Code: | Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550 90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO<br>Diagnostic Ultrasound Transducer, 21CFR 892.1570, 90-ITX |
| Primary Predicate<br>Device | K153424 E-CUBE i7 Diagnostic Ultrasound System |
| Device Description: | minisono product is an ultrasound imaging system for medical diagnosis.<br>This device is available for portable and tablet PC (Microsoft corp.,<br>Surface pro). Also, this innovative system platform provides optimal<br>patient diagnosis workflow with the tablet PC, optimal image quality. |
| | 1. Signal Mode:<br>B(2D) mode, M mode, Pulsed Wave Doppler(PWD) mode, Color Doppler,<br>Power Doppler, Tissue Harmonic Imaging |
| | 2. Combination Mode:<br>B/Color Doppler, B/PWD, B/Color Doppler/PWD |
Acoustic output track: Track 3
{6}------------------------------------------------
| | minisono C1-6 | minisono L3-12 |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Applicable<br>frequency | 1~6MHz | 3~12MHz |
| Intended Usage | Fetal,<br>Abdominal,<br>Pediatric,<br>Musculo-skeletal<br>(Conventional)<br>Musculo-skeletal<br>(Superficial)<br>Peripheral vessel<br>Urology (including<br>prostate) | Abdominal<br>Pediatric<br>Small Organ<br>Musculo-skeletal<br>(Conventional)<br>Musculo-skeletal<br>(Superficial)<br>Peripheral vessel |
| Applicable<br>mode | B mode,<br>M mode,<br>PWD mode,<br>Color Doppler<br>Power Doppler<br>Tissue Harmonic Imaging<br>Combined | B mode,<br>M mode,<br>PWD mode,<br>Color Doppler<br>Power Doppler<br>Tissue Harmonic Imaging<br>Combined |
| Scanning<br>depth(mm) | 300 | 100 |
| FOV | 60(°) | N/A |
| Steer Angle | N/A | Max 9(°) |
| Total number of<br>element | 128 | 128 |
Types of transducers compatible with the device:
Indications For Use:
The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications; Fetal; Abdominal; Pediatric, Small Organ (breast, testes, thyroid); Musculo-skeletal (Conventional); Musculo-skeletal (Superficial); Peripheral Vascular (PV); and Urology (including prostate).
#### Determination of Comparison with Predicate devices: Substantial : minisono and E-CUBE i7
## Equivalence:
| Feature | Proposed<br>minisono | Predicate<br>E-CUBE i7<br>(K153424) |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | The device is intended for use<br>by a qualified physician for the<br>evaluation of soft tissue and<br>blood flow in the clinical<br>applications; | The device is intended for use<br>by a qualified physician for the<br>evaluation of soft tissue and<br>blood flow in the clinical<br>applications; |
| | Fetal;<br>Abdominal;<br>Pediatric;<br>Small Organ (breast, testes,<br>thyroid); | Fetal;<br>Abdominal;<br>Pediatric;<br>Small Organ (breast, testes,<br>thyroid);<br>Adult Cephalic;<br>Trans-rectal(TR);<br>Trans-vaginal(TV); |
| | Musculo-<br>skeletal(Conventional);<br>Musculo-skeletal (Superficial); | Musculo-<br>skeletal(Conventional);<br>Musculo-skeletal (Superficial);<br>Cardiac (Adult); |
{7}------------------------------------------------
| | Peripheral Vessel (PV);<br>Urology (including prostate). | Cardiac (Pediatric);<br>Peripheral Vessel (PV);<br>Urology (including prostate). |
|---------------------|----------------------------------------------------------|----------------------------------------------------------------------------------|
| | | |
| Electrical<br>power | Voltage: 5Vdc, == 1A | Voltage: 19V, == 10.53A<br>Frequency: 50/60Hz<br>Power: 200W Max |
| Imaging | 2D (B) mode | 2D (B) mode |
| modes | M mode | M mode |
| | | Anatomical M |
| | Color Flow Doppler (CF) mode | Color Flow Doppler (CF) mode |
| | Power Doppler (PD) mode | Power Doppler (PD) mode |
| | Pulsed wave Doppler (PWD) | Directional PD<br>Pulsed wave Doppler (PWD) |
| | mode | mode |
| | | Continuous wave Doppler |
| | | (CWD) mode |
| | Tissue Harmonic Imaging | Tissue Harmonic Imaging |
| | mode | mode |
| Image | Xpeed™ | Xpeed™ |
| processing | SRI™ | Full SRI™ |
| technology | Spatial<br>Compounding<br>Image | Spatial Compounding<br>Image |
| | (SCI) | (SCI) |
| Software | | Panoramic |
| feature | | Needle Vision™ |
| | Needle Vision™ Plus | /Needle Vision™ Plus |
| | | Cube View™ |
| Thermal, | The minisono has been | The E-CUBE i7 has been |
| mechanical<br>and | designed to conform to the<br>following standards: | designed to conform to the<br>following standards: |
| electrical | - NEMA UD2, UD3 | - NEMA UD2, UD3 |
| safety | - AIUM Medical Ultrasound | - AIUM Medical Ultrasound |
| | Safety | Safety |
| | - IEC60601-1 | - IEC60601-1 |
| | - IEC60601-1-2 | - IEC60601-1-2 |
| | - IEC60601-2-37 | - IEC60601-2-37 |
Summary of Non-Clinical Tests:
minisono has been evaluated for biocompatibility, acoustic output as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. minisono and its application comply with voluntary standards as detailed in this premarket submission.
- IEC 60601-1:2005+AMD1:2012 (3.1ed, 2012-08-20 published), + Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC60601-1-2:2014 (4.0ed, 2014-02-25 published), Medical � Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests
- � IEC 60601-1-6:2010+AMD1:2013(3.0ed, 2013-10-28), Medical
{8}------------------------------------------------
electrical equipment - Part 1-6: General requirements forbasic safety and essential performance - Collateral standard: Usability
- � IEC60601-2-37:2007(2.0ed. 2007-08-09 published). Medical Electrical Equipment - Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment
- ISO10993-1:2009(5.0ed, � 2018-08 published), Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing
- � ાડળ 10993-5:2009(3.0ed, 2009-06 published), Biological evaluation of medical devices -- Part 5: Tests for in vitro cvtotoxicity
- � 10993-10:2010(3.0ed, 2010-08 published), ISO Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
- � ISO14971:2007(2.0ed, 2007-03 published), Application of risk management to medical devices
- AIUM/NEMA UD 2:2004 (R2009) Rev.3, 2009-08-21 published, � Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
The following quality management system measures were applied to the development of minisono:
- � Medical Device Risk Management
- � Requirements Reviews
- � Desian Reviews
- � Component Verification
- � Integration Review (System Verification)
- � Performance Testing (System Verification)
- � Safety Testing (Compliance Test)
- � Design Validation
Transducer materials and other patient contact materials are biocompatible.
## Summary of Clinical Tests:
The subject of this premarket submission, minisono, did not require clinical studies to support substantial equivalence.
## Discussion:
minisono is a new device and simple diagnostic device. minisono (minisono C1-6 & minisono L3-12) are composed system and USB cable. Software is installed to the tablet PC (Microsoft corp., Surface pro). minisono's functions of imaging/software are same as predicate E-CUBE i7 functions.
Also, the subject device is in conformance with applicable safety standards.
Therefore, the differences between minisono and the predicate devices would not affect the safety, effectiveness and essential performance.
- Conclusion: ALPINION MEDICAL SYSTEMS Co., Ltd. considers minisono to be as safe, as effective. Performance, technology and software are substantially equivalent to the predicate devices.
{9}------------------------------------------------
ALPINION MEDICAL SYSTEMS Co., Ltd. will update and include in this summary any other information deemed reasonably necessary by the FDA or the requirements will be published in guidance documents.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.